K203550 · Activ Surgical, Inc. · OWN · Apr 8, 2021 · Gastroenterology, Urology
Device Facts
Record ID
K203550
Device Name
ActivSight Intraoperative Imaging System
Applicant
Activ Surgical, Inc.
Product Code
OWN · Gastroenterology, Urology
Decision Date
Apr 8, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
ActivSight is intended to provide real-time endoscopic fluorescence and near infrared imaging. ActivSight enables surgeons to visually assess vessels, blood flow, and related tissue perfusion using fluorescence and near infrared imaging, and at least one of the major bile ducts (cystic duct, common bile duct, or common hepatic duct) using fluorescence, all during minimally invasive surgery. Fluorescence imaging of biliary ducts with ActivSight is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for stand-alone use for biliary duct visualization.
Device Story
ActivSight is an intraoperative imaging accessory for existing laparoscopic systems. It consists of an imaging module (optics/sensing) attached between a third-party laparoscope and camera, a light engine for laser fluorescence/near-infrared illumination, and a light cable. It processes video input from the third-party camera and its own module, outputting to a surgical monitor. Used in surgical environments by surgeons to assess blood flow, tissue perfusion, and biliary anatomy. It utilizes ICG fluorescence and laser speckle contrast imaging (LSCI) to provide real-time visual feedback. The system aids surgeons in intraoperative decision-making by providing perfusion and vessel visualization, potentially improving surgical outcomes. It includes a sterile drape and calibration target for each use.
Clinical Evidence
Bench testing included basic safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), human factors/usability (IEC 62366-1), software lifecycle (ANSI AAMI IEC 62304), laser safety (IEC 60825-1), and endoscopic equipment safety (IEC 60601-2-18). Clinical evidence includes an animal study comparing ActivSight's speckle laser and ICG fluorescence modes against the predicate's ICG mode for visualizing blood flow and perfusion.
Technological Characteristics
Components: Imaging module (optics/CMOS sensor), light engine (laser diodes), Y-shaped bifurcated light cable, stainless-steel sterilization tray. Sensing: IR-sensitive camera, 852 nm laser for speckle imaging. Connectivity: Interoperable with third-party laparoscopes/cameras. Sterilization: Steam sterilization for light cable and tray; disposable sterile drape and calibration target. Software: Class II device, software lifecycle per ANSI AAMI IEC 62304.
Indications for Use
Indicated for patients undergoing minimally invasive surgery requiring real-time endoscopic fluorescence and near-infrared imaging for visual assessment of vessels, blood flow, tissue perfusion, and major bile ducts (cystic, common bile, or common hepatic). Not for stand-alone biliary duct visualization.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Stryker Infrared Fluorescence (IRF) Imaging System (K142310)
K231344 — ActivSight Intraoperative Imaging System · Activ Surgical, Inc. · Aug 2, 2023
K241361 — Arthrex Synergy Vision Endoscopic Imaging System · Arthrex, Inc. · Jul 25, 2024
K142310 — Stryker Infrared Fluorescence (IRF) Imaging System · Stryker Endoscopy · Dec 2, 2014
K242513 — FloNavi Endoscopic Fluorescence Imaging System (OPTO-CAM214K, OPTO-CHD214KE, OPTOCHD214KH, OPTO-LED214K) · Guangdong Optomedic Technologies, Inc. · Oct 23, 2024
K162882 — KARL STORZ ICG Imaging System · KARL STORZ Endoscopy-America, Inc. · Jan 26, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure embracing a sphere. The FDA acronym is in a blue square, and the full name of the agency, "U.S. Food & Drug Administration," is in blue text.
Activ Surgical, Inc. Matthew King Director of Regulatory Affairs and Quality Assurance 840 Summer Street Boston, Massachusetts 02127
April 8, 2021
Re: K203550
Trade/Device Name: ActivSight Intraoperative Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ, OWN, DPT, Dated: March 8, 2021 Received: March 9, 2021
Dear Matthew King:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/5 description: The image shows the name "Neil R.P. Ogden" in a large, bold font. To the right of the name, there is the text "Digitally s" followed by "Neil R.P. O" and "Date: 202" and "19:43:42 -". The text on the right appears to be part of a digital signature.
Digitally signed by Neil R.P. Ogden Date: 2021.04.08 19:43:42 -04'00'
Neil R.P. Ogden Assistant Director. THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K203550
Device Name ActivSight
#### Indications for Use (Describe)
ActivSight is intended to provide real-time endoscopic fluorescence and near infrared imaging. ActivSight enables surgeons to visually assess vessels, blood flow, and related tissue perfusion using fluorescence and nearing, and at least one of the major bile duct, common bile duct or common hepatic duct) using fluorescence, all during minimally invasive surgery.
