← Product Code [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY) · K222839

# EndoWrist Stapler 30 System, EndoWrist Stapler 45 System, SureForm 45 System, SureForm 60 System, 8 mm SureForm 30 System (K222839)

_Intuitive Surgical, Inc. · NAY · Mar 16, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K222839

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY.md)
- **Decision Date:** Mar 16, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The EndoWrist Stapler 30 Staplers and Reloads, EndoWrist Stapler 45 Staplers and Reloads, SureForm 45 Staplers and Reloads, SureForm 60 Stapler and Reloads, and 8 mm SureForm 30 Staplers and Reloads all have the identical intended use to resect, transect and/or create anastomoses in surgery.

## Device Story

Fully wristed, articulating surgical staplers designed for use with da Vinci Xi and X Surgical Systems. Controlled by surgeon via Surgeon Console. Device places multiple staggered rows of titanium alloy staples in target tissue followed by tissue transection along the staple line. Single-use sterile reloads contain staples; color-coded for leg height and tissue gap. Used in OR by surgeons to perform resection, transection, and anastomosis. Provides mechanical tissue closure and cutting; benefits include precise robotic-assisted tissue management. Subject of this notification is the addition of a contraindication regarding use on necrotic, friable, or altered tissue.

## Clinical Evidence

No clinical data. Usability testing was conducted to demonstrate that surgeons can correctly select and safely use the device according to the revised labeling. Testing met all acceptance criteria.

## Technological Characteristics

Fully wristed, articulating surgical staplers. Materials: Titanium alloy (Ti3Al2.5V) staples. Energy source: Robotic-assisted mechanical actuation via da Vinci Xi/X systems. Form factor: Multi-row staple cartridges (reloads) with varying leg lengths (2.0mm to 4.6mm). Sterilization: Sterile, single-use reloads.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- EndoWrist Stapler 30 Curved-Tip Stapler, EndoWrist Stapler 30 Stapler, EndoWrist Stapler 30 Reloads ([K170508](/device/K170508.md))
- EndoWrist Stapler 45 Stapler, EndoWrist Stapler 45 Curved-Tip Stapler, EndoWrist Stapler 45 Reloads ([K170508](/device/K170508.md))
- SureForm 45 Staplers and Reloads ([K183224](/device/K183224.md) and [K190999](/device/K190999.md))
- SureForm 60 Stapler, SureForm 60 Reloads ([K173721](/device/K173721.md))
- 8 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler, 8 mm SureForm 30 Reloads ([K211997](/device/K211997.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 16, 2023

Intuitive Surgical Inc Amrit Jaggi Senior Regulatory Specialist 1266 Kifer Road Sunnyvale, California 94086

# Re: K222839

Trade/Device Name: EndoWrist Stapler 30 System, EndoWrist Stapler 45 System, SureForm 45 System, SureForm 60 System, 8 mm SureForm 30 System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY, GDW, GAG Dated: February 22, 2023 Received: February 22, 2023

Dear Amrit Jaggi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.03.16
15:07:08 -04'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known)

#### Device Name

Endo Wrist Stapler 45 System, Endo Wrist Stapler 30 System, SureForm 45 System, 8 mm SureForm 30 System

#### Indications for Use (Describe)

The Intuitive Surgical EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler Accessories (including the bladeless obturators) are intended to be used with a compatible da Vinci Surgical System for resection and/or creation of anastomoses in General, Thoracic, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

The Intuitive Surgical EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with a compatible da Vinci Surgical System for resection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pediativ use. The device can be used with staple line or tissue buttressing material.

The Intuitive Surgical SureForm 60 Reloads and accessories are intended to be used with a compatible da Vinci System for resection, transection of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

The Intuitive Surgical SureForm 45 Stapler, SureForm 45 Reloads and other Stapler accessories are intended to be used with a compatible da Vinci Surgical System for resection, and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

The Intuitive Surgical 8 mm SureForm 30 Curved-Tip and 8 mm SureForm 30 Reloads and accessories are intended to be used with a compatible da Vinci Surgical System for resection of vasculature and tissue, and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric Surgery.

| Type of Use (Select one or both, as applicable)                                                                         |                                                                                                                        |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "I". The font is sans-serif and the letters are evenly spaced. The image is simple and clean.

