The DePuy Synthes Porous Polyethylene Implants are intended for use in non-load bearing applications in craniofacial reconstruction, cosmetic surgery, and repair of craniofacial trauma. The DePuy Synthes Titanium Wire is indicated for use in soft tissue approximation and/or ligation, canthoplasty, canthopexy and/or canthal tendon repair. The DePuy Synthes Titanium Wire With Barb is indicated for use in soft tissue approximation and/or ligation, canthoplasty, canthopexy and/or canthal tendon repair.
Device Story
Portfolio includes porous polyethylene implants (with/without titanium) and titanium wires (with/without barbs). Implants used for non-load bearing craniofacial reconstruction, cosmetic surgery, and trauma repair. Wires used for soft tissue approximation, ligation, and canthal procedures. Devices are surgically implanted by physicians. Submission updates labeling to MR Conditional status. Safety in MR environment validated via non-clinical testing.
Clinical Evidence
No clinical data. Bench testing only. Non-clinical performance testing conducted per ASTM standards: displacement force (ASTM F2052-14), torque (ASTM F2213-06), RF heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07).
Technological Characteristics
Implants: Porous/smooth polyethylene sheets, optionally with titanium. Wires: Commercially pure titanium, monofilament, nonabsorbable, with or without stainless steel needles/barbs. MR Conditional status established via ASTM F2052-14, F2213-06, F2182-11a, and F2119-07.
Indications for Use
Indicated for patients requiring craniofacial reconstruction, cosmetic surgery, or repair of craniofacial trauma (implants); or soft tissue approximation, ligation, canthoplasty, canthopexy, or canthal tendon repair (wires).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized caduceus. To the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name in blue text.
March 13, 2019
Synthes (USA) Products LLC Fredy Varela Senior Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380
# Re: K183472
Trade/Device Name: DePuy Synthes Porous Polyethylene Implants and Titanium Wires Portfolio - MR Conditional Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM, GAQ Dated: December 11, 2018
Received: December 14, 2018
# Dear Mr. Varela:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by David Krause -David Krause -S Date: 2019.03.13 13:12:50 -04'00'
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### K183472
Device Name
DEPUY SYNTHES POROUS POLYETHYLENE IMPLANTS - MR CONDITIONAL
Indications for Use (Describe)
The DePuy Synthes Porous Polyethylene Implants are intended for use in non-load bearing applications in craniofacial reconstruction, cosmetic surgery, and repair of craniofacial trauma.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### K183472
Device Name
DEPUY SYNTHES TITANIUM WIRE - MR CONDITIONAL
Indications for Use (Describe)
The DePuy Synthes Titanium Wire is indicated for use in soft tissue approximation and/or ligation, canthoplasty, canthopexy and/or canthal tendon repair.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
### K183472
Device Name
DEPUY SYNTHES TITANIUM WIRE WITH BARB - MR CONDITIONAL
Indications for Use (Describe)
The DePuy Synthes Titanium Wire With Barb is indicated for use in soft tissue approximation and/or ligation, canthoplasty, canthopexy and/or canthal tendon repair.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary 1.
Date Prepared: March 7, 2019
#### 1.1. Submitter
## Primary Contact:
Fredy Varela Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: 610-719-5883 Email: fvarela2@its.jnj.com
### Alternate Contact:
Elizabeth Jacobs Regulatory Affairs Project Leader DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5768 Email: ejacob10@its.jnj.com
#### 1.2. Device
Name of Device: DEPUY SYNTHES POROUS POLYETHYLENE IMPLANTS – MR Conditional
Classification Name(s): Surgical Mesh
Regulatory Class: Class II; 878.3300
## Product Code(s): FTM
#### Predicate Devices 1.3.
- K051879 Synthes Porous Polythylene Implants
- K040364 Medpor Craniofacial Implants with Embedded Titanium
- K952677 Medpor Barrier Surgical Implant Biomaterial
- K832283 Medpor Surgical Implants
- K022665 ePor Porous HDPE Surgical Implants
#### Device Description 1.4.
The DePuy Synthes Porous Polyethylene Implants consist of porous/smooth implant sheets with or without titanium.
