← Product Code [NQQ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/NQQ) · K160878

# OPTIS Metallic Stent Optimization E.4 SW (K160878)

_Lightlab Imaging, Inc. · NQQ · Jun 29, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NQQ/K160878

## Device Facts

- **Applicant:** Lightlab Imaging, Inc.
- **Product Code:** [NQQ](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/NQQ.md)
- **Decision Date:** Jun 29, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The OPTIS™ Software with Dragonfly™ DUO or Dragonfly™ OPTIS™ Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly™ DUO or Dragonfly ™ OPTIS™ Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly™ DUO or Dragonfly™ OPTIS™ Imaging Cathesters is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The OPTIS™ Software will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.

## Device Story

Software upgrade for existing OCT systems (ILUMIEN OPTIS, OPTIS Mobile, OPTIS Integrated). Inputs: OCT imaging data from Dragonfly catheters; pressure waveform data from distal intracoronary and proximal aortic transducers. Processing: OCT image acquisition, scan conversion (polar to raster via bilinear interpolation), FFR computation (mean pressure values over heart cycles), and proprietary image processing for automatic segmentation of vessel lumen, side branches, guidewires, and metallic stent struts. Outputs: 2D/3D OCT images, stent apposition maps, stent roadmaps, and FFR values. Used in clinical settings by physicians for intra-procedural guidance during interventional procedures. Benefits: enhanced visualization of ostium geometry, stent strut apposition, and co-registration of OCT/angiography to assist in therapeutic decision-making.

## Clinical Evidence

Bench testing only. Software verification and validation conducted in compliance with IEC 62304:2006 and DICOM 2015b standards. No clinical data presented.

## Technological Characteristics

Software-based OCT imaging and FFR computation system. Runs on 64-bit Windows 7 Embedded. Uses bilinear interpolation for scan conversion. Integrates COTS libraries (Open Inventor, LEADTOOLS, SQL Server Express 2008). Connectivity via DICOM servers. Features automated segmentation algorithms for vessel structures and metallic stents. Compatible with Dragonfly DUO/OPTIS catheters.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- OPTIS™ Mobile System ([K152120](/device/K152120.md))

## Reference Devices

- ILUMIEN OPTIS ([K150878](/device/K150878.md))
- OPTIS Integrated OCT Systems ([K151286](/device/K151286.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is simple, clean, and easily recognizable.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 29, 2016

Lightlab Imaging, Inc. Padmini Suravaram Regulatory Affairs Specialist 4 Robbins Road Westford, MA 01866

Re: K160878

Trade/Device Name: Optis™ Metallic Stent Optimization E.4 SW Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: NOO Dated: April 18, 2016 Received: April 20, 2016

Dear Padmini Suravaram:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K160878

Device Name OPTIS Metallic Stent Optimization E.4 SW

#### Indications for Use (Describe)

The OPTIS™ Software with Dragonfly™ DUO or Dragonfly™ OPTIS™ Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly™ DUO or Dragonfly ™ OPTIS™ Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly™ DUO or Dragonfly™ OPTIS™ Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.

The OPTIS™ Software will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K160878

## Lightlab Imaging, Inc. OPTIS METALLIC STENT OPTIMIZATION E.4 SW 510(k) Summary (Per 21 CFR 807.92(c))

#### 1. SPONSOR/MANUFACTURER

LightLab Imaging, Inc. 4 Robbins Road Westford, MA 01886

| Contact Person: | Padmini Suravaram |
|-----------------|-------------------|
| Telephone:      | 978-577-3483      |

Date Prepared: 03/28/16

#### 2. DEVICE NAME

| Proprietary Name:    | OPTIS™ Metallic Stent Optimization E.4 SW |
|----------------------|-------------------------------------------|
| Common/Usual Name:   | Ultrasonic pulsed echo imaging system     |
| Classification Name: | Ultrasonic pulsed echo imaging system     |

#### 3. PREDICATE DEVICE

Predicate Device: Lightlab Imaging, Inc. OPTIS™ Mobile System, K152120

#### 4. DEVICE DESCRIPTION

The Metallic Stent Optimization SW E.4 is an OCT system software that allows update of the current software version of ILUMIEN OPTIS (K150878), OPTIS Mobile System (K152120) and OPTIS Integrated OCT Systems (K151286) on the field for the Stent Optimization Tool features. The Metallic Stent Optimization SW E.4 is also compatible with flush catheters compatible with the above mentioned OCT systems (Dragonfly OPTIS and Dragonfly DUO catheters).

The OCT system software is used to control the OCT imaging engine in order to acquire OCT images, it is used to make measurements based on acquired OCT images, it is used to maintain a database of OCT images and data, and it is used to review previously acquired OCT images and data.

{4}------------------------------------------------

# K160878

The OCT System software is also used to control the Aortic (AO) and PressureWire (PW) USB receivers to collect and store pressure waveform data for computing Fractional Flow Reserve (FFR). Similar to the OCT images, the FFR waveforms are stored in the database and are available for later review.

