MULTAPORT CANNULA REDUCER AND ACCESSORIES

K981941 · Dexide, Inc. · GCJ · Jul 29, 1998 · Gastroenterology, Urology

Device Facts

Record IDK981941
Device NameMULTAPORT CANNULA REDUCER AND ACCESSORIES
ApplicantDexide, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateJul 29, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MultAport reusable cannula is used with an AccuPort Seal or Orbital Reducer Seal and Trocar for creating a point of entry for laparoscopic instruments into the abdominal cavity. The AccuPort Seal is a single-use device used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments with diameters ranging from 5mm to 12mm. The Orbital Reducer is a single used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments of equal and smaller size.

Device Story

Device consists of reusable stainless steel cannula and single-use disposable reduction seal systems (AccuPort or Orbital Reducer). Used in laparoscopic surgery to create entry point for instruments into abdominal cavity. Cannula provides physical pathway; seal system maintains pneumoperitoneum during instrument insertion. Operated by surgeons in clinical/OR setting. Output is physical access port for surgical tools. Benefits patient by enabling minimally invasive laparoscopic procedures.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Materials: stainless steel (cannula), silicon rubber, thermoplastic rubber, and plastics (seal systems). Form factor: cannula with threaded seal interface. Sterilization: reusable cannula (methods specified in data insert).

