REALHAND HIGH DEXTERITY INSTRUMENTS

K072715 · Novare Surgical Systems, Inc. · GCJ · Oct 16, 2007 · Gastroenterology, Urology

Device Facts

Record IDK072715
Device NameREALHAND HIGH DEXTERITY INSTRUMENTS
ApplicantNovare Surgical Systems, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateOct 16, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The RealHand™ High Dexterity (HD) instruments are intended for grasping, mobilization, dissection, transection, suturing, and/or electrocautery of tissue under direct and endoscopic visualization.

Device Story

RealHand High Dexterity (HD) instruments are endoscopic surgical tools designed for tissue manipulation. The device features an articulating mechanism to provide high dexterity during procedures. Modifications include an added electroconductive pathway consisting of an external post for connection to a third-party generator via an insulated cable and an internal jumper to conduct current to the instrument shaft. These instruments are used by surgeons in clinical settings to perform hemostatic clamping, electrocautery, and tissue ligation. The device provides the surgeon with enhanced control for grasping, dissection, and suturing under endoscopic visualization. The output is the physical manipulation of tissue and the delivery of electrical current for cauterization, which assists in achieving hemostasis and tissue transection, potentially reducing surgical time and improving precision in minimally invasive procedures.

Clinical Evidence

Bench testing only. The device was evaluated using valid test methodologies to compare electrocautery function and mechanical performance against the original Novare device. Results confirmed that the modified instruments are equivalent in performance and functionality to the predicate devices.

Technological Characteristics

Endoscopic surgical instrument with added electroconductive pathway (post and internal jumper) for electrocautery. Compatible with third-party generators via insulated cables. Mechanical articulation for high dexterity. Class II device (Product Code GCJ).

Indications for Use

Indicated for grasping, mobilization, dissection, transection, suturing, and/or electrocautery of tissue during surgical procedures requiring direct or endoscopic visualization.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOVARE SURGICAL SYSTEMS, INC ummary of Safety Information SEPTEMBER 19, 2007 emarket Notification, Section 510(k) Safe Medical Devices Act of 1990, 21 CFR 807.92 Regulatory Authority: Ko72715 RealHand High Dexterity (HD) Instruments Device Name: Common Name(s): Endoscopic instruments Classification Name(s): Endoscope and Accessories 007 16 2007 Manufacturer: Novare Surgical Systems, Inc. 2954739 Reg. Number: 10440 Bubb Road, Suite A Address: Cupertino, CA 95014 Classification(s): Sec. 876, 1500 Endoscope and accessories (a) Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescope lamps, biliary pancreatoscopes. proctoscopes. resectoscopes. nephroscopes, sigmoidoscopes. ureteroscones. urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classifications in other parts of the device classification regulations. (b) Classification. (1) Class II (performance standards). (2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in 876.9. [48 FR 53023, Nov. 23, 1983, as amended at 61 FR 1122, Jan. 16, 1996, | Device Class: | Class II for the requested indications | |-----------------------|----------------------------------------| | Classification Panel: | Gastroenterology & Urology | | Product Code(s): | GCJ | ### Equivalent Predicate Device: Novare Surgical Systems, Inc. believes that the modified RealHand™ High Dexterity Instruments are substantially equivalent to the following legally marketed device: {1}------------------------------------------------ - Endolink "RealHand" Articulating Instruments, K071488 Novare Surgical Systems, Inc., and . Endolink "True Movement" Articulating Instruments, K043541 - Novare Surgical Systems, Inc. - Pajunk Modular Handled Instruments, K033249 . - . Endopath Endoscopic Instruments, K984240 ### Device Description: A new electroconductive pathway was added. This change consists of two basic modifications to the existing instruments: 1) a post was added to attach a third party generator via an insulated commercially available legally marketed cable. 2) an internal jumper was added to conduct current from the post to the internal shaft of the instrument. The substantial equivalence of the changes are based on a comparison of physical and functional attributes of the device contrasted with the Novare predicate and the indications of the Pajunk instruments. The place of manufacture has not changed. The physical appearance of the device, though slightly different, does not adversely affect the modified device. Functional expectations, material suitability, surgical technique and indications for use are just the same as the comparison devices placed in product code GCI by FDA. RealHand™ High Dexterity Instruments are substantially equivalent to the original instruments with respect to functionality and performance characteristics based the following criteria: 1) RealHand™ High Dexterity Instruments provide hemostatic clamping that allows electrocautery and ligature of tissue. The performance of the systems are essentially identical. 2) The RealHand™ High Dexterity Instruments were carefully evaluated and compared with the original Novare device. Documented testing established that the characteristics are equal to the original. 3) The original device was evaluated for proper function during use. The modified RealHand™ High Dexterity Instruments were evaluated using a valid test methodology, including electrocautery function and the test result was equivalent, and 4) the RealHand™ High Dexterity Instruments worked as intended. #### Company Contact: Mr. David Danitz Novare Surgical Systems, Inc. 10440 Bubb Road. Suite A Cupertino, CA 95014 408.873.3161 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## OCT 1 6 2007 Novare Surgical Systems, Inc % Buckman Company, Inc Mr. David w. Schlerf Vice President 1070 Concord Avenue, Suite 230 Concord, California 94520 Re: K072715 Trade/Device Name: RealHand™ High Dexterity (HD) Instruments Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: September 19, 2007 Received: September 25, 2007 Dear Mr. Schlerf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. David w. Schlerf This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, مع Mark N. Melkerson Director 16 Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Page 3 – Mr. David w. Schlerf cc: HFZ-401 DMC HFZ-401 Divic HFZ-404 510(k) Staff HFZ-410 DGRND/GSDB D.O. . . . D.O. f/t:RPW:kxl:10-04-07 2017 2017 2017 2014 07 2016 10 :RPW:kxl:10-04-07 and the comments of the comments of the comments of . OC Numbers: | Division of Enforcement A | 240-276-0115 | |-----------------------------------------------------------|--------------| | Dental, ENT and Ophthalmic Devices Branch | 240-276-0115 | | OB/GYN, Gastro. & Urology Devices Branch | 240-276-0115 | | General Hospital Devices Branch | 240-276-0115 | | General Surgery Devices Branch | 240-276-0115 | | Division of Enforcement B | 240-276-0120 | | Cardiovascular & Neurological Devices Branch | 240-276-0120 | | Orthopedic, Physical Medicine & Anesthesiology Devices Br | 240-276-0120 | : . . . . . . . . : : {5}------------------------------------------------ 510(k) Number :_ | צורגרס | צ Page _ 1 _ _ _ _ of of RealHandTM High Dexterity (HD) Instruments Device Name(s): # Indications for Use Statement(s): The RealHand™ High Dexterity (HD) instruments are intended for grasping, mobilization, dissection, transection, suturing, and/or electrocautery of tissue under direct and endoscopic visualization. Prescription Use -------X OR Over-The-Counter Use PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE) Rela R (Division Sign-Off) Division of General, Restorative, and Neurological Devices **510(k) Number** k072715 (Per 21 CFR 801.109) (Optional format 1-2-96) Cautery2.Doc ાં ર
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...