K101462 · Caldera Medical, Inc. · FTL · Aug 16, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101462
Device Name
ASCEND BLUE AC MESH; ASCEND BLUE PC MESH
Applicant
Caldera Medical, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Aug 16, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ascend® Blue AC and Ascend® Blue PC are indicated for repair of pelvic organ prolapse, including anterior, posterior, vaginal vault and uterine prolapse.
Device Story
Ascend® Blue Mesh is a surgical implant for pelvic organ prolapse repair. Available in anterior (AC) and posterior (PC) configurations, it consists of macroporous monofilament polypropylene warp knit mesh. The central portion features blue-tinted polypropylene fibers for intraoperative and postoperative visualization, while clear fibers form the remainder. The device includes polyethylene sleeves and sutures as a delivery mechanism, which are removed after implantation; only the mesh remains in the body. Used in inpatient or outpatient surgical settings by physicians, the device is implanted using reusable Caldera Medical introducers. The mesh provides direct support for the prolapsed organ, while lateral arms facilitate tension-free fixation. The blue tinting improves surgical visualization compared to the predicate device. Benefits include structural support for pelvic organs and improved intraoperative identification of the implant's central portion.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through bench testing, cadaver lab testing, and validation testing. Parameters assessed included mesh thickness, pore size, mesh density, flexural rigidity, tear resistance, burst strength, suture pullout, and pyrogen levels. Biocompatibility was evaluated per ISO 10993-1. The device also underwent ISTA Class C1 shipping challenge and sterilization validation.
Technological Characteristics
Materials: Macroporous monofilament polypropylene warp knit mesh, polyethylene sleeves, polyester braided suture. Features: Multi-zone construction with blue-tinted polypropylene fibers in the central portion. Energy: None (mechanical implant). Connectivity: None. Sterilization: SAL < 10^-6. Form factor: Pre-shaped, single-use implant. Color additive: 21 CFR 74.3045.
Indications for Use
Indicated for women requiring surgical repair of pelvic organ prolapse, including anterior, posterior, vaginal vault, and uterine prolapse.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Caldera Medical. The logo consists of a circular graphic on the left and the text "CALDERA MEDICAL" on the right. The graphic is a complex pattern of small circles or dots, creating a textured effect. The text is in a simple, sans-serif font.
K101462 Page 1/3
# 510(k) Summary
AUG 16 2010
Date of Summary:
· Submitter:
Address:
Contact:
Facsimile:
Phone:
June 16, 2010
Submitted by:
Caldera Medical, Inc. 28632 Roadside Drive, Suite 260 Agoura Hills, CA 91301 Vicki Gail, Manger QA/RA (818) 879-6555 x 102 (818) 276-8400
#### Device Name:
| Trade Names: | Ascend® |
|---------------------|----------|
| | Ascend® |
| | Ascend® |
| Common Name: | Surgical |
| Classification Name | Mesh, Su |
| Regulation No. | 878.3300 |
| Product Code | FTL |
Blue Mesh AC Blue Mesh PC Mesh rgical, Polymeric
510(k) Number K101462
Predicate Device: Ascend® Mesh (Caldera Medical, Inc.) K083722
Description of Device:
Ascend® Blue Mesh is designed to be used in the inpatient or outpatient surgery setting for use in women suffering from pelvic or accean and inpatient of outpatient of outpatible for used.
introducers, which are and in implanted using reusable Caldera Medical introducers, which are regal produced in is migranced using tedsable Calibra Medical
designess and for a transmission in Ascend® Blue Mesh is available in two different designs – one for anterior compartment defects (AC) and one for posterior compartment defects (PC).
Ascend® Blue Mesh AC design is composed of macroporous monofilament polypropylene warp knit mesh in a multi-zone construction, which includes blue polypropylene fibers in the central portion and clear polypropylene fibers on the remainder of the device. This formation improves the visualization of the central portion intraoperatively and postoperatively. The central portion provides direct support for the prolapsed organ, while the lateral arms will aid in delivery of the implant and provide tension free fixation. A visual reference of the product is provided below.
| Bladder neck end of mesh positioned at bladder neck | |
|----------------------------------------------------------------|------------------------|
| Superior Obturator arm | Superior Obturator arm |
| Inferior Obturator arm | Inferior Obturator arm |
| Apical arm | |
| Ascend <sup> o </sup> Blue AC mesh implant | |
{2}------------------------------------------------
K101462 Page 2/3
The Ascend® Blue product, PC configuration, consists of the same material components as the AC design: macroporous monofilament polypropylene warp knit mesh in a multizone construction, polyethylene sleeves, suture, and polyethylene tips.
Image /page/1/Figure/3 description: The image shows a diagram of an arm with a polyethylene sleeve cover and suture. The arm is depicted as a horizontal line with a thicker section in the middle. The polyethylene sleeve covers are shown on either side of the thicker section, and the suture is visible at the end of the arm. The image also includes the text "Arm" on the left side and "Arm with polyethylene sleeve covers and suture" on the right side.
Ascend® Blue PC mesh implant
Each arm is covered by a polyethylene sleeve to protect the arm and reduce tissue resistance during positioning. The suture, sleeves and tips are removed from the device once the unit has been surgically implanted, for this construction serves as a delivery mechanism only. The only component to remain within the body is a portion of mesh.
