TRAXIS CEMENT RESTRICTOR

K031318 · Spinal Concepts, Inc. · JDK · Aug 22, 2003 · General, Plastic Surgery

Device Facts

Record IDK031318
Device NameTRAXIS CEMENT RESTRICTOR
ApplicantSpinal Concepts, Inc.
Product CodeJDK · General, Plastic Surgery
Decision DateAug 22, 2003
DecisionSESU
Submission TypeSpecial
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TraXis Cement Restrictor System is intended for use as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement. This device is not intended for any spinal indications. The safety and effectiveness of this device when implanted in the spine have not been established.

Device Story

TraXis Cement Restrictor is an orthopedic implant used to restrict bone cement within the distal femoral canal during hip and knee replacement surgeries. The device is a modified version of the Cadence Cement Restrictor. It is manufactured from titanium alloy (ASTM F136) or PEEK OPTIMA (ASTM F2026). PEEK versions include press-fit tantalum wire (ASTM F560) markers for radiopacity. The device is available in various dimensions to accommodate different anatomical requirements. It is intended for use by orthopedic surgeons in a clinical/OR setting. By restricting cement flow, the device assists in the proper placement and containment of bone cement during joint arthroplasty, potentially improving the stability and longevity of the prosthetic implant.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Titanium alloy (ASTM F136) or PEEK OPTIMA (ASTM F2026). Radiopaque markers: Tantalum wire (ASTM F560). Form factor: Various lengths, widths, and heights. Function: Mechanical cement restriction. No software or energy source.

Indications for Use

Indicated for use as a cement restrictor in orthopedic surgeries, specifically femoral canal and tibial plateau procedures during hip stem and total knee replacement. Contraindicated for spinal use.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## AUG 2 2 2003 #### Spinal Concepts, Inc. TraXis™ Cement Restrictor # KO31318 .. #### 510(k) Summary | SUBMITTED BY | Spinal Concepts, Inc.<br>5301 Riata Park Court, Bldg. F<br>Austin, TX 78727 | |--------------------------------------|-----------------------------------------------------------------------------| | ESTABLISHMENT<br>REGISTRATION NUMBER | 1649384 | | CONTACT PERSON | Primary | Alternate | |---------------------|--------------------------------------------------|-----------------------------------------------------------------------| | | Lisa Peterson<br>Regulatory Affairs Specialist | David M. Hooper, Ph.D<br>Director, Clinical and<br>Regulatory Affairs | | | Phone: 512-533-1080<br>Fax: 512-249-6734 | Phone: 512-533-1038<br>Fax: 512-249-6734 | | DATE PREPARED | April 18, 2003 | | | CLASSIFICATION NAME | Prosthesis, Hip, Cement Restrictor | | | COMMON NAME | Cement Restrictor | | | PROPRIETARY NAME | Spinal Concepts Inc. TraXis™ Cement Restrictor | | | PREDICATE DEVICE | Spinal Concepts Inc. Cadence™ Cement Restrictor. | | #### DEVICE DESCRIPTION The Spinal Concepts Inc. TraXis Cement Restrictor is a modified version of the Spinal Concepts Cadence Cement Restrictor (K022218 and K023647). Both are indicated for use as cement restrictors in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement. TraXis is crafted from titanium alloy (ASTM F136) or PEEK OPTIMA™ (polyaryletheretherketone, ASTM F2026). As PEEK OPTIMA™ is radiolucent, radiographic markers are included in the distal and proximal ends of the PEEK implants. The markers consist of tantalum wires (ASTM F560) that are press-fit into small holes in the implant. TraXis is offered in various lengths, widths and heights. TraXis is intended for restriction of bone cement in the distal femoral canal. {1}------------------------------------------------ ## KO31 318 Page 2 of 2 #### INDICATIONS: The TraXis Cement Restrictor System is intended for use as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement. This device is not intended for any spinal indications. The safety and effectiveness of this device when implanted in the spine have not been established. #### MECHANICAL TEST DATA No mechanical tests were performed to support this application. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with flowing lines suggesting movement or connection. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 2 2 2003 Ms. Lisa Peterson Regulatory Affairs Specialist Spinal Concepts, Inc. 5301 Riata Park Court, Bldg. F Austin, TX 78727 Re: K031318 - Trade/Device Name: TraXis Cement Restrictor System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: JDK Dated: June 24, 2003 Received: June 26, 2003 Dear Ms. Peterson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling: THIS DEVICE IS NOT INTENDED FOR ANY SPINAL INDICATIONS. WARNING: > THE SAFETY AND EFFECTIVENESS OF THIS DEVICE WHEN IMPLANTED IN THE SPINE HAVE NOT BEEN ESTABLISHED. Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. {3}------------------------------------------------ Page 2- Ms. Lisa Peterson The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Daniel Schultz, M.D. Daniel Schultz, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### 510(k) Number (if known): Device Name: #### Spinal Concepts, Inc. TraXis™ Cement Restrictor Indications for Use: - The TraXis Cement Restrictor System is intended for use as a cement restrictor in orthopedic surgeries such as those involving the femoral canal and tibial plateau in hip stem and total knee replacement. This device is not intended for any spinal indications. The safety and effectiveness of this device when implanted in the spine have not been established. #### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: (Per 21 CFR 801.109) OR Over-The-Counter: (Optional Format 1-2-96) Mark N. Milham Dr. ision of General, Restorative and Neurological Devices 510(k) Number -
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...