K133198 · ETHICON, Inc. · FTL · May 30, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133198
Device Name
ULTRAPRO COMFORT PLUG
Applicant
ETHICON, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
May 30, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ULTRAPRO COMFORT PLUG™ Partially Absorbable Hernia Repair Device is indicated for reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as groin hernia defects.
Device Story
Sterile, single-use, partially absorbable hernia repair device for open extra-peritoneal groin hernia repair. Comprises 3D plug to fill defect and flat pre-shaped onlay patch to reinforce surrounding area. Plug constructed from undyed ULTRAPRO mesh with laminated dyed (D&C Violet No. 2) polydioxanone polymer film ribs for visibility/fixation. Mesh composed of approximately equal parts absorbable poliglecaprone 25 monofilament and non-absorbable polypropylene monofilament. Used by surgeons in clinical settings to provide permanent support during and after wound healing. Device provides mechanical reinforcement of soft tissue defects; visibility ribs facilitate insertion and fixation.
Clinical Evidence
No clinical data. Evidence consists of bench testing (dimensional measurement, shelf life) and animal model performance testing (mesh compression, tissue integration, tissue reaction, infection potentiation). Biocompatibility testing performed per ISO 10993-1:2009 (cytotoxicity, sensitization, systemic toxicity, pyrogenicity, subchronic toxicity, genotoxicity, local tissue response).
Technological Characteristics
Partially absorbable surgical mesh. Materials: poliglecaprone 25 monofilament, polypropylene monofilament, polydioxanone polymer film (dyed D&C Violet No. 2). 3D plug and flat onlay patch configuration. Sterilized, single-use. Complies with ISO 10993-1:2009 for biocompatibility.
Indications for Use
Indicated for reinforcement of soft tissue where weakness exists, specifically for groin hernia repair procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
## 510(K) SUMMARY
# MAY 3 0 2014
| Applicant: | Ethicon Inc.<br>P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876<br>USA<br>Phone: +1-908-218-3323<br>Fax: +1-908-218-2595 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Date: | May 12, 2014 |
| Contact Person: | Rey Librojo |
| Proprietary Device Name: | ULTRAPRO COMFORT PLUGTM<br>Partially Absorbable Hernia Repair Device |
| Common Device Name: | Surgical Mesh |
| Classification: | Class II<br>Regulation Number 878.3300 - Surgical Mesh, polymeric;<br>Product Code: FTL |
| Predicate Devices: | ULTRAPRO PLUG™ (ETHICON, Inc.) - K070224;<br>BARD Mesh Perfix™ Plug (Davol, Inc.) - K922916 |
| Manufacturer: | Johnson & Johnson MEDICAL GmbH<br>Robert-Koch-Strasse 1<br>22851 Norderstedt<br>Germany |
1 of 3
{1}------------------------------------------------
#### Description of the Device Subject to Premarket Notification:
ULTRAPRO COMFORT PLUG™ is a sterile, single-use partially absorbable hernia repair device designed for open extra-peritoneal abdominal wall repair, particularly groin hernia repair, to provide permanent support during and following wound healing. ULTRAPRO COMFORT PLUG™ consists of a three-dimensional plug, which fills the defect and a flat pre-shaped onlay patch, which lays on the defect to reinforce the weak area around the defect. The Plug is composed of an undyed ULTRAPRO™ Mesh. Dyed ribs, made from dyed (D&C Violet No. 2) polydioxanone polymer film, are laminated to the Plug mesh. The onlay patch is a pre-shaped, dyed ULTRAPRO™ Mesh. The ULTRAPRO™ Mesh is manufactured from approximately equal parts of absorbable poliglecaprone 25 monofilament fibers and non-absorbable polypropylene monofilament fibers. The device is available in different sizes.
### Indications for Use:
The ULTRAPRO COMFORT PLUG™ Partially Absorbable Hernia Repair Device is indicated for reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as groin hernia defects.
### Summary of Technological Characteristics of New Device to Predicate Devices:
#### Similarity:
The principle of operation and fundamental scientific technology of the proposed device are equivalent to the predicate devices. ULTRAPRO COMFORT PLUG™ and the predicate devices function in the same manner - they are designed as plug plus onlay patch devices for open extraperitoneal abdominal wall repair, particularly groin hernia repair. The polymer materials in the proposed device is identical to that of Ultrapro Plug, which is composed with approximately equal parts of absorbable poliglecaprone 25 monofilament fibers and non-absorbable, polypropylene monofilament fibers. The mash knitting structure of both plug and onlay patch of ULTRAPRO COMFORT PLUG™ is the same as the onlay patch material of ULTRAPRO Plug the proposed device has similar 3-D physical shapes as that of Bard Mesh Perfix Plug. Difference:
The plug material for proposed device is composed of undyed ULTRAPRO™ Mesh. Dyed ribs, made from dyed (D&C Violet No. 2) polydioxanone polymer film, are laminated to the Plug mesh to provide better visibility to the plug for ease of insertion into the defect and fixation.
### Performance Data:
ULTRAPRO COMFORT PLUG™, Partially Absorbable Hernia Repair Device underwent an extensive safety and performance testing program, including bench and animal testing, to support that the device meets the requirements as defined in user specifications, performs as intended, and is substantially equivalent to the predicate devices. The tests conducted include:
{2}------------------------------------------------
- Biocompatibility testing in accordance to the tests recommended in the ISO 10993-. 1:2009 standard including cytotoxicity, sentitization, acute systemic toxicity, pyrogenicity, subchronic toxicity, genotoxicity, and local tissue response and absorption study
- Bench testing including device dimensional measurement and comparison, and shelf life . testing
- Performance testing in animal model including efficacy study evaluating mesh . compression, tissue integration, and tissue reaction and infection potentiation study.
The ULTRAPRO COMFORT PLUG™ Partially Absorbable Hernia Repair Device has the same intended use, fundamental scientific technology, and principles of operation as its predicate devices. Performance data demonstrates that the device is as safe and as effective as the predicate devices for the intended use. Thus we conclude that the proposed device is substantially equivalent to the predicate devices under the Federal Food, Drug, and Cosmetic Act.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Asenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
May 30, 2014
Ethicon, Inc. Reynaldo Librojo Director, Regulatory Affairs Route 22 West, P.O. Box 151 Somerville, NJ 08876
Rc: K133198
Trade/Device Name: ULTRAPRO COMFORT PLUG Partially Absorbable Hernia Repair Device Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: May 22, 2014 Received: May 28, 2014
Dear Mr. Reynaldo Librojo,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enacument date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic room reat the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Librojo
CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
## David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Ethicon Inc. October 2013
#### INDICATIONS FOR USE STATEMENT 4
| 510(k) No (if known): | K133198 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | ULTRAPRO COMFORT PLUG <sup>TM</sup> |
| | Partially Absorbable Hernia Repair Device |
| Indications for Use: | The ULTRAPRO COMFORT PLUG <sup>TM</sup> Partially Absorbable Hernia Repair Device is indicated for reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as groin hernia defects. |
V Prescription Use __ (Part 2) CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Peter L. Hudson -S
Page 12 of 49
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.