ECHELON ENDOPATH Staple Line Reinforcement (ECH60R)
K252906 · Ethicon Endo-Surgery, LLC · OXC · Jan 9, 2026 · General, Plastic Surgery
Device Facts
Record ID
K252906
Device Name
ECHELON ENDOPATH Staple Line Reinforcement (ECH60R)
Applicant
Ethicon Endo-Surgery, LLC
Product Code
OXC · General, Plastic Surgery
Decision Date
Jan 9, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic, PCCP
Indications for Use
ECHELON ENDOPATH™ Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. ECHELON ENDOPATH™ Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel, and colorectal procedures.
Device Story
Absorbable staple line reinforcement (buttress) used as an adjunct to surgical staplers; provides support to soft tissue during healing. Device consists of Polyglactin 910 (Vicryl) knit, Polydioxanone (PDO) film, and Poloxamer blend adhesive. Supplied in a hand-held polycarbonate applicator for attachment to stapler anvil and reload jaws. During stapling, device is divided simultaneously with tissue. Used in surgical environments by surgeons. No changes to device design or materials; submission increases sterilization dose (25-40 kGy) and shelf-life (3 years).
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing (staple pull-through, shear force, release force), packaging verification, and biocompatibility testing. A GLP canine study evaluated device degradation at 40 kGy at 10, 28, and 120 days, confirming the degradation profile remains consistent with the predicate.
Technological Characteristics
Absorbable buttress composed of Polyglactin 910 (Vicryl) knit, Polydioxanone (PDO) film, and Poloxamer blend adhesive. Dimensions: 65.367mm x 10.4mm. Sterilization: E-Beam Irradiation (25-40 kGy). Biocompatibility per ISO 10993-1. Shelf life: 3 years. Absorption process: Hydrolysis (120 days).
Indications for Use
Indicated for patients undergoing surgical procedures requiring soft tissue transection or resection where staple line reinforcement is needed, including lung resection, bariatric, gastric, small bowel, and colorectal procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
ECHELON ENDOPATH Staple Line Reinforcement (K221343)
Reference Devices
ECHELON ENDOPATH Staple Line Reinforcement (K190937)
K030782 — SEAMGUARD BIOABSORBABLE STAPLE LINE REINFORCEMENT MATERIAL · W. L. Gore & Associates, Inc. · Apr 21, 2003
K190937 — Echelon Endopath Staple Line Reinforcement · Ethicon Endo-Surgery, LLC · Oct 14, 2019
K032865 — GORE SEAMGUARD BIOABSORBABLE STAPLE LINE REINFORCEMENT MATERIAL · W. L. Gore & Associates, Inc. · Nov 28, 2003
K043056 — GORE SEAMGUARD BIOABSORBABLE STAPLE LINE REINFORCEMENT MATERIAL · W. L. Gore & Associates, Inc. · Dec 22, 2004
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
January 09, 2026
Ethicon Endo-Surgery LLC
Roberta Franca
Regulatory Affairs, Senior Program Lead
475 Calle C, PR UNITED STATES
Guaynabo, PR 00696
Re: K252906
Trade/Device Name: ECHELON ENDOPATH Staple Line Reinforcement (ECH60R)
Regulation Number: 21 CFR 878.3300
Regulation Name: Surgical Mesh
Regulatory Class: Class II
Product Code: OXC
Dated: September 11, 2025
Received: September 12, 2025
Dear Roberta Franca:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252906 - Roberta Franca
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{2}
K252906 - Roberta Franca
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEK N. LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252906 | |
| Device Name ECHELON ENDOPATH™ Staple Line Reinforcement (ECH60R) | |
| Indications for Use (Describe) ECHELON ENDOPATH™ Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. ECHELON ENDOPATH™ Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel, and colorectal procedures. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Ethicon Endo-Surgery LLC
510(k) Premarket Notification
(Traditional)
Echelon Endopath™ Staple Line
Reinforcement K252906
# 510(k) Summary
# (K252906)
## I. SUBMITTER
Company: Ethicon Endo-Surgery, LLC
475 Calle C
Guaynabo, PR 00969
Contact: Roberta Cabral França
Regulatory Affairs Senior Program Lead
Ethicon Endo-Surgery, Inc.
Phone: (513) 337 8855
Email: rfranca@its.jnj.com
Date Prepared: December 12, 2025
## II. SUBJECT DEVICES
Trade Name: ECHELON ENDOPATH™ Staple Line Reinforcement
Product Code: ECH60R
Classification Name: Mesh, Surgical, absorbable, staple line reinforcement
Classification Regulation: 21 CFR 878.3300
Device Class: II
Product Code: OXC
## III. Predicate DEVICES
| Predicate Device 510(k) Number | Predicate Device Name | Model |
| --- | --- | --- |
| K221343 | ECHELON ENDOPATH Staple Line Reinforcement | ECH60R |
{5}
Ethicon Endo-Surgery LLC
510(k) Premarket Notification
(Traditional)
Echelon Endopath™ Staple Line
Reinforcement K252906
## Reference Device
| Reference Device
510(k)
Number | Reference Device Name | Model |
| --- | --- | --- |
| K190937 | ECHELON ENDOPATH Staple Line Reinforcement | ECH60R |
## IV. Device Description
Echelon Endopath™ Staple Line Reinforcement is a staple line reinforcement, also known as a buttress, for use in the surgical environment for the purpose of reinforcing a staple line.
