GORE-TEX SOFT TISSUE PATCH PLUS/DUALMESH PLUS BIOMATERIAL/MYCROMESH PLUS BIOMATERIAL/DUALMESH PLUS BIOMATERIAL WITH HOLE

K965038 · W. L. Gore & Associates, Inc. · FTL · Feb 6, 1997 · General, Plastic Surgery

Device Facts

Record IDK965038
Device NameGORE-TEX SOFT TISSUE PATCH PLUS/DUALMESH PLUS BIOMATERIAL/MYCROMESH PLUS BIOMATERIAL/DUALMESH PLUS BIOMATERIAL WITH HOLE
ApplicantW. L. Gore & Associates, Inc.
Product CodeFTL · General, Plastic Surgery
Decision DateFeb 6, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The devices are indicated for use in the reconstruction of hernias and soft/tissue deficiencies and for the temporary bridging of fascial defects.

Device Story

Inert, biocompatible expanded polytetrafluoroethylene (ePTFE) surgical mesh; loaded with antimicrobial preservative agents silver carbonate and chlorhexidine diacetate. Used by surgeons for hernia repair, soft tissue deficiency reconstruction, and temporary fascial defect bridging. Devices provide permanent or transient wound/defect support. Modification involves labeling update only; no changes to device design, materials, or performance characteristics.

Clinical Evidence

No clinical data; bench testing only. Substantial equivalence based on identical design and materials to previously cleared predicates.

Technological Characteristics

Expanded polytetrafluoroethylene (ePTFE) mesh; loaded with silver carbonate and chlorhexidine diacetate antimicrobial agents. No changes to materials, design, or performance characteristics from predicate devices.

Indications for Use

Indicated for patients requiring reconstruction of hernias, soft tissue deficiencies, or temporary bridging of fascial defects.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{1} K965038 FEB - 6 1997 # VI. Premarket Notification 510(k) Summary A. **Submitter:** W.L. Gore and Associates, Inc. P.O. Box 900 Flagstaff, AZ 86002-0900 **Contact:** R. Larry Pratt **Phone:** 520-779-2771 B. **Applicant Devices:** **Names of Devices with Modified Labeling** GORE-TEX® Soft Tissue Patch PLUS GORE-TEX® DualMesh PLUS Biomaterial GORE-TEX® MycroMesh PLUS Biomaterial GORE-TEX® DualMesh PLUS Biomaterial with Holes C. **Applicant Devices Description:** Inert, biocompatible, expanded polytetrafluoroethylene (ePTFE) loaded with the antimicrobial preservative agents silver carbonate and chlorhexidine diacetate. D. **Indications For Use:** The devices are indicated for use in the reconstruction of hernias and soft/tissue deficiencies and for the temporary bridging of fascial defects. E. **Predicate Devices:** The following devices, loaded with antimicrobial preservative agents, are cited as substantially equivalent predicate devices concerning possible adverse reactions, materials, design, biocompatibility, mechanical strength, manufacturing, packaging, sterilization, antimicrobial activity and preservative efficacy: GORE-TEX® Soft Tissue Patch PLUS GORE-TEX® DualMesh PLUS Biomaterial GORE-TEX® MycroMesh PLUS Biomaterial GORE-TEX® DualMesh PLUS Biomaterial with Holes page 10 of 11 {2} K965038 The following devices, without antimicrobial preservative agents, are cited as substantially equivalent predicate devices which contain the same indications for use statement: GORE-TEX® Soft Tissue Patch GORE-TEX® DualMesh Biomaterial GORE-TEX® MycroMesh Biomaterial GORE-TEX® DualMesh Biomaterial with Holes ## F. Technological Characteristics: This Premarket Notification submission is for a modification to the indications for use for the applicant devices. The applicant devices are not being changed in any way (except for labeling) as a result of this submission’s clearance. The materials, design and performance characteristics for the applicant devices have not changed from those of the predicate devices. There are no new technological characteristics due to the modification of the indications for use. ## G. Safety and Effectiveness Conclusions: This Premarket Notification concerns a modification to the indications for use for currently marketed, previously cleared surgical mesh devices. Therefore, the applicant devices are the same, and consequently, substantially equivalent to the predicate devices with regard to possible adverse reactions, materials, design, biocompatibility, preservative effectiveness, antimicrobial activity, mechanical strength, packaging, manufacturing process and sterilization process. The applicant devices and the predicate devices have the same intended use as prostheses for permanent or temporary wound or defect support. Each of the devices perform their equivalent clinical functions by incorporating biocompatible materials to permanently or transiently bridge or support a tissue defect. The modification of the indications for use for the applicant devices does not raise questions of safety or effectiveness that have not been previously addressed. GORE-TEX®, MycroMesh®, DualMesh®, Soft Tissue Patch PLUS™ are trademarks of W.L. Gore and Associates, Inc. page 11 of 11
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...