← Product Code [NQQ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/NQQ) · K251730

# LIA Console  (542-7) (K251730)

_Leadoptik, Inc. · NQQ · Dec 19, 2025 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NQQ/K251730

## Device Facts

- **Applicant:** Leadoptik, Inc.
- **Product Code:** [NQQ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/NQQ.md)
- **Decision Date:** Dec 19, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The LIA Console is indicated for use as an imaging tool in the evaluation of human tissue microstructure in the tracheobronchial tree by providing two-dimensional, cross-sectional, real-time depth visualization using Optical Coherence Tomography (OCT). The safety and effectiveness of this device for diagnostic analysis (i.e., differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated

## Device Story

LIA Console is a mobile Swept-Source Optical Coherence Tomography (SS-OCT) imaging system used in conjunction with the LIA-1 Catheter. The system includes a 24-inch touch display, a Catheter Driving Unit (CDU) with a rotary motor for 360-degree imaging, a stabilizing Catheter Driving Arm (CDA), and an image engine containing a laser source and computing unit. The device transmits infrared light to tissue via the catheter and processes backscattered optical signals to generate real-time, cross-sectional 2D images of tissue microstructure. Operated by clinicians in a clinical setting, the system provides visual feedback to assist in endobronchial biopsy procedures. The output is displayed on the console monitor for real-time evaluation of tissue structure. The device benefits patients by providing high-resolution visualization of the tracheobronchial tree to support procedural guidance.

## Clinical Evidence

No human clinical data. Evidence includes bench testing (OCT sensitivity, resolution, NURD, optical power, metalens profile) and a GLP porcine study (n=2) evaluating 32 fiducial lung nodules. The porcine study demonstrated 100% biopsy success rate and real-time imaging capability in airways 1mm-10mm in diameter with no complications.

## Technological Characteristics

Swept-Source OCT (SS-OCT) imaging system. Components: mobile console, articulating Catheter Driving Arm, Catheter Driving Unit with Fiber Optical Rotary Joint (FORJ). Laser: 1300 nm center wavelength, Class 1M Laser Product (IEC 60825-1). Connectivity: proprietary OCTICA software. Electrical safety: ANSI/AAMI ES 60601-1, IEC 60601-1-2. Ingress protection: IPX1. Service life: 3 years.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- NvisionVLE Imaging System ([K182616](/device/K182616.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 19, 2025

LEADOPTIK Inc.
% Lorry Weaver
Principal Consultant
Qserve Group
350 S. Main Street, Suite 309,
Doylestown, Pennsylvania 18901

Re: K251730
Trade/Device Name: LIA Console (542-7)
Regulation Number: 21 CFR 892.1560
Regulation Name: Ultrasonic Pulsed Echo Imaging System
Regulatory Class: Class II
Product Code: NQQ
Dated: June 5, 2025
Received: June 6, 2025

Dear Timothy Joiner:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251730 - Timothy Joiner
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K251730 - Timothy Joiner
Page 3

Sincerely,

JESSICA CARR -S

Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K251730  |   |
|  Device Name LIA Console (542-7)  |   |
|  Indications for Use (Describe) The LIA Console is indicated for use as an imaging tool in the evaluation of human tissue microstructure in the tracheobronchial tree by providing two-dimensional, cross-sectional, real-time depth visualization using Optical Coherence Tomography (OCT). The safety and effectiveness of this device for diagnostic analysis (i.e., differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

LEADOPTIK Inc.
LIA Console
Traditional 510(k)

# Section 5: 510(k) Summary – [21 CFR 807.92]

807.92(a)(1), (2), (3)

Date Prepared: December 12, 2025

Submitter's Name: LEADOPTIK, Inc.
2380 Qume Dr., STE A
San Jose, CA 95131
info@leadoptik.com

Official Correspondent: Lorry Weaver, Principal Consultant
Qserve Group US
+1 (916) 220-1137
lorry.weaver@qservegroup.com

Device Trade Name: LIA Console

Common Name: Optical Coherence Tomography (OCT) Imaging System

Review Panel: General &amp; Plastic Surgery/Radiology

Classification Name: Ultrasonic pulsed echo imaging system

Regulation No.: 892.1560

Classification Code: NQQ

Predicate Device: NvisionVLE Imaging System: 510(k) K182616
Classification Code: NQQ
Regulation: 892.1560

# Device Description 807.92(a)(4):

The LIA Console is indicated for use as an imaging tool in the evaluation of human tissue microstructure in the tracheobronchial tree by providing two-dimensional, cross-sectional, real-time depth visualization using Optical Coherence Tomography (OCT). The LIA Console is intended to be used only in conjunction with the LIA-1 Catheter in order to function as intended.