This submission is to increase the maximum sterilization dose from 25-30 kGy to 25-40 kGy and device Shelf-Life from 2 years to 3 years. There are no changes to the device itself associated with this submission.
Echelon Endopath™ Staple Line Reinforcement is to be used with surgical stapling devices. Surgical stapling devices place staggered rows of staples with a reinforcement material and simultaneously divide the tissue and the reinforcement material between the stapled rows.
The Subject Device is an absorbable staple line reinforcement material which is secured to both the stapler anvil and reload with a synthetic attachment material. The product consists of an applicator which includes the implantable device, one for each of the upper and lower stapler jaws. The implantable material consists of 3 materials: the Vicryl material, the Polydioxanone film and the attachment adhesive material.
Echelon Endopath™ Staple Line Reinforcement is an implanted material which works as an adjunct to surgical staples after transection, to provide support to soft tissue during the healing process. There are no modifications to the design of the Predicate device; and the materials of the Subject Device and Predicate Device are the same. Each unit is packaged sterile in separate pouch.
{6}
Ethicon Endo-Surgery LLC
510(k) Premarket Notification
(Traditional)
Echelon Endopath™ Staple Line
Reinforcement K252906
V. Subject Device Indications for Use
There are no changes to the intended use: “ECHELON ENDOPATH™ Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. ECHELON ENDOPATH™ Staple Line Reinforcement can be used for reinforcement of staple lines during lung resection and bariatric surgical procedures. The device can also be used for reinforcement of staple lines during gastric, small bowel and colorectal procedures”.
VI. Technological Comparison
| Characteristic | Predicate Device (K221343) | Subject Device (K252906) |
| --- | --- | --- |
| Product Descriptor Code | ECH60R | Same |
| How Supplied | Sterile, single patient use | Same |
| Biocompatibility of materials | Yes, guided by ISO 10993-1 | Same |
| Packaging | Foil | Same |
| Sterilization method | E-Beam Irradiation | Same |
| Sterility Assurance Level | SAL 10^{-6} | Same |
| Implanted Material Composition | Polyglactin 910 (Vicryl™ Knit)
laminated with Polydioxanone – (PDO Film)
Poloxamer blend (BAM)
No pigments or dyes | Same |
| Implant Absorption Process | Hydrolysis | Same |
| Implant Physical Dimensions | Length: 65.367 ± 0.30 mm
Width: 10.4 ± 0.30 mm | Same |
| Device Applicator Component | Hand-held polycarbonate applicator | Same |
| Absorption Rate | 120 days | Same |
| Storage Conditions | Room Temperature, cool dry place not to exceed 30° C | Same |
| Key Differences | | |
{7}
Ethicon Endo-Surgery LLC
510(k) Premarket Notification
(Traditional)
Echelon Endopath™ Staple Line
Reinforcement K252906
| Characteristic | Predicate Device (K221343) | Subject Device (K252906) |
| --- | --- | --- |
| Sterilization Dose (E-Beam Irradiation) | 25-30 kGy | 25-40 kGy |
| Shelf Life | 2 years | 3 years |
## VII. Non-Clinical and/or Clinical Tests
The non-clinical tests that have been submitted include:
- Device Performance Tests:
- Staple Pull Through
- Buttress Alignment in Applicator
- Reload Shear Force (device applied)
- Release force
- Packaging Verification
- Seal Strength
- Oxygen Content
- Biocompatibility
- Chemical Characterization
- Cytotoxicity
- Material Stability
- Acute Systemic Toxicity
- Sensitization
- Intracutaneous Irritation
- Pyrogenicity
- Genotoxicity
- Systemic Toxicity - Subcutaneous Implantation – 4 week
- Sub-chronic toxicity - 13 week
- Subcutaneous implantation endpoint related to device degradation: the Gastric Intended Use GLP Canine study conducted at 40 kGy, with timepoints at 10, 28, and 120 days is included in this submission. This study aligns with the complete degradation profile of the absorbable device with the increased sterilization dose. Also included in the submission is a retrospective comparison of implant studies for predicate and subject device at original and increased dose, as agreed upon with the FDA during the Pre-Submission Q231522.
## Clinical Tests
{8}
Ethicon Endo-Surgery LLC
510(k) Premarket Notification
(Traditional)
Echelon Endopath™ Staple Line
Reinforcement K252906
This premarket submission did not rely on the assessment of clinical performance data to demonstrate device performance and equivalence.
## VIII. Predetermined Change Control Plan
The predetermined change control plan (PCCP) for the device specifies anticipated modifications to the device packaging design per table below.
| Planned Modification | Test Methods and Validation Activities | Communication to Users, as needed |
| --- | --- | --- |
| Change in the device packaging that involves transitioning from existing flexible foil packaging to a new pre-formed high barrier flexible foil laminate tray design with moisture-absorbing and oxygen-scavenging materials due to material obsolescence. | Packaging design verification, package functionality and stability testing per Modification Protocol will confirm that packaging will continue to protect device from transit forces and maintain sterility. | Labeling will be updated to provide users with current device packaging illustrations as applicable. |
Implementation of the predetermined change control plan does not affect the substantial equivalence to the predicate.
## IX. Conclusion
In conclusion, based upon the Intended Use, Indications for Use, product information, performance testing, proposed modifications outlined in the PCCP, and standards compliance provided in this premarket notification, the Subject device (K252906) has been shown to be substantially equivalent to the Predicate devices and does not raise any new questions of safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.