K992556 · Tissue Science Laboratories, Plc · FTL · Feb 17, 2000 · General, Plastic Surgery
Device Facts
Record ID
K992556
Device Name
PERMACOL
Applicant
Tissue Science Laboratories, Plc
Product Code
FTL · General, Plastic Surgery
Decision Date
Feb 17, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Permacol™ is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes. It is specifically indicated for the repair of abdominal, inguinal, diaphragmatic, femoral, scrotal, umbilical and incisional hernias; colon, rectal, urethral and vaginal prolapse; muscle flap reinforcement; reconstruction of the pelvic floor and procedures such as sacrocolposuspension and urethral sling.
Device Story
Permacol™ is a sterile, acellular, crosslinked porcine dermal collagen and elastin flat sheet. Used as a surgical patch to reinforce or repair soft tissue membranes; applied by surgeons in clinical settings. Device provides mechanical support for hernia repairs, prolapse, and pelvic floor reconstruction. Biocompatible material integrates into host tissue. Clinical benefit includes ease of handling and successful repair of damaged soft tissue.
Clinical Evidence
Clinical experience in several hundred European patients; case histories for 60 patients undergoing various soft tissue repairs. Results demonstrated ease of handling, successful clinical outcomes in nearly all cases, and no product-related adverse effects. Supporting data includes 6-month rat implantation histological evaluation and standard biocompatibility testing (hemolysis, cytotoxicity, systemic toxicity, genotoxicity, sensitization, pyrogenicity).
Technological Characteristics
Acellular crosslinked porcine dermal collagen and elastin flat sheet. Sterile, moist, flexible, fibrous material. Biocompatibility testing performed per FDA memorandum #G95-1. Non-synthetic, biological mesh.
Indications for Use
Indicated for patients requiring soft tissue reinforcement or repair of damaged/ruptured membranes, including hernia repair (abdominal, inguinal, diaphragmatic, femoral, scrotal, umbilical, incisional), prolapse repair (colon, rectal, urethral, vaginal), muscle flap reinforcement, pelvic floor reconstruction, sacrocolposuspension, and urethral sling procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
GraftPatch®
Rapi-Seal™ Patch
Biosynthetic Surgical Mesh
Peri-Guard® family of products
Glycar Tissue Repair Patch
Submission Summary (Full Text)
{0}------------------------------------------------
Permacol™ Porcine Collagen Surgical Mesh 510(k) Premarket Notification Submission
FEB 1 7 2000
## 510(k) Summary
Sponsor: Tissue Science Laboratories, PLC Grevholme House Victoria Road Aldershot Hants GU11 1SJ United Kingdom
### Contact Person:
Dr. Allison F. Wren The Wren Group 7124 La Jolla Blvd La Jolla CA 92037 Tel: (619) 454-6260 Fax: (619) 454-2508 E-mail: TheWrenGrp@aol.com
or
Howard M. Holstein, Esq. Hogan & Hartson, LLP 555 Thirteenth Street, N.W. Washington, DC 20004 Tel: (202) 637-5813 Fax: (202) 637-5910 E-mail: hmholstein@hhlaw.com
- Submission July 31, 1999 Date:
- Device Name: Permacol™
- Classification: Polymeric Surgical Mesh
Predicate Devices:
GraftPatch® Rapi-Seal™ Patch Biosynthetic Surgical Mesh Peri-Guard® family of products Glycar Tissue Repair Patch
{1}------------------------------------------------
### Permacol™ Porcine Collagen Surgical Mesh 510(k) Premarket Notification Submission
Device Description: Permacol™ is a sterile, tough. off-white. moist. flexible, fibrous flat sheet solely comprised of acellular crosslinked porcine dermal collagen and elastin Intended Use: Permacol™ is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes. It is specifically indicated for use in the following types of soft tissue repair procedures: abdominal, inquinal, diaphragmatic, femoral, scrotal, . umbilical, and incisional hernias; . colon, rectal, urethral, and vaginal prolapse: muscle flap reinforcement; . . reconstruction of the pelvic floor; sacrocolposuspension; and . . urethral sling. Substantial Equivalence: Permacol™ is substantially equivalent to the predicate devices in terms of its intended use, specific indications, and performance. There are minor differences in the technological characteristics, but laboratory and animal testing as well as clinical experience have shown that these do not raise any new concerns of either safety or effectiveness. Safetv Evaluation: Standard biocompatibility testing was performed according to the FDA-modified matrix recommended in FDA memorandum #G95-1. The product passed all of the following tests: Hemolysis Cytotoxicity - elution method Acute Systemic Toxicity Implantation/Subchronic Toxicity Genotoxicity - Ames Mutagenesis Genotoxicity - Bone Marrow Micronucleus Genotoxicity - Mouse Lymphoma Mutagenesis Sensitization-Magnusson and Kligman Pyrogenicity
2
{2}------------------------------------------------
| | Permacol™ Porcine Collagen Surgical Mesh<br>510(k) Premarket Notification Submission |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | In addition to standard biocompatibility testing, a histological<br>evaluation of the material after implantation in rats for 6<br>months demonstrated that Permacol™ is safe and well<br>tolerated. |
| Clinical<br>Evaluation: | Clinical experience in several hundred patients in Europe<br>also has demonstrated that Permacol™ is safe and well<br>tolerated. Case histories for 60 patients undergoing a<br>variety of soft tissue repair procedures demonstrate that<br>Permacol™ is easy to handle, and successful clinical<br>outcomes were achieved in nearly all cases. No<br>product-related adverse effects were noted in any of the<br>procedures. |
| Conclusions<br>Drawn: | Permacol™ is biocompatible, safe and effective for its |
intended use.
\\\DC - 59177/1 - 0790126.01
The Wren Group on Behalf of Tissue Science Laboratories, PLC Aldershot, UK
:
.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized symbol that resembles three overlapping lines that form a shape similar to a bird in flight or a stylized human figure.
#### Public Health Service
# FEB 1 7 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Tissue Science Laboratories, PLC c/o Mr. Howard M. Holstein, Esq. Hogan & Hartson, LLP 555 Thirteenth Street, N.W. Washington, D.C. 20004
K992556 Re: Trade Name: Permacol™ Regulatory Class: II Product Code: FTL Dated: December 22, 1999 Received: December 22, 1999
### Dear Mr. Holstein:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Howard M. Holstein, Esq.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Neil K.R. Ogden
ದ್ದಾಗಿ Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
| 510(k) Number (if known): | K992556 |
|---------------------------|-----------|
| Device Name: | Permacol™ |
Indications for Use:
Permacol™ is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes. It is specifically indicated for the repair of abdominal, inguinal, diaphragmatic, femoral, scrotal, umbilical and incisional hernias; colon, rectal, urethral and vaginal prolapse; muscle flap reinforcement; reconstruction of the pelvic floor and procedures such as sacrocolposuspension and urethral sling.
| | <i>NRO for JZD</i> |
|-----------------------------|--------------------|
| (Division Sign-Off | |
| Division of General Devices | |
| 510(k) Number | K992556 |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
| Prescription Use | |
|----------------------|--|
| Use | |
| (Per 21 CFR 801.109) | |
Over-The-Counter
(Optional Format 1 - 2.96) (Form prepared by The Wren Group)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.