TIMESH ALSO KNOWN AS TIMESH-TC, MODELS 6000001 & 6000004
K031225 · Gfe Medizintechnik GmbH · FTL · Sep 29, 2003 · General, Plastic Surgery
Device Facts
Record ID
K031225
Device Name
TIMESH ALSO KNOWN AS TIMESH-TC, MODELS 6000001 & 6000004
Applicant
Gfe Medizintechnik GmbH
Product Code
FTL · General, Plastic Surgery
Decision Date
Sep 29, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TiMESH/TiMESH-TC is specifically indicated for laparoscopic and open surgery for the repair of direct or indirect inguinal, femoral, umbilical, and incisional hernias; rectal, vaginal, and apical prolapses; and parietal reinforcement of tissues and abdominal wall repair. The TiMESH is specifically indicated for open surgery for the repair of direct or indirect inguinal and femoral hernias, pariental reinforcement of tissues, and abdominal wall repair. The TiMESH has the same indications as a combination of the predicate devices. TIMESH is a prescriptive device and should only be used by a licensed physician.
Device Story
TiMESH/TiMESH-TC is a non-absorbable surgical mesh used for hernia repair and tissue reinforcement. Constructed from warp-knitted monofilament polypropylene fibers with a proprietary coating; the mesh provides strength, flexibility, and durability while facilitating tissue ingrowth. The material is elastic in all directions and can be cut to size without unraveling. Used by licensed physicians in laparoscopic or open surgical procedures. The device is implanted to provide structural support to the abdominal wall or pelvic floor, aiding in the repair of hernias and prolapses. It is supplied sterile via ethylene oxide gas.
Clinical Evidence
No clinical data. Evidence consists of bench testing conducted in accordance with the FDA Guidance Document for Surgical Mesh and biocompatibility testing per AAMI/ANSI/ISO 10993.
Technological Characteristics
Warp-knitted monofilament polypropylene mesh with proprietary coating. Sterilized via ethylene oxide gas. Complies with AAMI/ANSI/ISO 10993 for biocompatibility. Elastic in all directions; non-unraveling construction.
Indications for Use
Indicated for reinforcement of tissue during surgical repair, including laparoscopic and open repair of direct/indirect inguinal, femoral, umbilical, and incisional hernias; rectal, vaginal, and apical prolapses; and parietal reinforcement of tissues and abdominal wall repair. Prescription use only.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
# SEP 2 9 2003
## 510(k) Summary
K031225
page 1/3
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness is outlined below.
| Date Prepared: | April 12, 2003 (Revision 1 as of May 26, 2003) |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name of Submitter: | John E. Sawyer<br>Realistic Quality Solutions<br>C/O: GfE Medizintechnik GmbH<br>6254 South 1280 East<br>Salt Lake City, Utah 84121<br>(801) 685-7050 |
| | Realistic Quality Solutions has received authorization to submit this 510(k) on behalf of GfE Medizintechnik GmbH |
| Device Proprietary Name: | TiMESH also known as TIMESH-TC |
| Classification Name: | Surgical Mesh per 21 CFR 878.3300 |
| Common/Usual Names: | Mesh, Metallic<br>Mesh, Non-Metallic<br>Polypropylene (Absorbable Mesh)<br>Polypropylene, (Non-Absorbable) Mesh |
| Predicate Device: | The subject device has the same indications for use as the predicates. The main differences are; changes in the construction of the mesh and the coating of the mesh. |
| K00112<br>2 | Prolene Soft Mesh | Ethicon, Inc. |
|-------------|---------------------------------------------------|---------------------------------|
| K90565<br>5 | NonAbsorbable<br>Polypropylene Surgical<br>Mesh | United States<br>Surgical Corp. |
| K91552<br>6 | Coated SURGIPRO<br>Polypropylene Surgical<br>Mesh | United States<br>Surgical Corp. |
PAGE 2-1 Revised
10
{1}------------------------------------------------
#### 510(k) Summary (Continued)
K031225
Page 2/3
#### Device Description:
TiMESH, also known as TiMESH-TC, hereinafter known as TiMESH/TiMESH-TC" is a non-absorbable polypropylene mesh constructed of warp knitted fabric made from monofilament polypropylene fibers with all around proprietary coating. The TiMESH is constructed using a warp-knit (not woven, not knitted) process.
