K151286 · Lightlab Imaging, Inc. · NQQ · Aug 5, 2015 · Radiology
Device Facts
Record ID
K151286
Device Name
OPTIS Integrated System Mobile Workstation
Applicant
Lightlab Imaging, Inc.
Product Code
NQQ · Radiology
Decision Date
Aug 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The OPTIS Integrated System with Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The OPTIS Integrated System will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
Device Story
System performs optical coherence tomography (OCT) and fractional flow reserve (FFR) assessment. Near-infrared light emitted via fiber-optic catheter; reflected light collected/processed by imaging engine; converted to electrical signals; software generates real-time coronary artery images. FFR calculated from distal intracoronary and proximal aortic pressure transducer inputs. Used in cath labs by physicians/technicians. Mobile Workstation (MWS) accessory provides wireless remote control of OCT procedures and patient data entry outside sterile field. Output assists physicians in visualizing coronary anatomy and assessing physiological significance of lesions to guide interventional decisions.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-6, IEC 60601-2-18, and IEC 60601-1-2. Hardware and system-level design verification confirmed performance specifications met.
Indicated for patients who are candidates for transluminal interventional coronary procedures. Dragonfly catheters intended for vessels 2.0-3.5 mm diameter. Contraindicated for left main coronary artery or vessels with previous bypass procedures.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES - USA
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2015
Lightlab Imaging, Inc. % Erdie De Peralta Regulatory Director Lightlab Imaging,inc. 4 Robbins Road Westford, Massachusetts 01866
Re: K151286
Trade/Device Name: OPTIS Integrated System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: NOO Dated: May 12, 2015 Received: May 14, 2015
Dear Erdie De Peralta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 4. INDICATIONS FOR USE
510(k) Number (if known): K151286
Device Name: OPTIS Integrated System
Indications for Use:
The OPTIS Integrated System with Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The OPTIS Integrated System will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 510(K) SUMMARY ട്.
# For the LightLab Imaging, Inc. OPTIS Integrated System (per 21CFR 807.92)
## 1. SUBMITTER/510(K) HOLDER
LightLab Imaging, Inc. 4 Robbins Road Westford, MA 01886
| Contact Person: | Erdie De Peralta |
|-----------------|------------------|
| Telephone: | 978-577-3481 |
| Date Prepared: | 5th August 2015 |
## 2. DEVICE NAME
| Proprietary Name: | OPTIS Integrated System |
|----------------------------|---------------------------------------|
| Common/Usual Name: | OPTIS Integrated System |
| Classification Name: | Ultrasonic pulsed echo imaging system |
| Product Code: | NQQ |
| Classification Regulation: | 21 CFR 892.1560 |
#### 3. PREDICATE DEVICE
- OPTIS Integrated System, Dragonfly OPTIS Imaging Catheter manufactured by ● LightLab Imaging, Inc.K141769
#### 4. DEVICE DESCRIPTION
The OPTIS Integrated System performs optical coherence topography (OCT) and fractional flow reserve (FFR) procedures and provides images of the coronary arteries in patients who are candidates for transluminal interventional procedures. The device utilizes fiber-optic technology to emit near infrared light and receive light reflected from coronary tissue in order to produce high resolution, real time images. The imaging engine generates wavelength scanning light, which is guided to the DOC and the catheter. The reflection is collected and sent back to the engine. The engine processes the optical signal and converts it to electrical signal, which is then fed into the Host PC. The software application processes the signal and generates OCT images. The device is compatible with Dragonfly OPTIS, Dragonfly Duo and
{4}------------------------------------------------
PressureWire Aeris.
