The Augmenta Penile Implantable device intended for use in the cosmetic correction of soft tissue deformities, and is trimmed to length at the surgeon's discretion for a custom implant.
Device Story
Augmenta Penile Implant is a silicone-based, implantable device designed for cosmetic correction of soft tissue deformities. It is provided sterile for single-use. The device is available in 175 size variations to accommodate different anatomical dimensions (natural girth 5.5–13.0 cm; augmented girth 8.5–16.0 cm; length 8.0–18.0 cm). During surgery, the physician may trim the proximal end to customize the fit based on patient anatomy and surgical technique. The device functions as a physical implant to provide structural augmentation. It is intended for use by surgeons in a clinical setting. The implant provides a permanent cosmetic improvement for the patient.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance were established through biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, implantation, genotoxicity), extractables/leachables analysis, toxicological risk assessment, and product/packaging integrity testing (visual, seal strength, shipping, burst, sterilization validation, EO residuals, and image artifact evaluations).
Technological Characteristics
Silicone implant; hydrophilic coating. Dimensions: 8.0–18.0 cm length; 2.3–3.4 cm height; 0.54–0.8 cm thickness. Sterilization: Ethylene Oxide. Standards: ISO 10993-1 (biocompatibility), ASTM D4169-16 (shipping), ASTM F88/F88M-15 (seal strength), ASTM F1140/F1140M-13 (burst), ISO 11135:2014 (sterilization), ISO 10993-7:2008 (EO residuals), ASTM F2119-07 (image artifacts).
Indications for Use
Indicated for cosmetic correction of soft tissue deformities in patients requiring penile augmentation.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 30, 2022
Augmenta LLC % Lisa L. Pritchard, BSEEE Vice President, Regulatory, Quality, Clinical & Engineering DuVal & Associates, P.A. 1820 Medical Arts Building 825 Nicollet Mall Minneapolis, MN 55402
Re: K200073
Trade/Device Name: Augmenta Penile Implant Regulation Number: 21 CFR& 874,3620 Regulation Name: Ear, Nose, and Throat Synthetic Polymer Material Regulatory Class: II Product Code: MIB Dated: August 4, 2022 Received: August 4, 2022
Dear Lisa L. Pritchard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200073
Device Name Augmenta Penile Implant
Indications for Use (Describe)
The Augmenta Penile Implantable device intended for use in the cosmetic correction of soft tissue deformities, and is trimmed to length at the surgeon's discretion for a custom implant.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K200073 Page 1 of 3
Image /page/3/Picture/1 description: The image contains the logo for Augmenta. The logo consists of a stylized, abstract symbol in blue, resembling a curved line that forms a loop or a stylized letter 'a'. Below the symbol, the word "AUGMENTA" is written in bold, uppercase letters. The font appears to be sans-serif, and the overall design is clean and modern.
## 510(k) Summary
| 510(k) Owner: | Augmenta LLC<br>1315 St. Joseph Parkway<br>Suite 1700<br>Houston, TX 77002<br>Telephone: 713.205.4674<br>Contact: Robert J. Cornell, MD<br>Date prepared: September 28, 2022 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Trade Name: Augmenta Penile Implant<br>Common Name: Elastomer, Silicone Block<br>Classification Name: Ear, Nose, and Throat Synthetic Polymer Material<br>Regulation: 21 CFR §874.3620<br>Regulatory Classification: 2<br>Product Code: MIB |
| Predicate Device: | Pre-Formed Penile Silicone Block (K181387) |
## Device Description:
The Augmenta Penile Implant is an implantable device intended for use in the cosmetic correction of soft tissue deformities, and is trimmed to length at the surgeon's discretion for a custom implant. The Augmenta Penile Implant is made of silicone. The Augmenta Penile Implant comes in 175 sizes with a range of dimensions. The proximal end of the device may be trimmed by the physician to further customize the fit. The device is provided sterile and is intended for single-use only.
The predicate device has not been subject of a design-related recall.
| Measurement | Augmenta Size Range |
|--------------------------------------------------------|---------------------|
| Natural Girth Circumference (Inner<br>Circumference) | 5.5 – 13.0 cm |
| Augmented Girth Circumference (Outer<br>Circumference) | 8.5 – 16.0 cm |
| Ventral Gap | 1.5 – 3.0 cm |
Table 1: Augmenta Penile Implant Size Ranges
## Indications for Use
Total Length (L)
The Augmenta Penile Implant is an implantable device intended for use in the cosmetic correction of soft tissue deformities, and is trimmed to length at the surgeon's discretion for a custom implant.
