SurgiCount+ System

K232250 · Stryker Instruments · PBZ · Jan 11, 2024 · General Hospital

Device Facts

Record IDK232250
Device NameSurgiCount+ System
ApplicantStryker Instruments
Product CodePBZ · General Hospital
Decision DateJan 11, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2750
Device ClassClass 2
AttributesAI/ML

AI Performance

OutputAcceptanceObservedDev DSTest DS
Hemoglobin (Hb) mass estimation on surgical spongesLimits of agreement between actual hemoglobin mass and measured Hb mass within ±1.99 g HbLimits of agreement fell within the acceptance limit of ±1.99 g HbNot specifiedNon-clinical (bench) testing comparing software measurements to known Hb mass determined by a reference assay
Sponge recognitionFailure rate (sum of false images and failed detection) less than or equal to 6.5%Failure rate of 0.19%Not specifiedNon-clinical (bench) testing using a representative 18x18 inch sponge type

Indications for Use

The SurgiCount+ System Software is a multifunctional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+ AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. The Triton AI configuration is intended to be used with surgical sponges, software, hardware and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entent Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites. The Triton QBL (Quantification of Blood Loss) configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items.

Device Story

SurgiCount+ System is a multi-functional SaMD application for blood loss estimation and surgical sponge management. Inputs include images of blood-containing sponges (via mobile device camera) and RFID signals from tagged sponges. AI algorithms analyze images to estimate hemoglobin mass and blood volume; RFID reader (902-928 MHz) tracks sponge counts and locations. System components include Apple iPad Pro, RFID-tagged sponges, RFID reader, and Bluetooth scale. Used in operating rooms by clinical staff to track absorbent items and quantify blood loss. Output displayed on mobile device assists clinicians in monitoring surgical blood loss and ensuring all sponges are accounted for post-procedure. Benefits include improved accuracy in blood loss estimation and reduced risk of retained surgical items.

Clinical Evidence

Bench testing only. Hb algorithm validation showed limits of agreement within ±1.99 g Hb. Sponge Recognition Algorithm (SRA) demonstrated 0.19% failure rate. Human factors validation with 15 nurses confirmed safety and effectiveness in clinical scenarios. EMC testing per IEC 60601-1-2 and wireless coexistence per ANSI C63.27/AAMI TIR 69 performed.

Technological Characteristics

SaMD product utilizing mobile device camera for image processing and UHF RFID (902-928 MHz) for sponge tracking. Includes Bluetooth-enabled scale and RFID reader. Software utilizes fixed AI algorithms for image analysis. Complies with IEC 60601-1-2, ANSI C63.27, and AAMI TIR 69.

Indications for Use

Indicated for use as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items in clinical environments. Intended for use with validated RFID-tagged surgical sponges, hardware, and accessory devices.

Regulatory Classification

Identification

An image processing device for estimation of external blood loss is a device to be used as an aid in estimation of patient external blood loss. The device may include software and/or hardware that is used to process images capturing externally lost blood to estimate the hemoglobin mass and/or the blood volume present in the images.

