← Product Code [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL) · K132441

# BARD VENTRALEX HERNIA PATCH (K132441)

_C.R. Bard, Inc. · FTL · Dec 13, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K132441

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL.md)
- **Decision Date:** Dec 13, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Bard® Ventralex™ Hernia Patch is intended for use in all forms of hernia repair requiring reinforcement with a nonabsorbable support material. The small Bard® Ventralex™ Hernia Patch (4.3 cm / 1.7") is also intended to repair soft tissue deficiencies, including deficiencies caused by trocars.

## Device Story

Self-expanding, sterile hernia repair prosthesis; consists of two polypropylene mesh layers stitched with PTFE monofilament to an ePTFE sheet. Features SorbaFlex™ Memory Technology: an absorbable polydioxanone (PDO) monofilament recoil ring encased in a knitted polypropylene mesh tube. Device provides structural support for hernia defects; ePTFE layer minimizes tissue attachment; polypropylene mesh promotes tissue in-growth. Includes positioning strap and pocket for surgical placement/fixation. Used by surgeons in clinical settings for hernia repair. PDO ring provides memory for insertion/placement, degrading via hydrolysis in 6-8 months. Depth marker on strap assists in laparoscopic trocar positioning; excess strap material is trimmed post-fixation.

## Clinical Evidence

No clinical data. Evidence consists of bench testing and animal model simulated use. Testing evaluated deployment integrity, recoil performance (flatness after folding), and biocompatibility per ISO 10993. Resorption data for the PDO monofilament ring was adopted from the predicate Ventrio Hernia Patch (K081777).

## Technological Characteristics

Materials: Polypropylene monofilament mesh, ePTFE sheet, PTFE monofilament thread, absorbable polydioxanone (PDO) monofilament recoil ring. Construction: Knitted mesh, heat-sealed edges. Principle: Self-expanding mechanical support. Sterilization: Sterile. Connectivity: None. Software: None.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Bard® Ventralex™ Hernia Patch ([K021736](/device/K021736.md))
- Bard® Ventralex™ Hernia Patch ([K024008](/device/K024008.md))
- Bard® Ventrio™ Hernia Patch ([K081777](/device/K081777.md))
- Bard® Ventrio™ Hernia Patch ([K100229](/device/K100229.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

## AUGUST 27, 2013

This 510(k) Summary is provided per the requirements of section 807.92(c).

## Submitter Information:

| Submitter's Name: | Davol, Inc., Subsidiary of C. R. Bard, Inc. |
|-------------------|---------------------------------------------|
| Contact Person:   | Mariya Buharin                              |
|                   | Regulatory Affairs Specialist               |
| Address:          | 100 Crossings Boulevard                     |
|                   | Warwick, RI 02886                           |
| Telephone:        | (401) 825-8729                              |
| Fax:              | (401) 825-8765                              |
| Email:            | marie.buharin@crbard.com                    |
| Device Name:      |                                             |

Trade Name:

Common/Usual Name: Classification Name: Classification Code:

- Surgical Mesh - Mesh, Surgical, Polymeric - Class II, § 878.3300, Product Code FTL

Bard® Ventralex™ Hernia Patch

#### Predicate Device Names:

- . Bard® Ventralex™ Hernia Patch,
	- K021736 (Davol, Inc.), FDA cleared on 07/16/2002 o
	- K024008 (Davol, Inc.), FDA cleared on 02/20/2003 o
- Bard® Ventrio™ Hernia Patch, .
	- K081777 (Davol, Inc.), FDA cleared on 09/29/2008 o
	- K 100229 (Davol, Inc.), FDA cleared on 04/21/2010 o

### Device Description:

The proposed Ventralex Hernia Patch is a self-expanding, sterile prosthesis, containing two primary layers of polypropylene mesh stitched with polytetrafluoroethylene (PTFE) monofilament to an expanded polytetrafluoroethylene (ePTFE) sheet. The mesh component is non-absorbable; however, the device contains a fully absorbable recoil ring using SorbaFlex™ Memory Technology, an absorbable polydioxanone (PDO) monofilament.

