K113552 · Insightra Medical · FTL · Aug 23, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113552
Device Name
FREEDOM INGUINAL HERNIA IMPLANT
Applicant
Insightra Medical
Product Code
FTL · General, Plastic Surgery
Decision Date
Aug 23, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K113552 · Aug 23, 2012
FREEDOM INGUINAL HERNIA IMPLANT
Insightra Medical
Retrospective clinical experience/chart review of 61 patients
The sponsor reviewed retrospective clinical experience from 61 patients treated with custom-fabricated implants to demonstrate the safety and effectiveness of the dilation delivery technique and implant design.
Retrospective clinical experience (Group 1); Retrospective case series; Follow-up/Duration: Minimum of 3 years
Patients receiving custom implants fabricated from surgical polypropylene mesh; Sample Size: 61; Number of Sites: 1
Not applicable for this study
Safety and effectiveness (recurrence, complications, chronic discomfort)
Indications for Use
The Freedom Inguinal Hernia Implant is intended to be implanted to reinforce soft tissues where weakness for open repair of inguinal hernias exists.
Device Story
Freedom Inguinal Hernia Implant is a surgical mesh device for open inguinal hernia repair. Comprised of two polypropylene meshes and two polypropylene rings; multi-looped core fills hernia opening; underlay patch reinforces defect and prevents expulsion/rotation. Provided sterile with a delivery device used to compact the implant for insertion into the defect. Used in clinical settings by surgeons; no sutures required for fixation. Tissue contracts around implant to grip it in place. Provides mechanical reinforcement of soft tissue weakness. Benefits include effective hernia repair with reduced procedural complexity and no need for permanent suture fixation.
Clinical Evidence
Clinical evidence includes two prospective studies (n=61 and n=30) and an independent investigator study (n=24). Total cohort n=54 for specific clinical data. Primary endpoints: procedural success (successful deployment) and hernia recurrence at 1, 6, 12, 24, 36 months. Results showed no reports of tissue tearing, spermatic cord injury, dislodgements, or recurrence. Mean follow-up 25 months. Bench testing included in vitro, in vivo (survival study), and cadaver studies validating implantation technique and tissue incorporation. Biocompatibility testing confirmed compliance.
Technological Characteristics
Material: Polypropylene. Configuration: Two-part mesh (underlay patch and multi-looped core). Sterilization: Ethylene Oxide (EtO). Dimensions: Two sizes (25mm/60mm and 40mm/70mm). Delivery: Provided with a dedicated delivery device for compaction and insertion. No software or electronic components.
Indications for Use
Indicated for open repair of inguinal hernias in adult patients (18-85 years) with direct, indirect, or mixed inguinal hernias (NYHUS I, II, IIIa, IIIb). Contraindicated in patients with local/systemic infection, non-inguinal hernias, unstable angina, NYHA class IV, pregnancy, active drug use, immunosuppression, end-stage renal disease, abdominal ascites, or BMI > 35.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
# 11 3557
of 7
### 5 510(k) Summary
a. Submitter
AUG 2 3 2012
Applicant Name:
Insightra Medical 9200 Irvine Center Drive, Suite 200 Irvine, CA 92618
Contact Person: Wayne Noda Date Summary Prepare: Revised August 22, 2012
### b. Device Information
Trade Name: Freedom Inquinal Hernia Implant Common Name: Surgical mesh Classification Name: Surgical polymeric mesh (21 CFR 878.3300) Product Code: FTL
### c. Predicate Devices
Bard Perfix Plug (K922916) Ethicon Prolene Hernia System (K984220)
### d. Device Description
The Freedom Inguinal Hernia Implant is used to plug or patch a hernia opening. The implant is wholly manufactured from polypropylene. It is comprised of two polypropylene meshes and two small polypropylene rings. A multi looped shaped core fills the hernia opening while an underlay patch reinforces the defect and holds the implant in place. This underlay patch helps to prevent expulsion and rotation. Once the tissue defect is prepared, the implant is deployed into the defect area using the delivery device provided in the Freedom Inguinal Hernia Repair kit. The delivery device serves to compact the implant and provide access to the defect opening to aid implantation. Once deployed, the tissue contracts around the implant, gripping it in place. There are two different sizes of implants and corresponding delivery devices to accommodate different size defects. The implant and delivery device are provided sterile for single patient use.
### e. Indications for Use:
The Freedom Inguinal Hernia Implant is intended to reinforce soft tissues where weakness for open repair of inquinal hernias exists.
