K172545 · Sofregen Medical, Inc. · OXF · Sep 22, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172545
Device Name
SERI Contour
Applicant
Sofregen Medical, Inc.
Product Code
OXF · General, Plastic Surgery
Decision Date
Sep 22, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SERI® Contour is indicated for use as a transitory scaffold for soft tissue support and repair to reinforce deficiencies where weakness or voids exist that require the addition of material to obtain the desired surgical outcome. This includes reinforcement of soft tissue in plastic and reconstructive surgery and general soft tissue reconstruction.
Device Story
SERI® Contour is a knitted, multifilament, long-term bioresorbable scaffold derived from pure silk fibroin. Used in open or laparoscopic procedures, it provides immediate physical and mechanical stabilization of soft tissue defects. The device is flexible, tear-resistant, and cuttable, designed for delivery through laparoscopic trocars. It acts as a transitory scaffold, supporting tissue ingrowth while maintaining strength during the healing process. Surgeons use the mesh to reinforce soft tissue deficiencies; the increased pore size and fully knitted construction improve pliability and drapability compared to the predicate. The device is sterile and for single-use only.
Clinical Evidence
Bench testing and large animal model study. Bench testing verified burst strength exceeds maximum abdominal pressure. Large animal study evaluated tissue ingrowth and burst strength at 3, 6, and 12 months, demonstrating equivalent performance to the predicate scaffold through 1-year evaluation.
Technological Characteristics
Knitted, multifilament, long-term bioresorbable scaffold made of pure silk fibroin. Features increased pore size and fully knitted construction for enhanced elasticity and drapability. Sterile, single-use, supplied in various sizes. Compatible with open or laparoscopic delivery.
Indications for Use
Indicated for patients requiring soft tissue support and repair in plastic, reconstructive, or general surgery to reinforce tissue deficiencies, weaknesses, or voids.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2017
Sofregen Medical, Inc. Anh Hoang, Ph.D. Chief Scientific Officer 200 Boston Avenue. Suite 1100 Medford, MA 02155
Re: K172545
Trade/Device Name: SERI® Contour Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OXF Dated: August 14, 2017 Received: August 23, 2017
Dear Dr. Hoang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172545
Device Name SERI® Contour
Indications for Use (Describe)
SERI® Contour is indicated for use as a transitory scaffold for soft tissue support and repair to reinforce deficiencies where weakness or voids exist that require the addition of material to obtain the desired surgical outcome. This includes reinforcement of soft tissue in plastic and reconstructive surgery and general soft tissue reconstruction.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary for the SERI® Contour is provided below.
#### SUBMITTER 1.
Sofregen Medical, Inc. 200 Boston Avenue, Suite 1100 Medford, MA 02155
Contact Person: Anh Hoang, Ph.D. Phone: 781-874-2352 Email: anh@sofregenmedical.com Date Prepared: August 16, 2017
#### 2. DEVICE
Name of Device: SERI® Contour Common Name: Surgical Mesh Classification Regulation: 21 CFR 878.3300 Regulatory Class: II Product Code: OXF Panel: General and Plastic Surgery
#### PREDICATE DEVICE 3.
Predicate Device: SERI® Surgical Scaffold (K123128)
Reference Device: None
#### DEVICE DESCRIPTION 4.
SERI® Contour is a knitted, multifilament, bioengineered, long-term bioresorbable scaffold. It is derived from pure silk fibroin. The device is a mechanically strong and biocompatible protein mesh. SERI® Contour is a sterile, single use only product and is supplied in a variety of sizes. ready for use in open or laparoscopic procedures. The scaffold is flexible and well-suited for delivery through a laparoscopic trocar. It is tear resistant, with excellent suture retention, and can be cut in any direction. SERI® Contour provides immediate physical and mechanical stabilization of a tissue defect through its strength and porous (scaffold-like) construction.
#### INDICATION FOR USE 5.
SERI® Contour is indicated for use as a transitory scaffold for soft tissue support and repair to reinforce deficiencies where weakness or voids exist that require the addition of material to obtain the desired surgical outcome. This includes reinforcement of soft tissue in plastic and reconstructive surgery and general soft tissue reconstruction.
{4}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6.
The subject SERI® Contour has undergone a design change to increase the pore size and knit construction of the mesh compared to the predicate SERI® Surgical Scaffold (K123128). All other material and processing parameters are the same for the production and sterilization of Contour and Scaffold, independent of the pore size and knit pattern. The increased pore size of Contour and knitted construction in both material directions increase the mesh's elasticity that results in a more pliable and drapable surgical mesh. The design change updates handleability in response to clinical feedback, but maintain performance characteristics to accommodate soft tissue repair and support.
#### PERFORMANCE DATA 7.
The differences in surgical mesh specifications between Contour and Scaffold do not affect production, biocompatibility, and intended use of the product. Introducing greater pore area into the Contour design reduces overall density and burst strength of the mesh compared to the Scaffold predicate. Burst strength specifications for both devices were established to exceed peak forces in the abdomen. The fully knitted construction of Contour increases the elasticity and drapability of the mesh compared to Scaffold. When evaluated in a large animal model. Contour supported tissue ingrowth and equivalent burst strength to Scaffold at 3, 6, and 12month evaluation.
#### CONCLUSIONS 8.
The design change for SERI Contour introduces larger pores into a fully knitted mesh format, increasing both pliability and drapability requested from clinical feedback of SERI Scaffold. Independent of the mesh pattern, silk processing, mesh production, sterilization, biocompatibility, and intended use are identical to the predicate Scaffold surgical mesh. Burst stress for Contour was verified to perform in excess of the maximum calculated, worst case scenario of abdominal pressure. In vivo performance testing in a large animal model resulted in equivalent burst stress for Contour and Scaffold, confirming soft tissue formation through 1-year evaluation.
In conclusion, based on the same intended use and design change introducing larger pores in a fully knitted mesh, the subject SERI® Contour is substantially equivalent to the predicate SERI® Surgical Scaffold (K123128).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.