Fluorescence imaging of biliary ducts with ActivSight is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for stand-alone use for biliary duct visualization.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Activ Surgical. The word "ACTIV" is in large, blue, sans-serif font. Below that, the word "SURGICAL" is in a smaller, light gray, sans-serif font. The logo is simple and modern.
### 510(k) SUMMARY (21 CFR 807.92)
### 1. General Information
| Submitter Name: | Activ Surgical, Inc. |
|----------------------|-------------------------------------------------------------------------------------------------|
| Submitter Address: | 840 Summer Street, Suite 108<br>Boston, MA 02127 |
| Contact Person: | Matthew King<br>Director of Regulatory Affairs and Quality<br>Assurance<br>Activ Surgical, Inc. |
| Contact Information: | Email: mking@activsurgical.com<br>Phone: 603-459-9755<br>Fax: 617-977-0992 |
| Submission Type | Traditional 510(k) |
| Date Prepared | December 4, 2020 |
#### 2. Proposed Device
| Proprietary Name | ActivSight Intraoperative Imaging System |
|----------------------------|-------------------------------------------|
| Device Common Name | Light Source, Imaging Module, Light Cable |
| Device Model Number | 450-00001 |
| Device Classification Name | Confocal Optical Imaging |
| Regulation Number | 21 CFR 876.1500 |
| Product Code | OWN |
| Device Classification | II |
| Review Panel | General and Plastic Surgery |
| Premarket Review | General and Surgery Devices (DHT4A) |
### 3. Predicate Device and Reference Device
| | Predicate Device | Reference Device |
|-------------------------------|-------------------------------------------------------|--------------------------------|
| Proprietary Name | Stryker Infrared Fluorescence<br>(IRF) Imaging System | Perimed PeriCam PSI |
| 510(k) Number | K142310 | K120884 |
| Device Classification<br>Name | Confocal Optical Imaging | Extravascular blood flow probe |
| Regulation Number | 21 CFR 876.1500 | 21 CFR 870.2120 |
| Product Code | OWN | DPT |
| Device Classification | II | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Activ Surgical. The word "ACTIV" is in blue, with each letter in a bold, sans-serif font. Below "ACTIV" is the word "SURGICAL" in a smaller, light gray, sans-serif font. The logo is simple and modern.
### 4. Device Description:
The ActivSight Intraoperative Imaging System (ActivSight) is an accessory to existing commercial surgical laparoscope systems, including cameras and video processor units. ActivSight provides real-time endoscopic fluorescence and near-infrared imaging. These imaging features allow surgeons to visually assess vessels, blood flow, and tissue perfusion (using fluorescence and near-infrared imaging), and to visually assess at least one of the major bile ducts (cystic duct, common bile duct, or common hepatic duct) using fluorescence. Fluorescence imaging is enabled through use of any commercially available Indocyanine Green (ICG). These visualization features are available for surgeons to use during minimally invasive surgery. ActivSight is intended to be used in a surgical environment.
ActivSight consists of the following reusable components:
- 글 ActivSight Imaging Module, consisting of optics and sensing electronics. The imaging module attaches physically between the third-party laparoscope and the third-party imaging system camera. Reprocessing of this component requires cleaning and disinfection between uses.
- = ActivSight Light Engine, equipment consisting of system electronics that provide laser fluorescence and near-infrared illumination, processing of video input from both the third-party Camera Control Unit (CCI) and the ActivSight lmaging Module, and outputs video to the surgical monitor. This component does not require reprocessing between uses.
- ActivSight Light Cable, component consisting of a Y-shaped bifurcated light a cable that connects to the third party white-light source, the ActivSight Light Engine, and the third party laparoscope light post (providing white-light illumination of the surgical site through the laparoscope). Reprocessing of this component requires cleaning and steam sterilization.
- ActivSight Sterilization Tray, stainless-steel sterilization tray designed to a properly secure the ActivSight Light Cable for disinfection and sterilization. Reprocessing of this component is accomplished in its use- cleaning and steam sterilization.
ActivSight consists of the following disposable components:
- ActivSight Sterile Drape, a sterilized plastic drape that is provided for each use = to provide a sterile barrier between the imaging module and the patient.
