The Anchor Tissue retrieval system by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.
Device Story
Sterile, disposable retrieval pouch with dedicated introducer; used in laparoscopic surgery. Device consists of rip-stop nylon pouch, stainless steel arms, internal pusher rod, and handle. Surgeon inserts device into body cavity; deploys pouch via pusher rod mechanism; captures tissue; removes pouch. Introducer handle features spring-loaded button and stop mechanism for bag separation. Subject device allows for redeployment within a single patient procedure, unlike the predicate. Used in operating room by surgeons to facilitate tissue extraction; benefits patient by enabling safe removal of tissue through small laparoscopic incisions.
Clinical Evidence
Bench testing only. No clinical data. Performance verified through non-clinical bench and simulated use testing, including design verification and validation. Biocompatibility testing performed per ISO 10993-1:2009. Sterilization validated per ISO 11135:2014. Redeployment verification testing confirmed the device can be deployed multiple times within a single patient procedure.
Technological Characteristics
Materials: rip-stop nylon, stainless steel, polycarbonate. Deployment: automatic via pusher rod compression. Separation: push-button mechanism. Sterilization: Ethylene Oxide (ISO 11135:2014). Biocompatibility: ISO 10993-1:2009. Packaging: ISO 11607-1:2006. Single-patient use. No electronic or software components.
Indications for Use
Indicated for patients undergoing laparoscopic surgery requiring the encapture and removal of an organ or tissue from the body cavity.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2017
ConMed Corporation Ms. Rachelle Fitzgerald Senior Specialist, Regulatory Affairs 525 French Road Utica, New York 13502
Re: K172940
Trade/Device Name: ANCHOR Tissue Retrieval System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: September 25, 2017 Received: September 26, 2017
Dear Ms. Fitzgerald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely, Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K172940
Device Name
Anchor Tissue Retrieval System by CONMED
Indications for Use (Describe)
The Anchor Tissue retrieval system by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
Anchor Tissue Retrieval System™ by CONMED
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92. ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) number ______________________________________________________________________________________________________________________________________________
## A. Submitter
ConMed Corporation 525 French Road Utica, NY 13502
Establishment Registration: 1320894
## B. Company Contact
Rachelle Fitzgerald Senior Specialist, Regulatory Affairs T: (203) 799-2400 ext 8195 F: (203) 799-2401
### C. Device Name
| Proprietary Name: | Anchor Tissue Retrieval System™ by CONMED |
|--------------------|-------------------------------------------|
| Common Name: | Laparoscope, General & Plastic Surgery |
| Panel: | Gastroenterology/Urology |
| Product Code: | GCJ |
| Device Class: | II |
| Regulation Number: | 876.1500 |
### D. Predicate Device
| Primary Device Name: | Anchor Tissue Retrieval System |
|----------------------|--------------------------------|
| Company Name: | CONMED Corporation |
| 510(k): | K091930 |
This predicate has not been subject to a design-related recall.
### E. Device Description
The Anchor Tissue Retrieval System™ by CONMED is a sterile, disposable retrieval pouch for use with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery. The Anchor Tissue Retrieval System™ by CONMED consists of a rip stop nylon pouch, two stainless steel arms attached to internal pusher rod connected to a handle at the distal end to the patient. Around the internal pusher rod is the introducer and handle. The introducer handle
{4}------------------------------------------------
incorporates a spring loaded button and stop mechanism is built into the pusher rod. There is a braided drawstring which enters the introducer with the pusher rod and loops through the rip stop nylon pouch.
Differences between the predicate device and the Anchor Tissue Retrieval System™ by CONMED are limited to the additional feature that the Anchor Tissue Retrieval System™™ by CONMED is able to be redeployed within a single patient procedure.
#### F. Intended Use / Indications for Use
The Anchor Tissue retrieval system by CONMED is a sterile disposable pouch used with a dedicated introducer for the encapture and removal of an organ or tissue from the body cavity during laparoscopic surgery.
### G. Technological Characteristics
The Anchor Tissue Retrieval System™ by CONMED has the same technological characteristics as the predicate device in that the Anchor Tissue Retrieval System™ by CONMED has the same design, same deployment mechanism, and utilizes the same rip stop nylon pouch, introducer components, and external introducer as the predicate. As with the predicate, the Anchor Tissue Retrieval System™ by CONMED continues to function as a single patient/procedure use pouch with a dedicated introducer for the encapture and removal of the organ or tissue from the body cavity during laparoscopic surgery. The difference in the subject device from the predicate device is that the subject device has a new feature where, if the surgeon deems it appropriate, the Anchor Tissue Retrieval System™ by CONMED can be redeployed in the same single patient procedure. The Anchor Tissue Retrieval System™ by CONMED is safe and effective and substantially equivalent to the predicate as demonstrated by non-clinical performance testing for the same indications for use, target population, principles of operation, performance specifications, and standards for sterilization, packaging, and biocompatibility.
| Characteristic | Anchor Tissue Retrieval System™ by<br>CONMED | Predicate Device<br>K091930 | |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------|
| Intended Use | The Anchor Tissue retrieval system by<br>CONMED is a sterile disposable pouch used<br>with a dedicated introducer for the encapture<br>and removal of an organ or tissue from the<br>body cavity during laparoscopic surgery. | Same | |
| | Where Used | Operating room | Same |
| | Prescription Only | Yes | Same |
| | Design | Specimen bag and a delivery system<br>consisting of an introducer shaft deployment<br>handle | Same |
| | | Materials | Nylon, stainless steel, polycarbonate |
| Mechanism to deploy bag | | Automatic with compression of pusher rod | Same |
| Mechanism to separate bag<br>from introducer | Push button | Same | |
{5}------------------------------------------------
| Characteristic | Anchor Tissue Retrieval System™ by<br>CONMED | Predicate Device<br>K091930 |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Performance | There are no FDA performance standards for<br>this device. Device performance was<br>adequately tested through bench testing<br>methodologies. | Same |
| Sterilization | Ethylene Oxide per ISO 11135:2014 | Same |
| Biocompatibility | According to ISO 10993-1:2009 | Same |
## H. Performance Testing
Non-clinical bench and simulated use testing demonstrate the Anchor Tissue Retrieval System™ by CONMED is substantially equivalent to the predicate device with regard to indication for use, materials, technology, and performance. Design verification testing demonstrates devices comply with design specifications and applicable sections of ISO 11607-1:2006. ISO 11135:2014. ISO 10993-7:2008. Results from design validation testing performed demonstrate that the Anchor Tissue Retrieval System™ by CONMED conforms to user needs and the intended use. Risk management activities in accordance with ISO 14971 demonstrate the risks associated with the use of the Anchor Tissue Retrieval System™ by CONMED, including redeployment, are mitigated to an acceptable level. No new issues of safety of effectiveness were identified and redeployment has no impact on current device safety and effectiveness. Analyses of these activities conclude the benefits associated with the use of the Anchor Tissue Retrieval System™ by CONMED outweigh the residual risks. Material analysis and testing demonstrate the patient contacting materials are biocompatible and comply with the requirements of ISO 10993-1:2009. Redeployment verification testing determined that the Anchor Tissue Retrieval System™ by CONMED can be deployed more than once within a single patient procedure. Performance testing demonstrates that the performance of the Anchor Tissue Retrieval System™ by CONMED is substantially equivalent to the predicate device.
#### Substantial Equivalence -
There are no differences between the predicate device and the proposed device regarding design, intended use, principals of operation and technical characteristics. Supporting information per this premarket submission confirms that the Anchor Tissue Retrieval System™ by CONMED does not raise any new risks of safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.