GORE SEAMGUARD BIOABSORBABLE STAPLE LINE REINFORCEMENT MATERIAL
K043056 · W. L. Gore & Associates, Inc. · FTL · Dec 22, 2004 · General, Plastic Surgery
Device Facts
Record ID
K043056
Device Name
GORE SEAMGUARD BIOABSORBABLE STAPLE LINE REINFORCEMENT MATERIAL
Applicant
W. L. Gore & Associates, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Dec 22, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material can be used for reinforcement of staple lines during hysterectomy, lung resection, liver resection, bladder reconstruction, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, small bowel, and spleen procedures. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is also intended to be used for reinforcement of suture-lines and staple-lines (i.e., occlusion of the left atrial appendage during open chest procedures) during cardiac surgery.
Device Story
Bioabsorbable membrane sleeve with polyester braided suture pullcord; composed of microporous synthetic glycolide and trimethylene carbonate copolymer. Used with surgical stapling devices to reinforce staple lines and suture lines during various soft tissue surgical procedures. Applied to stapler cartridge and anvil jaws using Tyvek inserts. Operates as a physical reinforcement material to support tissue integrity during healing. Used by surgeons in clinical/OR settings. Benefits include reduced risk of leaks or bleeding at surgical sites.
Clinical Evidence
In vivo studies were conducted to support the expanded indications for use across various surgical procedures, including hysterectomy, liver resection, bladder reconstruction, bronchial, cardiac, esophagus, pancreas, and spleen procedures. No specific quantitative clinical metrics (e.g., sensitivity, specificity) are provided in the summary.
Technological Characteristics
Microporous membrane composed of synthetic bioabsorbable glycolide and trimethylene carbonate copolymer. Includes polyester braided suture pullcord. Supplied in sterile foil film laminate pouches with Tyvek inserts for application to surgical stapler jaws. Mechanical reinforcement device; no energy source or software.
Indications for Use
Indicated for patients undergoing surgical procedures requiring soft tissue transection or resection where staple line reinforcement is necessary, including hysterectomy, lung, liver, bladder, bronchial, bariatric, colon, colorectal, esophageal, gastric, mesenteric, pancreatic, splenic, and cardiac surgeries (e.g., left atrial appendage occlusion).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material (K030782 and K032865)
{0}------------------------------------------------
K043056
DEC 2 2 2004
510(k) Premarket Notification
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material
# 510(k) SUMMARY OF SUBSTANTIAL EQUIVALENCE
| Proprietary Name: | GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Staple Line Reinforcement Material |
| Classification Name: | Surgical Mesh |
| Device Classification: | Class II |
| Product Classification and Code: | 878.3300, FTL |
| Classification Panel: | General and Plastic Surgery Devices |
| Establishment Registration Number: | 2025240 |
| Contact Person: | Brandon Hansen<br>Regulatory Affairs<br>Medical Products Division<br>WL Gore & Associates, Inc.<br>3450 West Kiltie Lane<br>Flagstaff, AZ 86002-0500<br>Telephone: (928) 864-3784<br>Facsimile: (928) 864-4144<br>E-mail: bhansen@wlgore.com |
# Performance Standards
Performance standards do not currently exist for these devices. None established under Section 514.
{1}------------------------------------------------
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material
### Device Description
The GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Material consists of a bioabsorbable membrane formed into a sleeve with use of a polyester braided suture pullcord. The bioabsorbable material is comprised of a microporous structure of synthetic bioabsorbable glycolide and trimethylene carbonate copolymer. The GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Material is to be used with surgical stapling devices.
The GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Material is supplied in sterile, foil film laminate pouches. Pouches contain the necessary material for the cartridge jaw and the anvil jaw of the surgical stapler. Tyvek® inserts facilitate placement of the reinforcement material onto the stapler jaws.
#### Indication for Use
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material can be used for reinforcement of staple lines during hysterectomy, lung resection, liver resection, bladder reconstruction, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, small bowel, and spleen procedures.
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is also intended to be used for reinforcement of suture-lines and staple-lines (i.e., occlusion of the left atrial appendage during open chest procedures) during cardiac surgery.
## Substantially Equivalent Devices
In WL Gore & Associates' opinion, the GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Material is believed to be substantially equivalent to the following predicate devices currently in interstate commerce with respect to comparable features, materials of construction, and labeling.
- SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material (WL . Gore & Associates, Inc., Flagstaff, AZ) - K030782 and K032865
- GORE-TEX® Vascular Graft (WL Gore & Associates, Flagstaff, AZ) -. K991683
#### Summary of Studies
In vivo studies support the modification to the "Indication for Use" of the GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Material for reinforcement of staple lines during hysterectomy, liver resection, bladder reconstruction, bronchial, cardiac, esophagus, pancreas, and spleen surgical procedures.
{2}------------------------------------------------
Koy. in.
510(k) Premarket Notification
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material
# Conclusion (Statement of Equivalence)
Through data and information presented, numerous similarities support a I hrough and of substantial equivalence, and therefore market clearance of the WL determination of succeant.com Bioabsorbable Staple Line Reinforcement Material through this 510(k) Premarket Notification.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white emblem. The emblem is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. The bottom half of the circle contains a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 22 2004
Mr. Brandon Hansen Regulatory Affairs Medical Products Division W.L. Gore & Associates, Inc. 3450 West Kiltie Lane Flagstaff, Arizona 86002
Re: K043056
K043020
Trade/Device Name: Gore Seamguard Bioresorbable Staple Line Reinforcement Material Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: FTL Dated: November 4, 2004 Received: November 5, 2004
Dear Mr. Hansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device w & nowed above and have determined the device is substantially equivalent (for the indications forcherous as over and his sure) to legally marketed predicate devices marketed in interstate for tase stated in the encreases of the enactment date of the Medical Device Amendments, or to conniner of they 2011-11-11 accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). Tou may, therefore, market the device, subject to the general controls provisions of the Act. The I ou mayy are so provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Extre rate of 7 more of 1 (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Brandon Hansen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Gore Seamgaurd Bioresorbable Staple Line Reinforce ment Device Name:__________________________________________________________________________________________________________________________________________________________________ Material
Indications For Use:
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is indicated for use in surgical procedures in which soft tissue transection or resection with staple line ass in vargent is needed. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material can be used for reinforcement of staple lines during hysterectomy, lung resection, liver resection, bladder reconstruction, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, small bowel, and spleen procedures.
GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is also intended to be used for reinforcement of suture-lines and staple-lines (i.e., occlusion of the left atrial appendage during open chest procedures) during cardiac surgery.
Miriam C. Provost
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number_KO43056
AND/OR
Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.