K150237 · Lightlab Imaging, Inc. · NQQ · May 5, 2015 · Radiology
Device Facts
Record ID
K150237
Device Name
ILUMIEN, Drageonfly OPTIS Imaging Catheter
Applicant
Lightlab Imaging, Inc.
Product Code
NQQ · Radiology
Decision Date
May 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The ILUMIEN with C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure. The ILUMIEN will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
Device Story
System comprises cart-mounted imaging engine, display monitors, and sterile single-use Dragonfly OPTIS intravascular catheter. Catheter features rotating fiber optic core emitting near-infrared light; reflected light processed to construct OCT images. System also accepts radio frequency signals from distal intracoronary and proximal aortic pressure transducers to calculate Fractional Flow Reserve (FFR). Used in clinical settings by physicians to visualize coronary anatomy and assess physiological significance of lesions. FFR output assists clinical decision-making regarding therapeutic intervention. Patient benefits from high-resolution imaging and physiological assessment to guide interventional procedures.
Clinical Evidence
Bench testing only. Software verification and validation conducted per FDA guidance. Design verification and validation performed in compliance with internal design controls. No clinical trial data presented.
Technological Characteristics
OCT imaging system using near-infrared light. Catheter includes stainless steel torque wire, TiO2 doped window, and RFID. System includes isolation transformer, USB receivers for pressure signals, and GUI. Connectivity: DICOM integration. Standards: UL 60601-1, IEC 60601-1-2, IEC 60825-1, 21 CFR 1040.10, FCC Part 15.
Indications for Use
Indicated for patients undergoing transluminal interventional procedures requiring coronary artery imaging (vessels 2.0-3.5 mm diameter). Contraindicated for left main coronary artery or vessels with previous bypass procedures.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with a wavy line below them.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2015
Lightlab Imaging, Inc. Jeffrey Roberts Principal Regulatory Affairs Specialist 4 Robbins Road Westford, Massachusetts 01886
Re: K150237
Trade/Device Name: ILUMIEN with Dragonfly OPTIS Imaging Catheter Regulation Number: 21 CFR 892.1960 Regulation Name: Radiographic Intensifying Screen Regulatory Class: Class II Product Code: NOO. DOO Dated: March 3, 2015 Received: March 6, 2015
Dear Jeffrey Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### 3. INDICATIONS FOR USE
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) Not yet assigned
Device Name
ILUMIEN with Dragonfly OPTIS Imaging Catheter
#### Indications for Use (Describe)
The ILUMIEN with C7 Dragonfly, Dragonfly OPTIS Imaging Catheter is intended for the intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The II.UMIEN will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
| Type of Use (Select one or both, as applicable) | <span> <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14) Page 1 of 1
{3}------------------------------------------------
# Lightlab Imaging, Inc. ILUMIEN AND DragonflyTM Optis™ Imaging Catheter
# 510(k) Summary (per 21 CFR 807.92(c))
### 1. SPONSOR/MANUFACTURER
LightLab Imaging, Inc. 4 Robbins Road Westford, MA 01886
| Contact Person: | Jeffrey Roberts |
|-----------------|-----------------|
| Telephone: | 978-577-3451 |
1/29/15 Date Prepared:
# 2. DEVICE NAME
| Proprietary Name: | ILUMIEN |
|----------------------|---------------------------------------|
| Common/Usual Name: | Ultrasonic pulsed echo imaging system |
| Classification Name: | Ultrasonic pulsed echo imaging system |
| Proprietary Name: | The Dragonfly OPTIS Imaging Catheter |
| Common/Usual Name: | Diagnostic Intravascular Catheter |
| Classification Name: | Diagnostic Intravascular Catheter |
# 3. PREDICATE DEVICE
| Predicate Device: | Lightlab Imaging, Inc. ILUMIEN, K11120 |
|-------------------|----------------------------------------|
|-------------------|----------------------------------------|
### 4. DEVICE DESCRIPTION
The ILUMIEN is a cart-mounted computer and Imaging Engine (or optical engine) placed inside an ergonomically designed mobile cart with a mains power cable. It also includes the Drive-motor and Optical Controller (DOC), which provides the interconnection between the ILUMIEN System and the Dragonfly Catheter. The cart is
{4}------------------------------------------------
equipped with two display monitors (one for the console operator, and the other for the physician), as well as a keyboard and mouse.
The cart also contains an isolation transformer for electrical safety and includes two AO and PW USB Receivers which accept the distal intracoronary and proximal aortic pressure signals and status information from the AO Interface Unit and PressureWire® Aeris (K080813) respectively, and communicate the FFR data for display on the ILUMIEN system.
The Dragonfly OPTIS Imaging Catheter is a sterile, single-use intravascular catheter consisting of a catheter body external sheath and an internal rotating fiber optic imaging core. The external sheath serves two primary functions: 1) to facilitate placement of the device into the coronary artery and 2) to cover and protect the inner rotating fiber optic imaging core.