LIA Console consists of the following main parts:

1. Monitor: 24-inch multi-touch display monitor allowing software user interface.
2. Catheter Driving Unit (CDU): Electromechanical and optical interface between the LIA Console and the LIA-1 Catheter. A rotary motor inside drives the LIA-1 Catheter to form 360-degree side view real-time imaging.
3. Catheter Driving Arm (CDA): A supporting arm allowing CDU positioning, mobility and flexibility during the procedure.
4. Image Engine: The main body of the LIA Console contains laser source, computing unit, data acquisition and power distribution module and other components.

LEADOPTIK, Inc. 2380 Qume Dr., STE A San Jose, CA 95131

{5}

LEADOPTIK Inc.
LIA Console
Traditional 510(k)

5. Caster: Four caster wheels installed at the base provide mobility for the LIA Console. Each caster is equipped with a brake that can lock the LIA Console in place as needed.
6. OCTICA Software: A proprietary GUI-based software that controls data acquisition, rotary motor and other hardware components to enable real-time OCT imaging of tissue microstructure.

## Indications for Use [Intended Use 807.92(a)(5)]:

The LIA Console is indicated for use as an imaging tool in the evaluation of human tissue microstructure in the tracheobronchial tree by providing two-dimensional, cross-sectional, real-time depth visualization using Optical Coherence Tomography (OCT). The safety and effectiveness of this device for diagnostic analysis (i.e., differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.

## Technological Characteristics 807.92(a)(6):

The LIA Console is an imaging tool for the evaluation of human tissue microstructure using Optical Coherence Tomography (OCT). The system is not intended for diagnostic differentiation. Swept Source Optical Coherence Tomography (SS-OCT) is the method for real-time cross-sectional imaging, supporting tissue evaluation.

The LIA Console consists of (i) an imaging console, (ii) a catheter driving arm, and (iii) proprietary LIA Real-Time Imagine Software used to acquire, process, and visualize OCT images. It is used in conjunction with a compatible LIA-1 Catheter during the endobronchial biopsy procedure.

The mobile imaging console contains a light source, interferometer, and computing unit. It processes optical signals received from the LIA-1 Catheter to generate and display real-time images on the monitor.

The Catheter Driving Arm (CDA) includes an articulating stabilization arm and a driving module including an optical receptacle to which the LIA-1 is connected. The driving module drives the imaging stylet to form a 360-degree 2D image. The arm stabilizes the imaging probe throughout the procedure.

The Catheter Driving Unit (CDU) serves as the optomechanical interface between the LIA-1 Catheter and the imaging engine. It includes a Fiber Optical Rotary Joint (FORJ) that transmits optical signals to the LIA-1 Catheter and receives the backscattered signal after interaction with tissue. The FORJ, in conjunction with a motor, also rotates LIA-1 Catheter's optical probe to generate a 360-degree image.

As shown in Table 1 below, the LIA Console is substantially equivalent to the predicate devices, as it has the same intended use and its technological characteristics are similar to, and substantially equivalent to, the cited predicates.

LEADOPTIK, Inc. 2380 Qume Dr., STE A San Jose, CA 95131

{6}

LEADOPTIK Inc.
LIA Console
Traditional 510(k)

Table 1. Predicate Comparison

|  Devices | Subject Device
LIA Console
(K251730) | Predicate Device
NvisionVLE® Imaging System
(K182616)  |
| --- | --- | --- |
|  Regulation
Product
Code | 892.1560
NQQ | 892.1560
NQQ  |
|  Indications
for Use | The LIA Console is indicated for use as an imaging tool in the evaluation of human tissue microstructure in the tracheobronchial tree by providing two-dimensional, cross-sectional, real-time depth visualization using Optical Coherence Tomography (OCT).

The safety and effectiveness of this device for diagnostic analysis (i.e., differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated. | The NvisionVLE® Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross-sectional, real-time depth visualization and may be used to mark areas of tissue. The software provides segmentation and display of common imaging features, including hyper-reflective surface, layering, and hypo-reflective structures.

