← Product Code [FTM](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM) · K162965

# SurgiMend MP Collagen Matrix for Soft Tissue Reconstruction (K162965)

_Tei Biosciences, Inc. · FTM · Feb 16, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM/K162965

## Device Facts

- **Applicant:** Tei Biosciences, Inc.
- **Product Code:** [FTM](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM.md)
- **Decision Date:** Feb 16, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

SurgiMend is intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue membranes. SurgiMend is specifically indicated for: - Plastic and reconstructive surgery - Muscle flap reinforcement - Hernia repair including abdominal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias.

## Device Story

SurgiMend MP is an acellular bovine dermal collagen matrix used for soft tissue reinforcement and repair. It is provided in solid, fenestrated, or perforated configurations (2-3 mm holes in staggered array) to facilitate fluid flow. The device is implanted by surgeons during procedures such as hernia repair or plastic/reconstructive surgery. It acts as a scaffold to support tissue integration and revascularization. The device is supplied sterile in various sizes and thicknesses. It functions as a surgical mesh to provide mechanical support to weakened or damaged soft tissue membranes.

## Clinical Evidence

Bench testing and animal studies only. Mechanical testing (tensile strength, suture pull-out, tear resistance, ball burst) performed per FDA guidance. Animal models included 12-week rat hernia repair, 4-week rat intra-abdominal, and 2-week rat intra-muscle implant models. Endpoints included mechanical strength, revascularization, and tissue integration. Results demonstrated that properties of the perforated configuration remain equivalent to the non-perforated predicate.

## Technological Characteristics

Acellular bovine dermal collagen matrix. Available in solid, fenestrated, or perforated (2-3 mm holes) configurations. Sterile. Mechanical properties include tensile strength, suture pull-out, tear resistance, and ball burst. No electronic or software components.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- SurgiMend® Collagen Matrix for Soft Tissue Reconstruction ([K083898](/device/K083898.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, overlaid on three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2017

Integra LifeSciences Corporation c/o Ms. Kavita Amin Specialist, Regulatory Affairs 311 Enterprise Drive Plainsboro, New Jersey 08536

Re: K162965

Trade/Device Name: SurgiMend MP Collagen Matrix for Soft Tissue Reconstruction Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM Dated: January 13, 2017 Received: January 17, 2017

Dear Ms. Amin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162965

#### Device Name

SurgiMend® MP Collagen Matrix for Soft Tissue Reconstruction

Indications for Use (Describe)

SurgiMend is intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue membranes.

SurgiMend is specifically indicated for:

- · Plastic and reconstructive surgery
- · Muscle flap reinforcement
- · Hernia repair including abdominal, femoral, diaphragmatic, scrotal, umbilical, and incisional hernias.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

### SurgiMend MP Collagen Matrix for Soft Tissue Reconstruction

### Submitter's name and address:

TEI Biosciences Inc. 7 Elkins Street Boston, MA 02127 USA

#### Contact person and telephone number:

Kavita Amin Specialist, Regulatory Affairs Telephone: 609-750-7827

### Date the Summary was prepared: 13 January 2017

#### Name of the device:

| Trade name:          | SurgiMend® MP Collagen Matrix for Soft Tissue<br>Reconstruction |
|----------------------|-----------------------------------------------------------------|
| Common Name:         | Surgical Mesh                                                   |
| Classification Name: | Mesh, Surgical (21 CFR 878.3300)                                |
| Product Code:        | FTM                                                             |

### Predicate Device:

SurgiMend® MP is substantially equivalent in function and intended use to its predicate device as detailed in Table 1.

### Table 1: Predicate Device

| 510(k)<br>Number | Product<br>Code | Trade Name                                                   | Manufacturer            |
|------------------|-----------------|--------------------------------------------------------------|-------------------------|
| K083898          | FTM             | SurgiMend® Collagen Matrix for<br>Soft Tissue Reconstruction | TEI Biosciences<br>Inc. |

### Device Description:

SurgiMend is an acellular dermal tissue derived from bovine dermis. The device is available in solid, fenestrated, and perforated configurations. The device is supplied sterile in a variety of sizes, shapes, and thicknesses.

#### Indication for Use:

SurgiMend is intended for implantation to reinforce soft tissue where weakness exists and for surgical repair of damaged or ruptured soft tissue membranes:

SurgiMend is specifically indicated for:

- Plastic and reconstructive surgery ●
- Muscle flap reinforcement ●
- Hernia repair including abdominal, inguinal, femoral, diaphragmatic, scrotal, ● umbilical, and incisional hernias.

{4}------------------------------------------------

## Substantial Equivalence Comparison:

The modified device, SurgiMend MP, utilizes the same design, material, manufacturing process, packaging and sterilization parameters as the predicate device (K083898). The proposed device, SurgiMend MP, is offered in perforated configuration with holes of 2-3 mm in diameter in a staggered array. The addition of defined holes in this new configuration allows for better fluid flow through the device upon implantation.

#### Testing and Test Results:

The modified device, SurgiMend MP, is comprised of the same materials, processed, packaged and sterilized by the same method as the predicate device (K083898). The biocompatibility testing conducted on SurgiMend Collagen Matrix for Soft Tissue Reconstruction (K083898) product confirms the biological safety of the SurgiMend MP.

Performance testing for SurgiMend MP is the same as the predicate device. Mechanical testing, including tensile strength, suture pull-out, tear resistance, and ball burst were conducted in compliance with FDA Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh, March 2, 1999. All test results were acceptable.

The implantation studies were conducted in three different animal models (12 week rat hernia repair study, 4 week rat intra-abdominal, 2 week rat intra-muscle implant model) with the primary endpoint of mechanical strength to support soft tissue reinforcement. The studies also also characterized secondary endpoints of revascularization and tissue integration of perforated test and non-perforated control configurations. These studies demonstrate the basic properties of the subject device after implantation remain unchanged from the predicate. The result demonstrates the subject device is substantially equivalent to the predicate device.

#### Conclusion:

The modified device, SurgiMend MP, is substantially equivalent to the commercially available marketed device. SurgiMend. The modifications expressed in this 510(k) Premarket Notification do not change the intended use, nor alter the fundamental scientific technology of the device, and do not raise any new issues of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM/K162965](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM/K162965)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