### Intended Use of Device:
The Ascend® Blue AC (anterior compartment) and Ascend® Blue PC (posterior compartment) are indicated for repair of pelvic organ prolapse, including anterior, vaginal vault and uterine prolapse.
### Technological Characteristics Comparison
The Ascend® Blue Mesh is a modification of the predicate mesh device comprised of the same knit pattern, sleeve and suture materials. The Ascend® Blue Mesh has the same intended use and does not change the fundamental scientific technology of the predicate device.
| feature | Ascend® Mesh, K083722 | Ascend® Blue Mesh |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Ascend® AC and Ascend® PC are<br>indicated for repair of pelvic organ<br>prolapse, including anterior,<br>posterior, vaginal vault and uterine<br>prolapse. | same |
| Design/ Materials | Pre-shaped, sterile, single use<br>implant consisting of knitted large-<br>pore monofilament polypropylene<br>clear mesh with clear polyester<br>heatshrink formed tips | Change: Central mesh is woven<br>with blue tinted polypropylene<br>yarn for improved visualization.<br>Change: Color coded molded tips<br>replace heat-shrink formed tips.<br>Mesh weave pattern is unchanged. |
| Mesh color additive | none | 21 CFR 74.3045 |
| Sterility | SAL $< 10-6$ | same |
| Operating Principle | Supporting the pelvic region for<br>treatment of pelvic organ prolapse | same |
| Shelf Life | 60 months | 18 months |
{3}------------------------------------------------
K101462 Page 3/3
Image /page/2/Picture/1 description: The image shows the words "CALDERA MEDICAL" in a stylized font. The letters are composed of small dots, giving the text a textured appearance. The overall impression is of a logo or brand name, possibly for a medical company.
| feature | Ascend® Mesh, K083722 | Ascend® Blue Mesh |
|---------------------------|----------------------------------------------------------------------------------------------------------|-------------------|
| Packaging and<br>labeling | Tyvek inner and outer pouch,<br>packaging card, IFU, labeling,<br>packaging box and tamper-proof<br>seal | same |
| Product<br>Configuration | AC - anterior compartment<br>PC - posterior compartment | same |
| Design | Pre-shaped, sterile, single use<br>product | same |
| Mesh | Polypropylene monofilament | same |
| Sleeve | Polyethylene | same |
| Suture | Polyester braided suture,<br>size 0 | same |
| Junction | Polyester | PE |
#### Performance Summary
In accordance with FDA's Guidance for the Preparation of a Premarket Notification Application for Surgical Mesh, the results of bench, cadaver lab and validation testing has shown the Ascendo Blue Mesh to be equivalent to the predicate device.
The results of bench and cadaver testing demonstrated equivalent mesh characteristics, junction strength, and device function based upon its intended use to the predicate device.
The following parameters were assessed measured against the predicate device and found to be substantially equivalanet: Mesh thickness, pore size, mesh density area, flexural rigidity, tear resistance, burst strength, suture pullout, pyrogen levels.
In addition, the new Ascend® Blue Mesh has passed all testing requirements and demonstrated substantial equivalence to the predicate device in terms of aging, ISO 10993-1 tissue implant biocompatibility, ISTA Class C1 shipping challenge and sterilization.
### Summary of Substantial Equivalence
The Ascend® Blue Mesh is safe and effective for its intended use and is substantially equivalent to the predicate device. Ascend® Mesh, also a product of Caldera Medical.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a black and white drawing of a bird inside of a circle. The bird is facing to the right and has three lines that make up its body. The bird's head is on the left side of the circle, and its tail is on the right side. The circle is made up of small dots.
#### Public Health Service
Food and Drug Administration 10903 Now Hampshire Avenue Document Control Room - WO66-1 (6 Silver Spring, MI) 20993-0002
Caldera Medical, Inc. % Ms. Vicki Gail Manager, QA/RA 28632 Roadside Drive. Suite 260 Agoura Hills, California 91301-6099
Re: K101462
Trade/Device Name: Ascend® Blue Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: 11 Product Code: FTL Dated: July 20, 2010 Received: July 22, 2010
Dear Ms. Gail:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
ANG 1 6 Ping
{5}------------------------------------------------
Page 2 - Ms. Vicki Gail
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please the Colleged Marger for sont DA/CentersOffices/CDRH/Offices/ucm 1 5809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Prair.)
907 07) - F 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYour/Industry/default.htm.
Sincerely yours.
ely yours,
For
N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Caldera Medical. The logo consists of a circular graphic on the left and the text "CALDERA MEDICAL" on the right. The circular graphic appears to be a stylized representation of a flower or sunburst. The text is in all caps and appears to be a sans-serif font.
# Statement of Indications For Use
# Indications For Use
510 (k) Number (if known):
Device Name: Ascend® Blue
Indications for Use:
Ascend® Blue AC and Ascend® Blue PC are indicated for repair of pelvic organ prolapse, including anterior, posterior, vaginal vault and uterine prolapse.
Prescription Use -- X --(Part 21 CFR 801 Subpart D) AND/OR
Over the Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dail Knane for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101462
Ascend® Blue Caldera Medical, Inc. Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.