#### Technological Characteristics and Comparison to Predicate Devices f.
The Freedom Inguinal Hernia Implant is of similar design and materials commonly used in other marketed hernia repair meshes.
{1}------------------------------------------------
U11 3532
2 of 7
Table 1 provides a comparison of the Freedom Inguinal Hernia Implant with the predicate hernia repair implants. This table illustrates the substantial equivalence of the subject device with the predicates.
| | Freedom Inguinal Hernia<br>Implant | Ethicon Prolene Hernia<br>System | Bard Perfix Plug |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| 510(k) Clearance | This submission | K984220 | K922916 |
| Indications for Use | The Freedom Inguinal<br>Hernia Implant is intended<br>to be implanted to reinforce<br>soft tissues where<br>weakness for open repair of<br>inguinal hernias exists. | The PROLENE Hernia<br>System is indicated for<br>the repair of abdominal<br>wall hernia defects | This product is<br>indicated for the<br>repair of groin<br>hernia defects |
| Shape | Image: Freedom Inguinal Hernia Implant Shape | Image: Ethicon Prolene Hernia System Shape | Image: Bard Perfix Plug Shape |
| Texture | Mesh | Mesh | Mesh |
| Material | Polypropylene | Polypropylene | Polypropylene |
| Configuration | Underlay mesh patch,<br>looped shaped mesh core | Underlay mesh patch,<br>cylindrical mesh core,<br>Onlay mesh patch | Conical multi layer<br>mesh core, Onlay<br>mesh patch |
| How is the implant<br>secured? | Inserted into the hernia; no<br>sutures necessary | Inserted into the hernia;<br>no sutures necessary | Inserted into the<br>hernia; no sutures<br>necessary |
| Number of models | 2 | 6 | 4 |
| Size range (smallest to<br>largest) | 25 mm core & 60 mm dia.<br>underlay patch<br>40 mm core & 70 mm dia.<br>Underlay patch | Onlay patch - 44mm x<br>100mm to 55mm x<br>125mm<br>Core Connector (for all<br>sizes) -<br>19mm dia x 13mm<br>height<br>Underlay patch-75mm<br>to 100mm diameter | 25 x 34 mm to<br>41 x 50 mm;<br>preshaped onlay<br>patch same for all<br>size plugs |
| Sterilization | EtO | EtO | EtO |
Table 1 – Comparison Table for Freedom Inguinal Hernia Implant
{2}------------------------------------------------
### g. Pre-Clinical Test Results:
In vitro, in vivo and biocompatibility tests were performed on the Freedom Inguinal Hernia Implant. The in vitro tests showed that the device meets pre-determined acceptance criteria or specifications that were based on the clinical demands placed on the device. The in vivo testing validated the implantation procedure and a survival study was performed that demonstrated the tissue incorporation of the implant. Results of the in vivo study were published in the journal of Artificial Organ (3). The biocompatibility testing showed the Freedom Inquinal Hernia Implant complies with the requirements for this device classification.
Cadaver studies were performed during the development of both the Freedom Inguinal Hernia Implant as well as the delivery device. These studies showed that the implant could be successfully deployed in the proper location and further validated the implantation technique. These results were considered in the final design configurations.
### h. Results of Clinical Experience:
Two groups of data from relevant clinical experience were collected and reviewed. The first group of patients (n=61) received custom implants that were fabricated from surgical polypropylene mesh by the implanting surgeon. This was a single center, single surgeon clinical experience. In these patients, the surgeon used both stainless steel and rigid plastic tubes to provide the slight dilation of the defect and access required to deliver the implant. Patients were followed for a minimum of 3 years after the date of implantation. One patient was lost to follow up due to unrelated heart failure death after 14 months. The results demonstrated that defect dilation with tubes and treatment with flower shaped implants was safe and effective for the treatment of inguinal hernias. After 3 years, there have been no reports of unresolved post operative complications, recurrences, or chronic discomfort in these patients. The diameters of the dilation tubes were the diameters of the Insightra Medical Delivery Device, that is subject of this premarket notification.
The second group of patients reviewed (n=30) received the Insightra manufactured implant that is the subject of this submission. The implant was delivered using standard open surgical techniques using a previous version of the Freedom Inquinal Hernia Delivery System. This delivery system was identical dimensionally to the delivery system that is included in this premarket notification. Follow-up ranged from perioperative to one year post-implantation. Of these 30 patients, 27 were followed for a year or more with a mean followup of 25 months (ranging 11 to 36 months). There were no reports of unresolved perioperative complications, tissue injury or spermatic cord complications associated with this device.
