← Product Code [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL) · K962401

# GORE-TEX DUALMESH PLUS BIOMATERIAL WITH HOLES (K962401)

_W. L. Gore & Associates, Inc. · FTL · Sep 3, 1996 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K962401

## Device Facts

- **Applicant:** W. L. Gore & Associates, Inc.
- **Product Code:** [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL.md)
- **Decision Date:** Sep 3, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

An implantable surgical mesh prosthesis indicated for the reconstruction of hernias and soft tissue deficiencies.

## Device Story

Implantable surgical mesh prosthesis; composed of expanded polytetrafluoroethylene (ePTFE) with antimicrobial preservative agents; flat sheet configuration with macropores to optimize rapid tissue fixation. Used by surgeons for hernia repair and soft tissue deficiency reconstruction. Device integrates features of previously cleared predicate devices; mechanical strength verified via bench testing; provides structural support for tissue repair.

## Clinical Evidence

Bench testing only. Mechanical strength values (suture pull-out force, tensile force to break) compared to predicate devices; results demonstrate substantial equivalence.

## Technological Characteristics

Material: expanded polytetrafluoroethylene (ePTFE). Features: macropores for tissue fixation, antimicrobial preservative agents. Sterilization: same methods as predicates. Mechanical properties: suture pull-out force (7.7-8.2 kg/5 pins), tensile force to break (8.3 kg/0.5").

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- GORE-TEX® DualMesh PLUS Biomaterial
- GORE-TEX® DualMesh Biomaterial with Holes

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SEP 3 1996
K962401

VI. Premarket Notification 510(k) Summary

A. Submitted By: W.L. Gore &amp; Associates, Inc.
P.O. Box 900
Flagstaff, AZ 86002-0900

Contact: R. Larry Pratt
Regulatory Affairs
Phone: 520-779-2771
Fax: 520-779-3480

B. Device Name: GORE-TEX® DualMesh PLUS Biomaterial with Holes

C. Applicant Device Description:

Biocompatible, expanded polytetrafluoroethylene (ePTFE) with antimicrobial preservative agents in a flat sheet configuration of various length and width dimensions. Macropores are placed in the device in order to optimize rapid tissue fixation.

D. Indications For Use:

An implantable surgical mesh prosthesis indicated for the reconstruction of hernias and soft tissue deficiencies.

E. Predicate Devices:

Gore’s surgical mesh products, GORE-TEX® DualMesh PLUS Biomaterial and GORE-TEX® DualMesh Biomaterial with Holes, are cited as predicate devices which have been found to be substantially equivalent through the premarket notification process.

F. Technological Characteristics:

The applicant device integrates the clinically successful attributes of GORE-TEX® DualMesh PLUS Biomaterial and GORE-TEX® DualMesh Biomaterial with Holes into a single device through the manufacture of macropores into GORE-TEX® DualMesh PLUS Biomaterial.

page 14 of 16

{1}

The applicant device is manufactured using the same inert, biocompatible ePTFE material as the predicate devices. The antimicrobial preservative agents loaded into the applicant device are the same, and are loaded in the same amounts, as for the previously cleared GORE-TEX® DualMesh PLUS Biomaterial. The macropores manufactured into the applicant device have the same dimensions and are created using the same process as the macropores in the previously cleared GORE-TEX® DualMesh Biomaterial with Holes.

Mechanical strength values of the applicant device are substantially equivalent to the mechanical strength values of the predicate devices and are sufficient for its indicated uses.

|  Subject | Applicant Device | Mean | Std Dev  |
| --- | --- | --- | --- |
|  Suture Pull-Out Force (kg/5 pins) |  |  |   |
|  Low Rate |  | 7.7 | 0.5  |
|  High Rate |  | 8.2 | 0.7  |
|  Tensile Force to Break (kg/0.5") |  | 8.3 | 0.3  |

G. Safety and Effectiveness Conclusions:

The applicant device is indicated for use in clinical applications which have previously been cleared for the predicate devices.

The applicant device is manufactured of the same inert, biocompatible expanded PTFE and the same antimicrobial preservative agents as the previously cleared GORE-TEX® DualMesh PLUS Biomaterial.

Macropores are created in the applicant device utilizing the same manufacturing process and technology as used for creating macropores in the previously cleared GORE-TEX® DualMesh Biomaterial with Holes.

The packaging materials and process used for the applicant device do not differ from those used for the predicate devices. The applicant device is sterilized using the same sterilization methods and utilizes the same post-sterilization release criteria as the predicate devices.

{2}

The applicant device has been shown via bench testing to have mechanical strength values substantially equivalent to the mechanical strength values of the predicate devices.

Combining clinical features from each of the previously cleared predicate devices into GORE-TEX® DualMesh PLUS Biomaterial with Holes does not compromise the safety or effectiveness of the applicant device. The manufacturing of macropores into the predicate GORE-TEX® DualMesh PLUS Biomaterial does not raise new types of safety or effectiveness questions. The descriptive and mechanical strength information contained within this Premarket Notification document are precise enough to demonstrate the substantial equivalency of GORE-TEX® DualMesh PLUS Biomaterial with Holes.

GORE-TEX and DualMesh are trademarks of W.L. Gore &amp; Associates.

page 16 of 16

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K962401](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K962401)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
