K120025 · Aspide Medical · FTL · Jun 26, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120025
Device Name
SURGIMESH XB - SKIRTED
Applicant
Aspide Medical
Product Code
FTL · General, Plastic Surgery
Decision Date
Jun 26, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SURGIMESH® XB - Skirted mesh is used for the reinforcement of tissues during surgical repair. The SURGIMESH®XB-Skirted is indicated for the reconstruction of hernias and soft tissue deficiencies and for the temporary bridging of fascial defects. The silicone layer minimizes the tissue attachment to the mesh in the case of direct contact with the viscera.
Device Story
SURGIMESH®XB-Skirted is a non-absorbable, synthetic surgical mesh used for tissue reinforcement during hernia repair and soft tissue reconstruction. Device consists of a non-woven, non-knitted polypropylene layer (SURGIMESH®XB) and a skirted polypropylene component (SURGIMESH®WN), joined by a PVDF thread and coated on one side with silicone. Silicone layer serves to minimize tissue attachment when in direct contact with viscera. Device is supplied sterile in various anatomic forms for surgeon use. It functions as a mechanical scaffold for tissue repair. Clinical benefit derived from structural support of fascial defects and reduced visceral adhesion risk.
Clinical Evidence
Bench testing only. Biocompatibility (ISO 10993-1), sterilization (ISO 10993-7, 11137-1, 14937, USP 28), packaging (ISO 11607), and structural characterization (ISO 5084, 3801, 9073-3/4/7, 13934-1, 13938-1) confirmed device specifications and performance are substantially equivalent to predicates.
Technological Characteristics
Non-absorbable, synthetic mesh. Materials: Polypropylene (non-woven/non-knitted), silicone coating, PVDF thread. Form: Anatomic shapes. Sterilization: Validated per ISO 11137-1/14937. Standards: ISO 10993-1 (biocompatibility), ISO 11607 (packaging), ISO 5084/3801/9073/13934/13938 (structural characterization).
Indications for Use
Indicated for reconstruction of hernias, soft tissue deficiencies, and temporary bridging of fascial defects in patients requiring surgical tissue reinforcement.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Covidien PARIETEX COMPOSITE OPEN ("PCO") SKIRT Mesh (K110816)
Submission Summary (Full Text)
{1}------------------------------------------------
K120025 page 1/3
Image /page/0/Picture/22 description: The image shows the logo for Aspide Medical. The word "ASPIDE" is in a large, bold font, with the word "Medical" in a smaller font underneath. Above the word "ASPIDE" is a graphic of a grid of squares.
## PREMARKET NOTIFICATION 510(k) SURGICAL MESH: SURGIMESH®XB - SKIRTED
#### 510(k) Summary
JUN 2 6 2012
#### SURGIMESH®XB - Skirted
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
ASPIDE MEDICAL 246 allée Lavoisier 42350 LA TALAUDIERE (FRANCE) Tel: +33 4 77 53 16 59 Fax: +33 4 77 53 01 97
Contact Person; Mr, William Wiecek
Date Prepared: May 4, 2012
#### Name of Device and Name/Address of Sponsor
SURGIMESH®XB - Skirted
ASPIDE MEDICAL 246 allée Lavoisier 42350 LA TALAUDIERE (FRANCE)
#### Common or Usual Name
Polymeric Surgical Mesh
#### Classification Name
Surgical Mesh
#### Predicate Devices
(1) Apide Medical's SURGIMESH®XB (K072974); (2) Aspide Medical's SURGIMESH®WN (K061445); (3) Gore's DUALMESH (K992189); and (4) Covidien's PARIETEX COMPOSITE OPEN ("PCO") SKIRT Mesh (K110816).
### Intended Use / Indications for Use
The SURGIMESH® XB - Skirted mesh is used for the reinforcement of tissues during surgical repair.
The SURGIMESH®XB-Skirted is indicated for the reconstruction of hernias and soft tissue deficiencies and for the temporary bridging of fascial defects. The silicone layer minimizes the tissue attachment to the mesh in the case of direct contact with the viscera.
{2}------------------------------------------------
K120025 Page 2/3
Image /page/1/Picture/1 description: The image contains a logo with the word "ASPIDE" in a bold, sans-serif font, with the word underlined. Below "ASPIDE" is the word "Médical" in a smaller, italicized font. Above the word "ASPIDE" is a graphic consisting of a series of small, filled squares arranged in a diagonal line, creating a visual element above the text.
