COMPOSIX L/P MESH WITH ECHO PS POSITIONING SYSTEM VENTRALIGHT ST MESH WITH ECHO PS POSITIONING SYSTEM
K122436 · C.R. Bard, Inc. · FTL · Nov 2, 2012 · General, Plastic Surgery
Device Facts
Record ID
K122436
Device Name
COMPOSIX L/P MESH WITH ECHO PS POSITIONING SYSTEM VENTRALIGHT ST MESH WITH ECHO PS POSITIONING SYSTEM
Applicant
C.R. Bard, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Nov 2, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ventralight™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The Echo PS™ Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair. Composix™ L/P Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects. The Echo PS™ Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
Device Story
Device consists of surgical mesh (Ventralight™ ST or Composix™ L/P) pre-attached to Echo PS™ Positioning System (nylon balloon with polycarbonate connectors). Used in laparoscopic surgery by surgeons to facilitate unrolling, positioning, and placement of mesh. Modification includes addition of anchor to inflation tube for improved connection and white markings for identification of tube cutting point. Introducer tool and inflation assembly included as accessories. Device provides mechanical support for soft tissue reconstruction; positioning system aids in laparoscopic delivery. Benefits include improved handling and placement of prosthetic mesh during minimally invasive procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing and in vivo simulated use experiments demonstrating that the modified device design meets product specifications and intended use. Biocompatibility testing was performed in accordance with ISO 10993-1 standards.
Technological Characteristics
Ventralight™ ST: PP/PGA co-knitted mesh with HA/CMC/PEG hydrogel coating. Composix™ L/P: PP mesh and ePTFE layer stitched with PTFE monofilament. Positioning system: Nylon balloon, polycarbonate connectors. Modifications: Inflation tube anchor and visual markings. Biocompatibility: ISO 10993-1 compliant. Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients requiring reconstruction of soft tissue deficiencies, including hernia repair and chest wall defects. The Echo PS™ Positioning System is indicated for use during laparoscopic hernia repair to facilitate delivery of soft tissue prosthetics.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Ventralight™ ST Mesh with Echo PS™ Positioning System (K110820)
Composix™ L/P Mesh with Echo PS™ Positioning System (K102766)
Submission Summary (Full Text)
{0}------------------------------------------------
K122436 Page 1/3
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
NOV 2 2012
This 510(k) Summary is provided per the requirements of section 807.92(c). The information listed below applies to and is identical for both devices being modified as part of thirobundly of submission.
#### Submitter Information:
| Submitter's Name: | Davol, Inc., Subsidiary of C. R. Bard, Inc. |
|-------------------|---------------------------------------------|
| Contact Person: | Keti Sino |
| | Regulatory Affairs Specialist |
| Address: | 100 Crossings Boulevard |
| | Warwick, RI 02886 |
| Telephone: | (401) 825-8575 |
| Fax: | (401) 825-8765 |
| Email: | Keti.Sino@crbard.com |
**Device Name:**
| Trade Names: | Ventralight™ ST Mesh with Echo PST™ Positioning System<br>Composix™ L/P Mesh with Echo PST™ Positioning System |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Names: | - Surgical Mesh<br>- Endoscope & Accessories<br>- Mesh Deployment Balloon |
| Classification Names: | - Mesh, Surgical, Polymeric<br>- Mesh Deployment Balloon<br>- Laparoscope, General & Plastic Surgery |
| Classification Codes:<br>Subsequent Codes: | - Class II, § 878.3300, Product Code FTL<br>- Class II, § 878.3300, Product Code OQL<br>- Class II, § 876.1500, Product Code GCJ |
#### Predicate Device Names:
- Ventralight™ ST Mesh with Echo PSTM Positioning System, K110820 (Davol Inc.), . FDA cleared on 04/01/2011
- Composix™ L/P Mesh with Echo PS™ Positioning System, K102766 (Davol Inc.), . FDA cleared on 12/16/2010
#### Device Descriptions:
The proposed Ventralight™ ST Mesh with Echo PSTM Positioning System and Composix™ L/P Mesh with Echo PSTM Positioning System devices both comprise of a hernia repair mesh with a pre-attached mesh positioning system (mesh deployment balloon). A brief description of the two (2) mesh components is as follows:
PREMARKET NOTIFICATION FOR THE ECHO PSTM POSITIONING SYSTEM DEVICES
### SECTION 8
{1}------------------------------------------------
K122436 Page 2/3
- The Ventralight™ ST Mesh is a low profile, sterile, single use device indicated for 0 use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. This device is co-knitted using polypropylene (PP) and polyglycolic acid (PGA) fibers to result in a two-sided mesh with a PP surface and a PGA surface. The mesh is coated on the PGA surface with a bioresorbable, chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC) and polyethylene glycol (PEG) based hydrogel that is resorbed from the site in less than 30 days.
