KARL STORZ ICG Imaging System

K171238 · KARL STORZ Endoscopy-America, Inc. · OWN · May 25, 2017 · Gastroenterology, Urology

Device Facts

Record IDK171238
Device NameKARL STORZ ICG Imaging System
ApplicantKARL STORZ Endoscopy-America, Inc.
Product CodeOWN · Gastroenterology, Urology
Decision DateMay 25, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The KARL STORZ ICG Imaging System is intended to provide real-time visible and near-infrared fluorescence imaging. The KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. The KARL STORZ VITOM II ICG System is intended for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.

Device Story

System provides real-time visible (VIS) and near-infrared (NIR) fluorescence imaging during surgery. Inputs: VIS and NIR light signals captured via endoscope or VITOM II telescope after ICG dye administration. Operation: Light source emits VIS/NIR illumination; color video camera head captures fluorescence; Image1S Camera Control Unit (CCU) processes signals for display on medical-grade HD monitor. Used in OR by surgeons for minimally invasive, plastic, micro-, and reconstructive procedures. Output: Magnified VIS/NIR video feed. Clinical utility: Enables intraoperative visual assessment of blood flow, tissue perfusion, and biliary duct anatomy; serves as adjunctive tool for tissue-transfer circulation evaluation. Benefits: Enhanced visualization of perfusion and anatomical structures to guide surgical decision-making.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including verification and validation of design specifications, cleaning/sterilization summaries, risk evaluation, and biological evaluation.

Technological Characteristics

System includes 5mm/10mm HOPKINS ICG/NIR endoscopes, VITOM II ICG/NIR telescope, light source with foot switch, color video camera head, and Image1S CCU. Connectivity via DVI-D or 3G-SDI. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 8600-1/3/4/5/6, ISO 14971, ISO 10993.

Indications for Use

Indicated for surgeons performing minimally invasive surgery or plastic, micro-, and reconstructive surgical procedures requiring real-time visible and near-infrared fluorescence imaging for visual assessment of blood flow, tissue perfusion, and biliary duct visualization (cystic, common bile, or common hepatic ducts).