The mesh is constructed from knitted polypropylene monofilament and forms a strong, porous, support material. The top layer of mesh has a mesh strap and forms a positioning pocket with the second mesh layer. The propylene mesh side promotes tissue in-growth and repairs the defect. The strap and pocket facilitate placement, positioning and fixation of the device. For a reference point, the small Ventralex Hernia Patch is constructed with a depth marker on the positioning strap. The depth marker is polypropylene monofilament dyed with a blue colorant to indicate the position of the device relative to the end of the laparoscopic trocar sleeve. After fixation to the abdominal wall, excess positioning strap material above the fixation line and at the level of the fascia is cut off and discarded.

# PREMARKET NOTIFICATION FOR THE BARD® VENTRALEX™ HERNIA PATCH

{1}------------------------------------------------

A layer of ePTFE is stitched with PTFE monofilament thread and edge heat sealed to the polypropylene mesh. The ePTFE layer minimizes tissue attachment to the device. The SorbaFlex Memory Technology is comprised of an absorbable PDO monofilament welded to form a ring. The PDO monofilament recoil ring provides memory and stability to the device, facilitating ease of initial insertion, proper placement, and fixation of the device. The PDO monofilament fully degrades by means of hydrolysis in vivo in 6 - 8 months. The PDO monofilament ring is placed within a mesh tube constructed from a knitted polypropylene monofilament. The purpose of the mesh tube is to contain the PDO monofilament recoil ring during the degradation process. Intended Use:

The Ventralex Hernia Patch is intended for use in all forms of hernia repair requiring reinforcement with a non-absorbable support material. The small Bard Ventralex Hernia Patch (4.3 cm / 1.7") is also intended to repair soft tissue deficiencies, including deficiencies caused by trocars.

# Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use:

The proposed Ventralex™ Hernia Patch has the same indications for use, technological characteristics, and performance characteristics as the original Ventralex Hernia Patch, cleared via K021736 and K024008, with the exception of the material type of the recoil ring and the containment mesh tube in which it is encased.

In this submission. Davol proposes modifying the material type of the recoil ring used in the Ventralex Hernia Patch. In the proposed device, the non-absorbable PET recoil ring will be replaced with an absorbable PDO monofilament recoil ring encased in a containment mesh tube The proposed PDO monofilament ring and containment mesh tube are identical to the PDO monofilament ring and containment mesh tube used in the previously cleared Ventria Hernia Patch (K081777/K 100229).

## Performance Data:

Biocompatibility testing previously conducted for the Ventrio Hernia Patch (as presented via K081777) is also applicable to the proposed device and thus will not be repeated. All testing conducted to date (as per the requirements of ISO 10993) indicate that the device is biocompatible as per its intended use.

Bench testing results and in vivo simulated use experiments demonstrate that the proposed device design meets product specifications and intended uses In support of this submission, the proposed Ventralex Hernia Patch was deployed in a laparoscopic deployment model to evaluate the integrity of the PDO monofilament recoil ring and mesh tube when in a folded position. Additionally, the proposed Ventralex Hernia Patch underwent bench testing to ensure that the device recoiled to a flat position after folding. Whole system simulated use testing was conducted in an animal model to ensure that the proposed Ventralex Hernia Patch met key user needs. In vivo and in vitro resorption studies were performed and provided in support of the Ventrio Hernia Patch via K081777. These resorption studies were adopted for the proposed Ventralex Hernia Patch.

All test results provided in this submission support the safety and effectiveness of the device for its intended use and demonstrate that the proposed device is substantially equivalent to its predicates. .

## PREMARKET NOTIFICATION FOR THE BARD® VENTRALEX™ HERNIA PATCH

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

C.R. Bard Incorporated Ms. Mariya Buharin Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886

December 13, 2013

Re: K132441

Trade/Device Name: Bard® Ventralex™ Hernia Patch Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: November 15, 2013 Received: November 18, 2013

Dear Ms. Buharin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Mariya Buharin

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the · Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours. Joshua ©. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATION FOR USE STATEMENT

510(k) Number (if known):

Not known

Device Name:

-------------------

Bard® Ventralex™ Hernia Patch

The Bard® Ventralex™ Hernia Patch is intended for use in all forms of hernia repair requiring reinforcement with a nonabsorbable support material. The small Bard® Ventralex™ Hernia Patch (4.3 cm / 1.7") is also intended to repair soft tissue deficiencies, including deficiencies caused by trocars.

Prescription Use ਮ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# David Krause -S

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K132441

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K132441](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K132441)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
