GRAFT PATCH

K970561 · Organogenesis, Inc. · FTL · Aug 1, 1997 · General, Plastic Surgery

Device Facts

Record IDK970561
Device NameGRAFT PATCH
ApplicantOrganogenesis, Inc.
Product CodeFTL · General, Plastic Surgery
Decision DateAug 1, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The GraftPatch® is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists.

Device Story

GraftPatch® is a surgical mesh used for soft tissue reinforcement in general surgery, including hernia repairs. The device consists of heat-laminated layers of crosslinked porcine collagen. It is intended for use by surgeons in an operating room setting. The device provides mechanical support to weakened tissue areas. Substantial equivalence was established through descriptive comparison of design, materials, and intended use, supported by animal implant studies and biocompatibility testing.

Clinical Evidence

No human clinical data provided. Evidence consists of animal implant studies to confirm functionality and biocompatibility testing conducted in conformance with Tripartite and ISO 10993 recommendations.

Technological Characteristics

Material: heat-laminated layers of crosslinked porcine collagen. Form factor: surgical mesh. Biocompatibility: tested per Tripartite and ISO 10993 standards.

Indications for Use

Indicated for use in general surgical procedures for reinforcement of soft tissue where weakness exists, such as hernia repairs. Prescription use only.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K97C561 510(k) SUMMARY Organogenesis Inc. GraftPatch® AUG - 1 1997 ## 1. DATE PREPARED January 23, 1997 ## 2. SUBMITTER Organogenesis Inc. 150 Dan Road Canton, Massachusetts 02021 ## 3. CONTACT Joel T. Cademartori, P.E. Vice President of Regulatory Affairs, Quality Assurance and Control Organogenesis Inc. 150 Dan Road Canton, Massachusetts 02021 Telephone: (617) 575-0775 Facsimile: (617) 575-0440 ## 4. DEVICE NAME GraftPatch® ## 5. DEVICE CLASSIFICATION Class II (21 CFR 878.3300) Product Code: 79 FTM Organogenesis Inc. - GraftPatch® 510(k) 1/23/97 Page F-1 {1} # 6. DEVICE DESCRIPTION AND COMPARISON TO PREDICATE PRODUCTS GraftPatch® is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists, e.g., hernia repairs. The device is composed of heat-laminated layers of crosslinked porcine collagen. GraftPatch® is similar in design, materials, function, and intended use to other surgical meshes which are currently in U.S. commercial distribution. Examples of other surgical meshes which have been cleared for commercial distribution include the Supple Periguard® by Bio-Vascular, Inc. (K923657), and Vicryl surgical mesh by Ethicon, Inc. (K810218). Determination of substantial equivalence for this product was based on descriptive information about the design, materials, and intended use of the device. Animal implant studies were performed to confirm the functionality of the device. Biocompatibility testing was conducted on the device material in conformance with Tripartite and ISO 10993 recommendations. Results of all testing demonstrated that GraftPatch® Surgical Mesh is both suitable for its intended use and substantially equivalent to the predicate devices. Organogenesis Inc. - GraftPatch® 510(k) 1/23/97 Page F-2 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Joel T. Cademartori, P.E. Vice President of Regulatory Affairs Organogenesis Inc. 150 Dan Road Canton, Massachusetts 02021 Re: K970561 Trade Name: GraftPatch® Regulatory Class: II Product Code: FTL Dated: June 23, 1997 Received: Jun 25, 1997 AUG - 1 1997 Dear Mr. Cademartori: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for {3} Page 2 - Mr. Joel T. Cademartori, P.E. devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} 510(k) Number (if known): K970561 Device Name: **GraftPatch®** Indications For Use: The GraftPatch® is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) (Division Sign-Off) Division of General Restorative Devices 510(k) Number K970561 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96) Organogenesis Inc. - GraftPatch® 510(k) 1/23/97
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...