## PREMARKET NOTIFICATION 510(k) SURGICAL MESH: SURGIMESH®XB - SKIRTED
### Technological Characteristics
The SURGIMESH®XB - Skirted surgical mesh is a non-absorbable, synthetic mesh, made of non-knitted, non-woven fibers of polypropylene, one surface of which is coated with silicone. The use of SURGIMESH®XB - Skirted mesh provides reinforcement of soft tissues. On the opposite side, the silicone layer minimizes tissue attachment to the mesh in case of direct contact with the viscera. The SURGIMESH®XB - Skirted mesh is supplied sterile and is available in anatomic forms in order to meet individual surgeons' needs.
More specifically, the SURGIMESH® XB - Skirted mesh is composed of a layer of non-woven, non-knitted material made from polypropylene (i.e., SURGIMESH®XB), a skirted part made from polypropylene (i.e., SURGIMESH®WN) linked with the first layer by sewing a PVDF thread into the two parts with a layer of silicone (equivalent to SURGIMESH® XB product material).
#### Performance Data
Preclinical testing was conducted. Biocompatibility, product structure, and final product specifications were all tested. In all instances, the SURGIMESH®XB -Skirted functioned as intended and the results observed were as expected. Specifically, the company conducted the following performance testing:
- Biocompatibility testing in accordance with ISO 10993-1 standards ● were conducted and results demonstrated that the device is biocompatible per these standards;
- Sterilization validation testing in accordance with ISO 10993-7, ISO . 11137-1, ISO 14937, and USP 28 and results demonstrated that the device is sterile per these standards;
- Product packaging testing in accordance with ISO 11607 and results . demonstrated that the device packaging has the appropriate sealing characteristics:
- The device structure and product characterization testing was . performed in accordance with ISO 5084, ISO 3801, ISO 9073-3, ISO 9073-4, ISO 9073-7, ISO 13934-1 and ISO 13938-1 and results demonstrated the SURGIMESH®XB - Skirted specifications are substantially similar to the identified predicate device specifications.
### Substantial Equivalence
The SURGIMESH®XB - Skirted is substantially equivalent to: (1) Apide Medical's SURGIMESH®XB (K072974); (2) Aspide Medical's SURGIMESH®WN (K061445); (3) Gore's DUALMESH (K992189); and (4) Covidien's PARIETEX COMPOSITE OPEN ("PCO") SKIRT Mesh (K110816).
The SURGIMESH®XB - Skirted has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the SURGIMESH®XB - Skirted and its
{3}------------------------------------------------
K120025 page 3/3
Image /page/2/Picture/1 description: The image shows the logo for "ASPIDE Médical". The word "ASPIDE" is in a bold, sans-serif font, and there is a line underneath it. Below "ASPIDE" is the word "Médical" in a smaller font. Above the word "ASPIDE" is a graphic of a square made up of smaller squares.
## PREMARKET NOTIFICATION 510(k) SURGICAL MESH: SURGIMESH®XB - SKIRTED
predicate devices raise no new issues of safety or effectiveness. The SURGIMESH®XB - Skirted mesh's mechanical and material characteristics are substantially equivalent to its predicate devices. Thus, the SURGIMESH®XB -Skirted is substantially equivalent.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, clean design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
JUN 2 6 2012
Aspide Medical % Mr. John J. Smith Partner, Hogan Lovells US LLP 555 Thirteenth St, NW Columbia Square Washington, District of Columbia 20004-1109
Re: K120025
Trade/Device Name: SURGIMESH®XB-Skirted mesh Regulation Number: CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL, OXJ Dated: June 19, 2012 Received: June 19, 2012
#### Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. John J. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Eric L. Keith
£ / Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
120025
510(k) **Number (if known):**
Device Name: SURGIMESH®XB - Skirted mesh
Indications for Use:
The SURGIMESH®XB - Skirted mesh is intended for use in the reinforcement of tissues during surgical repair.
The SURGIMESH®XB-Skirted is indicated for the reconstruction of hernias and soft tissue deficiencies and for the temporary bridging of fascial defects. The silicone layer minimizes the tissue attachment to the mesh in the case of direct contact with the viscera
Prescription Use X (Per 21 C.F.R. 801.109) AND/OR
Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David Krane for UMM
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120025
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.