OVERSTITCH ENDOSCOPIC SUTURE SYSTEM

K081853 · Apollo Endosurgery, Inc. · OCW · Aug 18, 2008 · Gastroenterology, Urology

Device Facts

Record IDK081853
Device NameOVERSTITCH ENDOSCOPIC SUTURE SYSTEM
ApplicantApollo Endosurgery, Inc.
Product CodeOCW · Gastroenterology, Urology
Decision DateAug 18, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Apollo Endosurgery OverStitch Endoscopic Suture System is intended for endoscopic placement of suture(s) and approximation of soft tissue.

Device Story

The OverStitch Endoscopic Suturing System is a mechanical device used by physicians to perform tissue apposition within the gastrointestinal tract and peritoneal cavity. The system enables endoscopic placement of sutures and soft tissue approximation. A key feature allows the surgeon to reload the suture without removing the endoscope. The device is operated by a physician in a clinical setting. It functions as a mechanical tool for suturing, providing a benefit by facilitating endoscopic tissue repair and closure without the need for more invasive surgical procedures.

Clinical Evidence

Bench testing only. Testing included mechanical strength, bond strength, suture operation, suture securement, and visual/dimensional inspection. Performance was compared to commercially available predicate devices, demonstrating equivalence.

Technological Characteristics

Mechanical endoscopic suturing system. Materials are biocompatible and previously cleared for similar use. The device features a reloadable suture mechanism. No software, electronics, or energy sources are utilized.

Indications for Use

Indicated for endoscopic placement of suture(s) and approximation of soft tissue in the gastrointestinal tract and peritoneal cavity.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081853 page 1/2 #### 510(k) Summary AUG 1 8 2008 General Information Classification Class II Trade Name OverStitch Submitter Apollo Endosurgery, Inc. 7000 Bee Caves Road Suite 350 Austin, Texas 78746 Tel: (512) 328-9990 Contact Dennis Mc Williams President & CEO ### Intended Use The Apollo Endosurgery OverStitch Endoscopic Suture System is intended for endoscopic placement of suture(s) and approximation of soft tissue. ## Predicate Devices | K061770 | Endoscopic Suturing System (ESS)<br>Ethicon Endo Surgery | |---------|----------------------------------------------------------| | K003956 | EndoCinch Suturing System<br>Davol / Bard | | K972911 | Auto Suture-Endo Stitch<br>Tyco Healthcare | #### Device Description The Apollo Endosurgery OverStitch Endoscopic Suture System provides physicians the ability to perform several different types of tissue apposition within the Gastrointestinal {1}------------------------------------------------ K081893 page 2/2 (GI) Tract and peritoneal cavity. Additionally, the system allows the surgeon to 'reload' the suture without the need for removing the endoscope. ### Materials All materials used in the manufacture of the OverStitch are suitable for this use and have been used in numerous previously cleared products. #### Testing Product testing was conducted to evaluate conformance to product specification. Testing included mechanical strength testing, bond strength, suture operation, suture securement and visual / dimensional inspection. Suture placement and securement testing was performed by comparing the OverStitch to commercially available predicate products. The products were used per their respective Instructions for Use. The results showed the OverStitch was equivalent to the predicate devices. ### Summary of Substantial Equivalence The OverStitch is equivalent to the predicate products. The indications for use, basic overall function, methods of manufacturing, and materials used are substantially equivalent. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is depicted in black, and the text is also in black. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Apollo Endosurgery, Inc. % Mr. Dennis McWilliams President & CEO 7000 Bee Caves Road, Suite 350 Austin, Texas 78746 AUG 1 8 2008 Re: K081853 Trade/Device Name: OverStitch Endoscopic Suturing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCW, HCF Dated: June 26, 2008 Received: June 30, 2008 Dear Mr. McWilliams: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced a bove and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 8C7); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ #### Page 2 - Mr. Dennis Mc Williams This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Pretmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Millerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K081853 # Indications for Use 510(k) Number (if known): Device Name: Indications for Use: OverStitch Endoscopic Suturing System The Apollo Endosurgery OverStitch Endoscopic Suture System is intended for endoscopic placement of suture(s) and approximation of soft tissue. Prescription Use X (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of Device Evaluation (ODE) (Division Sign-Off) Restorative. (Division Sign-On) Restorative, Division of Neurological Devices 510(k) Number Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...