PERMAMESH-D HYDROXYLAPATITE MATRIX

K960354 · Ceramed Corp. · LZK · Mar 20, 1996 · General, Plastic Surgery

Device Facts

Record IDK960354
Device NamePERMAMESH-D HYDROXYLAPATITE MATRIX
ApplicantCeramed Corp.
Product CodeLZK · General, Plastic Surgery
Decision DateMar 20, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3550
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Facial (malar) restoration and augmentation.

Device Story

PermaMesh-D is a synthetic hydroxylapatite (HA) implant for malar restoration and augmentation. Device consists of high-purity, radiopaque, 1000-micron diameter HA particles organized into a flexible, woven, cloth-like sheet using absorbable suture. Implanted by surgeons in clinical settings to restore or augment facial bone structure. Provides a biocompatible, mineral-based scaffold for surgical use.

Clinical Evidence

Bench testing only. Material composition verified via X-ray diffraction; calcium to phosphorus ratio confirmed at 1.69; compliance with ASTM F1185 standards for ceramic hydroxylapatite.

Technological Characteristics

Synthetic hydroxylapatite (HA) beads; 1000-micron particle size; woven into flexible sheet using absorbable suture. Radiopaque. Material conforms to ASTM F1185 (Standard Specification for Composition of Ceramic Hydroxylapatite for Surgical Implants).

Indications for Use

Indicated for patients requiring facial (malar) restoration and augmentation.

Regulatory Classification

Identification

A chin prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the chin.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K960 354 1 of 1 510(k) SUMMARY January 22, 1996 MAR 20 1996 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. 1. Submitter: CeraMed Corporation 12860 West Cedar Drive Lakewood, CO 80228 (303) 985-0800 Contact Person: Barbara A. Watson 2. Device Name: PermaMesh-D Hydroxylapatite Matrix, 1000 microns Classification Name: Malar implant 3. Predicate Device: MEDPOR® Surgical Implant 4. Device Description: PermaMesh-D is a synthetic form of hydroxylapatite, the major mineral component of tooth enamel and bone, produced in the form of a woven sheet. It is manufactured as high purity, radiopaque, rounded particles sized at 1000 microns diameter and organized into a flat, flexible clothlike form by means of absorbable suture. 5. Intended Use: Facial (malar) restoration and augmentation. 6. Comparison of Product Characteristics: PermaMesh-D consists of 100% synthetic hydroxylapatite beads strung on absorbable suture. X-ray diffraction shows PermaMesh-D beads to be 100% HA. The hydroxylapatite component of PermaMesh-D conforms to ASTM Standard # F1185, "Standard Specification for Composition of Ceramic Hydroxylapatite for Surgical Implants", for trace elements. Typical calcium to phosphorus mole ratio is 1.69.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...