- 5. 510(K) SUMMARY K222839
Image /page/3/Picture/3 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the "I" in the word. The font is sans-serif and the letters are evenly spaced.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "U". The font is sans-serif and the letters are evenly spaced.

# 510(k) Summary As Required by 21 CFR 807.92(c))

| Submitter:        | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Amrit Jaggi<br>Senior Regulatory Specialist<br>Phone Number: 949-690-8799<br>Fax Number: 408-523-8907                                                                                                                                                                                                                                                                                                                                                          |
| Date Prepared:    | March 13, 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name:       | EndoWristTM Stapler 30 Curved-Tip Stapler, EndoWrist Stapler 30<br>Stapler, EndoWrist Stapler 30 Reloads<br>EndoWrist Stapler 45 Stapler, EndoWrist Stapler 45 Curved-Tip<br>Stapler, EndoWrist Stapler 45 Reloads<br>SureFormTM 45 Stapler Curved-Tip, SureForm 45 Stapler, SureForm<br>45 Reloads<br>SureForm 60 Stapler, SureForm 60 Reloads<br>8 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler,<br>8 mm SureForm 30 Reloads                  |
| Common Name:      | System, surgical, computer controlled instrument                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification:   | Class II<br>21 CFR 876.1500, Endoscope and Accessories<br>21 CRF 878.4750, Implantable Staple<br>21 CFR 878.4740, Surgical Stapler                                                                                                                                                                                                                                                                                                                             |
| Product Codes:    | NAY (Endoscope and accessories)<br>GDW (Implantable Staple)<br>GAG (Stapler, Surgical)                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device: | EndoWristTM Stapler 30 Curved-Tip Stapler, EndoWrist Stapler 30<br>Stapler, EndoWrist Stapler 30 Reloads (K170508)<br>EndoWrist Stapler 45 Stapler, EndoWrist Stapler 45 Curved-Tip<br>Stapler, EndoWrist Stapler 45 Reloads (K170508)<br>SureForm 45 Staplers and Reloads (K183224 and K190999)<br>SureForm 60 Stapler, SureForm 60 Reloads (K173721)<br>8 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler,<br>8 mm SureForm 30 Reloads (K211997) |

Device Description

Image /page/4/Picture/5 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "U". The font is sans-serif and the letters are evenly spaced.

{5}------------------------------------------------

# INTUITIVE

The Intuitive Surgical EndoWrist Stapler 30 Staplers, EndoWrist Stapler 45 Staplers, SureForm 45 Staplers, SureForm 60 Stapler, and 8 mm SureForm Staplers are fully wristed, articulating, surgical staplers and are designed for use exclusively with the Intuitive Surgical da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200 Systems). The staplers are controlled by the surgeon using the Surgeon Console of the IS4000/IS4200 Systems. They are intended for resection, transection and/or creation of anastomoses in surgery. The staplers achieve their intended use by placing multiple staggered rows of implantable staples in the target tissue (stapling) followed by cutting of the target tissue along the middle of the staple line (transection). The Reloads consist of a single-use cartridge that contains multiple staggered rows of implantable titanium alloy (Ti3A12.5V) staples. The reloads are sterile, single use devices that are offered in various configurations by product family. Each color represents a different staple leg height and tissue gap for use with various tissue thicknesses. Table 1Reload Specifications outlines the specifications of the reloads.

