Senhance Surgical System

K223095 · Asensus Surgical, Inc. · NAY · Mar 16, 2023 · Gastroenterology, Urology

Device Facts

Record IDK223095
Device NameSenhance Surgical System
ApplicantAsensus Surgical, Inc.
Product CodeNAY · Gastroenterology, Urology
Decision DateMar 16, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Senhance Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization, and retraction. The Senhance Surgical System is intended for use in general laparoscopic surgical procedures and laparoscopic gynecological surgery. The system is indicated for adult and pediatric use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use. Use of the device is limited to patients two (2) years of age and older and a weight equal to or above 10kg, who are suitable to be subjected to a conventional endoscopic technique.

Device Story

Multi-arm, console-based robotic system for laparoscopic surgery; enables remote tele-operation of instruments and endoscopes. Surgeon console (cockpit) features manipulators/handles and video monitor; Intelligent Surgical Unit (ISU) serves as communication hub between cockpit and manipulator arms. System inputs: surgeon hand movements; optional eye-tracking for camera control. System outputs: robotic arm movement; endoscopic video display; optional force feedback (tactile sensory input) to surgeon handles. Used in operating rooms by trained physicians. Benefits: provides robotic assistance for tissue manipulation (grasping, cutting, suturing, etc.) while maintaining laparoscopic approach; force feedback aids in sensing tissue elasticity.

Clinical Evidence

Retrospective clinical data review of 32 pediatric patients (ages 2-18, weight 12-117kg) undergoing various general/gynecological procedures. Results: 1 intra-operative complication, 5 post-operative complications, 4 conversions to standard laparoscopy, 0 conversions to open surgery. Compared against systematic literature review of 7,660 pediatric procedures (robotic, laparoscopic, open). Metrics: Length of stay (3.88 days), surgical complication rate (3.13%), conversion rate (12.50%), 30-day readmission (3.13%), 30-day re-operation (6.25%), mortality (0%), operative time (141.47 min). Data demonstrates safety and effectiveness comparable to alternative techniques.

Technological Characteristics

Multi-arm robotic system; tele-operation principle. Components: surgeon console, manipulator arms, ISU, laparoscopic instruments. Features: force feedback (tactile), eye-tracking camera control. Materials similar to traditional laparoscopic instrumentation. Connectivity: ISU-based communication hub. Software: robotic control algorithms. Sterilization: compatible with standard laparoscopic instrument sterilization methods.