This construction along with the proprietary coating allows the TiMESH to have the strength, flexibility, as well as durability and surgical adaptability to allow for necessary tissue ingrowth.
The TiMESH with its current construction provides for elasticity in all directions. The construction of the mesh permits the use of any size mesh without it's unraveling. The wetability allows for the proper adaptation of the mesh to various stresses that can be encountered in the body. The product is stenlized by using ethylene oxide gas.
#### Intended Use:
The TiMESH/TiMESH-TC is specifically indicated for laparoscopic and open surgery for the repair of direct or indirect inguinal, femoral, umbilical, and incisional hernias; rectal, vaginal, and apical prolapses; and parietal reinforcement of tissues and abdominal wall repair. The TiMESH is specifically indicated for open surgery for the repair of direct or indirect inguinal and femoral hernias, pariental reinforcement of tissues, and abdominal wall repair. The TiMESH has the same indications as a combination of the predicate devices. TIMESH is a prescriptive device and should only be used by a licensed physician.
#### Indications Statement:
The TiMESH/TiMESH-TC The TiMESH, also known as TiMESH-TC, along with Ethicon Inc. Prolene Soft Mesh, U.S.S.C.'s NonAbsorbable Polypropylene Surgical Mesh and Coated SURGIPRO Polypropylene Surgical Mesh are intended to be used for the reinforcement of tissue during surgical repair. Thus, the TiMESH also known as TiMESH-TC mesh and all of the predicates have the same intended use. The TiMESH is specifically indicated for laparoscopic and open surgery for the repair of direct or indirect inguinal, femoral, umbilical, and incisional hernias; rectal, vaginal, and apical prolapses; and parietal reinforcement of tissues and abdominal wall repair.
#### PAGE 2-2 (Revised)
{2}------------------------------------------------
K031225
Page 3/3
## 510(k) Summary (Continued)
| <b>Indications Statement Cont'd:</b> | The TIMESH is specifically indicated for open surgery for the repair of direct or indirect inguinal and femoral hernias, parietal reinforcement of tissues, and abdominal wall repair. The TIMESH has the same indications as a combination of the predicate devices. TIMESH is a prescriptive device and should only be used by a licensed physician. |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Technological Characteristics:</b> | GfE Medizintechnik GmbH believes that the subject device is substantially equivalent to the predicate devices. The TIMESH/TIMESH-TC is constructed of the same materials as the predicate devices except for the coating. The Coating process allows the TIMESH to be more flexible than the predicates. |
| <b>Performance Data:</b> | The patient contact materials used in this device are common materials used in other medical devices as well as the predicate devices and have a long history of biocompatibility. This device complies with the requirements of AAMI/ANSI/ISO 10993, Biological Evaluation of Medical Devices". In addition, bench testing was conducted in accordance with the FDA Guidance Document "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh". |
| <b>Conclusions:</b> | Based on the information contained herein, we conclude that the changes are minor and that the subject device is substantially equivalent to the predicate devices. |
PAGE 2-3
Revised
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The text is written in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W(1)66-G609 Silver Spring, MI) 20993-0002
GfE Medizintechnik GmbH % Mr. John Sawyer President Realistic Quality Solutions 6254 South 1280 East Salt Lake City, Utah 84121
JUN 2 0 2012
Re: K031225
Trade/Device Name: TiMESH also known as TiMESH-TC Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL, OXJ Dated: August 8, 2003 Received: August 12, 2003
Dear Mr. Sawyer:
This letter corrects our substantially equivalent letter of September 29, 2003
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
### Page 2 - Mr. John Sawyer
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Jon
Mark N. Melkerson
Don No Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 510(k): K031225
TIMESH Device Name:
- The TiMESH is intended for the reinforcement of Indications for Use: tissue during surgical repair. Specifically, the TiMESH is indicated for laparoscopic and open surgery for the repair of direct inguinal hernias, indirect inguinal hernias, femoral hernias, umbilical hernias, incisional hernias, parietal reinforcement of tissues and abdominal wall repair.
Prescription Use X (Per 21 CFR 801 Subpart D)
AND/OR Over the Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kune for MM
(Division Sign-Off) Division of Surgical. Orthopedic, and Restorative Devices
510(k) Number K031225
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.