MVM 3.3 MM MICROENDOSCOPE

K972504 · KARL STORZ Endoscopy-America, Inc. · GCJ · Aug 11, 1997 · Gastroenterology, Urology

Device Facts

Record IDK972504
Device NameMVM 3.3 MM MICROENDOSCOPE
ApplicantKARL STORZ Endoscopy-America, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateAug 11, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

This instrument is indicated for use during general endoscopic and laparoscopic surgical procedures, and, using additional accessories, to perform-variousdiagnostic and therapeutic procedures.

Device Story

The KSEA MVM 3.3 mm Microendoscope is a rigid, panoramic telescope utilizing rod lens technology. It consists of an endoscope module and a video module that connects to a camera control unit. Used by surgeons and physicians in clinical settings for endoscopic and laparoscopic procedures, the device provides visualization of the surgical site. The surgeon views the output via the connected camera control unit to guide surgical interventions. The device facilitates minimally invasive access, potentially reducing patient trauma compared to open surgery.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design and intended use.

Technological Characteristics

Rigid, panoramic telescope; rod lens technology; surgical grade stainless steel construction; video module interface for camera control unit; 3.3 mm diameter.

Indications for Use

Indicated for use by qualified surgeons and physicians during general endoscopic and laparoscopic surgical procedures, including various diagnostic and therapeutic procedures when used with additional accessories.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS included in this document is accurate and complete to the best of KSEA's knowledge. ×972504 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data | Applicant: | Karl Storz Endoscopy - America, Inc.<br>600 Corporate Pointe Drive<br>Culver City, CA 90230<br>(310) 558-1500 | AUG 1 1 1997 | |------------------------|---------------------------------------------------------------------------------------------------------------|--------------| | Contact: | Kevin Kennan<br>Regulatory Affairs Specialist | | | Device Identification: | Common Name:<br>Video Endoscope<br><br>Trade Name: (optional)<br>Karl Storz MVM 3.3 mm Microendoscope | | Indication: The KSEA MVM 3.3 mm Microendoscope is designed to be used by qualified surgeons and physicians for general endoscopic and laparoscopic surgical procedures. Device Description: The KSEA MVM 3.3 mm Microendoscope is comprised of a video module, and a endoscope module. The endoscope module is a rigid, panoramic telescope which utilizes rod lens technology. The video module is an imaging device which connects to a camera control unit. The body contact portions of the KSEA MVM 3.3 mm Microendoscope are composed of surgical grade stainless steel, which is commonly used in medical devices for a wide range of applications and has a long history of biocompatibility for human use. Substantial Equivalence: The KSEA MVM 3.3 mm Microendoscope is substantially equivalent to the predicate devices since the basic features, design and intended uses are the same. The minor differences between the KSEA MVM 3.3 mm Microendoscope and the predicate devices raise no new issues of safety and effectiveness, as these design differences have no affect on the performance, function or intended use of the devices. Signed: gned: Kevin Kennan Regulatory Affairs Specialist Regulatory Affairs Specialist {1}------------------------------------------------ Image /page/1/Picture/10 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Kevin Kennan Regulatory Affairs Specialist Karl Storz Endoscopy-America, Inc. ......... 600 Corporate Pointe Culver City, California 90230-7600 AUG 11 1997 Re: K972504 Trade Name: MVM 3.3 mm Microendoscope Regulatory Class: II Product Code: GCJ Dated: July 1, 1997 Received: July 3, 1997 Dear Mr. Kennan: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). Y-ou may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ## Page 2 - Mr. Kevin Kennan This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoscopy. The logo is white text on a black background. The word "STORZ" is in large, bold letters, with the "O" stylized as a circle with a dot in the center. Below "STORZ" is the text "Karl Storz Endoscopy" in a smaller font. 510(k) Number (if known): Not yet assigned K972504 Device Name: MVM 3.3 mm Microendoscope Indications for Use: This instrument is indicated for use during general endoscopic and laparoscopic surgical procedures, and, using additional accessories, to perform-variousdiagnostic and therapeutic procedures. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General Restorative Devices 510(k) Number K972504 | Prescription Use: (Per 21 CFR 801.109) | <div>✓</div> | OR Over-The-Counter Use: | |----------------------------------------|--------------|--------------------------| |----------------------------------------|--------------|--------------------------| (Optional Format 1-2-96) 000003
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...