K220069 · Olympus Medical Systems Corp. · GCJ · Mar 9, 2022 · Gastroenterology, Urology
Device Facts
Record ID
K220069
Device Name
4K UHD LCD Monitor OEV321UH
Applicant
Olympus Medical Systems Corp.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Mar 9, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The 4K UHD LCD Monitor is intended to provide 4K 2D color video displays of images from endoscopic camera systems and other compatible medical imaging systems. The 4K UHD LCD Monitor is a wide-screen, high-definition, medical grade monitor for real-time use during endoscopic procedures and is suitable for use in hospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environments.
Device Story
The 4K UHD LCD Monitor (OEV321UH) is a medical-grade, wide-screen display for real-time visualization of endoscopic and laparoscopic camera feeds. It accepts various input signals (12G-SDI, 3G-SDI, DisplayPort, HDMI, DVI-D) and transforms them into 4K 2D color video via an LED-backlit LCD panel (IPS). The device operates by controlling liquid crystal apertures based on input signals to render images. It is intended for use by physicians and assistants in clinical environments such as operating rooms and surgical centers. The monitor provides high-definition visual feedback to assist clinicians during minimally invasive procedures, facilitating real-time decision-making. It features multiple gamma curve settings (including DICOM and HLG) and color space options to optimize image quality for surgical viewing.
Clinical Evidence
No clinical or animal testing was performed. Substantial equivalence is supported by bench testing, including electrical safety (ANSI AAMI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), and software verification/validation.
Technological Characteristics
LCD panel (IPS) with LED backlight. Dimensions: 753.9 x 476.3 x 79.2 mm. Resolution: 3840 x 2160 pixels. Inputs: 12G-SDI, 3G-SDI, DisplayPort, HDMI, DVI-D. Outputs: 12G-SDI, 3G-SDI, CloneOUT, DC power. Gamma curves: 1.8, 2.0, 2.2, 2.4, 2.6, DICOM, Endoscope, HLG. Standards: ANSI AAMI ES 60601-1, IEC 60601-1-2, ISO 14971.
Indications for Use
Indicated for 4K 2D color video display of images from endoscopic/laparoscopic and other compatible medical imaging systems during endoscopic/laparoscopic procedures in hospital operating rooms, surgical centers, clinics, and doctors' offices.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
March 9, 2022
Olympus Medical Systems Corp. % Elizabeth Greene Program Manager Olympus Corporation of the Americas 3500 Corporate Parkway Center Valley, Pennsylvania 18034-0610
Re: K220069
Trade/Device Name: 4K UHD LCD Monitor, Model Number: OEV321UH Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: December 28, 2021 Received: January 10, 2022
Dear Elizabeth Greene:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220069
Device Name 4K UHD LCD Monitor, Model Number: OEV321UH
Indications for Use (Describe)
The 4K UHD LCD Monitor is intended to provide 4K 2D color video displays of images from endoscopic camera systems and other compatible medical imaging systems. The 4K UHD LCD Monitor is a wide-screen, highdefinition, medical grade monitor for real-time use during endoscopic procedures and is suitable for use in hospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environments.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <b> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. There is a registered trademark symbol to the right of the letter "S".
# 510(k) Summary
For
# 4K UHD LCD MONITOR, Model Number: OEV321UH
| General Information | |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan<br>192-8507<br>Phone: (+81) 42-642-2694<br>Fax: (+81) 42-642-2307<br>Establishment Registration Number: 8010047 |
| Manufacturer: | Shirakawa Olympus Co., Ltd.<br>3-1 Okamiyama, Odakura, Nishigo-mura,<br>Nishishirakawa-gun, Fukushima 961-8061,<br>Japan<br>Phone: (+81) 248-27-2239<br>Fax: (+81) 248-27-2429<br>Establishment Registration Number: 3002808148<br>Sony Global Manufacturing & Operations<br>Corporation<br>Kosai Site<br>554 Sakaijuku, Kosai-shi, Sizuoka 431-0496,<br>Japan<br>Phone: +(1) 201-358-4945<br>Establishment Registration Number: 3004421482 |
| 510(k) Submitter: | Olympus Corporation of the Americas<br>3500 Corporate Parkway<br>Center Valley, PA 18034-0610 |
| Establishment Registration Number: | 2429304 |
| Contact Person: | Elizabeth Greene<br>Program Manager<br>Fax: 484-896-7128<br>Mobile: (561) 774-1483<br>Email: elizabeth.greene@olymus.com |
| Date Prepared: | December 28, 2021 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. A registered trademark symbol is located to the right of the word.
