← Product Code [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL) · K021736

# BARD VENTRALEX HERNIA PATCH, MODELS 0010302, 0010301 (K021736)

_C.R. Bard, Inc. · FTL · Jul 16, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K021736

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [FTL](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL.md)
- **Decision Date:** Jul 16, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Ventralex Patch is intended for use in all hernia repairs requiring reinforcement with a nonabsorbable support material.

## Device Story

Ventralex Patch is a self-expanding, three-layer surgical mesh for hernia repair. Device consists of two layers of polypropylene mesh and one layer of expanded polytetrafluoroethylene (ePTFE). A monofilament PET polymer ring is captured between polypropylene layers for stability. Includes a temporary polypropylene positioning strap and pocket to facilitate placement and fixation; strap is removed and discarded post-fixation. Used by surgeons in clinical settings for hernia reinforcement. Device provides nonabsorbable support to the hernia site, aiding in tissue repair and stability.

## Clinical Evidence

Bench testing only. Biocompatibility testing confirmed materials are non-toxic and non-sensitizing. Laboratory testing demonstrated that the design and materials meet established specifications for consistent performance.

## Technological Characteristics

Three-layer surgical mesh. Materials: Polypropylene mesh, expanded polytetrafluoroethylene (ePTFE), monofilament PET polymer ring, PTFE monofilament stitching. Construction: Interlocking PTFE stitch pattern; peripheral edge of polypropylene mesh heat-sealed to ePTFE layer. Self-expanding design. Non-absorbable.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Bard Composix Kugel Mesh ([K021736](/device/K021736.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K02 1736

# JUL 1 6 2002

### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS FOR THE VENTRALEX PATCH

#### A. Submitter Information

Submitter's Name: Address:

Telephone: Fax: Contact Person: Date of Preparation: Davol, Inc. Subsidiary of C. R. Bard, Inc. 100 Sockanossett Crossroad Cranston, RI 02920 401-463-7000 ext. 2263 401-463-3845 Brian A. Kanerviko May 1, 2002

#### B. Device Name

Ventralex Patch

#### C. Predicate Device Name

Trade name: Bard Composix Kugel Mesh (Davol Inc.)

#### D. Device Description

The proposed Ventralex Patch is a self-expanding, three laver device. Two layers consist of polypropylene mesh. The top layer of polypropylene mesh forms a positioning strap and pocket. The purpose of the strap and pocket is to facilitate placement, positioning and fixation of the device. After the device has been properly placed and attached, the positioning strap must be removed and properly discarded. The monofilament PET polymer "ring" is captured between the two layers of polypropylene mesh and adds stability to the device enabling greater simplicity and assurance in the proper placement. The third layer of the device is a single layer of expanded polytetrafluoroethylene (ePTFE) that is attached to the polypropylene mesh with an interlocking PTFE stitch pattern. The peripheral edge of the polypropylene mesh is heat sealed to the ePTFE layer.

#### E. Intended Use

The Ventralex Patch is intended for use in all hernia repairs requiring reinforcement with a nonabsorbable support material.

### F. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use

{1}------------------------------------------------

The Ventralex Patch and the predicate Composix Kugel Mesh are both intended for the use in all forms of hernia repair requiring reinforcement with a nonabsorbable support material.

Technologically, the proposed device has the same the same materials, and similar physical attributes and manufacturing methods.

The proposed device is designed to form a strap and pocket. The predicate device is designed to form just a pocket. The strap will be made of polypropylene. The purpose of the strap is to facilitate placement, positioning and fixation of the device. After the device has been properly placed and attached, the positioning strap must be removed and properly discarded.

## G. Performance Data

Biocompatibility and bench testing have been completed and support the safety and effectiveness of the Ventralex Patch for its intended use.

As previously mentioned, the Composix Kugel predicate and the proposed Ventralex Patch devices use the same materials and similar manufacturing methods for production. The Ventralex Patch and the Composix Kugel Mesh have a single layer of expanded polytetrafluoroethylene (ePTFE) and is attached to the polypropylene mesh with an interlocking stitch using polytetrafluoroethylene (PTFE) monofilament. The peripheral edge of the polypropylene mesh is heat sealed to the ePTFE layer.

The biocompatibility test results show that the material used in the design and manufacture of the device are non-toxic and non-sensitizing to biological tissues consistent with their intended use. Laboratory test results demonstrate that the materials chosen and the design utilized in manufacturing the Ventralex Patch will meet the established specifications necessary for consistent performance during their intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 6 2002

C. R. Bard, Inc. Brian A. Kanerviko 100 Sockanossett Crossroad Cranston, Rhode Island 02920

Re: K021736

Trade Name: Ventralex Patch Regulation Number: 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: FTL Dated: May 23, 2002 Received: May 28, 2002

Dear Mr. Kanerviko:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 Mr. Brian A. Kanerviko

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

A. Styt Murdu

1. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

1021736

510(k) Number (if known):

Device Name: Ventralex Patch

Indications for Use:

•••••

Intended for the use in all forms of hernia repair requiring reinforcement with nonabsorbable support material.

(Please do not write below this line -- Continue on another page if needed)

..............................................................................................................................................................................

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use<br>(Per 21 CFR 801.109) |  |
|------------------------------------------|--|
|------------------------------------------|--|

OR

| Over-the Counter Use |  |
|----------------------|--|
|----------------------|--|

(Optional Format 1-2-96)
  

(Division Sign-Off)
Division of General, Restorative
and Neurological Devices

| Davol Inc.           | May 23, 2002  | CONFIDENTIAL |
|----------------------|---------------|--------------|
| 510(k) for Ventralex | 510(k) Number | K021736      |

009

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K021736](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/FTL/K021736)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