Indications for Use

Indicated for adult and pediatric patients (≥2 years, ≥10kg) undergoing general laparoscopic or gynecological surgical procedures, provided they are suitable for conventional endoscopic techniques.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. March 16, 2023 Asensus Surgical, Inc. Casey Hinckley Regulatory Affairs Manager 1 TW Alexander Drive, Suite 160 Durham, North Carolina 27703 Re: K223095 Trade/Device Name: Senhance Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY Dated: February 13, 2023 Received: February 15, 2023 Dear Casey Hinckley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.03.16 14:59:30 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223095 Device Name Senhance® Surgical System #### Indications for Use (Describe) The Senhance Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization, and retraction. The Senhance Surgical System is intended for use in general laparoscopic surgical procedures and laparoscopic gynecological surgery. The system is indicated for adult and pediatric use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use. Use of the device is limited to patients two (2) years of age and older and a weight equal to or above 10kg, who are suitable to be subjected to a conventional endoscopic technique. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY #### Senhance Surgical System #### [In accordance with 21 CFR 807.92] 510(k) Sponsor: Asensus Surgical, Inc. 1 TW Alexander Drive, Suite 160 Address: Durham, NC 27703 Casey Hinckley Contact Person: Regulatory Affairs Manager Contact Information: Email: chinckley@asensus.com Phone: 801.310.5491 Date Summary Prepared: March 16, 2023 Proprietary (Trade) Name: Senhance® Surgical System Common Name: System, Surgical, Computer Controlled Instrument Class II Classification: Classification Advisory General and Plastic Surgery Committee: Regulation Number: 21 CFR 876.1500, Endoscope and Accessories Product Codes: NAY (System, Surgical, Computer Controlled Instrument) Primary Predicate Device: Senhance® Surgical System (K220889) Secondary Predicate Device: Intuitive Surgical da Vinci Si Surgical System (K171699) {4}------------------------------------------------ ### Device Description: The Senhance Surgical System is a multi-arm, console-based robotic system that allows a surgical team to perform laparoscopic surgery in the abdomen and pelvis in a manner similar to a manual laparoscopic approach. Each robotic arm can hold either a laparoscopic surgical instrument or an endoscope to facilitate a surgeon remotely operating the instrument from the cockpit. More specifically, the Senhance Surgical System consists of: a surgeon console (cockpit), which provides remote manipulators or handles to allow the surgeon to maneuver the surgical instruments and a video monitor to display the endoscopic signal; manipulator arms, which hold and maneuver the instruments and endoscope based on inputs from the surgeon: Intelligent Surgical Unit (ISU), which is the system communication hub, connecting the cockpit and manipulator arms; and instruments, which manipulate the tissue of interest. In addition, force feedback provides an optional tactile sensory input to the surgeon control handles to give a sense of tissue elasticity. An eye tracking feature provides the surgeon an optional method to control the endoscope from the cockpit, rather than using the surgeon control handles. The ISU allows for three additional methods of camera control, in addition to the optional eye tracking method. The purpose of this submission is to expand the indications for use to include pediatric patients, two (2) years of age and older and a weight equal to or above 10kg. ### Intended Use/ Indications for Use: The Senhance Surgical System is intended to assist in the accurate control of laparoscopic instruments for visualization and endoscopic manipulation of tissue including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, mobilization, and retraction. The Senhance Surgical System is intended for use in general laparoscopic surgical procedures and laparoscopic gynecological surgery. The system is indicated for adult and pediatric use. It is intended for use by trained physicians in an operating room environment in accordance with the instructions for use. Use of the device is limited to patients two (2) years of age and older and a weight equal to or above 10kg, who are suitable to be subjected to a conventional endoscopic technique. # Summary of Technological Characteristics: The subject device has the same technological characteristics as the primary predicate device, the Senhance Surgical System (K220889), and similar characteristics to the secondary predicate. Both the subject and predicate devices involve robotically assisted tele-operation as the primary technological principle. It is based on the accurate translation of user inputs to robotically assisted outputs. It involves the use of endoscopic instrumentation for manipulation of tissue and vessels in the insufflated body cavity. {5}------------------------------------------------ The Senhance Surgical System consists of: a surgeon console (cockpit), which provides remote manipulators or handles to allow the surgeon to maneuver the surgical instruments and a video monitor display the endoscopic signal; manipulator arms, which hold and maneuver the instruments and endoscopic based on inputs from the surgeon; Intelligent Surgical Unit (ISU), which is the system communication hub, connecting the cockpit and manipulator arms; and instruments, which manipulate the tissue of interest. In addition, force feedback provides optional tactile sensory input to the surgeon control handles to qive a sense of tissue elasticity. An eye tracking feature provides the surgeon an optional method to control the endoscope from the cockpit, rather than using the surgeon control handles. The ISU allows for three additional methods of camera control, in addition to the optional eye tracking method. The Senhance instruments are similar in design and materials to traditional laparoscopic instrumentation. Since there are no technological differences between the subject Senhance system and the primary predicate, no different questions of safety or effectiveness have been raised. # Performance Data: There have been no changes to the device since the previous clearance (K220889), only change is to expand the indications for use to allow pediatric use. Extensive bench testing was conducted on the previously cleared Senhance system, and