← Product Code [LZK](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/LZK) · K982763

# SEARE BIOMEDICAL MALAR IMPLANTS (K982763)

_Seare Biomedical Corp. · LZK · Sep 30, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/LZK/K982763

## Device Facts

- **Applicant:** Seare Biomedical Corp.
- **Product Code:** [LZK](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/LZK.md)
- **Decision Date:** Sep 30, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3550
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Seare Biomedical Malar Implants are intended to be used to augment or reconstruct the maxilla for cosmetic or reconstructive surgery.

## Device Story

Seare Biomedical Malar Implants are crescent-shaped, convex silicone elastomer rubber implants designed for surgical implantation to augment or reconstruct the maxilla. Manufactured in mirror-image pairs, the implants feature varying degrees of surface texturing and porosity. They are produced via standard molding techniques and provided in multiple sizes and styles, differing by length and projection. The implants are intended for use by surgeons in cosmetic or reconstructive procedures. They are supplied in both sterile and non-sterile formats. The device provides a physical scaffold to enhance or restore maxillary contour, benefiting patients undergoing facial reconstructive or aesthetic surgery.

## Clinical Evidence

No clinical data.

## Technological Characteristics

Silicone elastomer rubber; crescent-shaped, convex geometry; varying surface texturing and porosity; manufactured via standard molding techniques; provided sterile or non-sterile.

## Regulatory Identification

A chin prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the chin.

## Predicate Devices

- Allied Biomedical Malar Implants (1997)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

<982763

Image /page/0/Picture/1 description: The image shows the date "SEP 30 1998" in a simple, sans-serif font. The letters and numbers are printed in black ink against a white background. The date is formatted with the month abbreviated, followed by the day and then the year.

Image /page/0/Picture/2 description: The image shows a circular shape with a pattern inside. The pattern consists of dark and light areas that are intertwined. The dark areas appear to be curved and winding, creating a sense of movement within the circle. The overall impression is abstract and textured.

## SEARE BIOMEDICAL CORPORATION

## 510(k) Summary

| Contact Information: | Seare Biomedical Corporation<br>3190 Chula Vista Circle<br>Salt Lake City, Utah 84121<br>Telephone: 1(801) 355-5533<br>Facsimile: 1(801) 942-1999 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name:          | Seare Biomedical Malar Implants                                                                                                                   |
| Common Name:         | Silicone Elastomer Malar Implants                                                                                                                 |
| Classification Name: | Implant, Malar                                                                                                                                    |

Substantial Equivalence: The Seare Biomedical Malar Implant configurations are substantially equivalent in material, function, performance, and design to the Allied Biomedical Malar Implants manufactured and marketed by Allied Biomedical 1997

Device Description: Seare Biomedical Malar implants are crescent shaped convex silicone elastomer rubber implants made from specially formulated silicone elastomers designed for implantation. They are manufactured in pairs with a mirror image left and right. Surface characteristics will varying degrees of texturing and porosity. Seare Biomedical Malar Implants are intended to be used to augment or reconstruct the maxilla for cosmetic or reconstructive surgery. The Seare Biomedical Malar Implants will be produced using standard manufacturing molding techniques. The Seare Biomedical Malars will be available in many sizes and styles, all of which are very similar - differing only by a few millimeters in length and projection. The Searc Biomedical Malar Implants will be provided sterile and nonsterile.

Indications For Use: Seare Biomedical Malar Implants are intented to be used to augment or reconstruct the maxilla for cosmetic or reconstructive surgery.

Predicate Devices: The Seare Biomecical Malar Implant configurations are substantially equivalent in material, function, performance, and design to the Allied Biomedical Malar Implants manufactured and marketed by Allied Biomedical. The products have identical indications for use and are offered in the same exact size and options.

Clinical Tests: None

Adverse S&E Information: None

William f. Seone, Jr., M.D.

William J. Seare, Jr. M.D. President & C.E.O.

8/4/98

Date

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure embracing a globe, signifying the department's mission to protect the health of all Americans.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 3 0 1998

William J. Seare, M.D. President & C.E.O. Seare Biomedical Corporation 3190 Chula Vista Circle Salt Lake City, Utah 84121

Re: K982763 Seare Biomedical Malar Implants Trade Name: Regulatory Class: II Product Code: LZK Dated: August 04, 1998 Received: August 06, 1998

Dear Dr. Seare:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. તિ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Dr. William J. Seare

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M

Cella M. Witten, Ph.D., M.D. Director

Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: Seare Biomedical Malar Implants

Indications For Use:

Seare Biomedical Malar Implants are intended to be used to augment or reconstruct the maxilla for cosmetic or reconstructive surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

co'eez

(Division Sign-Off)
Division of General Restorative Devices K982763
510(k) Number

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use***_***_

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/LZK/K982763](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/LZK/K982763)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
