PERMAMESH HYDROXYLAPATITE MATRIX
Device Facts
| Record ID | K960362 |
|---|---|
| Device Name | PERMAMESH HYDROXYLAPATITE MATRIX |
| Applicant | Ceramed Corp. |
| Product Code | LZK · General, Plastic Surgery |
| Decision Date | Mar 20, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.3550 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
Facial restoration and augmentation.
Device Story
Synthetic hydroxylapatite (HA) woven sheet; high-purity, radiopaque, 1000-micron rounded HA particles organized into flexible cloth-like form using absorbable suture. Used for facial restoration and augmentation. Implanted by surgeons to provide structural support or contouring in facial areas.
Clinical Evidence
No clinical data provided; equivalence based on material characterization and chemical analysis.
Technological Characteristics
100% synthetic hydroxylapatite beads; 1000-micron particle size; absorbable suture carrier. Material conforms to ASTM F1185 for trace elements; calcium-to-phosphorus mole ratio 1.69. Radiopaque.
Indications for Use
Indicated for patients requiring facial restoration and augmentation.
Regulatory Classification
Identification
A chin prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the chin.
Predicate Devices
- MEDPOR® Surgical Implant
- OsteoGraf/D-700 (Originally OsteoGraf/AR)
Related Devices
- K960349 — PERMARIDGE HYDROXYLAPATITE MATRIX · Ceramed Corp. · Mar 13, 1996
- K960354 — PERMAMESH-D HYDROXYLAPATITE MATRIX · Ceramed Corp. · Mar 20, 1996
- K960361 — PERMAMESH-D HYDROXYLAPATITE MATRIX · Ceramed Corp. · Apr 15, 1996
- K022665 — POROUS HDPE SURGICAL IMPLANTS, TO BE DISTRIBUTED UNDER VARIOUS TRADEMARKS INCLUDING EPOR, BIOPOR, MINOPOR & P-HDPE · Hohenstein Enterprises, Inc. · Oct 9, 2002
- K141880 — BIOPOR AOC POROUS POLYETHYLENE, CEREPOR · Ceremed , Inc. · Apr 16, 2015