K100229 · C.R. Bard, Inc. · FTL · Apr 21, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100229
Device Name
VENTRIO HERNIA PATCH
Applicant
C.R. Bard, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Apr 21, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ventrio Hernia Patch is indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias.
Device Story
Ventrio Hernia Patch is a sterile, single-use surgical mesh for soft tissue reconstruction and hernia repair. Device consists of two layers of monofilament polypropylene mesh stitched to an ePTFE sheet, forming a positioning pocket. An absorbable polydioxanone (PDO) recoil ring, utilizing AbsorbaFlex Memory Technology, provides structural memory and stability to facilitate insertion, placement, and fixation. The PDO ring fully absorbs in vivo within 6-8 months. Used by surgeons in clinical settings for hernia repair; the polypropylene mesh promotes tissue in-growth, while the ePTFE layer minimizes tissue attachment. The device is deployed during surgery to provide mechanical reinforcement to the abdominal wall or other soft tissue defects.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing and a 4-week post-implantation porcine model study. Bench testing included linear verification, recoil, ball burst strength, suture pullout, laparoscopic deployment, and PDO containment. The porcine study evaluated tissue attachment, mesh contracture, tissue in-growth, and host inflammatory/fibrotic response, demonstrating performance equivalent to the predicate.
Technological Characteristics
Materials: monofilament polypropylene mesh, ePTFE sheet, polydioxanone (PDO) monofilament (dyed with D&C Violet No. 2). Principle: mechanical reinforcement of soft tissue. Form factor: multi-layer patch with absorbable recoil ring. Sterilization: sterile, single-use. Connectivity: none. Software: none.
Indications for Use
Indicated for reconstruction of soft tissue deficiencies, such as hernia repair, in patients requiring surgical mesh.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
APR 2 1 2010
K100229
1/3
# 8. 510(k) Summary
# 510(k) SUMMARY
This 510(k) Summary is provided per the requirements of section 807.92(c).
| 510(k) Number: | TBD |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Owner Name: | Davol Inc., Subsidiary of C. R. Bard, Inc. |
| Address: | 100 Crossings Boulevard<br>Warwick, RI 02886 |
| Contact Person: | Michelle Godin<br>Regulatory Affairs Associate<br>Tel: 401-825-8589<br>Fax: 401-825-8765<br>Email address: michelle.godin@crbard.com |
| Alternate Contact: | Thomas Hutchinson<br>Vice President, Quality Assurance, Regulatory and Clinical<br>Affairs<br>Tel: 401-825-8409<br>Fax: 401-825-8765<br>Email address: tom.hutchinson@crbard.com |
| Date Prepared: | January 25, 2010 |
| Device Trade Name: | Ventrio Hernia Patch |
| Device Common Name: | Surgical Mesh |
| Class: | Class II, 21 CFR 878.3300, Product Code FTL |
Predicate Device(s):
- Ventrio Hernia Patch (K081777) .
# Device Description:
The Ventrio Hernia Patch is a sterile, single use device indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. This device contains two primary layers of monofilament polypropylene mesh stitched with PTFE
{1}------------------------------------------------
monofilament to an ePTFE sheet to form a positioning pocket. The device also contains an absorbable recoil ring using AbsorbaFlex Memory Technology, which provides memory and stability to the device and facilitates ease of initial insertion, proper placement, and fixation of the device. The AbsorbaFlex Memory Technology is comprised of an absorbable polydioxanone (PDO) monofilament, which fully absorbs in vivo between 6-8 months. The polydioxanone monofilament is dyed violet using D & C Violet No. 2.
## Indication for Use:
The Ventrio Hernia Patch is indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias.
#### Comparison to Predicate Devices:
The Ventrio Hernia Patch has the same intended use and fundamental scientific technology as the predicate device. All technological characteristics of the Ventrio Hernia Patch are the same as the predicate device including packaging, biocompatibility, sterilization, and labeling. The technological characteristics include polypropylene mesh layers to allow for tissue in-growth, an ePTFE layer to minimize tissue attachment, and a PDO monofilament ring to provide stability to the device. Two minor differences exist between the predicate device and the proposed device. The first difference is that the anterior polypropylene slit mesh layer for all product sizes is now constructed of a lighter weight, larger pore knitted polypropylene monofilament which is identical in the polypropylene material composition. The knitted polypropylene monofilament differs in diameter (.006 in to .004 in) and knit design (pore size and knit pattern) only. The second difference is that the PDO monofilament ring is contained within a knitted polypropylene mesh tube for the 4 larger sizes while the PDO monofilament ring is contained between two stitched layers of knitted polypropylene mesh for the small sizes. The purpose of the containment mesh is to contain the PDO ring during the degradation process; there is no difference in functionality of the polypropylene mesh tube versus the two stitched layers of knitted mesh. The knitted polypropylene mesh tube and the two stitched layers of knitted polypropylene mesh are the same material composition and only differ in
{2}------------------------------------------------
K106229
3/
monofilament diameter (.006 in to .005 in) and knit design (pore size and knit pattern). Where minor technological differences exist between the proposed device and the predicate device, performance testing demonstrates that these differences do not adversely affect the safety and effectiveness of the proposed device.
## Summary of Non-Clinical Testing:
Bench testing and animal studies were performed on the Ventrio Hernia Patch to support substantial equivalence. Bench testing on the proposed device included linear verification, visual inspections, recoil testing, ball burst strength testing, suture pullout testing, laparoscopic deployment testing, PDO containment testing, and mesh/fixation holding strength. Predicate device testing included PDO weld and monofilament testing, seal strength testing, package qualification testing, and biocompatibility testing as there were no PDO, packaging, or material changes to the device. A 4 week post-implantation study in a porcine model was completed on the predicate device and proposed device to evaluate tissue attachment, mesh contracture, tissue in-growth, and host inflammatory and fibrotic response for both devices. Results of testing demonstrate that the Ventrio Hernia Patch design meets product specifications and intended uses.
## Statement of Equivalence:
The Ventrio Hernia Patch has the same indications for use and technological characteristics as the predicate device. Based on this and the design and engineering data provided in the Premarket Notification, the proposed Ventrio Hernia Patch has been shown to be substantially equivalent to the cleared Ventrio Hernia Patch.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized image of an eagle. The eagle is depicted with its wings spread, symbolizing protection and service, and is a common emblem used by the U.S. government.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR 2 1 2010
C.R. Bard, Inc. % Davol, Inc. Ms. Michelle Godin, MS, RAC 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K100229
Trade/Device Name: Ventrio Hernia Patch Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL Dated: March 19, 2010 Received: March 22, 2010
### Dear Ms. Godin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Michelle Godin, MS, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Barbara Bryson
Mark N. Melker Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 7. Statement of Indications for Use
#### Indications for Use
510(k) Number (if known):
Device Name: Ventrio Hernia Patch
#### Indications for Use:
The Ventrio Hernia Patch is indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C) AND/OR
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
for DXK
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K100229
Special 510(k) Submission for the Bard Ventrio Hernia Patch
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.