| Reloads                           | No. of rows | No. of Staples | Unformed<br>Staple Leg<br>Length |
|-----------------------------------|-------------|----------------|----------------------------------|
| EndoWrist Stapler 45 White Reload | 6 row       | 66 staples     | 2.5 mm                           |
| EndoWrist Stapler 45 Blue Reload  | 6 row       | 66 staples     | 3.5 mm                           |
| EndoWrist Stapler 45 Green Reload | 4 row       | 44 staples     | 4.3 mm                           |
| EndoWrist Stapler 30 Gray Reload  | 6 row       | 48 staples     | 2.0 mm                           |
| EndoWrist Stapler 30 White Reload | 6 row       | 48 staples     | 2.5 mm                           |
| EndoWrist Stapler 30 Green Reload | 6 row       | 48 staples     | 4.3 mm                           |
| EndoWrist Stapler 30 Blue Reload  | 6 row       | 48 staples     | 3.5 mm                           |
| SureForm 60 White Reload          | 6 row       | 90 staples     | 2.5 mm                           |
| SureForm 60 Blue Reload           | 6 row       | 90 staples     | 3.5 mm                           |
| SureForm 60 Green Reload          | 6 row       | 90 staples     | 4.3 mm                           |
| SureForm 60 Black Reload          | 6 row       | 90 staples     | 4.6 mm                           |
| SureForm 45 Gray Reload           | 6 row       | 66 staples     | 2.0 mm                           |
| SureForm 45 White Reload          | 6 row       | 66 staples     | 2.5 mm                           |
| SureForm 45 Blue Reload           | 6 row       | 66 staples     | 3.5 mm                           |
| SureForm 45 Green Reload          | 6 row       | 66 staples     | 4.3 mm                           |
| SureForm 45 Black Reload          | 6 row       | 66 staples     | 4.6 mm                           |
| 8 mm SureForm 30 Gray Reload      | 4 row       | 34 staples     | 2.0 mm                           |
| 8 mm SureForm 30 White Reload     | 4 row       | 34 staples     | 2.5 mm                           |
| 8 mm SureForm 30 Blue Reload      | 4 row       | 34 staples     | 3.5 mm                           |

#### Table 1 Reload Specifications

### Intended Use:

The EndoWrist Stapler 30 Staplers and Reloads, EndoWrist Stapler 45 Staplers and Reloads, SureForm 45 Staplers and Reloads, SureForm 60 Stapler and Reloads, and 8 mm SureForm 30 Staplers and Reloads all have the identical intended use to resect, transect and/or create anastomoses in surgery.

{6}------------------------------------------------

# Indications for Use

The Intuitive Surgical EndoWrist Stapler 45 Reloads and other Stapler Accessories (including the bladeless obturators) are intended to be used with a compatible da Vinci Surgical System for resection, transection of anastomoses in General, Thoracic, Gynecologic and Urologic surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

The Intuitive Surgical EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with a compatible da Vinci Surgical System for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pediatric use. The device can be used with staple line or tissue buttressing material.

The Intuitive Surgical SureForm 60, SureForm 60 Reloads and accessories are intended to be used with a compatible da Vinci System for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

The Intuitive Surgical SureForm 45 Stapler, SureForm 45 Reloads and other Stapler accessories are intended to be used with a compatible da Vinci Surgical System for resection, transection, and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery. The device can be used with staple line or tissue buttressing material (natural or synthetic).

The Intuitive Surgical 8 mm SureForm 30 Curved-Tip and 8 mm SureForm 30 Reloads and accessories are intended to be used with a compatible da Vinci Surgical System for resection, transection of vasculature and tissue, and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric Surgery.

## Comparison of Technological Characteristics

The subject devices are identical to their predicate devices in terms of design, technology, and performance specifications.

### Performance Data:

Usability testing was conducted to demonstrate that surgeons can correctly select and safely use the device, according to the revised labeling. Testing met all acceptance criteria.

## Summary:

The subject of this 510(k) is to add a contraindication to warn users against the use of these devices on tissues that are necrotic, friable, or have altered integrity. Based on the intended

Image /page/6/Picture/14 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "U". The font is simple and sans-serif. The word is centered in the image.

{7}------------------------------------------------

# INTUITIVE

use, indications for use, technological characteristics, and performance data, the subject devices EndoWrist Stapler 30 Staplers and Reloads, EndoWrist Staplers and Reloads, SureForm 45 Staplers and Reloads, SureForm 60 Staplers and Reloads, and 8 mm SureForm 30 Staplers and Reloads are substantially equivalent to the currently marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K222839](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K222839)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