The OPTIS Integrated System configuration has the following components integrated into a single catheterization lab.
| Component | Location |
|------------------------------------|---------------------------------------------------------------------|
| Laser and<br>Engine | Contained within the system cabinet (M5a) |
| PC Embedded<br>Software | Contained within the system cabinet |
| DOC | Located on the table side,<br>connected directly to the DOC Holster |
| Monitor<br>Keyboard<br>Video Mouse | Located in the control room |
| System<br>Cabinet | Located within the control room or technical closet |
| Remoting<br>Cable | The cable connects the DOC Holster to the system cabinet |
| DOC Holster | Located table side |
| Tableside<br>Controller | Located table side |
The OPTIS Integrated System Mobile Workstation (MWS) is an accessory for the OPTIS Integrated System. The MWS is a mobile terminal that mirrors the functionality of the control room monitor, keyboard and mouse that is part of the OPTIS Integrated System. The Control room monitor, keyboard and mouse of OPTIS Integrated System is a desktop setup located in the control room for operating system controls.
The MWS consists of two major subassemblies:
- MWS console is comprised of a cart, keyboard, mouse, monitor, wireless . Keyboard/Video/Mouse (KVM) receiver, cabling, branding labels and compliance label.
- MWS transmitter is comprised of a wireless KVM transmitter, video splitter, power supplies and cabling. All items are shipped together in a single carton.
# 5. INTENDED USE
The OPTIS Integrated System with Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly DUO or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly
{5}------------------------------------------------
DUO or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The OPTIS Integrated System will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
## 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
# OPTIS Integrated System Mobile Workstation
The OPTIS Integrated System with Mobile Workstation accessory is equivalent to the OPTIS Integrated System predicate device in that both devices perform optical coherence tomography (OCT) and fractional flow reserve (FFR) procedures and provide images of the coronary arteries in patients who are candidates for transluminal interventional procedures. Both devices utilizes fiber-optic technology to emit near infrared light and receive light reflected from coronary tissue in order to produce high resolution, real time images. The imaging engine in both devices generates wavelength scanning light, which is guided to the Digital Optical Coupler (DOC) and the catheter. The reflection is collected and sent back to the engine. The engine processes the optical signal and converts it to electrical signal, which is then fed into the Host PC. The software application processes the signal and generates OCT images. Both devices are compatible with Dragonfly Duo and Dragonfly OPTIS imaging catheters, in addition to the PressureWire Aeris.
The Mobile Workstation accessory has been added to the OPTIS Integrated System predicate device as an option to the OPTIS Integrated System providing an additional control point to input patient demographics and to control the OCT procedure, when needed, from anywhere in the procedure room outside the sterile field. The Mobile Workstation will communicate with the OPTIS Integrated System via a wireless communication interface capable of supporting video signal transmission and a USB keyboard and mouse interface. The Mobile Workstation functionality can be broken into two major functional blocks. The transmitter connects to the System Cabinet, and the Mobile Station itself which will be powered via AC mains and will otherwise be free of physical connections to the OPTIS Integrated System.
# 7. PERFORMANCE TESTING
The OPTIS Integrated System Mobile Workstation has been tested and is in compliance with IEC 60601-1:2005 + A1: 2012 Medical electrical equipment - Part 1: General requirements
K151286
{6}------------------------------------------------
for basic safety and essential performance, IEC 60601-1-6:2010 + A1: 2013 Medical electrical equipment - Part 1-6: General requirements for safety - Collateral Standard: Usability, IEC 60601-2-18:2009 Medical equipment - Part 2: Particular requirements for the basic safety and essential performance of endoscopic equipment, and IEC 60601-1-2 Ed. 3, Electromagnetic emissions and immunity requirements for medical electrical equipment - Group 1 Equipment. Class A for non-life supporting equipment.
In addition to the Electrical Safety Testing performed, Hardware Design Verification, System Level Design Verification and Transport & Storage Conditions Verification was conducted to FDA regulations, standards, guidance document requirements and internal design control procedures. The results of this testing conclude the OPTIS Integrated System and Mobile Workstation have met these requirements. The Design Verification activities demonstrate that OPTIS Integrated System and the Mobile Workstation comply with the defined design and performance specifications. The addition of Mobile Workstation does not affect the safety, efficacy and performance of OPTIS Integrated System. The results of this testing concludes the OPTIS Integrated System with Mobile Workstation is determined to be safe and effective and is substantially equivalent to the OPTIS Integrated System predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.