Indications for Use

Indicated for patients undergoing laparoscopic surgery requiring a pathway for instrument entry into the abdominal cavity while maintaining pneumoperitoneum.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K981941 ## F: 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 510(k) SUMMARY | 1) | Submitter: | Dexide, Inc.<br>7509 Flagstone Drive<br>Fort Worth, TX 76118-6995<br>Phone No.: (817) 589-1454 | | |----|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | | Contact Person: | Lynette Caldwell<br>Director of Quality Assurance | | | | Date Prepared: | Friday, May 29, 1998 | | | 2) | Name of Device: | Cannula, Reducer and Accessories. | | | | Common Name: | Trocar Cannula | | | 3) | Predicate Devices: | Dexide, Inc. - K912980<br>Aesculap - K942053 | | | 4) | Description of Device: | The "candidate device" is a reusable, stainless steel Cannula<br>and disposable, single-use reduction seal system that threads<br>into the Cannula. The reducer system is composed of silicon<br>rubber, thermoplastic rubber, and plastics. | | | | | Accessories:<br>Cannulas & Reduction Systems | | | 5) | Intended Use: | The Cannulas are to be used with a Dexide seal package and<br>spike for creating a point of entry for laparoscopic instruments<br>into the abdominal cavity. | | | 6) | Technological characteristics of this device are comparable to the predicate device in that<br>predicate device(s) is also used to provide a pathway or point of entry for laparoscopic<br>instruments into the abdominal cavity. | | | - {2}------------------------------------------------ ## SECTION G: - - ## MANUFACTURER'S STATEMENT OF SUBSTANTIAL EQUIVALENCE (To be provided with 510(k) notification for tier II devices) The Cannulas arc to be used with a Dexide seal STATEMENT OF INDICATIONS FOR USE: STATEMENT OF INDICATIONS I OF entry for laparoscopic instruments into the abdominal cavity. Provides a pathway for entry of laparoscopic instruments of equal size and smaller CLAIMS: into the abdominal cavity. The Cannula is reusable and the seal package is single-use disposable. into the addominal cavity. The Cannula secommended sterilization methods are listed on the data insert. This notification contains all of the information required by 21 CFR 807.87. This notification of the "DRAERD Premarket Notification 510(k) Reviewer's Screening Checklist" is attached. The candidate device conforms to the following voluntary and mandatory standards: There are no mandatory existing performance standards that Dexide, Inc is aware. The candidate device has the same technological characteristics as a legally marketed predicate I he candidate device has the the materials, and mode of action are equivalent. If device. Specifically, the reallers, open.contents of the kit are either pre-Amendment devices or have been cleared for marketing through previous 510(k)'s. The kit contains no drug or biologic products. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ The above statements are accurate representations of this 510(k) Premarket Notification and of the device this firm intends to market. All data and information submitted in this Premarket Notification is truthful and accurate and no material fact has been omitted (21 CFR 807.87(J). | MANUFACTURER: | Dexide, Inc. | |-------------------------|--------------------------------------------------------------------------------------------------------------------------| | OFFICIAL CORRESPONDENT: | <div style="display:flex; flex-direction:column;">Lynette Caldwell (signature)<br/>Lynette Caldwell (printed name)</div> | | TITLE: | Director of Quality Assurance | | DATE: | May 29, 1998 | {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circle's perimeter. Public Health Service Food and Drug Administration 9200 Corporato Boulevard Rockville MD 20850 Inf 2 9 1998 Ms. Lynette Caldwell ·Director of Quality Assurance Dexide, Inc. 7509 Flagstone Drive Fort Worth, Texas 76118 K981941 Re: R981941 Trade Name: Multaport Cannula Reducer and Accessories Regulatory Class: II Product Code: GCJ Dated: May 29, 1998 Received: June 2, 1998 Dear Ms. Caldwell: We have reviewed your Section 510(k) notification of intent to market the device referenced in the indication bit a baland in incientially equivalent (for the indications fo We have reviewed your Section 3 iv(k) noundedour of them users of the indications for above and we have determined the device is substantially equivalent (for to above and we have determined the uce not in interstate comments on to de use stated in the enclosure) to devices intaked bevice Americans, or to devices that May 28, 1976, the enactment adaly of the Meetical Ferient Food, Drug, and have been reclassitied in accordance with the provisions of the general controls Cosmetic Act (Act). You may, therefore, market the device, subject to the generals for Cosmetic Act (Act). You may, thereone, thanks and to to the Act include requirements for provisions of the Act. The general controls provisions of the Activel and provisions of the Act. The gelectares, good manifacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III n and the many of the may of any cases additional controls. Existing major (Premarket Approval), it may be subject to such additions (existing major (Premarket Approval), it itlar be sugges to sales of Federal Regulations. Title 21, regulations affecting your device can be found in the costsmes compliance with the regulations affecting your device can or round in the other sessmes compliance with the Parts 800 to 895. A substantally equivalians set forth in the Quality System current Good Manufacturing Practice requirement, as set fort in the Quality System current Good Manufacturing Fraction regulation (21 CFR Part 820) and that, Regulation (QS) for Medical Devices: "God and Drug. Administration (FDA) will verify such assumptions. Failure to comply with the CMP regulation may result in regulation such assumptions. Fallure to comply with the SHIP regenents concerning your device in the action. In addition, FDA may ploinst turner and to your premarket notification submission does Federal Register. Please note: this response to your set of the Act for Federal Register. Please note. under sections 531 through 542 of the Act for not affect any obligation you might have union securements or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Ms. Lynette Caldwell This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ | Page | 1 of 1 | |------|--------| |------|--------| 510(k) Number (if known): K981941 Device Name: Cannula, Reducer and Accessories ________________________________________________________________________________________________________________________________ Indications For Use: MultAport Cannulas: MultAport Califords: The MultAport reusable cannula is used with an AccuPort Seal or Orbital Reducer Seal and Trocar The Multi port reasure callining as assaroscopic instruments into the abdominal cavity. AccuPort 5-12mm Seal: Accur on 9 Tellin Gad. The AccuPort Seal is a single-use device used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments with diameters ranging from 5mm to 12mm. Orbital Reducer Seal: The Orbital Reducer is a single used during laparoscopy with MultAport Cannulas to maintain pneumoperitoneum while accommodating instruments of equal and smaller size. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-----------------------------------------|----------| | Division of General Restorative Devices | | | 510(k) Number | 12981941 | | Prescription Use | OR | Over-The-Counter Use | |----------------------|----|--------------------------| | (Per 21 CFR 801.109) | | (Optional Format 1-2-96) |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...