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 25, 2017 KARL STORZ Endoscopy- America, Inc. Winkie Wong Senior Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245 Re: K171238 > Trade/Device Name: KARL STORZ ICG Imaging System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OWN Dated: April 26, 2017 Received: April 27, 2017 Dear Winkie Wong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171238 Device Name KARL STORZ ICG Imaging System Indications for Use (Describe) The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging. The KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization. The KARL STORZ VITOM II ICG System is intended for capturing fluorescent mages for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light. Type of Use (*Select one or both, as applicable*) | <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for STORZ. The logo is in blue and consists of the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. ## 510(k) Summary This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge. | Submitter: | KARL STORZ Endoscopy-America, Inc. | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 2151 E. Grand Avenue<br>EI Segundo, CA 90245 | | Contact: | Winkie Wong<br>Senior Regulatory Affairs Specialist<br>Phone: (424) 218-8379<br>Fax: (424) 218-8519 | | Date of<br>Preparation: | April 26, 2017 | | Type of 510(k)<br>Submission: | Special | | Device<br>Identification: | Trade Name: KARL STORZ ICG Imaging System | | | Classification Name: Confocal Optical Imaging | | Regulatory Class: | II | | Product Code: | OWN | | Regulation: | 21 CFR part 876.1500 (Endoscope and Accessories) | | Predicate<br>Device(s): | Primary Predicate Device: KARL STORZ Endoscopic ICG System<br>(K62882) | | | These predicate devices have not been subject to a design-related<br>recall. | | Device Description: | The KARL STORZ ICG Imaging System is used to provide real-time<br>high definition (HD) endoscopic or telescope images of visible<br>(VIS) and near-infrared (NIR) indocyanine green (ICG) dye<br>fluorescence during minimally invasive surgery as well as plastic,<br>micro- and reconstructive surgical procedures. | | | The overall system includes a 5mm &10mm HOPKINS ICG/NIR<br>Endoscope for use in minimally invasive procedures and a VITOM<br>II ICG/NIR Telescope for use in plastic, micro- and reconstructive<br>surgical procedures for VIS and NIR illumination and imaging, a<br>light source with foot switch for emission of VIS and NIR<br>illumination, a color video camera head capable of capturing both | | | VIS and NIR imaging, and KARL STORZ ICG Kit. Additional<br>accessories used with the KARL STORZ ICG Imaging System<br>include two standard fiber-optic light cables for transmission of<br>VIS and NIR light and the Image1S Camera Control Unit (CCU). | | | The KARL STORZ ICG Imaging System can be used with any<br>medical grade HD monitor with a DVI-D or 3G-SDI input. | | Intended Use: | The KARL STORZ ICG Imaging System is intended to provide real-<br>time visible and near-infrared fluorescence imaging. | | Indications For<br>Use: | The KARL STORZ ICG Imaging System is intended to provide real-<br>time visible (VIS) and near-infrared (NIR) fluorescence imaging. | | | The KARL STORZ Endoscopic ICG System enables surgeons to<br>perform minimally invasive surgery using standard endoscopic<br>visible light as well as visual assessment of vessels, blood flow and<br>related tissue perfusion, and at least one of the major extra-<br>hepatic bile ducts (cystic duct, common bile duct and common<br>hepatic duct), using near infrared imaging. Fluorescence imaging<br>of biliary ducts with the KARL STORZ Endoscopic ICG System is<br>intended for use with standard of care white light and, when<br>indicated, intraoperative cholangiography. The device is not<br>intended for standalone use for biliary duct visualization. | | | The KARL STORZ VITOM II ICG System is intended for capturing<br>and viewing fluorescent images for the visual assessment of blood<br>flow, as an adjunctive method for the evaluation of tissue<br>perfusion, and related tissue-transfer circulation in tissue and free<br>flaps used in plastic, micro- and reconstructive surgical<br>procedures. The VITOM II ICG System is intended to provide a<br>magnified view of the surgical field in standard white light. | | Technological<br>Characteristics: | The KARL STORZ ICG Imaging System is modification of and<br>substantially equivalent to the KARL STORZ ICG Imaging System<br>(K162882) in terms of its indications for use, design technology,<br>and performance specifications. | | | The main difference between the subject and the primary<br>predicate device is the addition of a 5mm ICG Endoscope to the<br>KARL STORZ Endoscopic ICG System to be used in minimal<br>invasive surgery. | | | Bench and comparative testing were used to demonstrate | | | substantial equivalence to the primary predicate device. | | | Therefore, the differences between the subject device and the<br>predicate device do not raise any new issues of safety and<br>effectiveness. | | Non-Clinical<br>Performance Data: | There are no performance standards or special controls<br>developed under Section 514 of the FD&C Act for endoscopes.<br>However, the KARL STORZ ICG Imaging System follows the FDA<br>recognized consensus standards and is tested according to the<br>following standard:<br>• IEC 60601-1<br>• IEC 60601-1-2<br>• IEC 60601-2-18<br>• ISO 8600-1<br>• ISO 8600-3<br>• ISO 8600-4<br>• ISO 8600-5<br>• ISO 8600-6<br>• ISO 14971<br>• ISO 10993<br>Additional bench testing was performed to ensure the device met<br>its design specifications.<br>The bench testing performed verified and validated that the KARL<br>STORZ ICG Imaging System has met all its design specification and<br>is substantially equivalent to its predicate devices. | | Clinical<br>Performance Data: | Clinical testing was not required to demonstrate the substantial<br>equivalence to the predicate devices. Non-clinical bench testing<br>was sufficient to assess safety and effectiveness and to establish<br>the substantial equivalence of the modifications. | | Conclusion: | The conclusions drawn from the non-clinical tests such as the<br>cleaning and sterilization summary, the risk evaluation on the<br>modification of the system and biological evaluation summary<br>demonstrated that the subject device is as safe as and as effective<br>as the predicate device. As such, we concluded that the substantial<br>equivalence of the subject and the predicate devices has been met,<br>and the differences between the subject and the predicate devices<br>do not raise new or different questions of safety and effectiveness. | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and consists of the word "STORZ" in large, bold letters above the words "KARL STORZ - ENDOSKOPE" in smaller letters. The "O" in STORZ is stylized as a circle with a smaller circle inside it. {5}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...