In exceptional cases, an additional flat polypropylene mesh has been deployed to cover the inquinal floor, over the Freedom Inquinal implant that obliterates the hernia opening. This further reinforces the inguinal canal at the hernial protrusion when the groin structures appear obviously weak. The onlay flat mesh is not sutured or fixed to the inguinal structures, instead only connected to the central core of the Freedom Inguinal implant with a single polypropylene suture. Previous experiences of using this procedural variant produced, with the exception of some very seldom reported discomfort during the early postoperative stage, no unusual symptoms reported by patients who received the onlay patch in addition to the Freedom Inguinal implant.
{3}------------------------------------------------
For all patients in both groups reviewed with at least one year followup (n= 78) who received this dilation delivery technique, there were no reports of tissue tearing injury or spermatic cord injury. There have been no reports of dislodgements, unresolved post operative complications, chronic discomfort or recurrence of hernia. For the twenty seven patients in the second group who were followed for a year or more and received the Insightra Medical manufactured implant, there were no reports of spermatic cord injury or compression complications due to device rotation or inappropriate sizing.
The second study group by Amato (n=30) is summarized along with a third independent investigator sponsored study by Petrella, in the device labeling. Descriptive tables are included in the device "Technique Guide" as follows;
These clinical studies followed cohorts of patients (total n=54) that met the following inclusion/exclusion criteria:
| Inclusion Criteria: | Exclusion Criteria: |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • Scheduled to undergo routine inguinal<br>hernia repair • Competent to give consent • Clinically relevant inguinal hernia<br>(classification: NYHUS I, II, IIIa, IIIb) • Male or female • Over 18 years old to 85 years old • Life expectancy of at least 12-months • Diagnosed with direct, indirect or mixed<br>inguinal hernia, unilateral or bilateral • Primary or recurrent hernia | • Signs of obvious local or systemic<br>infection • Hernia was not in the inguinal area • Presenting with unstable angina or<br>NYHA class of IV • Pregnant • Active drug user • Immunosuppression, prednisone>15<br>mg/day, active chemotherapy • End stage renal disease • Abdominal ascites • Skin infection in area of surgical field • BMI > 35 |
Both studies used the following objectives:
| Primary Efficacy Objectives | Secondary Complications<br>Objectives |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • Procedural success is defined as the ability<br>to successfully deploy the implant into the<br>inguinal hernia defect<br>• Freedom from hernia recurrence at the time<br>points of 1, 6, 12, 24, 36 months | • Collection of data on any<br>perioperative/postoperative<br>complications (1, 6, 12, 24, 36<br>months)<br>• Bleeding, swelling or hematoma<br>from hernia defect dilation or<br>improper delivery or implantation<br>technique<br>• Seroma, infection/abscess, testicular<br>or spermatic cord injury in males,<br>wound complications, symptomatic<br>pain or chronic pain syndrome. |
{4}------------------------------------------------
.
:
| | Prospective study of a single operator<br>case series of a modified plug<br>technique and new 3D hernia implant |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Amato Study | |
| Number of centers/investigators | 1 center/ 1 investigators |
| Study Enrollment Period | June 2009 to August 2011 |
| Number of patients | 30 |
| Number of implants* | 25mm: n=12 40mm: n=22<br>*Four cases were double unilateral<br>hernias |
| Patient demographics | Male = 30/Female = 0 |
| | BMI (Kg/m2) = 28.4 (21.7 – 34.9) |
| | Age = 50.23 mean, range 23-82 |
| | Hernia types: Indirect=17, Direct=13 |
| | Hernia size: Mean 32 mm (range 20 –<br>37.7) |
| | Hernia type: |
| | NYHUS Type 1 = 5 |
| | NYHUS Type 2 = 13 |
| | NYHUS Type 3a = 12 |
| | NYHUS Type 3b = 4 |
| | Anesthesia: Local = 63%<br>Spinal = 17%<br>General = 20% |
| Operative procedure specifics | Procedure duration: Mean 29 minutes<br>(range 22 – 38) |
| | Defect measurement: Intra operative<br>measurement of the dissected flaccid<br>defect using surgical scale for implant<br>sizing |
| | |
| Patient Follow up Outcomes/adverse<br>events | Follow up: Mean = 25.23 months (range<br>11.0-36.5) |
| | Outcomes/Adverse: no long term<br>complications or recurrences at this time |
| | In two exceptional cases, an additional flat<br>polypropylene mesh was deployed to cover the<br>inguinal floor, over the Freedom Inguinal<br>implant that obliterated the hernia opening.<br>This further reinforced the inguinal canal at the<br>hernial protrusion when the groin structures<br>appeared obviously weak. |
| Additional Outcomes – Two patients in this<br>cohort received onlay mesh patches | Patient #1 – Male, 82 yrs, Double unilateral<br>hernia – Rt Indirect, Rt Direct, 25mm and<br>40mm implants used, Nyhus 2 & 3a. BMI: 29.<br>Comorbidities: Myocardial ischemic disease,<br>hypertension. Non recurrent hernias. Follow<br>up = 18.2 mos. |
| | Patient #2 – Male, 75 yrs, Lf Mixed hernia,<br>40mm implant, Nyhus 3b. BMI: 26.<br>Comorbidities: BPCO, myocardial<br>insufficiency, hypertension. Non recurrent<br>hernia. Follow up = 18.2 |
| | Results of these two procedures to date have<br>had no complications or unusual symptoms. |
| | Patient #1 reported occasional discomfort |
| | during the early postoperative stage but self<br>resolved after 5 months post op. |
| Ultrasound findings in cohort of patients | Full and stable obliteration of twelve<br>patients examined at 3, 6 and 12 months. |
. ·
:
:
1
.