8.0 - 18.0 cm
{5}------------------------------------------------
## Comparison of Technological Characteristics with the Predicate Device
The Augmenta Penile Implant has similar technological characteristics as the predicate device, the Preformed Penile Silicone Block cleared via K181387. A comparison of the devices is provided in Table 2.
| Device Comparison | Subject Device<br>Augmenta Penile Implant | Predicate Device<br>Pre-formed Penile Silicone Block<br>(Penuma)<br>(K181387) |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Augmenta Penile Implant | Pre-Formed Penile Silicone Block<br>(Penuma) |
| Manufacturer | Augmenta, LCC | International Medical Devices, Inc |
| FDA Product Code | MIB | MIB |
| Indications for Use | The Augmenta Penile Implant is<br>intended for use in the cosmetic<br>correction of soft tissue<br>deformities, and is trimmed to<br>length at the surgeon's discretion<br>for a custom implant. | The Pre-Formed Penile Silicone<br>Block is intended for use in the<br>cosmetic correction of soft tissue<br>deformities, and is contoured at<br>the surgeon's discretion to create a<br>custom implant. |
| Principles of Operation | The Augmenta Penile Implant may<br>be used in a variety of surgical<br>techniques. Therefore, the surgeon<br>is best advised to use the method<br>that his/her own practice and<br>discretion dictate to be best for the<br>patient. | The Penuma device may be used in<br>a variety of surgical techniques.<br>Therefore, the surgeon is best<br>advised to use the method that<br>his/her own practice and discretion<br>dictate to be best for the patient. |
| Device Material | Silicone | Medical grade silicone with an<br>embedded polyester mesh. |
| Device Coating | Hydrophilic coating | No coating |
| Device Offerings | 175 device offerings<br>Augmented Girth: 8.5 - 16.0 cm<br>Natural Girth: 5.5 – 13.0 cm<br>Ventral Gap: 1.5 — 3.0 cm<br>Length: 8.0 – 18.0 cm | 3 device offerings (L, XL, XXL) |
| Device Length | 8.0 - 18.0 cm length | 12 – 18 cm length |
| Device Size | Calculated:1<br>Height: 2.3 - 3.4 cm<br>Thickness: 0.54 - 0.8 cm | Height: 2.0 - 3.5 cm<br>Thickness: 0.5 - 1.1 cm |
| Condition of Use | Single Use | Single Use |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide |
#### Table 2: Comparison with Predicate Device
1Height and thickness are calculated measurements for Augmenta; specifications are for natural girth (inner circumference), augmented girth (outer circumference), and ventral gap.
## Non-Clinical Performance Data
No FDA performance standards have been promulgated that are applicable to the Augmenta Penile Implant.
{6}------------------------------------------------
The Augmenta Penile Implant is considered an implantable medical device with permanent (greater than 30 day) tissue/bone contact. The following assessments demonstrated safety in accordance with ISO 10993-1:
- Cytotoxicity
- Sensitization
- Irritation/intracutaneous reactivity
- Acute Systemic Toxicity
- Material-Mediated Pyrogenicity ●
- Implantation
- Genotoxicity
- . Extractables and Leachables
- Toxicology Risk Assessment
Additional non-clinical performance testing included evaluations of product and packaging integrity and acceptability of the sterilization processing These included visual integrity, seal strength, shipping, burst test, sterilization validation, EO residual, and image artifact evaluations. These assessments were conducted in accordance with the following recognized standards:
- . ASTM D4169-16
- ASTM F88/F88M-15
- ASTM F1140/F1140M-13
- ISO 111135: 2014
- ISO 10993-7: 2008
- F2119-07 (2013)
## Conclusion
The biocompatibility testing, extractable chemical analysis, toxicological risk assessment, product and packaging integrity, sterilization assessments, and image artifact testing conducted provides evidence that the Augmenta Penile Implant performs comparable to the legally marketed predicate device. The Augmenta Penile Implant has the same intended use as the predicate device. The data provided supports substantial equivalence of the Augmenta Penile Implant to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.