- -ActivSight Calibration Target, a sterilized paper imaging target containing a checkerboard pattern for calibration of the ActivSight imaging module prior to each use of the device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Activ Surgical. The word "ACTIV" is in large, blue, sans-serif font. Below that, in a smaller, light gray, sans-serif font, is the word "SURGICAL". The logo is simple and modern.
### 5. Indications for Use
The ActivSight Intraoperative Imaging System (ActivSight) is intended to provide realtime endoscopic fluorescence and near infrared imaging. ActivSight enables surgeons to visually assess vessels, blood flow, and related tissue perfusion using fluorescence and near infrared imaging, and at least one of the major bile ducts (cystic duct, common bile duct, or common hepatic duct) using fluorescence, all during minimally invasive surgery.
Fluorescence imaging of biliary ducts with ActivSight is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for stand-alone use for biliary duct visualization.
### 6. Performance Data
Safety and performance of the ActivSight Intraoperative Imaging System has been evaluated and verified in accordance with design specifications and applicable performance standards, including risk management, basic safety and essential performance, electromagnetic disturbances, human factors and usability, software lifecycle processes, safety of laser products. Activ Surgical has performed design validation in an animal study, comparatively visualizing blood flow and perfusion in specific structures with both ActivSight Perfusion (speckle laser) and ActivSight ICG (fluorescence) modes against the predicate's ICG mode. The following performance testing have also been conducted.
- Basic safety and essential performance testing completed in accordance with . IEC 60601-1:2005/R(2012) and A1:2012: Medical Electrical Equipment- Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1:2005. MOD). Testing indicates that the proposed device conforms to this standard.
- Electrical safety and electromagnetic compatibility testing completed in . accordance with IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral Standard: Electromagnetic disturbances: Requirements and tests. Testing indicates that the proposed device conforms to this standard.
- Human factors and usability testing completed in accordance with IEC 62366-1: . 2015 Medical Devices - Application of Usability Engineering to Medical Devices-Evaluation and reporting. Testing indicates that the proposed device presents no unacceptable risk to the user during its intended use, and that the device conforms to this standard.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Activ Surgical. The word "ACTIV" is in a large, sans-serif font and is colored blue. Below "ACTIV" is the word "SURGICAL" in a smaller, sans-serif font and is colored gray. The logo is simple and modern.
- Software development and software life cycle processes competed in accordance with ANSI AAMI IEC 62304:2006/A1:2016 Medical Device Software-Software Life Cycle Processes, Evaluation, Testing, and Reporting. Testing indicates that the proposed device conforms to this standard.
- · Laser safety testing completed in accordance with /EC 60825-1 Ed. 3.0 b:2014 Safety of Laser Products- Part 1: Equipment Classification and Requirements. Testing indicates that the proposed device conforms to this standard.
- Safety of endoscopic equipment testing completed in accordance with IEC . 60601-2-18: Edition 3.0 2009-08 Medical Electrical Equipment- Part 2-18: Particular Requirements for the Basic Safety and Essential Performance of Endoscopic Equipment. Testing indicates that the proposed device conforms to this standard.
The results of these studies demonstrate that the proposed device is as safe and effective as the predicate.
# 7. Substantial Equivalence
The ActivSight Intraoperative Imaging System (subject) is substantially equivalent to the Stryker® Infrared Fluorescence (IRF) Imaging System (predicate).
The subject device and predicate device have nearly identical Indications for Use. Both devices provide real-time endoscopic near-infrared fluorescence imaging during minimally invasive surgery. Both are indicated to enable surgeons visually assess blood flow, tissue perfusion, and at least one of the major bile ducts (cystic duct, common bile duct, or common hepatic duct). Fluorescence imaging of biliary ducts with both devices is intended for use with standard care white light, and when indicated, intraoperative cholangiography.
Both devices achieve their intended use through use of an FDA approved contrast drug agent (indocyanine green (ICG)).
Neither device is intended for stand-alone use for biliary duct visualization.
Differences between the two devices are ActivSight's capability to visualize blood flow and perfusion through speckle laser illumination, and ActivSight does not include laparoscopic equipment or a white light source. Laser speckle illumination is achieved through near-infrared laser diodes, imaging hardware, and imaging software that together display blood flow and perfusion without the aid of an imaging agent, such as ICG. ActivSight is interoperable with the existing laparoscopic equipment of the Stryker Infrared Fluorescence (IRF) Imaging System.
CONFIDENTIAL
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Activ Surgical. The logo is blue and features the word "ACTIV" in large, bold letters. Below the word "ACTIV" is the word "SURGICAL" in smaller, thinner letters. The logo is simple and modern.