The inner rotating fiber optic imaging core emits near infrared light to the tissue and receives reflected light. It is driven by a stainless steel torque wire visible under fluoroscopy and pulled back through the window tube of the external sheath by the DOC. The emitted and returned reflected light are combined and processed by the ILUMIEN Optis System software to construct an OCT image. The patient is never exposed to moving parts as the external sheath completely covers the rotating imaging core.
# 5. INDICATION FOR USE
### ILUMIEN
The ILUMIEN with C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The C7 Dragonfly, Dragonfly DUO, or Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary artery or in a target vessel which has undergone a previous bypass procedure.
The ILUMIEN will further acquire radio frequency signal outputs from both a distal intracoronary pressure transducer and a proximal aortic pressure transducer to determine the physiological parameter, Fractional Flow Reserve (FFR). The physician may use the FFR parameter, along with knowledge of patient history, medical expertise and clinical judgment to determine if therapeutic intervention is indicated.
{5}------------------------------------------------
#### Dragonfly OPTIS Imaging Catheter
The Dragonfly OPTIS Imaging Catheter with the OCT Imaging System is intended for the imaging of coronary arteries and is indicated in patients who are candidates for transluminal interventional procedures. The Dragonfly OPTIS Imaging Catheter is intended for use in vessels 2.0 to 3.5 mm in diameter. The Dragonfly OPTIS Imaging Catheter is not intended for use in the left main coronary.
# 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE ILUMIEN
The ILUMIEN is equivalent to the predicate device in terms of hardware and firmware components. They both contain a DOC which provides the interconnection between the ILUMIEN and the optical imaging catheters that emit near-infrared light to produce high-resolution real-time images. This process is accomplished for both the ILUMIEN and the predicate device through a graphical user interface (GUI) and software control to obtain Optical Coherence Tomography (OCT) imaging modality and fractional flow reserve (FFR) measurements.
The ILUMIEN represents an upgrade to the predicate device in terms of performance through the same hardware and firmware design and technological characteristics. The software has been upgraded to revision D.2 for the following features:
- A curtain GUI interface. ●
- Dragonfly OPTIS Imaging Catheter Support ●
- Continuous calibration for TiO2 doped catheter window ●
- DICOM modality integration ●
- Manual pull back triggering ●
- Improved acquisition workflow display ●
- Resting Pd / Pa
### Dragonfly OPTIS Imaging Catheter
The Dragonfly OPTIS Imaging Catheter is equivalent to the predicate device in terms of hardware components and operational use. They both are comprised of a catheter body external sheath and internal rotating fiber optic imaging core which emits near infrared light to the tissue and receives reflected light. They both are driven by a stainless steel torque wire by the DOC which is connected to the ILUMIEN OCT Imaging System. Thev are both purged through the central catheter with 100% contrast media prior to use.
{6}------------------------------------------------
In both the Dragonfly OPTIS Imaging Catheter and the predate device emitted and returned reflected light are combined and processed by the ILUMIEN software to construct an OCT image.
The Dragonfly OPTIS Imaging Catheter represents an upgrade to the predicate device in terms of performance through the same operational characteristics, and fundamental technological characteristics to include the following:
- TiO2 Doped window
- Flexible proximal end
- Improved break away joint
- Improved purge tube ●
- Dual lumen tip
- High speed proximal end
- 155um fiber ●
- RFID
# 7. PERFORMANCE TESTING
The ILUMIEN hardware and firmware is unchanged however previous testing including compliance with UL Standard No 60601-1, Medical Electrical Equipment Part I: General Requirements for Safety, IEC 60601-1-2 Ed. 2.1, Electromagnetic emissions and immunity requirements for medical electrical equipment - Group 1 Equipment, Class B for non-life supporting equipment, EN 60601-1-2:2007, Electromagnetic emissions and immunity requirements for medical electrical equipment - Group 1 Equipment, Class B for non-life supporting equipment, IEC 60825-1, 2nd, Ed., 2007, SAFETY OF LASER PRODUCTS - Part 1: Equipment classification and requirements, DICOM Standard (PS 3.2-2008), 21 CFR 1040.10, Performance Standards for Light-Emitting Products, Laser Products, and CFR 47 FCC Part 15 Subpart B Class B emissions requirements (USA)
Software verification and validation was conducted to FDA regulations, standards and guidance document requirements. The results of this testing conclude the software has met these requirements. Design verification and validation was also performed on the ILUMIEN and Dragonfly OPTIS Imaging Catheter in compliance with internal design control procedures which included bench testing. The results of this testing concludes the ILUMIEN and Dragonfly OPTIS Imaging Catheter is determined to be safe and effective and is substantially equivalent to the predicate ILUMIEN device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.