# Device Description
| Device Name: | 4K UHD LCD Monitor |
|----------------------|-----------------------------|
| Model Name: | OEV321UH |
| Generic/Common Name: | LCD Monitor |
| Regulation Number: | 876.1500 |
| Regulatory Class: | Class II |
| Classification Name: | Endoscope and accessories |
| Product Codes: | GCJ |
| Review Panel: | General and Plastic Surgery |
# Predicate Device
| Device Name | 510(k) Submitter | 510(k) No. |
|----------------------------|-------------------------------|------------|
| Sony LMD-X310S LCD Monitor | Sony Electronics Incorporated | K150377 |
### Indications for Use
The 4K UHD LCD Monitor is intended to provide 4K 2D color video displays of images from endoscopic/laparoscopic camera systems and other compatible medical imaging systems. The 4K UHD LCD Monitor is a wide-screen, high-definition, medical grade monitor for real-time use during endoscopic procedures and is suitable for use in hospital operating rooms, surgical centers, clinics, doctors' offices and similar medical environments.
### Principle of Operation
This monitor displays color video images that are output from medical imaging systems on the LCD (liquid crystal display) panel.
Liquid crystal and color filters are laid on the front of the flat light source (backlight) of the LCD panel. The LCD panel displays images by controlling the aperture of the liquid crystal according to input signals.
### Comparison of Technological Characteristics
Table 5-1 compares 4K UHD LCD Monitor OEV321UH to the predicate device with respect to intended use, technological characteristics, and principle of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
#### Table 5-1: Comparison of the technological characteristics of 4K UHD LCD Monitor to predicate device Eeature/Technological Subigat Daviga Dradionto Davico
| Feature/Technological<br>Characteristics | Subject Device | Predicate Device |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory | | |
| Device Name | Olympus 4K UHD LCD Monitor<br>OEV321UH | Sony LCD Monitor<br>LMD-X310S |
| Feature/Technological<br>Characteristics | Subject Device | Predicate Device |
| Regulatory Decision | This submission | K150377 |
| Product Code | Same as predicate | GCJ |
| Regulatory Class | Same as predicate | II |
| Regulation Number | Same as predicate | 876.1500 |
| Regulation Name | Same as predicate | Endoscope and accessories |
| Classification Panel | Same as predicate | General and Plastic Surgery |
| Indications for Use | The 4K UHD LCD Monitor is<br>intended to provide 4K 2D color<br>video displays of images from<br>endoscopic/laparoscopic camera<br>systems and other compatible<br>medical imaging systems. The 4K<br>UHD LCD Monitor is a wide-<br>screen, high-definition, medical<br>grade monitor for real-time use<br>during endoscopic/laparoscopic<br>procedures and is suitable for use in<br>hospital operating rooms, surgical<br>centers, clinics, doctors' offices and<br>similar medical environments. | The LCD Monitor is intended to<br>provide 4K 2D color video displays<br>of images from<br>endoscopic/laparoscopic camera<br>systems and other compatible<br>medical imaging systems.<br>The LCD Monitor is a wide-screen,<br>high-definition, medical grade<br>monitor for real-time use during<br>minimally invasive surgical<br>procedures and is suitable for use in<br>hospital operating rooms, surgical<br>centers, clinics, doctors' offices,<br>and similar medical environments. |
| Mode of Action | Same as predicate | This monitor displays color video<br>images that are output from medical<br>imaging systems on the LCD (liquid<br>crystal display) panel.<br>Liquid crystal and color filters are<br>laid on the front of the flat light<br>source (backlight) of the LCD panel.<br>The LCD panel displays images by<br>controlling the aperture of the liquid<br>crystal according to input signals. |
| Intended Environment | Same as predicate | Hospital operating rooms, surgical<br>centers, clinics, doctors' offices and<br>similar medical environments |
| Intended Users | Same as predicate | Doctors and Assistants |
| System Parameters and Specifications | | |
| Power | Same as predicate | AC 100-240V/ 50-60Hz |
| Dimensions (excluding<br>max. protrusions) | $753.9 \times 476.3 \times 79.2{\rm mm}$ | $753.8 \times 456.4 \times 69.3{\rm mm}$ |
| Display Dimension | Same as predicate | 2D |
| Input Signals | 12G-SDI1, 12G-SDI2, 3G-SDI,<br>DisplayPort, HDMI, DVI-D, DC<br>IN | DVI-D, HDMI, 3G/HD/SD-SDI |
| Output Signals | 12G-SDI1, 12G-SDI2, 3G-SDI,<br>CloneOUT, +5V DC OUT, +12V<br>DC OUT | DVI-D, 3G/HD/SD-SDI |
| Display Device | Same as predicate | LCD panel (IPS) |
| Backlight Device | Same as predicate | LED |
| Feature/Technological<br>Characteristics | Subject Device | Predicate Device |
| Viewing Angle | Same as predicate | Right>89[deg] (CR>10)<br>Left>89[deg] (CR>10)<br>Up>89[deg] (CR>10)<br>Down>89[deg] (CR>10) |
| Active Screen Size | 697(H)×392(V) mm | 697.958(H)×368.064(V) mm |
| Resolution | 3840 × 2160 pixels | 4096 × 2160 pixels |
| Luminance | ≥280[cd/m²] | ≥550[cd/m²] |
| Primary Colors | Same as predicate | RGB |
| Gamma Curve | 1.8, 2.0, 2.2, 2.4, 2.6, DICOM, Endoscope, HLG | 1.8, 2.0, 2.2, 2.4, 2.6, DICOM, Endoscope, Highlight |
| Color Space | Same as predicate | Auto, BT.709, BT.2020, Native |
| Refresh Rate | Same as predicate | 50/60Hz |
| Frame Rate | Same as predicate | 50/60 fps |
| Display Format | Same as predicate | Normal<br>Multi display<br>Flip display |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "K220069" in a bold, serif font. The text is black and the background is white. The letters and numbers are evenly spaced and aligned horizontally.
