← Product Code [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL) · K070018

# UNIVERSAL SURGICAL MESH, MODEL 100 (K070018)

_Generic Medical Device, Inc. · FTL · Apr 27, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K070018

## Device Facts

- **Applicant:** Generic Medical Device, Inc.
- **Product Code:** [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL.md)
- **Decision Date:** Apr 27, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Universal Surgical Mesh is intended for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

## Device Story

Universal Surgical Mesh is a nonabsorbable synthetic surgical mesh; constructed of knitted filaments of extruded polypropylene; provides strength, durability, and bi-directional elasticity for adaptation to body stresses; porous structure facilitates tissue in-growth; cuttable to desired shape/size without unraveling; used by surgeons in clinical settings for hernia or fascial defect repair; provides reinforcement or bridging to support tissue healing.

## Clinical Evidence

No clinical data; no non-clinical laboratory studies conducted.

## Technological Characteristics

Knitted filaments of extruded polypropylene; bi-directional elastic construction; porous; nonabsorbable synthetic material.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- PROLENE Polypropylene Mesh Nonabsorbable Synthetic ([K962530](/device/K962530.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

APR 2 7 2007

# 510(K) Summary

## Introduction

This 510(k) summary documentation is intended to comply with requirements of 21 CFR § 807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.

## 510K Submitted By

Generic Medical Device, Inc. 2201-34th Ave NW Gig Harbor, WA 98335

## USA Contact Person

Monica R. Montanez, RAC, CQA Consultant Tel: 713-523-9630 / Fax: 713-521-2770

## Date Prepared

December 28, 2006

## Trade Name of Device

Universal Surgical Mesh

#### Proprietary Name:

Universal Surgical Mesh Common Name:

Surgical Mesh

## Classification Name:

Surgical Mesh Class II: General & Plastic Surgery 21 CFR 878.3300, Product Code FTL.

Predicate Devices: K962530- PROLENE Polypropylene Mesh Nonabsorbable Synthetic Manufactured by: Ethicon, Inc. Somerville, New Jersey 08876-0151

510(k) Classification Class II

{1}------------------------------------------------

## Device Description

Universal Surgical Mesh is constructed of knitted filaments of extruded polypropylene identical in composition to that used in PROLENE Polypropylene Mesh Nonabsorbable Synthetic Surgical Mesh (ETHICON, INC). The mesh affords excellent strength, durability and surgical adaptability, with sufficient porosity for necessary tissue in-growth.

Universal Surgical Mesh in knitted by a process that interlinks each fiber junction and provides for elasticity in both directions. This construction permits the mesh to be cut into any desired shape or size without unraveling. The bi-directional elastic property allows adaptation to various stresses encountered in the body.

## Intended Use

The Universal Surgical Mesh is intended for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

## Comparison to Predicate Devices

Universal Surgical Mesh has the same intended use and same technological characteristics as the predicate device PROLENE Polypropylene Mesh Nonabsorbable Synthetic.

## Clinical/Non-Clinical Studies

The company did not conduct, nor depend on, clinical studies or non-clinical laboratory studies in order to establish substantial equivalence as there is no change to the clinical intended use as compared to the two predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Generic Medical Device, Inc. % General Medical Device Company Ms. Monica R. Montanez, RAC, CQA Consultant 3906 Roseland Street Houston, Texas 77006

APR 2 7 2007

Re: K070018

Trade/Device Name: Universal Surgical Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL Dated: April 14, 2007 Received: April 17, 2007

Dear Ms. Montanez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

## Page 2 - Ms. Monica R. Montanez, RAC, CQA

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Universal Surgical Mesh

510(k) Number (if known):

Device Name: Universal Surgical Mesh

Indications For Use: The Universal Surgical Mesh is indicated for the repair of hemia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Opl

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number 1000718

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K070018](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K070018)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
