DURALASTIC ANATOMICAL NASAL IMPLANTS

K973573 · Hanson Medical, Inc. · FZE · Nov 13, 1997 · General, Plastic Surgery

Device Facts

Record IDK973573
Device NameDURALASTIC ANATOMICAL NASAL IMPLANTS
ApplicantHanson Medical, Inc.
Product CodeFZE · General, Plastic Surgery
Decision DateNov 13, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3680
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Duralastic Nasal implants are intended for use in augmentation and reconstruction of the nasal contour during rhinoplasty. They are intended for insertion via an intraoral or nasal sill incision.

Device Story

Duralastic Nasal Implants are silicone elastomer devices used for nasal contour augmentation or reconstruction during rhinoplasty. Implants are 'L' shaped or straight with a concave/convex dorsal aspect, extending from the alar cartilage to the radix. They are implanted by a physician via intraoral or nasal sill incision. The device provides structural support to the nasal anatomy to improve contour. Implants are provided sterile or non-sterile; non-sterile versions require a validated heat cycle by the user. The device is manufactured from platinum-cured dimethyl polysiloxane silicone elastomer, available in white or flesh tones using biocompatible pigments.

Clinical Evidence

No clinical data provided. Biocompatibility, mutagenicity, carcinogenicity, and cytotoxicity were established via reference to material master files (MAF-562 and MAF-612).

Technological Characteristics

Materials: Platinum-cured dimethyl polysiloxane (LSR-30, MED 4211), Titanium Oxide, Iron Oxide. Physical properties: 30 Shore A durometer, 650% elongation, 950 PSI tensile strength, 150 PSI tear strength (Die C), 1.12 specific gravity. Surface: Smooth and textured. Sterilization: Gamma radiation (2.5-4.2 Megarads) per ANSI/AAMI/ISO 1137-1994 or validated heat cycle.

Indications for Use

Indicated for patients undergoing rhinoplasty requiring augmentation or reconstruction of the nasal contour.

Regulatory Classification

Identification

A nose prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the nasal dorsum.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Hanson Medical Inc. 19325 58th PL NE Seattle WA 98155 (425) 481 2185 510(k) Submission Nasal Implants K9735 73 NOV 1 3 1997 #### TAB H #### 510 (k) SUMMARY # PRODUCT DESCRIPTION Duralastic Nasal implants are "L" shaped or straight with concave convex dorsal aspect which reached from the alar catilage to the These silicone elastomer rubber implants are made from radix. specially formulated silicone elastomers designed for implantation. The LSR 30 Implant Grade elastomer is Masterfiled at FDA and has for biocompatibility, mutagenicity, thoroughly tested been carcinogenicity, and cytotoxicity. These referenced material characterizations are found in Applied Silicone's Master File MAF-562 and Nusil Technologies's MAF 612 for their MED 4211 Silicone Rubber -Unrestricted. The Duralastic Nasal Implants will be provided sterile and nonsterile. Several biocompatible pigments are used to make the white and flesh tone versions of these implants. ## SUBSTANTIAL EQUIVALENCE Duralastic Nasal Implants are SE to Allied Biomedical Nasal Implants, K971481. They will be produced in the exact sizes and shapes as the Allied Predicate device. #### INTENDED USE Duralastic Nasal implants are intended for use in augmentation and reconstruction of the nasal contour during rhinoplasty. They are intended for insertion via an intraoral or nasal sill incision. #### PHYSICAL AND CHEMICAL PROPERTIES The Duralastic Nasal implants are manufactured from Applied Silicone's LSR-30 part # 40029 which is a platinum cured dimethyl polysiloxane system. All chemical characterizations are found in Applied Masterfile MAF-562. Titanium Oxide and Iron Oxide pigments are added to these implants when ordered by the physician. These pigments are tested for cytotoxicity. The proposed Duralastic Nasal Implants will also be manufactured utilizing Nusil Technologies' MED 4211 Unrestricted Silicone Rubber Masterfiled at FDA under Nusil MAF 612. The physical properties are: Durometer 30 Shore A, Elongation 650%, Tensile Strength 950 PSI, tear strength Tear Die C 150 PSI, {2}------------------------------------------------ # 510(k) Summary Continued Specific Gravity 1.12, Modulus 300 PSI at 200% Elongation, Surface Smooth and Textured. # STERILIZATION CYCLE Duralastic Nasal Implants are sterilized via gamma radiation cycles of 2.5 - 4.2 Meqarads. The validation of this cycle will be designed and validated by a private contractor. Sterigenics Corporation will be the contract gamma sterilizer. The validation will use Method 1 Testing as defined in the ANSI/AAMI/ISO 1137-1994 "Sterilization of Health Care Products -- Requirements for Validation and Routine Control - Radiation Sterilization." In the event the products are shipped nonsterile a validated heat cycle will be recommended to the user. ## PACKAGE DESCRIPTION The proposed Duralastic Nasal Implants will be double peel pouched in Tyvec pouches and provided sterile and non sterile. These pouches will be in turned packaged in corrugated boxes for shipping. {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Kathy Richardson Regulatory Affairs Hanson Medical Inc. 19325 58th Place NE Seattle, Washington 98155 NOV 1 3 1997 Re: K973573 Trade Name: Duralastic Anatomical Nasal Implants Regulatory Class: II Product Code: ESR Dated: September 15, 1997 Received: September 19, 1997 Dear Ms. Richardson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Ms. Kathy Richardson This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Pcolle Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 of 1 510(k) Number - K4973573 S Device Name: Silicone Elastomer Nasal Implant Indications For Use: To augment or reconstruct the nasal contour Concurrence of CDRH, Office of Device Evaluation (ODE) | | (Division Sign-Off) | |---------------|-----------------------------------------| | | Division of General Restorative Devices | | 510(k) Number | K973573 | | Presecription Use<br>(Per 21 CFR 801.109) | OR | Over The Counter Use<br>(Optional Format 1-2-96) | |-------------------------------------------|----|--------------------------------------------------| |-------------------------------------------|----|--------------------------------------------------| R.O.C. 2011
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...