K093799 · Tornier, Inc. · FTL · Apr 22, 2010 · General, Plastic Surgery
Device Facts
Record ID
K093799
Device Name
TORNIER SURGICAL MESH
Applicant
Tornier, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Apr 22, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Tornier® Surgical Mesh is intended for use where temporary wound support is required to reinforce soft tissues where weakness exists or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. Tornier® Surgical Mesh is also intended for reinforcement of soft tissues, in conjunction with sutures and/or suture anchors during tendon repair surgery; including the reinforcement of rotator cuff, patellar, Achilles, biceps and quadriceps tendons. Tornier® Surgical Mesh is not intended for use as a replacement for normal body structures or to provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, or quadriceps tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide mechanical strength for the tendon repair.
Device Story
Tornier Surgical Mesh is a bi-layer, synthetic, absorbable, woven fabric composed of poly(4-hydroxybutyrate) fibers. Designed for surgical reinforcement of soft tissues where weakness exists or for hernia/fascial defect repair. Used in tendon repair surgeries as a reinforcing material in conjunction with sutures and bone anchors. Supplied sterile for single-patient use; cut to desired shape/size by surgeon. Does not provide full mechanical strength for tendon repair; primary mechanical support remains provided by sutures and anchors. Benefits include temporary wound support and reinforcement of weakened tissues.
Clinical Evidence
Bench testing only. No clinical data provided. Biocompatibility testing conducted per ISO 10993 and performance testing consistent with FDA guidance for surgical mesh.
Technological Characteristics
Bi-layer, synthetic, absorbable, woven fabric. Material: poly(4-hydroxybutyrate) fibers. Form factor: cut-to-size mesh. Sterilization: sterile. No software or electronic components.
Indications for Use
Indicated for patients requiring temporary soft tissue reinforcement for hernia repair, fascial defects, or as an adjunct to sutures/anchors in tendon repair (rotator cuff, patellar, Achilles, biceps, quadriceps). Not for use as a structural replacement for tendons.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
K093799
. 1 of 2
#### 510(k) SUMMARY (Per 21 CFR 807.92)
## General Company Information
4.1
# APR 2 2 2010
- Tornier, Inc. Name: Contact: Howard Schrayer Regulatory Affairs Consultant
- 3601 W. 76th Street Address: Suite 200 Edina, MN 55435
- Telephone: (978) 232 - 9997 (978) 232 - 9998 Fax:
- Date Prepared December 9, 2009
#### General Device Information
- Product Name: Tornier® Surgical Mesh
- Classification: "Absorbable Surgical Mesh" Product code: FTL - Class II
# Predicate Device
| Tepha, Inc. | TephaFLEX™ Surgical Mesh<br>[510(k) Number K070894] |
|----------------|--------------------------------------------------------------------------------------|
| Artimplant AB | Artelon® SportMesh™ Tissue Reinforcement<br>Surgical Mesh<br>[510(k) Number K071887] |
| LifeCell Corp. | LTM-RC Surgical Mesh<br>[510(k) K071986] |
{1}------------------------------------------------
#### Description
Tornier® Surgical Mesh is a bi-layer, synthetic absorbable reinforced woven fabric made from poly(4-hydroxybutyrate) fibers. The construction permits the mesh to be cut into any desired shape or size without unraveling. The device is supplied sterile, for single-patient use for the reinforcement of soft tissue where weakness exists.
#### Intended Use (Indications)
Tornier® Surgical Mesh is intended for use where temporary wound support is required to reinforce soft tissues where weakness exists or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
Tornier® Surgical Mesh is also intended for reinforcement of soft tissues, in conjunction with sutures and/or suture anchors during tendon repair surgery; including the reinforcement of rotator cuff, patellar, Achilles, biceps and quadriceps tendons.
Tornier® Surgical Mesh is not intended for use as a replacement for normal body structures or to provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, or quadriceps tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide mechanical strength for the tendon repair.
## Substantial Equivalence
This submission supports the position that the Tornier™ Surgical Mesh System is substantially equivalent to previously cleared devices, including those listed above. Substantial equivalence has been demonstrated by conduct of applicable biocompatibility testing as outlined in ISO10993. In addition, Tornier has conducted testing consistent with the requirements of the Guidance Document entitled: "Guidance for the Preparation of a Premarket Notification for a Surqical Mesh" (March 2. 1999) A number of predicate devices list the same range of clinical uses.
#### Conclusions
Tornier, Inc. believes that the information provided establishes that similar legally marketed devices have been used for the same clinical applications as the Tornier™ Surgical Mesh System. The materials from which the Tornier device is fabricated have an established history of use, and the devices have been tested in accordance with applicable FDA guidelines.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and appears to be in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# APR 22 2010
Tornier, Inc. % Mr. Howard L. Schrayer 100 Cummings Center, Suite 444C Beverly, Massachusetts 01915
Re: K093799
Trade/Device Name: Tornier® Surgical Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL Dated: April 14, 2010 Received: April 16, 2010
Dear Mr. Schrayer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Howard L. Schrayer
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark N. Melk Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K093799
#### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: Tornier® Surgical Mesh
Indications For Use:
Tornier® Surgical Mesh is intended for use where temporary wound support is required to reinforce soft tissues where weakness exists or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
Tornier® Surgical Mesh is also intended for reinforcement of soft tissues, in conjunction with sutures and/or suture anchors during tendon repair surgery; including the reinforcement of rotator cuff, patellar, Achilles, biceps and quadriceps tendons.
Tornier® Surgical Mesh is not intended for use as a replacement for normal body structures or to provide the full mechanical strength to support the rotator cuff, patellar, Achilles, biceps, or quadriceps tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide mechanical strength for the tendon repair.
Prescription Use (Part 21 CFR 801 Subpart D) ANDIOR
over-The-Counter Use 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE .. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for DKK
Dististince of broical, Orthopedic, and Restor. Tive Devices
510(k) Number: K093799
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.