← Product Code [OEW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OEW) · K101979

# EXPLORER LIVER (PASSIVE TRACKING) (K101979)

_Pathfinder Therapeutics, Inc. · OEW · Feb 9, 2011 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94neurological-surgical-devices/OEW/K101979

## Device Facts

- **Applicant:** Pathfinder Therapeutics, Inc.
- **Product Code:** [OEW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OEW.md)
- **Decision Date:** Feb 9, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Explorer ™ Liver --- Passive Tracking device is indicated for open liver surgical procedures where image-guidance may be appropriate and where the patient can tolerate long apneic periods under general anesthesia.

## Device Story

Explorer Liver - Passive Tracking is an image-guided surgery system for open liver procedures. It maps surgical instrument positions onto preoperative MRI or CT images to assist physicians in tumor localization. The system uses an optical position sensor (Polaris Spectra or Vicra) to track passive markers on a reference, a probe, a laser range scanner, and surgical instrument adapters. The software platform, running on a PC, processes these inputs to display 3D spatial information on an LCD monitor. Physicians use this visual guidance to navigate anatomy during surgery. The system replaces active tracking with passive tracking, requiring changes to hardware and software localizer components to recognize unique rigid body designs of passive instruments. It benefits patients by providing accurate anatomical localization during complex liver surgeries.

## Clinical Evidence

Bench testing only. Validation and verification studies were conducted to evaluate performance characteristics, demonstrating the device's ability to safely and accurately perform its intended use with effectiveness similar to the predicate active tracking system.

## Technological Characteristics

System consists of a PC-based software platform, optical position sensor (Polaris Spectra or Vicra), passive tracked reference, probe, laser range scanner, and instrument adapters. Operates via optical tracking of passive markers. Connectivity via PC. Software utilizes a passive localizer component to recognize unique rigid body designs of tracked instruments.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Explorer (previously known as Linasys) Liver Image Guided Surgery System ([K071063](/device/K071063.md) & [K093494](/device/K093494.md))
- Medtronic Navigation StealthStation and its Passive Instrument Option ([K954276](/device/K954276.md) & [K972398](/device/K972398.md))
- InnerOptic InVision System ([K083728](/device/K083728.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K101979

Page 1 of 2

FEB - 9 2011

# 510(k) Summary (as required by 21 CFR 807.92(c)]

#### l. Submitted by:

Pathfinder Therapeutics, Inc. 2969 Armory Dr. Suite #100A Nashville, TN 37204

#### ll. Contact Person:

### Senhu Li

Senior Scientist Telephone: (615) 783-0094 Fax: (615) 783-0554 E-mail: senhul@pathsurg.com

#### lll. Date 510(k) Summary Prepared:

July 7th, 2010

#### Name of the Device: IV.

Explorer TM Liver --- Passive Tracking Proprietary Name: Common / Usual Name: Computer-assisted, image-guided stereotaxic system Classification Name: Stereotaxic Instrument (per 21 CFR 882.4560)

#### V. Substantial Equivalence:

The technological characteristics and indications for use of the Explorer™ Liver --- Passive Tracking are the same or similar to those found in the predicate devices. The patient contact components and component materials in both the new and predicate devices are equivalent. The packaging materials, packaging configurations, sterilization methods, and sterility assurance levels are also equivalent. The Explorer "" Liver --- Passive Tracking is substantially equivalent to the following FDA cleared frame stereotaxic systems:

- Explorer™ (previously known as Linasys) Liver Image Guided Surgery System [K071063 & 1. k093494]
- Medtronic Navigation StealthStation and its Passive Instrument Option [K954276 & K972398] 2.
- 3. InnerOptic InVision System {K083728]

{1}------------------------------------------------

K101979

#### VI. Device Description:

The Explorer" Liver --- Passive Tracking is an image-guided surgery medical device specifically designed to aid physicians during open liver procedures. The device is capable of mapping the current surgical position of tracked instruments onto preoperative, patient-specific MRI or CT medical images. These images can then be used as a guide by the physician for more accurate localization of tumors and other surrounding anatomic structures during liver surgery. The Explorer™ Liver --- Passive Tracking system consists of eight (8) components which are listed below:

- An image-guided surgery software platform installed on a personal computer (PC) (1)
- (2) An optical position sensor --- the Polaris Spectra or Vicra that can accurately localize the tracked devices listed below in passive tracking mode
	- . A passive tracked abdominal reference (used for definition of coordinate and instrument calibration)
	- . A passive tracked Localization Pen Probe
	- . A passive tracked Laser Range Scanner (LRS)
	- Two passive tracked adapters (can be attached onto up to two rigid surgical instruments ● simultaneously and used for their localization in 3-D surgical space)
- (3) An LCD display monitor.

#### Performance Data: VII.

Validation and verification studies through bench tests were conducted to evaluate the performance characteristics of the Explorer™ Liver --- Passive Tracking system. The results of these studies demonstrate that the device is capable of safely and accurately performing the stated intended use. The results also show similar effectiveness to the Explorer™ (active tracking) [K071063 & k093494].

#### VIII. Indications For Use:

The Explorer™ Liver --- Passive Tracking device is indicated for open liver surgical procedures where image-guidance may be appropriate and where the patient can tolerate long apneic periods under general anesthesia.

#### Device Modifications through this special 510(K) submission IX.

The modification on the subject device (Explorer --- Passive Tracking) over the previous cleared device (K071063 & k093494) through this special 510(K) submission is changing the tracking mode from active to passive. This device modification resulted in subsequent changes to the hardware, including the position sensor and the design of the tracked instruments. By replacing the active Localizer component in the software system with a passive Localizer component, the subject device is allowed to recognize each passive tracked instrument with its unique rigid body design.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Pathfinder Therapeutics, Inc. % Dr. Senhu Li 2969 Armory Drive, Suite 100A Nashville, Tennessee 37204

FEB - 9 201

Re: K101979

Trade/Device Name: Explorer™ Liver - Passive Tracking Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OEW Dated: January 14, 2011 Received: January 18, 2011

Dear Dr. Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Dr. Senhu Li

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

A.J. B. Rita
for

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K101979

## Indications for Use

510(k) Number (if known):

Device Name: Explorer ™ Liver --- Passive Tracking

## Indications For Use:

The Explorer ™ Liver --- Passive Tracking device is indicated for open liver surgical procedures where image-guidance may be appropriate and where the patient can tolerate long apneic periods under general anesthesia.

Prescription Use × AND/OR CFR 801 Subpart D)

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(Part 21

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R. Buntin Sur. Mxa

(Division Sign (Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number K101979

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94neurological-surgical-devices/OEW/K101979](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94neurological-surgical-devices/OEW/K101979)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