Image /page/5/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. A registered trademark symbol is located to the right of the word.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The font is sans-serif and the letters are closely spaced together. A registered trademark symbol is located to the right of the word.
### Traditional 510(k) Notification 4K UHD LCD Monitor, Model Number: OEV321UH
# Compliance to Voluntary Standards
The following voluntary standards have been applied to the subject devices respectively:
- ANSI AAMI ES 60601-1:2005/(R)2012 and A1:2012
- IEC 60601-1-2: 2014
- ISO 14971:2007/2019 ●
# Summary of Performance Testing
The following performance testing was conducted in support of the substantial equivalence determination.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC performance testing for the 4K UHD LCD Monitor OEV321UH is confirmed to be in compliance with the relevant requirements as noted below.
- ANSI AAMI ES 60601-1:2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 4)
# Software Verification and Validation Testing
Software testing has been performed and documented in compliance with the FDA guidance "Guidance for the Content of Premarket Submissions for Software contained in medical devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. Below the word is a yellow line. The registered trademark symbol is located to the right of the word.
### Risk Analysis
Risk analysis for the subject device was conducted in accordance with established in-house acceptance criteria based on ISO 14971. The design verification tests, and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
In the risk management process, the human factors validation testing was also performed in accordance with the FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices". In terms of human factors, an assessment of applicable adverse events along with a review of the overall risk analysis was conducted. These assessments confirmed that there was no unacceptable user-related residual risk for the 4K UHD LCD Monitor OEV321UH.
# Animal and Clinical Testing
Animal and Clinical testing was not applicable and not performed.
# Substantial Equivalence
Olympus has determined that the 4K UHD LCD Monitor OEV321UH is substantially equivalent to the legally marketed predicate device, Sony LCD Monitor LMD-X310S (K150377) for the following reasons:
- . same intended use;
- technological characteristics (design, materials, and operations) are similar or identical to the predicate devices; and
- does not introduce any new or novel treatments or standard of care that differs from ● predicate devices in commercial use.
The intended use, principles of operation, fundamental technology of the 4K UHD LCD Monitor OEV321UH are identical to the predicate device. The differences in system parameters, and specifications include device dimensions, input and output signals, active screen size, resolution, luminance, and gamma curve. The 4K UHD LCD Monitor OEV321UH is a general use medical monitor and is not intended to be used in limited procedures. The difference in indications for use between the 4K UHD LCD Monitor OEV321UH and Sony LCD Monitor LMD-X310S is not a change from single use labeling to reusable, is not a change from prescription (Rx) use to over the counter (OTC) use. Further, this change does not describe a new disease, condition, or patient population that the device is intended in diagnosing, treating, preventing, curing, or mitigating. A risk-based assessment of these differences did not identify any new risks or significantly modified existing risks, or raise new or different questions with respect to safety and effectiveness.
The 4K UHD LCD Monitor OEV321UH has been verified and validated to be equivalent in electrical performance for displaying images from endoscopic camera systems and other compatible medical imaging systems, when compared to the predicate. As the electrical safety and electromagnetic compatibility test results demonstrate equivalent performance, Olympus believe there are no new concerns or modified existing risks regarding safety and effectiveness of the subject device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. A registered trademark symbol is located to the right of the word. The font is sans-serif and the letters are closely spaced together.
# Conclusion
In summary, the Olympus 4K UHD LCD Monitor OEV321UH is substantially equivalent to the predicate device and raise no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.