K122048 · Rotation Medical, Inc. · FTM · Jan 8, 2013 · General, Plastic Surgery
Device Facts
Record ID
K122048
Device Name
COLLAGEN TENDON SHEET-D
Applicant
Rotation Medical, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Jan 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Collagen Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Device Story
Collagen Tendon Sheet-D is a resorbable type I collagen matrix designed to provide a protective layer between injured tendons and surrounding tissue. The device is a soft, pliable, porous sheet that is hydrated before use. It features a small amount of FD&C Blue #2 on the perimeter of the upper surface to provide visual contrast during surgical fixation. The device is provided sterile for single use. It is intended for use by surgeons in the management of tendon injuries where tendon tissue remains intact. By providing a physical barrier, the device aids in the protection of the injured tendon site.
Clinical Evidence
No clinical data. Evidence consists of bench testing and in vivo biocompatibility studies. Biocompatibility testing included cytotoxicity, sensitization, systemic toxicity, hemolysis, and genotoxicity, all passing FDA Blue Book Memorandum G95-1 and ISO 10993-1 standards. Mechanical characterization (density, strength, stiffness, tear resistance) and chemical characterization (SDS-PAGE, collagen typing, residual testing, viral inactivation) demonstrated equivalence to the predicate.
Technological Characteristics
Resorbable type I collagen matrix. Contains FD&C Blue #2 dye for contrast. Provided sterile, non-pyrogenic, single-use. Mechanical properties include density, strength, stiffness, and tear resistance. Evaluated per FDA Guidance for the Preparation of a Premarket Notification for a Surgical Mesh.
Indications for Use
Indicated for the management and protection of tendon injuries in patients with no substantial loss of tendon tissue.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}------------------------------------------------
K122048
Image /page/0/Picture/1 description: The image shows the logo for Rotation Medical. The logo consists of the word "rotation" in a bold, sans-serif font, with the word "MEDICAL" in a smaller, sans-serif font underneath. Above the word "rotation" is a semi-circle of circles.
1/2
# 510(k) Summary
JAN 0 8 2013
:
## Applicant Information
Applicant Name: Applicant Address:
Telephone: Fax: Contact Person: Rotation Medical, Inc. 15350 25th Avenue North, Suite 100 Plymouth, MN 55447 763-746-7502 763-746-7501 Jeff Sims Vice President, Clinical Programs and Regulatory Affairs January 7, 2013
Date Prepared:
Name of Device
Device Common Name: Device Trade Name: -Device Classification Name: Tendon Protector Collagen Tendon Sheet-D Mesh, Surgical 878.3300 Class II FTM
## Legally Marketed Devices to Which Substantial Equivalence is Claimed
Predicate Device(s):
Collagen Tendon Sheet, K112423 Rotation Medical, Inc.
## Description of the Device
Collagen Tendon Sheet-D is a resorbable type I collagen matrix that provides a layer of collagen over injured tendons. Collagen Tendon Sheet-D is designed to provide a layer between the tendon and the surrounding tissue. When hydrated, Collagen Tendon Sheet-D is an easy-touse, soft, pliable, nonfriable, porous collagen sheet. Collagen Tendon Sheet-D is provided sterile, non-pyrogenic, for single use only, in a variety of sizes, in double peel packages.
#### Intended Use
Collagen Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
{2}------------------------------------------------
K122048
2/2
Image /page/1/Picture/1 description: The image shows the logo for Rotation Medical. The logo consists of the word "rotation" in a stylized font with a line underneath it. Below the word "rotation" is the word "MEDICAL" in a smaller, sans-serif font. Above the word "rotation" is a semi-circle made of circles.
## Summary/Comparison of Technical Characteristics
Collagen Tendon Sheet-D is the same product as its predicate Collagen Tendon Sheet (K112423) in all regards excepting the addition of a small amount of FD&C Blue #2 on the perimeter of the upper surface to provide contrast with surrounding tissue during fixation.
Collagen Tendon Sheet-D has been evaluated in a number of in vitro and in vivo tests to assess its safety/biocompatibility, including: cytotoxicity (agarose overlay, liquid and extract methods), tto ourotyplooompanisms reactivity, systemic toxicity, hemolysis, genotoxicity (bacterial and Schildtion, intraductions routing) , and muscle implantation tissue response tests. The device passed all applicable FDA Blue Book Memorandum G95-1 and ISO 10993-1 testing for the biological evaluation of medical devices.
Collagen Tendon Sheet-D and its predicate share base material that has been characterized for chemical composition and purity using SDS-PAGE analysis, collagen typing, residual testing and viral inactivation (four appropriate viruses selected for assessment). Mechanical characterization included density, strength, stiffness and tear resistance to demonstrate substantial equivalence. Testing was conducted in accordance with FDA's Guidance for the Preparation of a Premarket Notification for a Surgical Mesh.
#### Conclusion of Non-clinical Studies
The results of the in vitro product characterization studies, and in vitro and in vivo biocompatibility studies, demonstrate that Collagen Tendon Sheet-D is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Rotation Medical, Incorporated % Mr. Jeff Sims Vice President, Clinical Programs and Regulatory Affairs 15350 25th Avenue North, Suite 100 Plymouth, Minnesota 55447
Re: K122048
Trade/Device Name: Collagen Tendon Sheet-D Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM. OWY Dated: December 06, 2012 Received: December 11, 2012
January 8, 2013
## Dear Mr. Sims:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced aboye and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Jeff Sims
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm] 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K12-2048
Confidential Rotation Medical, Inc.
# Indications for Use
510(k) Number (if known):
Device Name: _ Collagen Tendon Sheet-D
Indications for Use:
Collagen Tendon Sheet-D is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kime for MXM
(Division Sign-Off)
Division of Sure
510(k) Numb
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.