The OPTIS™ Metallic Stent Optimization E.4 SW software runs on 64-bit Microsoft Windows 7 Embedded Operating System only. The software also contains Intel JPEG Library, Microsoft MPEG-4 Video Codec, Microsoft Video 1 codec, AMD FirePro V4900 Display Adapter Driver, Intel 82579LM/82583V Gigabit Ethernet Controller Driver, Microsoft Foundation Class Library, Active Template Library, Microsoft Standard C++ Library, LEADTOOLS Medical Imaging Suite SDK, Microsoft SQL Server Express 2008 Service Pack 2 (SP2), Open Inventor Library, Haru PDF Library, IDEA Video Driver, iWRAP Bluetooth Stack and Intel Integrated Performance Primitives commercial off the shelf (COTS) software.

The E.4 software introduces the following new features:

- . 3D Flythrough, 3D Bifurcation (Bifurcation Visualization)- Enhance assessment of ostium geometry with optimized 3D views
- Cross Frame Angio (Stent Roadmap) Display stent roadmap with . OCT/angiography coregistrations
- Stent Display (Quantification of Stent Strut Apposition, Metallic Stent . Visualization) and 3D Display Options (Guide Wire Detection, Side Branch Detection) - Provide intra-procedural guidance by automatic detection and 3D visualization of BMS struts, side branches, and guidewires

### INDICATIONS FOR USE

The OPTIS™ Software with Dragonfly™ DUO or Dragonfly™ OPTIS™ Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly™ DUO or Dragonfly ™ OPTIS™ Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The DragonflyTM DUO or DragonflyTM OPTIS™ Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.

The OPTIS™ Software will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise 

{5}------------------------------------------------

and clinical judgment to determine if therapeutic intervention is indicated.

## 5. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The OPTIS™ Metallic Stent Optimization E.4 SW is substantially equivalent in terms of product design, materials of construction, operational and technological features, clinical use, and target population.

Both the proposed OPTIS™ Metallic Stent Optimization E.4 SW and the predicate device OCT System Software perform OCT Image acquisition and FFR. This application is preformed through the System OCT Software. The OPTIS application software and firmware is responsible for controlling the OCT system hardware for the purposes of acquiring raw OCT image data in polar format and scan converting it to raster format for display on the screen using a bilinear interpolation algorithm. This data is stored in a proprietary variant of the TIFF format. The software is also responsible for controlling the FFR hardware for the purposes of acquiring raw aortic and distal pressure waveforms, computing the mean pressure values over a user-defined number of heart cycles, and computing the resulting Fractional Flow Reserve and/or Resting Pd/Pa parameters from the mean pressure waveforms. For both modalities, the software provides storage and retrieval functionality including database storage of the corresponding patient demographic information using the commercial off the shelf SQL Server Express database. The software also enables storage of the OCT and FFR data in a variety of file formats on external DICOM servers.

In addition to providing the basic OCT and FFR modalities, the software contains a number of proprietary OCT image processing algorithms for automatically calibrating the images and segmenting the vessel lumen, side branches, guide wires and metallic stent struts, using a commercial off the shelf volume rendering library (Open Inventor) to provide an interactive 3D display of the image data and segmented objects in a variety of formats. The software also contains a number of proprietary angio image processing algorithms to track the Dragonfly Duo and OPTIS lens marker to provide a complete two dimensional Co-Registration between each OCT frame and each angio frame from simultaneously acquired image sequences.

The OPTIS™ Metallic Stent Optimization E.4 SW represents an upgrade to the predicate device OCT System software in terms of offering the following features:

- . Metallic Stent optimization and Stent Apposition Mapping
- . Stent Roadmap on Angio Co-registration
- Enhanced 3D Views ●

{6}------------------------------------------------

- Lumen Profile default is on ●
- Side-branch display o
- . Guide wire display
- . OCT/Angio export
- . Rotational stabilization improvement
- . Intravascular OCT (IVOCT) DICOM file format
- . Improved Coregistration Workflow
- . Improved OCT Frame Indicator on Angio
- . Bookmarks Displayed on Angio
- . New Home Screen
- Improved DICOM Export Workflow .
- XA added to DICOM Modality Worklist Query .
- On-Screen Display of Lumen Dimensions o
- Scrolling of Full Screen 3D Display

#### 6. PERFORMANCE TESTING

The OPTIS™ Metallic Stent Optimization E.4 SW has been developed and tested in compliance with IEC 62304: 2006 and DICOM Standard: 2015b. Software verification and validation was conducted to FDA regulations, standards and guidance document requirements. The results of this testing conclude the software has met these requirements. Design verification and validation was also performed in compliance with internal design control procedures.

The results of this testing conclude the OPTIS™ Metallic Stent Optimization E.4 SW is determined to be safe and effective and is substantially equivalent to the predicate OPTIS Mobile System software.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NQQ/K160878](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NQQ/K160878)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