:
:
.
{5}------------------------------------------------
| Petrella/Venditti Study | Prospective single arm study of the<br>ProFlor Inguinal hernia Device |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number of centers/investigators | 1 center/ 2 investigators |
| Study Enrollment Period | Dec 2011 to present |
| Number of patients | 24 enrolled; Target enrollment = 80 |
| Number of implants | 25mm: n=5 40mm: n=19 |
| Patient demographics | Male = 22/Female = 2<br>$BMI (Kg/m^2) = 25.2 (19,4 - 35,2)$<br>Age = 52.3 mean, range (31 - 73)<br>Hernia types: Indirect=20, Direct=4<br>Hernia size: 15mm40mm<br>NYHUS Type 2 = 20<br>NYHUS Type 3a = 4 |
| Operative procedure specifics | Local anesthesia<br>Procedure duration: 24 min mean, (20-30)<br>Defect measurement: Intra operative<br>measurement of the dissected flaccid<br>defect using surgical scale for implant<br>sizing |
| Patient Follow up Outcomes/adverse<br>events | Follow up: Mean = 4.5 months (range 3.0-<br>6.0)<br>Outcomes/Adverse: None observed |
| Ultrasound findings in cohort of patients | Ultrasound scans were not a requirement<br>of the study and were not performed for<br>this cohort |
#### 510(k) Summary Conclusion: i.
In summary, the Insightra Freedom Inquinal Hernia Implant path through the 510(k) requlatory analysis is as follows:
- The Insightra Freedom Inquinal Hernia Implant and the Bard Perfix Plug and Ethicon Prolene Hernia System predicate devices, are all intended for use in the repair of inguinal hernia defects.
- The Insightra Freedom Inguinal Hernia Implant has similar key technological . characteristics as the Bard Perfix Plug and Ethicon Prolene Hernia System devices. All three are manufactured from polypropylene mesh, all use mesh patch reinforcements, and all are inserted into the hernia defect without the need for suture;
- Although the Insightra Freedom Inguinal Hernia Implant and predicate devices have a . minor technological difference in terms of the method of delivery, this difference does not raise any new types of safety or effectiveness questions. In addition, valid scientific methods exist for assessing the effects of the differences in delivery methods. Preclinical and clinical data demonstrate that these devices are substantially equivalent.
{6}------------------------------------------------
V1113552
7 of 7
Therefore, the Insightra Freedom Inguinal Hernia Implant meets the criteria for substantial equivalence to the Bard Perfix Plug and Ethicon Prolene Hernia System predicate devices.
:
.
.
·
.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wing segments, representing health, services, and people. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Insightra Medical % Mr. Wayne Noda Chief Technology Officer 9200 Irvine Center Drive, Suite 200 Irvine, California 92618
AUG 2 3 2012
## Re: K113552
Trade/Device Name: Freedom Inguinal Hernia Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: August 20, 2012 Received: August 21, 2012
Dear Mr. Noda:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Mr. Wayne Noda
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Sincerely, yours,
Erin S. Keith
- Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use
510(k) Number (K113552):
Device Name: Freedom Inguinal Hernia Implant
Indications for Use:
The Freedom Inguinal Hernia Implant is intended to be implanted to reinforce soft tissues where weakness for open repair of inguinal hernias exists.
Prescription Use _ × AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Krane for MM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113552
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.