K974098 · ETHICON, Inc. · OTN · Jan 28, 1998 · General, Plastic Surgery
Device Facts
Record ID
K974098
Device Name
PUBOUURETHRAL SLING
Applicant
ETHICON, Inc.
Product Code
OTN · General, Plastic Surgery
Decision Date
Jan 28, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TVT device is a sterile, single-use device intended to be used as a pubourethral sling indicated for treatment of stress urinary incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The TVT Introducer and Rigid Catheter Guide accessories are intended to facilitate placement of the TVT device. The accessories, available separately, are provided non-sterile and are reusable.
Device Story
The Tension Free Vaginal Tape (TVT) System is a surgical device used to treat female stress urinary incontinence. It consists of a PROLENE polypropylene mesh tape covered by a polyethylene sheath, attached to two stainless steel needles. Accessories include a stainless steel TVT Introducer and a stainless steel Rigid Catheter Guide. The system is used in a surgical setting by a physician to create a pubourethral sling. The introducer facilitates the passage of the tape from the vagina to the abdominal skin. The device provides mechanical support to the urethra to restore continence. The tape is intended for single use, while accessories are reusable.
Clinical Evidence
Clinical evaluations were conducted to demonstrate that the TVT System functions as intended. No specific metrics or sample sizes provided.
Technological Characteristics
Materials: PROLENE polypropylene mesh, polyethylene sheath, stainless steel needles/introducer/catheter guide. Principle: Pubourethral sling for mechanical support. Form factor: Mesh tape with delivery needles and accessories. Sterilization: Device is sterile; accessories are non-sterile.
Indications for Use
Indicated for female patients with stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
ProteGen Sling Collagen Impregnated Material
Submission Summary (Full Text)
{0}------------------------------------------------
.���:
# K974098
JAN 28 1998
#### SECTION 7
## SUMMARY OF SAFETY AND EFFECTIVENESS
| 510(k) Summary<br>of Safety and<br>Effectiveness | Information supporting claims of substantial equivalence, as<br>defined under the Federal Food, Drug and Cosmetic Act, respecting<br>safety and effectiveness is summarized below. For the convenience of<br>the Reviewer, this summary is formatted in accordance with the<br>Agency's final rule "...510(k) Summaries and 510(k) Statements..." (21<br>CFR 807) and can be used to provide a substantial equivalence<br>summary to anyone requesting it from the Agency. | |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| | New Device<br>Name: | Tension Free Vaginal Tape (TVT) System |
| | Predicate Device<br>Name: | ProteGen Sling Collagen Impregnated Material |
| | 510(K) SUMMARY | |
| Device Description | The Tension Free Vaginal Tape (TVT) System is comprised of three<br>components; the device (TVT device) and its accessories (TVT<br>Introducer and TVT Rigid Catheter Guide). Each is available separately<br>for use at the surgical site. The TVT device is composed of PROLENE<br>polypropylene mesh (tape). The mesh is covered with a polyethylene<br>sheath with a slit in the middle. Both the mesh and sheath are attached<br>to two (2) stainless steel needles. The TVT Introducer (accessory) is<br>made of stainless steel. It is composed of three (3) parts; handle,<br>threaded staft and rubber O-ring. The introducer functions to facilitate<br>passage of the TVT device from the vagina to the abdominal skin. The<br>TVT Rigid Catheter Guide is made of stainless steel and used to add<br>rigidity to the Foley Catheter during the surgical procedure. | |
| Intended Use | The TVT device is intended to be used as a pubourethral sling for<br>treatment of stress urinary incontinence (SUI), for female urinary<br>incontinence resulting from urethral hypermobility and/or intrinsic<br>sphincter deficiency. | |
| Indications<br>Statement | The TVT device is a sterile, single-use device intended to be<br>used as a pubourethral sling indicated for treatment of stress urinarv<br>incontinence, for female urinary incontinence resulting from urethral<br>hypermobility and/or intrinsic sphincter deficiency. The TVT Introducer<br>and Rigid Catheter Guide accessories are intended to facilitate<br>placement of the TVT device. The accessories, available separately, are<br>provided non-sterile and are reusable. | |
| Technological<br>Characteristics | Technologically both the new device and predicate device are the<br>same (i.e. both are meshes that provide pubourethral support).<br>Additionally, both devices utilize accessories for use in the surgical<br>procedure. Any differences between the two devices do not raise new<br>questions of safety and effectiveness. | |
| Performance Data | Results of clinical evaluations were used to show that the TVT System<br>functioned as clinically intended. Sufficient data has been gathered<br>from clinical testing to assess that the TVT System performs as<br>clinically intended. | |
| Conclusions | Based on the 510(k) summaries and 510(k) statements (21 CFR 807)<br>and the information provided herein, we conclude that the modified<br>device is substantially equivalent to the existing legally marketed device<br>under the Federal Food, Drug and Cosmetic Act. | |
| Contact | Gregory R. Jones<br>Director<br>Regulatory Affairs<br>ETHICON, Inc.<br>Rt. #22 West<br>Somerville, NJ 08876-0151 | |
| Date | October 28, 1997 | |
Continued on next page
{1}------------------------------------------------
K974098
### SUMMARY OF SAFETY AND EFFECTIVENESS, Continued
Tension Free Vaginal Tape (TVT) System
ETHICON, Inc.
:44:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of what appears to be an abstract human figure or symbol, possibly representing health or well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Gregory R. Jones Director, Regulatory Affairs Ethicon, Inc. P.O. Box 151 SOMERVILLE NJ 08876
SEP 2 8 2012
Re: K974098 Trade/Device Name: Tension Free Vaginal Tape (TVT) System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTN Dated: October 29, 1997 Received: October 30, 1997
Dear Mr. Jones:
This letter corrects our substantially equivalent letter of January 28, 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin R. Fuchs
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
.
.
#### INDICATION FOR USE
510(k) Number (if known):
Device Name:
Indications for Use:
Tension Free Vaginal Tape (IVI) System
The TVT device is a sterile, single-use device intended to be used as a pubourethral sling indicated for treatment of stress urinary incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The TVT Introducer and Rigid Catheter Guide accessories are intended to facilitate placement of the TVT device. The accessories, available separately, are provided nonsterile and are reusable.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | <div>OR</div> <div>Over-The Counter Use</div> |
|------------------------------------------|-----------------------------------------------|
|------------------------------------------|-----------------------------------------------|
(Optional Format 1-2-9G)
| (Division Sign-Off) | |
|---------------------|---------|
| Device | |
| 510(k) Number | k974020 |
Tension Free Vaginal Tape (TVT) System ETHICON, Inc.