The NvisionVLE® Imaging System is intended to provide an image of tissue microstructure. The safety and effectiveness of this device for diagnostic analysis (i.e. differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.  |
|  Device
Description | The LIA Console is indicated for use as an imaging tool in the evaluation of human tissue microstructure in the tracheobronchial tree by providing two-dimensional, cross-sectional, real-time depth visualization using Optical Coherence Tomography (OCT).

LIA Console is intended to be used only in conjunction with the LIA-1 Catheter in order to function as intended.

LIA Console consists of the following main parts:
(1) Monitor: 24-inch multi-touch display monitor allowing software user interface.
(2) Catheter Driving Unit (CDU): Electromechanical and optical interface between the LIA Console and the LIA-1 Catheter. A rotary motor inside drives the LIA-1 Catheter to form 360-degree side view real-time imaging. (3) Catheter Driving Arm (CDA): A supporting arm allowing CDU positioning, mobility and flexibility during the procedure. (4) Image Engine: | The NinePoint Medical NvisionVLE® Imaging System is a high-resolution volumetric imaging system based on optical coherence tomography (OCT). In an analogous fashion to ultrasound imagery, OCT images are formed from the time delay and magnitude of the signal reflected from the tissue of interest. The NvisionVLE Imaging System employs an advanced form of OCT known as sweptsource OCT (SS-OCT), or Optical Frequency Domain Imaging (OFDI), in combination with a scanning optical probe to acquire high-resolution, cross-sectional, real-time imagery of tissue called Volumetric Laser Endomicroscopy (VLE).

In addition to the imaging capability, the device provides a means of marking areas of tissue with an additionally integrated 1470 nm laser. The ability to create temporary  |

LEADOPTIK, Inc. 2380 Qume Dr., STE A San Jose, CA 95131

{7}

LEADOPTIK Inc.
LIA Console
Traditional 510(k)

|   | The main body of the LIA Console contains laser source, computing unit, data acquisition and power distribution module and other components. (5) Caster: Four caster wheels installed at the base provide mobility for the LIA Console. Each caster is equipped with a brake that can lock the LIA Console in place as needed. (6) OCTICA Software: A proprietary GUI software that controls data acquisition, rotary motor and other hardware to enable real-time OCT imaging of tissue microstructure. | laser marks directly on tissue enables a clinician to place visual reference marks on tissue regions of clinical interest immediately following their identification via VLE. The device consists of the following five main components and accessories: (i) a mobile NvisionVLE Console with an integrated computer and two touch-screen interfaces; (ii) proprietary NvisionVLE Software used to acquire, process, and visualize VLE images; (iii) a single-use, sterile NvisionVLE Marking Probe that is inserted through the working channel of an endoscope; (iv) a single-use, sterile NvisionVLE Inflation System that is used to inflate the Marking Probe's balloon to facilitate placement; and (v) a Probe Lock Accessory to prevent longitudinal motion of the Marking Probe within the endoscope.  |
| --- | --- | --- |
|  Imaging Modality | Swept-Source OCT (SS-OCT) | Swept-Source OCT (SS-OCT)  |
|  System Components | Imaging System reusable electrical mechanical unit. Disposable imaging catheter. | Imaging System reusable electrical mechanical unit. Disposable imaging catheter.  |
|  Primary Function s | Delivers energy (infrared light) to the tissue. Measures the depth and pattern of reflections from the tissue from the return near infrared light to create high-resolution, real-time images. Stores images for evaluation and review | Delivers energy (infrared light) to the tissue. Measures the depth and pattern of reflections from the tissue from the return near infrared light to create high-resolution, real-time images. Stores images for evaluation and review  |
|  Software | OCTICA – IEC 62304 | Proprietary, artificial intelligent technology  |
|  Image Creation, Display and Storage | Process reflected optical signals to construct images. Display images. Store images. | Process reflected optical signals to construct images. Volumetric Laser Endomicroscopy (VLE), Display images. Store images.  |
|  Imaging Probe | Single use | Single use  |
|  A-line rate | 100 kHz | 20 kHz  |
|  Laser Center Wavelength | 1300 nm | 1305 nm  |
|  Console Feature s | Mobile Console | Mobile Console  |
|  Safety | ES60601-1:2005(R)2012 | IEC 60601-1  |
|  EMC | IEC 60601-1-2 | IEC 60601-1-2  |
|  Optical/laser safety | IEC 60825-1:2014 to be a Class 1M Laser Product | IEC 60825-1  |

LEADOPTIK, Inc. 2380 Qume Dr., STE A San Jose, CA 95131
Page 4 of 6

{8}

LEADOPTIK Inc.
LIA Console
Traditional 510(k)

# Summary of Non-Clinical Test - Performance and Safety Testing 807.92(b)(1):

LIA Console was evaluated and found to meet performance standards and requirements for general safety, electrical, electromagnetic, transportation and software development.

