MODIFICATION TO HEDROCEL TRABECULAR METAL RECONSTRUCTION SYSTEM
K032282 · Implex Corp. · EZX · Aug 21, 2003 · General, Plastic Surgery
Device Facts
Record ID
K032282
Device Name
MODIFICATION TO HEDROCEL TRABECULAR METAL RECONSTRUCTION SYSTEM
Applicant
Implex Corp.
Product Code
EZX · General, Plastic Surgery
Decision Date
Aug 21, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Hedrocel Trabecular Metal Reconstruction System is indicated for use in reinforcing weak and/or deficient bony tissues in orthopaedic surgical procedures such as pelvic reconstruction, acetabular reconstruction, cement restriction, and in long bone procedures such as femoral and humeral reconstruction. When used in long bone procedures, ancillary fixation, such as plates and screws, must be used. The Hedrocel Trabecular Metal Reconstruction System may also be used with bone graft.
Device Story
Hedrocel Trabecular Metal Reconstruction System consists of porous tantalum components used to reinforce weak or deficient bone. Device is manufactured from Hedrocel porous tantalum, characterized by 80% porosity and interconnecting pores (~0.5mm diameter). System components are provided in various cross-sections and heights. Used by surgeons in orthopaedic procedures (pelvic, acetabular, femoral, humeral reconstruction, cement restriction). In long bone applications, device requires ancillary fixation (plates/screws) and may be used in conjunction with bone graft. Device provides structural support to weakened bony tissues.
Clinical Evidence
Bench testing only. Mechanical test data from MAF #920 demonstrate sufficient strength for the indicated use in supporting weakened bony structures. No clinical data provided.
Technological Characteristics
Material: Hedrocel porous tantalum (80% porous, ~0.5mm diameter interconnecting pores). Form factor: Components with various cross-sections and heights. Energy source: None (passive implant). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring reinforcement of weak or deficient bony tissues during orthopaedic surgical procedures, including pelvic, acetabular, femoral, and humeral reconstruction, or for cement restriction. Ancillary fixation required for long bone procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Hedrocel Trabecular Metal Reconstruction System (K010378)
Related Devices
K023882 — THE HEDROCEL TRABECULAR METAL RECONSTRUCTION SYSTEM · Implex Corp. · Feb 19, 2003
K033384 — THE HEDROCEL TRABECULAR METAL RECONSTRUCTION SYSTEM, MODEL 06-165-ZZZZ · Implex Corp. · Nov 19, 2003
K032344 — THE HEDROCEL TRABECULAR METAL RECONSTRUCTION SYSTEM MODEL 10-330 · Implex Corp. · Oct 2, 2003
K040756 — THE TRABECULAR METAL RECONSTRUCTION SYSTEM, MODEL 10-340-XXXXX · Implex Corp. · Jul 14, 2004
K050101 — THE TRABECULAR METAL BONE STABILIZATION IMPLANT, MODEL 99-11197-XX-10 · Zimmer Trabecular · Apr 15, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## The Hedrocel Trabecular Metal Reconstructive System
| Submitter Name | Implex Corp. |
|------------------------------------|-------------------------------------------------------|
| And Address: | 80 Commerce Drive<br>Allendale, New Jersey 07401-1600 |
| Contact Person: | Marci Halevi |
| Phone Number: | (201) 818 - 1800, X 507 |
| Fax Number: | (973) 829 - 0825 |
| Date Prepared: | July 22, 2003 |
| Device Trade Name: | The Hedrocel Trabecular Metal Reconstructive System |
| Device Common Name: | Surgical Mesh |
| Classification Number<br>and Name: | 21 CFR § 878.3300<br>Surgical Mesh |
Substantial The term "substantial equivalence" as used in this 510(k) Equivalence: notification is limited to the definition of substantial equivalence found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
Device Hedrocel Trabecular Metal Reconstruction The System ાં તે Description: manufactured wholly of Hedrocel porous tantalum. Hedrocel porous tantalum is 80% porous with fully interconnecting pores that are about 0.5mm in diameter. This line extension adds components with a variety of cross-sections and heights to the Hedrocel Trabecular Metal Reconstruction System.
MATERIALS: Tantalum (Hedrocel porous tantalum)
- Indications for The Hedrocel Trabecular Metal Reconstruction System is indicated for use in reinforcing weak and/or deficient bony tissues in orthopaedic Use: surgical procedures such as pelvic reconstruction, acetabular reconstruction, cement restriction, and in long bone procedures such as femoral and humeral reconstruction. When used in long bone procedures, ancillary fixation, such as plates and screws, must be used. The Hedrocel Trabecular Metal Reconstruction System may also be used with bone graft.
{1}------------------------------------------------
032282
p. 2/2
Device Technological Characteristics & Comparison to Predicate Device: A comparison of device technological characteristics and properties demonstrates that the device is substantial equivalent to the cited predicate devices.
The Hedrocel Trabecular Metal Reconstructive System was tested per Performance FDA quidance documents and applicable standards for K010378, as Data: referenced in the predicate 023882. In addition, mechanical test data found in MAF #920 indicate the Hedrocel porous tantalum possesses sufficient strength for the indicated use. These results indicate that the subject device will perform as indicated for use in support of weakened bony structures.
- The Hedrocel Trabecular Metal Reconstructive System is substantially Conclusion: equivalent to the cited predicate devices identified in this premarket notification.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird with outstretched wings, rendered in a simple, bold design.
AUG 2 1 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Marci Halevi Manager of Regulatory Affairs Implex Corporation 80 Commerce Drive Allendale, New Jersey 07401-1600
Re: K032282
Trade/Device Name: The Hedrocel Trabecular Metal Reconstruction System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: EZX Dated: July 23, 2003 Received: July 24, 2003
Dear Ms. Halevi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Marci Halevi
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
sincerely yours,
Mark M Millenn
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known) :
032282
Device Name:
## The Hedrocel Trabecular Metal Reconstruction System
Indications For Use :
The Hedrocel Trabecular Metal Reconstruction System is indicated for use in reinforcing weak and/or deficient bony tissues in orthopaedic surgical procedures such as pelvic reconstruction, acetabular reconstruction, cement restriction, and in long bone procedures such as femoral and humeral reconstruction. When used in long bone procedures, ancillary fixation, such as plates and screws, must be used. The Hedrocel Trabecular Metal Reconstruction System may also be used with bone graft.
Mark n Milken
(Division Sign-C Division of Ge I, Restorative and Neurologi
510(k) Number.
) Number K032282
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED )
Concurrence of CDRH; Office of Device Evaluation (ODE)
Prescripti on Use (Per 21 CFR 801.109) OR . . .
Over - The -Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.