these data remain applicable to the subject device. The previously collected data demonstrated compatibility, mechanical integrity, functionality, reliability, and safe use, addressing verification of key device functions including video signal, endoscope compatibility, surgical instruments, and adapters, as well as validation of the system's camera control and force feedback features. The recognized consensus standards for the predicate Senhance system (K220889) are still applicable as there have been no changes to the technological characteristics or principles of operation of the device. # Clinical Data: To demonstrate that the subject device is safe and effective for the expanded indications for pediatric patients, the company has collected real-world evidence on the Senhance Surgical System for the pediatric population. The data demonstrates favorable performance and safety results for the proposed indication. A retrospective clinical data review was performed for the thirty-two (32) pediatric patients that underwent surgery with the Senhance system. There was only one intra-operative complication; it was resolved without the need for conversion and with no post-operative consequences. There were total of five (5) post-operative complications and four (4) conversions to standard laparoscopy. No conversions to open surgery were necessary. {6}------------------------------------------------ | Age Cohort | Age Summary | Weight Summary | Surgical Procedures Performed | |------------|--------------------------------------------------|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 2-12 | Min – 2<br>Max – 12<br>Median – 5<br>Mean – 5 | Min – 12 kg<br>Max – 38 kg<br>Median – 20 kg<br>Mean – 21.25 kg | Hernia inguinalis - 10 Nissen fundoplication - 1 Ladd's procedure - 1 Ileostomie - 1 Hellor-Dor - 1 Appendectomy - 1 | | 13-18 | Min – 13<br>Max – 18<br>Median – 15<br>Mean – 15 | Min – 40 kg<br>Max – 117 kg<br>Median – 59 kg<br>Mean – 62 kg | Hernia inguinalis - 1 Nissen fundoplication - 3 Cholecystectomy - 4 Coecostomy with chait - 2 Ileocoecal resection - 2 Proctocolectomie with ileopouchanale anastomose (IPAA) - 1 Appendectomy - 2 | Table 1. Age Cohort and Surgical Procedures Performed for Retrospective Clinical Data ### Systematic Literature Review: Pediatric data collected from the hospital were compared with the results from peer-reviewed research publications describing the clinical outcomes for more than 7660 pediatric procedures using three alternative surgical techniques: laparoscopic, open and robotically assisted surgery. | | Senhance System | Literature Review Averages | | | |------------------------------------|-----------------|----------------------------|-------------|-------| | | | Robotic | Laparoscopy | Open | | Length of Stay (days) | 3.88 | 2.63 | 2.75 | 3.3 | | Surgical Complication (%) | 3.13 | 1.91 | 5.91 | 1.65 | | Estimated Blood Loss<br>(ml) | 10.78 | 22.74 | 18.2 | 28.34 | | Conversion Rate (%) | 12.50 | 2.55 | 1.04 | N/A | | 30 Day Readmission<br>Rates (%) | 3.13 | 6.83 | 5 | 7.14 | | 30 Day Re-operation<br>Rates (%) | 6.25 | 7.04 | 4 | 7 | | Mortality (%) | 0 | 0.50 | 0 | N/A | | Post-operation<br>Complication (%) | 15.62 | 16.90 | 9.46 | 6.92 | | Operative Time (min) | 141.47 | 195 | 158 | 183 | Table 2. Retrospective Clinical Data Compared to Literature Review Data {7}------------------------------------------------ # Conclusion: Operative times and complication rates were comparable between the retrospective clinical data with the Senhance system and the systematic literature review. Overall, the comparison of the Senhance system pediatric data with the published literature data from other surgical techniques demonstrates that the Senhance system is safe and effective for clinical use in the pediatric population based on the following endpoints: - Length of Hospital Stay - Intraoperative Complication Rates - Estimated Blood Loss (EBL) Volumes and Blood Transfusion Rates ● - Conversion Rates ● - Readmission Rates ● - Reoperation Rates - Mortality Rates - Postoperative Complication Rates ● - Operative Times In conclusion, the clinical analysis of the subject Senhance Surgical System demonstrates that the device is as safe and effective as its predicate for patients who are at least two (2) years of age and 10kg or higher in weight. {8}------------------------------------------------ Image /page/8/Figure/0 description: The image is titled "Figure 1. Search criteria and flowchart for literature search". The image is a title for a figure, and it indicates that the figure will show the search criteria and flowchart for a literature search. The figure will likely contain a visual representation of the steps involved in conducting a literature search, as well as the criteria used to select relevant sources. Image /page/8/Figure/1 description: This image shows two flow charts, one for Pubmed and one for Embase, detailing the process of identifying and selecting studies for a qualitative synthesis. For Pubmed, 286 records were identified through database searching, with an additional 76 records identified. After removing duplicates, 210 records remained, which were then screened, leading to 188 records being excluded. For Embase, 357 records were identified through database searching, with an additional 28 records identified; after removing duplicates, 329 records remained, which were then screened, leading to 249 records being excluded. The following filters were applied to narrow the searches to relevant publications: #### Inclusion Criteria: US or EU study, to reduce variation in surgical method. Pediatric and young adults (≤21 years) patients only Studies on robotic assisted laparoscopy LOE ≤3b Studies on gynecology (OR) urology (OR) general surgery Data collected 2000 - Present. Study is an RCT on robotic assisted laparoscopy and/or comparative study reporting on robotic assisted pediatric cases versus minimally invasive/laparoscopic pediatric surgery) Exclusion Criteria Non-US or EU study Meta-Analysis/SLRs (summary data from meta-analysis/SLRs were excluded, however, the references from the meta-analysis/SLRs were analyzed for inclusion as selected articles) Publications not on robotic assisted laparoscopic procedure Publication is an HTA that was not published in a peer reviewed journal. Follow up study, missing original surgery data. Adult only or adult data not reported separately. Adult only or adult data not reported Animal studies Animal studies
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...