← Product Code [FTM](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM) · K050337

# SIS FISTULA PLUG (K050337)

_Cook Biotech, Inc. · FTM · Mar 9, 2005 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM/K050337

## Device Facts

- **Applicant:** Cook Biotech, Inc.
- **Product Code:** [FTM](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM.md)
- **Decision Date:** Mar 9, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The SIS Fistula Plug is for implantation to reinforce soft tissue where a rolled configuration is required, for repair of anal, rectal, and enterocutaneous fistulas. The device is supplied sterile and is intended for one-time use.

## Device Story

The SIS Fistula Plug is a surgical mesh device manufactured from porcine small intestinal submucosa (SIS). It is provided in a lyophilized (dried) state and configured as a roll for implantation. The device is intended for use by surgeons to reinforce soft tissue during the repair of anal, rectal, and enterocutaneous fistulas. By providing a scaffold for tissue reinforcement, the plug supports the repair process. It is a single-use, sterile device.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility, viral inactivation, and mechanical performance evaluations.

## Technological Characteristics

Material: Porcine small intestinal submucosa (SIS). Form factor: Rolled configuration, lyophilized (dried). Sterilization: Sterile. Regulatory class: II (21 CFR 878.3300).

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- SURGISTS of Tissue Graft (Cook Biotech Incorporated)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K050337 SIS Fistula Plug

37

/

## 9. 510(K) Summary

| Submitted By: | Mark Bleyer, President    |
|---------------|---------------------------|
|               | Cook Biotech Incorporated |
|               | 1425 Innovation Place     |
|               | West Lafayette, IN 47906  |
|               | (765) 497-3355            |
|               | March 2, 2005             |

### Names of Device:

SIS Fistula Plug Trade Name: Surgical Mesh Common/Usual Name: Proposed classification name: Surgical Mesh 21 CFR 878.3300 (FTM) Class II

#### Intended Use:

The SIS Fistula Plug is for implantation to reinforce soft tissue where a rolled configuration is a The SIS Fistula Flug is for implanation to remreros securers is supplied sterile and is intended for one-time use.

#### Predicate Devices:

Predicate Devices:
The SIS Fistula Plug is similar to predicate devices, including the SURGISTS of Tissue Graft The SIS Fistula Frights Shiftin's to productions are results and manufactured by Cook Biotech Incorporated.

#### Device Description:

The SIS Fistula Plug is manufactured from porcine small intestinal submucosa (SIS) and is I he SISTIStula I lug is manufactured from portiguration. The device is packaged in a lyophilized (dried) state, and supplied sterile in a sealed double pouch system.

#### Substantial Equivalence:

Substannal Equirateneer
The SIS Fistula Plug is similar with respect to intended use, materials and technological characteristics to the above predicate devices in terms of 510(k) substantial equivalence as shown through bench and biocompatibility testing and clinical experience.

#### Discussion of Tests and Test Results:

The material comprising the SIS Fistula Plug was subjected to extensive biocompatibility testing, viral inactivation testing, and mechanical testing. Outcomes show the device to be Vilal mactivation tosting, und meetialited. The material, and mechanical, and mechanical characteristics to be sufficient.

#### Conclusions Drawn from the Tests:

Conclusions Drawn from the Fistula Plug provide evidence of its suitability for use in Outcomes from the evaluation of the USD 1.0.2.1.2018 10:41:48 predicate devices in terms of intended use and technological characteristics.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human figures or birds in flight. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 9 - 2005

Mr. Mark Bleyer President Cook Biotech Incorporated 1425 Innovation Place West Lafayette, Indiana 47906

Re: K050337

Trade/Device Name: SIS Fistula Plug Frade/Don Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: FTM Dated: February 9, 2005 Received: February 10, 2005

Dear Mr. Bleyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section >10(t) prematic is substantally equivalent (for the indications
referenced above and have determined the device is substanted in interstate referenced above and have decemined the acres to to case and case of the states for use stated in the enclosure) to tegally manced produced Device Amendments, or to
commerce prior to May 28, 1976, the enactment da provisions of the Federal Food. Drue commerce prior to May 28, 1976, the enactively of the Federal Food. Drug.
devices that have been reclassified in accordance with the provisions of the Federal (PMA) devices that have been reclassified in accordance what as proval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act. The and Cosmetic Act (Act) that do not require approval controls provisions of the Act. The
 You may, therefore, market the device, subject to the general consumer station, ligl You may, therefore, market the device, subject to the gents for annual registration, listing of
general controls provisions of the Act include requires misbranding and general controls provisions of the Act merade requirements of the management misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it class III (PMA). it If your device is classified (see above) mo entire cass if epenlations affecting your device can
may be subject to such additional controls. Existing major regulations affec may be subject to such additional controls. Existing major vegator 99 in addion. FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 be found in the Code of Pederal Regalations, vour device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean Please be advised that FDA s issuance of a substantal option with other requirements of the Act
that FDA has made a determination that your device arther Fequirements You that FDA has made a determination mar your devres on other Federal agencics. You must or any Federal statutes and regulations administration to: registration and listing (21)
comply with all the Act's requirements, including, but not anglisements as set comply with all the Act s requirements, mendants wardica requirements as set
CFR Part 807); labeling (21 CFR Part 801); end if englicably. the clectronic CFR Part 807); labeling (21 CFK Farl 601); good manatatagon; and if applicable, the clectronic
forth in the quality systems (QS) regulation (2) CFR Patl CHP 1000-1050 forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000

{2}------------------------------------------------

Page 2 - Mr. Mark Bleyer

This letter will allow you to begin marketing your device as described in your Section 510(k)
10 legally This letter will allow you to begin marketing your device of your device of your device to a legally
premarket notification. The FDA finding of substantial equivalence of yo premarket notification. The PDA Innung of substantal equivalent of this, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r the are and and and 10 10 10 10 10 10 10 10 10 10 100 100 100 100 100 100 If you desire specific advice for your device of our acount of the regulation entitled.
contact the Office of Compliance at (240) 276-0115. Also, please note the regulation contact the Office of Compliance at (240) 270-01-15. Fran 807.97). You may obtain
"Misbranding by reference to premarket notification of Small "Misbranding by reference to premarket normical " (1782 Act from the Division of Small
other general information on your responsibilities under (1800) 638-204 other general intormation on your responsionities and effect number (800) 638-2041 or Manufacturers, International and Consumer Assistance at to very of the news and index.html.

Sincerely yours,

Miriam C. Provost, Ph.D.

ಕ್ಕಿ 

Acting Director Division of General, Restorative and Neurological Devices ()ffice of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known): K050337 Device Name: SIS Fistula Plug Indications for Use:

The SIS Fistula Plug is for implantation to reinforce soft tissue
r The SIS Fistula Flug is for implined, for repair of anal,
where a rolled configuration is required, for repair of supplied where a rolled colliguration for roquired, The device is supplied
rectal, and enterocutaneous fistulas. The device is supplied rectal, and is intended for one-time use.

Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (Fall 21 CFR 001 Baspare)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number_KO50332

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM/K050337](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTM/K050337)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