# Electrical Safety and EMC

LIA Console meets the requirements of the relevant elements of the following electrical safety and EMC standards.

- ANSI/AAMI ES 60601-1:2005 [Including Amendment 2 (2021)] &amp; EN/IEC60601-1-6:2013 Medical electrical equipment, Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014/AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

# Packaging / Distribution Testing

LIA Console has been tested and follows ASTM D4169 Standard Practice for Performance Testing of Shipping Containers and Systems.

# Laser Classification

LIA Console has been determined under IEC 60825-1:2014 to be a Class 1M Laser Product.

# Fluid Ingress Testing

LIA Console has been tested under IEC 60529, Edition 2.2 - 2013-08 and classified as IPX1.

# Service Life Testing

The service life of the LIA Console is expected to be 3 years from the date of installation.

# Software Testing

LIA Console and Application Software (OCTICA) were developed and tested in compliance with IEC 62304. Software verification and validation were conducted to FDA regulations, standards, and guidance document requirements. The results of this testing conclude the software has met these requirements.

# Bench Testing

LEADOPTIK Inc. performed a series of bench tests to demonstrate LIA Console meets its performance specifications using production-equivalent, finished, product. Comprehensive verification and validation activities were successfully completed,

LEADOPTIK, Inc. 2380 Qume Dr., STE A San Jose, CA 95131

{9}

LEADOPTIK Inc.
LIA Console
Traditional 510(k)

raising no new issues of safety or effectiveness. Tests included, but not limited to, OCT sensitivity characterization, service life, fluid ingress, lateral and axial resolution, NURD measurement, optical power efficiency, metalens field profile analysis. All testing passed the acceptance criteria.

# Summary of Pre-clinical Tests 807.92(b)(1):

A GLP porcine study with Fiducial Lung Nodule (FLN, ≤20mm) was conducted to evaluate human tissue microstructure in the tracheobronchial tree. Thirty two FLNs were administrated into airway system of two pigs (16 FLNs per animal). The data show that LIA System (LIA Console and LIA-1 Catheter) can provide real-time imaging of tissue microstructure in the tracheobronchial tree with diameters ranging from 1mm-10mm. Additionally, the data show the biopsy success rate using the LIA System was 100% for all 32 FLNs. There were no complications found during the study.

Therefore, the GLP porcine study concludes LIA System is substantially equivalent to the predicate device and meets the intended use.

# Summary of Clinical Tests 807.92(b)(2):

Clinical performance testing was not required as simulated use validation demonstrated intended use performance.

# Conclusion 807.92(b)(3):

LIA Console and the predicate device, NvisionVLE® Imaging System (K182616) are both devices that "project a beam into body tissue to determine the depth or location of the tissue interfaces" according to regulation 892.1560. Both devices are specifically Optical Coherence Tomography (OCT) imaging systems within product code NQQ.

The indications for use and key technological characteristics are substantially equivalent employing swept source Optical Coherence Tomography (SS-OCT) imaging modalities and ensuring comparable real-time tissue visualization for clinical applications

LIA Console was evaluated for electrical, optical, mechanical, and EMC safety and found to conform to applicable mandatory medical device safety standards.

The differences between the devices do not raise new and/or different questions of safety and effectiveness. LIA Console conforms to applicable safety standards and performance data in a simulated clinical workflow (including in K251402, LIA-1 Catheter) validates the intended use. The predicate comparison table (Table 1) and performance testing provided in this 510(k) is sufficient to demonstrate that the LIA Console is substantially equivalent to the legally marketed predicate device, NvisionVLE® Imaging System cleared under 510(k) K182616.

LEADOPTIK, Inc. 2380 Qume Dr., STE A San Jose, CA 95131

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NQQ/K251730](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NQQ/K251730)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