# Table 5.1: Comparison to Predicate
| | ActivSight Intraoperative<br>Imaging System | Stryker Infrared<br>Fluorescence (IRF) Imaging<br>System (K142310) - Predicate | Perimed PeriCam<br>PSI (K120884) -<br>Reference Device | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Intended Use | Confocal Optical Imaging<br>(OWN) | Confocal Optical Imaging<br>(OWN) | Probe, Blood-Flow,<br>Extravascular (DPT) | |
| Indications<br>for Use | The ActivSight<br>Intraoperative Imaging<br>System (ActivSight) is<br>intended to provide real-<br>time endoscopic<br>fluorescence and near<br>infrared imaging. ActivSight<br>enables surgeons to<br>visually assess vessels,<br>blood flow, and related<br>tissue perfusion using<br>fluorescence and near<br>infrared imaging, and at<br>least one of the major bile<br>ducts (cystic duct, common<br>bile duct, or common<br>hepatic duct) using<br>fluorescence, all during<br>minimally invasive surgery.<br><br>Fluorescence imaging of<br>biliary ducts with ActivSight<br>is intended for use with<br>standard of care white light,<br>and when indicated,<br>intraoperative<br>cholangiography. The<br>device is not intended for<br>stand-alone use for biliary<br>duct visualization. | The Stryker® IRF Light Source<br>and SafeLight Cable are<br>indicated for use to provide<br>real-time endoscopic visible<br>and near-infrared fluorescence<br>imaging. The Stryker® IRF<br>Light Source and SafeLight<br>Cable enable surgeons to<br>perform minimally invasive<br>surgery using standard<br>endoscope visible light as well<br>as visual assessment of<br>vessels, blood flow and related<br>tissue perfusion, and at least<br>one of the major extra-hepatic<br>bile ducts (cystic duct, common<br>bile duct and common hepatic<br>duct), using near infrared<br>imaging.<br><br>Fluorescence imaging of biliary<br>ducts with the Stryker® IRF<br>Light Source and SafeLight<br>Cable are intended for use with<br>standard of care white light<br>and, when indicated,<br>intraoperative<br>cholangiography. The devices<br>are not intended for standalone<br>use for biliary duct visualization. | The PeriCam PSI is<br>intended for non-<br>invasive two-<br>dimensional imaging<br>of peripheral tissue<br>blood perfusion. | |
| Technological Characteristics | | | | |
| | ActivSight Intraoperative Imaging<br>System<br>ActivSight<br>Perfusion | Stryker Infrared<br>Fluorescence<br>(IRF) Imaging<br>System<br>(K142310) -<br>Predicate | Perimed<br>PeriCam PSI<br>(K120884) -<br>Reference<br>Device | |
| Mode of Action | Laser Speckle | Fluorescence | Laser Speckle | |
| Contrast Agent | None | ICG | None | |
| Information<br>Presentation<br>Format | Heatmap signal<br>overlayed on<br>red/green/blue<br>(RGB) or heatmap<br>signal only | Green Signal<br>overlayed on<br>red/green/blue<br>(RGB) | Green signal<br>overlayed on<br>grayscale | Heatmap signal<br>only with option<br>for adjacent<br>color image. |
| Light Source(s) | Coherent 852 nm<br>laser | 808nm laser | 808nm laser | 785nm laser |
| Light Source<br>Power Rating | Class 3R out of<br>Light Cable, class 1<br>out of scope | Class 3R out of<br>Light Cable, class<br>1 out of scope | Class 1M | Class 1 |
| Light Capture | IR sensitive camera | IR sensitive<br>camera | IR sensitive<br>camera | IR sensitive<br>camera |
| Active Area of<br>imaging sensor | 1/2.9" CMOS,<br>4.98x3.74mm | 1/2.9" CMOS,<br>4.98x3.74mm | 1/3" CMOS,<br>4.8x3.6mm | Unknown |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Activ Surgical. The word "ACTIV" is in large, blue, sans-serif font. Below that, in a smaller, light gray, sans-serif font, is the word "SURGICAL". The logo is simple and modern.
### 8. Conclusion
Based on the information provided in this premarket notification, and based on a comparison to the indications for use, performance testing, and technological characteristics of the predicate, the ActivSight Intraoperative Imaging System is shown to raise no new questions of safety and efficacy, and is substantially equivalent to the Stryker Infrared Fluorescence (IRF) Imaging System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.