K073164 · Covidien UK Manufacturing, Ltd. · FTL · Mar 5, 2008 · General, Plastic Surgery
Device Facts
Record ID
K073164
Device Name
IVS TUNNELLER DEVICES
Applicant
Covidien UK Manufacturing, Ltd.
Product Code
FTL · General, Plastic Surgery
Decision Date
Mar 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IVS Tunneller™ Intra-Vaginal Sling placement device is intended to be used in females to position a polypropylene mesh for the treatment of genuine stress urinary incontinence (SUI), mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency, and vaginal vault prolapse. The Obturator IVS Tunneller™ Intra-Vaginal Sling placement device is intended to be used in females to position a polypropylene mesh for the treatment of genuine stress urinary incontinence (SUI) and for mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
Device Story
The IVS Tunneller™ and Obturator IVS Tunneller™ are surgical instruments used to position polypropylene mesh tape for treating female urinary incontinence and vaginal vault prolapse. The device consists of a curved stainless steel introducer with a delta wing polypropylene handle, a polypropylene stylette with an eyelet for threading the mesh, and a non-absorbable, porous polypropylene mesh tape. During surgery, the introducer creates a pathway; the stylette's atraumatic conical tip performs blunt tissue dissection and muscle deflection. The mesh is threaded through the stylette eyelet and positioned to support the urethra or vaginal vault. The device is used by surgeons in a clinical/OR setting. The mesh remains in the patient to provide structural support, potentially alleviating incontinence symptoms or prolapse.
Indicated for females with genuine stress urinary incontinence (SUI), mixed incontinence due to urethral hypermobility or intrinsic sphincter deficiency, and vaginal vault prolapse (IVS Tunneller version only).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
K073164
(pg 1 of 2)
### 510(k) Summary of Safety and Effectiveness
| SUBMITTER: | Covidien UK Manufacturing Ltd.<br>20 Garryduff Road, Ballymoney<br>County Antrim, Ireland BT53 7AP<br>Phone: 44 28 2766 3234 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Sharon Alexander<br>Senior Associate, Regulatory Affairs<br>Covidien<br>60 Middletown Avenue<br>North Haven, CT 06473 USA<br>Phone: (203) 492-6060 |
| DATE PREPARED: | March 4, 2008 |
| | MAR - 5 2008 |
| TRADE/PROPRIETARY NAME: | IVS Tunneller™ Device |
| COMMON/USUAL NAME: | Intra-Vaginal Sling Placement Device |
| CLASSIFICATION NAME: | Mesh, Surgical, Polymeric |
| PREDICATE DEVICE(S): | IVS Tunneller™ Intra-Vaginal Sling (K010035)<br>Parietene® Polypropylene Mesh (K991400)<br>Surgipro™ Polypropylene Surgical Mesh (K915526) |
| DEVICE DESCRIPTION: | The IVS Tunneller™ Intra-Vaginal Sling placement device is<br>comprised of three components: a stainless steel introducer or outer<br>sheath, a polypropylene stylette, and a polypropylene mesh tape. The<br>introducer is a curved stainless steel tube with a delta wing<br>polypropylene handle, which is used to create a pathway for the<br>correct placement of the polypropylene mesh tape. The polypropylene<br>stylette has an eyelet for threading the polypropylene mesh tape and<br>an atraumatic conical tip which allows for blunt tissue dissection and<br>muscle deflection during the passage of the introducer. The<br>polypropylene mesh tape is a non-absorbable, inert, porous, surgical<br>mesh tape, created from either monofilament or multifilament yarns,<br>and is supplied sterile. |
| INTENDED USE: | The IVS Tunneller™ Intra-Vaginal Sling placement device is intended<br>to be used in females to position a polypropylene mesh for the<br>treatment of genuine stress urinary incontinence (SUI), mixed<br>incontinence resulting from urethral hypermobility or intrinsic sphincter<br>deficiency, and vaginal vault prolapse. |
| TECHNOLOGICAL<br>CHARACTERISTICS: | The fundamental scientific technology of the IVS Tunneller™<br>Intra-Vaginal Sling placement device is substantially equivalent to the<br>predicate IVS Tunneller™ device. |
| MATERIALS: | All components of the IVS Tunneller™ Intra-Vaginal Sling placement<br>device have been evaluated in accordance with ISO Standard 10993-<br>1. |
| PERFORMANCE DATA: | The performance of the subject IVS Tunneller™ device is<br>substantially equivalent to the currently marketed predicate IVS<br>Tunneller™ device. |
ының қалымыздың жасандары мен жасандығы ««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««««»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»»