Special Controls

In combination with the general controls of the Food, Drug & Cosmetic Act, the Image processing device for estimation of external blood loss is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. Demonstration of the performance characteristics must include a comparison to a scientifically valid alternative method for measuring deposited hemoglobin mass. The following use conditions must be tested: (i) Lighting conditions; (ii) Range of expected hemoglobin concentrations; (iii) Range of expected blood volume absorption; and (iv) Presence of other non-sanguineous fluids ( *e.g.,* saline irrigation fluid).(2) Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by intended users of the device. (3) Appropriate analysis and non-clinical testing must validate the electromagnetic compatibility (EMC) and wireless performance of the device. (4) Appropriate software verification, validation, and hazard analysis must be performed. (5) Software display must include an estimate of the cumulative error associated with estimated blood loss values. (6) Labeling must include: (i) Warnings, cautions, and limitations needed for safe use of the device; (ii) A detailed summary of the performance testing pertinent to use of the device, including a description of the bias and variance the device exhibited during testing; (iii) The validated surgical materials, range of hemoglobin mass, software, hardware, and accessories that the device is intended to be used with; and (iv) EMC and wireless technology instructions and information.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 11, 2024 Stryker Instruments Ms. Patti Arndt Principal Specialist. Regulatory Affairs 1941 Stryker Way Portage, Michigan 49002 ## Re: K232250 Trade/Device Name: SurgiCount+ System Regulation Number: 21 CFR 880.2750 Regulation Name: Image Processing Device For Estimation Of External Blood Loss Regulatory Class: Class II Product Code: PBZ Dated: July 28, 2023 Received: November 13, 2023 Dear Ms. Arndt: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). # Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2024.01.11 11 -05:52:45 -05:52:45 -05:52:45 -05:00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232250 Device Name SurgiCount+ System ## Indications for Use (Describe) The SurgiCount+ System Software is a multifunctional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other absorbent items. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+ AI and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. The Triton AI configuration is intended to be used with surgical sponges, software, hardware and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entent Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites. The Triton QBL (Quantification of Blood Loss) configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items. | Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="border: 1px solid black;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="border: 1px solid black;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "stryker" in bold, black letters. The font is sans-serif and the letters are closely spaced together. A small registered trademark symbol is located to the right of the letter "r". # 510(k) Summary #### -Submitter Stryker Instruments 1941 Stryker Way Portage, MI 49002 Phone: 269-323-7700, 1-800-253-3210 Email: Kim.Wiersema@stryker.com #### II. Contact Person Ms. Patti L. Arndt Principal Specialist, Regulatory Affairs 1941 Stryker Way Portage, MI 49002 Phone: 269-888-0233 Email: Patti.Arndt@stryker.com Date Prepared: January 9, 2024 #### lll. Device | Name of Device: | SurgiCount+ System | |-----------------------|------------------------------------------------------------------------------------| | Common or Usual Name: | External Blood Loss Estimation System | | Classification Name: | Image Processing Device for Estimation of External<br>Blood Loss (21 CFR 880.2750) | | Regulatory Class: | II | | Product Code: | PBZ | | IV. Predicate Device | Pixel 3 Sponge System<br>K163507 | The predicate device has not been subject to a design related recall. No reference devices were used in this submission. #### V. Device Description The SurgiCount+ [SC+] System Software is a multi-functional software application that is intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges and other surgical substrates. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton Quantitative Blood Loss [QBL]. Additionally, combined workflows (SC+AI and SC+QBL) are provided for {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and the letters are closely spaced together. There is a registered trademark symbol to the right of the letter "r". The word is black and the background is white. use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. Two of the five software configurations (Triton Al and SC+Al) are Class II functions. The remaining configurations are Class I functions. The Class II Triton Al software configuration estimates the hemoglobin (Hb) mass contained on used surgical sponges using an AI algorithm that analyzes the image of each sponge. It also calculates an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The Triton Al software configuration's sponge counting functionality has been