{6}------------------------------------------------
# 1.5. Indications for Use
The DePuy Synthes Porous Polyethylene Implants are intended for use in non-load bearing applications in craniofacial reconstruction, cosmetic surgery, and repair of craniofacial trauma.
#### Substantial Equivalence 1.6.
The purpose of this submission is to add MR Conditional information to the device labeling for the DEPUY SYNTHES POROUS POLYETHYLENE IMPLANTS. The intended use and technological characteristics of the devices remain unchanged.
#### 1.7. Performance Testing
Non-clinical testing results are provided to support the conditional safety of the DEPUY SYNTHES POROUS POLYETHYLENE IMPLANT AND TITANIUM WIRE PORTFOLIO in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14), torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07).
#### Conclusion 1.8.
The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{7}------------------------------------------------
#### 1. 510(k) Summary
Date Prepared: March 7, 2019
#### 1.1. Submitter
Primary Contact: Fredy Varela Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: 610-719-5883 Email: fvarela2@its.jnj.com
Alternate Contact: Elizabeth Jacobs Regulatory Affairs Project Leader DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5768 Email: ejacob10@its.jnj.com
#### 1.2. Device
Name of Device: DEPUY SYNTHES TITANIUM WIRE – MR Conditional
Classification Name(s): Suture, Nonabsorbable, Steel, Monofilament and Multifilament, Sterile
Regulatory Class: Class II; 878.4495
Product Code(s): GAQ
#### 1.3. Predicate Devices
K041333 - Synthes Titanium Wire Ethilon™ Nylon Suture
#### Device Description 1.4.
The DePuy Synthes Titanium Wire is a nonabsorbable, monofilament, sterile surgical wire composed of commercially pure titanium wire is available in a length of 500mm with different gauge sizes, and is available with or without a permanently attached stainless steel needle.
#### Indications for Use 1.5.
DePuy Synthes Titanium Wire is indicated for use in soft tissue approximation and/or ligation, canthoplasty, canthopexy and/or canthal tendon repair.
{8}------------------------------------------------
# 1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DEPUY SYNTHES TITANIUM WIRE. The intended use and technological characteristics of the devices remain unchanged.
#### 1.7. Performance Testing
Non-clinical testing results are provided to support the conditional safety of the DEPUY SYNTHES POROUS POLYETHYLENE IMPLANT AND TITANIUM WIRE PORTFOLIO in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14), torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07).
#### Conclusion 1.8.
The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{9}------------------------------------------------
#### 1. 510(k) Summary
Date Prepared: March 7, 2019
#### Submitter 1.1.
## Primary Contact:
Fredy Varela Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: 610-719-5883 Email: fvarela2@its.jnj.com
## Alternate Contact:
Elizabeth Jacobs Regulatory Affairs Project Leader DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5768 Email: ejacob10@its.jnj.com
#### 1.2. Device
Name of Device: DEPUY SYNTHES TITANIUM WIRE WITH BARB – MR Conditional
Classification Name(s): Suture, Nonabsorbable, Steel, Monofilament and Multifilament, Sterile
Regulatory Class: Class II; 878.4495
## Product Code(s): GAQ
- 1.3. Predicate Devices
K042606 – Synthes Titanium Wire K041333 – Synthes Titanium Wire
#### 1.4. Device Description
The DePuy Synthes Titanium Wire with Barb is a nonabsorbable, monofilament, sterile surgical wire with an attached barb. The 28 gauge titanium wire with barb is available in a length of 538mm, and has a permanently attached stainless steel needle.
#### 1.5. Indications for Use
The Depuy Synthes Titanium Wire With Barb is indicated for use in soft tissue approximation and/or ligation, canthoplasty, canthopexy and/or canthal tendon repair.
{10}------------------------------------------------
# 1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DEPUY SYNTHES TITANIUM WIRE WITH BARB. The intended use and technological characteristics of the devices remain unchanged.
#### Performance Testing 1.7.
Non-clinical testing results are provided to support the conditional safety of the DEPUY SYNTHES POROUS POLYETHYLENE IMPLANT AND TITANIUM WIRE PORTFOLIO in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14), torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07).
#### Conclusion 1.8.
The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.