- The Composix™ L/P Mesh is a low profile, nonabsorbable, sterile prosthesis O indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects. It is constructed of one layer of large pore polypropylene mesh and one layer of expanded polytetrafluoroethylene (ePTFE) stitched together with PTFE monofilament.
Each of the meshes described above will be sold pre-attached to the Echo PSTM Positioning System. The positioning system is composed of a nylon balloon that is preattached to the mesh via use of polycarbonate connectors, and is designed to help facilitate laparoscopic deployment, including unrolling, positioning, and placement of the prostheses.
Both of the proposed devices will also include an Introducer Tool and inflation assembly accessory designed to aid the user in laparoscopic introduction and inflation of the mesh/positioning system assembly. Both of these accessories in the proposed device are identical to those cleared via K110820 and K102766.
#### Intended Uses:
- The Ventralight™ ST Mesh is indicated for use in the reconstruction of soft tissue o deficiencies, such as for the repair of hernias. The Echo PS™ Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
- Composix™ L/P Mesh is indicated for use in the reconstruction of soft tissue o deficiencies, such as for the repair of hernias and chest wall defects. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
## Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Uses:
The Ventralight™ ST Mesh with Echo PSTM Positioning System has the same indication: reconstruction of soft tissues deficiencies such as the repair of hernias as well as to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair. Further, the proposed product has the same physical attributes and performance characteristics as the predicate device.
The Composix™ L/P Mesh with Echo PS™ Positioning System has the same indication: reconstruction of soft tissues deficiencies such as the repair of hernias as well as to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair. Fruther, the proposed product has the same physical attributes and performance characteristics as the predicate device.
PREMARKET NOTIFICATION FOR THE ECHO PSTM POSITIONING SYSTEM DEVICES
**SECTION 8**
{2}------------------------------------------------
K122436 page 3/3
The Introducer Tool and inflation assembly accessories to be included with either mesh configuration also maintain the same intended uses as their cleared predicates.
The change described in this submission is a modification solely to the design of the Echo PSTM Positioning System which is sold pre-attached to both meshes and is identical in both products, with no change to either of the permanently implantable neemesh components or related accessories included with the device.
The inflation tube component of the positioning system will be modified to include an anchor which will improve the connection between the tube and associated inflation assembly. This same inflation tube component of the positioning system is also being modified to include white markings which more clearly identify the cutting point of the tube to the user.
#### Performance Data:
Biocompatibility testing in accordance with ISO 10993-1 standards was conducted to assess the design modification on the proposed Ventralight™ ST Mesh with Echo PSTM Positioning System and Composix™ L/P Mesh with Echo PSTM Positioning System and the results indicate that the devices are biocompatible per these standards.
Bench testing results and in vivo simulated use experiments demonstrate that the proposed device design meets product specifications and intended uses.
All test results provided in this submission support the safety and effectiveness of the device for its intended use and demonstrate that the proposed device is substantially equivalent to its predicate devices.
PREMARKET NOTIFICATION FOR THE ECHO PSTM POSITIONING SYSTEM DEVICES
SECTION 8
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter Dated: November 2, 2012
C.R. Bard, Incorporated % Ms. Keti Sino Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K122436
Trade Name: Ventralight" ST Mesh with Echo PS" Positioning System, Composix" L/P Mesh with Echo PS™ Positioning System Regulation Number: 21 CFR 878.3300. 21 CFR 878.1500 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: FTL, OQL, GCJ Dated: October 05, 2012 Received: October 09, 2012
Dear Ms. Sino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Keti Sino
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATION FOR USE STATEMENT
510(k) Number (if known):
Not known
Device Name:
## Ventralight™ ST Mesh with Echo PSTM Positioning System
Ventralight™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dániel Kinefu MM
(Division Sign-Off) (Division of Surgical, Orthopedie, and Restorative Devices
510(k) Number K122436
23
{6}------------------------------------------------
### INDICATION FOR USE STATEMENT
510(k) Number (if known):
Not known
Device Name:
## Composix™ L/P Mesh with Echo PSTM Positioning System
Composix™ L/P Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias and chest wall defects. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
Prescription Use X (Part 21 CFR 801 Subpart D)
#### AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Knaebel MM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K122436
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.