K073164 Response
March 4, 2008
のおいたいですが、2006年には、2005年には、2005年には、2005年には、2005年には、100%、2005年に、100%、2005年に、1700年、2002
{1}------------------------------------------------
K073164
(pg 1 of 2)
## 510(k) Summary of Safety and Effectiveness
| SUBMITTER: | Covidien UK Manufacturing Ltd.<br>20 Garryduff Road, Ballymoney<br>County Antrim, Ireland BT53 7AP<br>Phone: 44 28 2766 3234 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Sharon Alexander<br>Senior Associate, Regulatory Affairs<br>Covidien<br>60 Middletown Avenue<br>North Haven, CT 06473 USA<br>Phone: (203) 492-6060 |
| DATE PREPARED: | March 4, 2008 |
| TRADE/PROPRIETARY NAME: | Obturator IVS Tunneller™ Device |
| COMMON/USUAL NAME: | Intra-Vaginal Sling Placement Device |
| CLASSIFICATION NAME: | Mesh, Surgical, Polymeric |
| PREDICATE DEVICE(S): | IVS Tunneller™ Intra-Vaginal Sling (K010035)<br>Parietene® Polypropylene Mesh (K991400)<br>Surgipro™ Polypropylene Surgical Mesh (K915526) |
| DEVICE DESCRIPTION: | The Obturator IVS Tunneller™ Intra-Vaginal Sling placement device<br>is comprised of three components: a stainless steel introducer or<br>outer sheath, a polypropylene stylette, and a polypropylene mesh<br>tape. The introducer is a curved stainless steel tube with a delta wing<br>polypropylene handle, which is used to create a pathway for the<br>correct placement of the polypropylene mesh tape. The polypropylene<br>stylette has an eyelet for threading the polypropylene mesh tape and<br>an atraumatic conical tip which allows for blunt tissue dissection and<br>muscle deflection during the passage of the introducer. The<br>polypropylene mesh tape is a non-absorbable, inert, porous, surgical<br>mesh tape, created from either monofilament or multifilament yarns,<br>and is supplied sterile. |
| INTENDED USE: | The Obturator IVS Tunneller™ Intra-Vaginal Sling placement device<br>is intended to be used in females to position a polypropylene mesh for<br>the treatment of genuine stress urinary incontinence (SUI) and for<br>mixed incontinence resulting from urethral hypermobility or intrinsic<br>sphincter deficiency. |
| TECHNOLOGICAL<br>CHARACTERISTICS: | The fundamental scientific technology of the Obturator IVS<br>Tunneller™ Intra-Vaginal Sling placement device is substantially<br>equivalent to the predicate IVS Tunneller™ device. |
| MATERIALS: | All components of the Obturator IVS Tunneller™ Intra-Vaginal Sling<br>placement device have been evaluated in accordance with ISO<br>Standard 10993-1. |
| PERFORMANCE DATA: | The performance of the subject Obturator IVS Tunneller™ device is<br>substantially equivalent to the currently marketed predicate IVS<br>Tunneller™ device. |
:
.
March 4, 2008
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 5 2008
Covidien UK Manufacturing Ltd. % Covidien Ms. Sharon Alexander Senior Associate, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473
Re: K073164
Trade/Device Name: Obturator IVS Tunneller™ Intra-Vaginal Sling Placement Device Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL Dated: February 21, 2008 Received: February 22, 2008
Dear Ms. Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Sharon Alexander
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Presmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark M. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 2073164
#### Indications For Use
510(k) Number (if known): K073164
Device Name: IVS Tunneller™ Intra-Vaginal Sling Placement Device
Indications For Use:
The IVS Tunneller™ Intra-Vaginal Sling placement device is intended to be used in females to position a polypropylene mesh for the treatment of genuine stress urinary incontinence (SUI), mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency, and vaginal vault prolapse.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Gelman for man
Division of General. Restorative. and Neurological Devices
510(k) Number
{5}------------------------------------------------
K073164
(pg 2 of 2)
### Indications For Use
510(k) Number (if known): K073164
Device Name: Obturator IVS Tunneller™ Intra-Vaginal Sling Placement Device
Indications For Use:
The Obturator IVS Tunneller™ Intra-Vaginal Sling placement device is intended to be used in females to position a polypropylene mesh for the treatment of genuine stress urinary incontinence (SUI) and for mixed incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nehardl for mxm
(Division Sign Off)
Division of General, Restorative, and Neurological Devices
510(k) Number K073164
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.