modified to enhance the product's surgical sponge counting/management functionality, compared to the predicate device. New workflow steps allow users to scan, identify, and count RFIDtagged surgical sponges and other absorbent items, and to locate missing surgical sponges inside the operating room and, noninvasively, in surgical sites. The SaMD product includes the following nonmedical device and Class I consumable and hardware accessories: a mobile device (Apple iPad Pro), RFID-tagged surgical sponges/absorbent items, an RFID reader, a bluetooth-enabled scale, and a stand (or optional wall mount) that houses the hardware accessories and connects the accessory devices to electrical power. #### VI. Indications for Use The SurgiCount+ System Software is a multifunctional software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponqes and other absorbent items. The system incorporates three distinct software configurations: Triton AI, SC+ Sponge Counting, and Triton QBL. Additionally, combined workflows (SC+ Al and SC+QBL) are provided for use in clinical environments that require both the tracking of absorbent surgical items and estimation of patient blood loss. The Triton Al configuration is intended to be used with surgical sponges, software, hardware, and accessory devices that have been validated for use with the application to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The configuration is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The SC+ Sponge Counting configuration is intended for use as an adjunctive technology for augmenting the manual process of counting, displaying, and recording the number of RFID-tagged absorbent articles used during surgical procedures, and providing a non-invasive means of locating RFID-tagged absorbent articles within an operating room and surgical sites. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are closely spaced together, and the word is in black. There is a registered trademark symbol to the upper right of the word. The background is white. The Triton QBL (Quantification of Blood Loss) configuration is intended to be used to record the weight of used surgical sponges and other absorbent items in order to calculate the quantity of fluid volume on the sponges/absorbent items. #### VII. Comparison of Technology Characteristics with the Predicate Device The Indications for Use Statement has been updated but is unchanged with respect to its representation of the product's intended use. New indications for use (i.e., associated medical conditions/diseases/patient populations) were not proposed. The modified product remains an adjunctive technology that is intended to be used to estimate external patient blood loss and in the management of surgical sponges. Kev changes include a product name change and an updated description of the modified SaMD product, neither of which impact the safety and effectiveness of the device. Essential device functions for the SaMD product include the processing of images for the estimation of external patient blood loss and the management of surgical sponges. The SurgiCount+ System uses the same technology as the predicate device to estimate blood loss; its sponge counting functionality, however, is enhanced. Image Processing for the Estimation of External Blood Loss: Using a qualified mobile device camera, the SaMD product instructs the user to scan/image blood-containing surgical sponges and other absorbent items used during a medical or surgical procedure. The Application then processes the images using proprietary Al Algorithms. The algorithms generate the same output for a given input (are fixed) and have been trained using machine learning techniques to recognize the sponges and to estimate Hb mass and blood loss volume on the imaged substrates. Sponge Counting and Management: The SurgiCount+ System's sponge counting function uses ultra-high frequency (UHF) RFID technology to scan, count, and locate RFID-tagged surgical sponges. The handheld, reusable, and wireless RFID reader emits an electromagnetic field that activates items labeled with passive RFID tags within its range. The reader's firmware uses a defined communication protocol, and its antenna transmits RF energy between 902-928 MHz. The Application collects and manages the data received from the reader. By comparing the RFID-tagged absorbent articles "counted in" to a surqical procedure to the articles "counted out," the Application is able to display the number and identity of any missing absorbent articles. The predicate software system "counts" sponges by displaying the number of sponges imaged by the device for the primary purpose of blood loss estimation. The predicate and subject devices are identical with respect to their intended use and essential performance. As shown in the following table, the key difference between the two devices is the augmentation of the predicate system's sponge counting and management functionality. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the first letter, which is uppercase. A registered trademark symbol is located to the right of the word. | Key Feature/<br>Characteristic | Predicate<br>Device<br>Pixel 3<br>Sponge<br>System<br>(K163507) | Subject<br>Device<br>SurgiCount+<br>System<br>(K232250) | Comparison | |---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Type: | SaMD<br>Product | SaMD<br>Product | Same | | Intended Use: | Image-based<br>estimation of<br>external<br>patient blood<br>loss and<br>surgical<br>sponge<br>management | lmage-based<br>estimation of<br>external<br>patient blood<br>loss and<br>surgical<br>sponge<br>management | Same | | Clinical Workflows | Single<br>Workflow | Multifunctional<br>Software<br>Device with<br>Five<br>Workflows | Same<br>Predicate device<br>functionality is<br>replicated or<br>augmented in the<br>modified device<br>(Triton Al and<br>SC+Al<br>Workflows). | | Essential Function | Estimation of<br>Hb/Blood<br>Volume on<br>Imaged<br>Surgical<br>Sponges | Estimation of<br>Hb/Blood<br>Volume on<br>Imaged<br>Surgical<br>Sponges | Same<br>The incorporated<br>Al Algorithm is<br>identical across<br>the two systems.<br>The subject<br>device meets the<br>same accuracy<br>specification as<br>developed and<br>required for the<br>predicate system. | | Key Feature/<br>Characteristic | Predicate<br>Device<br>Pixel 3<br>Sponge<br>System<br>(K163507) | Subject<br>Device<br>SurgiCount+<br>System<br>(K232250) | Comparison | | Implementation of<br>Sponge<br>Counting/Management<br>Functionality | Device<br>counts and<br>displays the<br>number of<br>imaged<br>sponges | System uses<br>RFID-tagged<br>surgical<br>substrates<br>and an RFID<br>bar code<br>reader to<br>count/manage<br>and locate<br>surgical<br>sponges | Similar<br>Modifications<br>augment/enhance<br>the basic sponge<br>counting<br>functionality<br>provided by the<br>predicate system | {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the "S", which is uppercase. A registered trademark symbol is located to the right of the "r" in "stryker". The text is black and the background is white. #### VIII. Performance Data Extensive software verification and validation activities were completed for the SurqiCount+ System. All test cases passed and fully traced to the product's Risk Analysis and Software Design Specification. Unit, regression, and system-level testing also was conducted and confirmed that the software functioned as designed. Cybersecurity design planning and testing ensured that the software was architected and designed to minimize cybersecurity threats. Non-clinical (bench) testing was conducted to validate the performance of the SurqiCount+ System and demonstrated compliance with the product's Special Controls. (1) Hemoglobin (Hb) Algorithm Performance Validation Testing was conducted to evaluate the accuracy of the Hb algorithm in estimating the Hb mass on surgical sponges, compared to the known Hb mass on each sponge as determined by a reference assay. It was shown that the limits of agreement between the actual hemoglobin mass and sHbL measured by the SurgiCount+ software fell within the acceptance limit of ±1.99 q Hb. (2) Sponge Recognition Algorithm (SRA) Performance Validation Test results demonstrated that the SRA had a failure rate (sum of false images and failed detection) of 0.19%, when used to detect a representative 18x18 inch sponge type. This was well below the acceptance criterion of less than or equal to 6.5%. {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase except for the "S", which is uppercase. A small registered trademark symbol is located to the right of the "r" in "stryker". The word is in black and the background is white. # (3) Electromagnetic Compatibility (EMC)Testing The SurgiCount+ System conforms with IEC 60601-1-2 and demonstrates electromagnetic compatibility (EMC) safety and effectiveness in the hospital environment. - (4) Wireless Coexistence Testing Wireless coexistence testing was conducted on the system in accordance with ANSI C63.27 and AAMI TIR 69. The SurgiCount+ System functioned as designed in the presence of Wi-Fi and Bluetooth interferers that were no closer than 30 cm away from the SurgiCount+ System's iPad. - (5) Human Factors Validation Fifteen nurses participated in a complete, end-to-end, usability validation of the SurqiCount+ System. The validation study included the SurgiCount+ System's critical tasks and use scenarios, the facilitators' evaluation of the users' performance working through clinically realistic scenarios, users' input on usability and safety of the system, and residual use-error risk analysis. The validation study concluded that the SurgiCount+ System is reasonably safe and effective for its intended users and use environment and that all use risks were effectively mitigated. No significant residual or new usability risks were found. #### IX. Conclusions The SurgiCount+ System has the same intended use as the predicate Triton Sponge System. Both the subject and predicate devices are used to estimate blood loss on used surgical sponges and to count/manage surgical substrates. The SurgiCount+ System shares the same technological characteristics as the predicate device and includes new UFH RFID technology that enhances the product's ability to identify, count, and locate surgical sponges. Software and non-clinical performance testing raised no new questions of safety or effectiveness and support the substantial equivalence of the modified device to its legally marketed predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...