8 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler, SureForm 30 Reloads
K211997 · Intuitive Surgical, Inc. · NAY · Dec 10, 2021 · Gastroenterology, Urology
Device Facts
Record ID
K211997
Device Name
8 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler, SureForm 30 Reloads
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Dec 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Intuitive Surgical 8 mm SureForm 30 staplers and reloads and accessories are intended to be used with a compatible da Vinci Surgical system for resection of vasculature and tissue and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.
Device Story
Disposable, fully wristed, articulating surgical staplers; designed for use with da Vinci Xi and X Surgical Systems (IS4000/IS4200). Surgeon controls device via Surgeon Console. Device places multiple staggered rows of titanium alloy (Ti3Al2.5V) staples into target tissue; cuts tissue along middle of staple line. Reloads available in three configurations (gray, white, blue) for varying tissue thicknesses. Used in OR by surgeons for resection, transection, and anastomosis. Provides consistent staple formation and tissue transection; benefits include reduced manual effort and precise control during robotic-assisted procedures.
Clinical Evidence
No human clinical data. Bench testing verified physical, mechanical, electrical, and user interface requirements. Animal validation (acute and chronic) included staple line performance, hemostasis, leak onset pressure, and torque evaluation in canine and porcine models. Chronic studies (lung lobectomy, lung wedge resection, gastrectomy, small bowel anastomosis, nephrectomy) demonstrated well-healed staple lines and no intra-operative leaks or bleeding at terminal procedures, confirming performance equivalent to predicate.
Technological Characteristics
Disposable, wristed, articulating stapler; titanium alloy (Ti3Al2.5V) staples; EO sterilization. Compatible with da Vinci Xi/X systems. Reloads feature 4 staggered rows (2 per side of cut line). Dimensions: 8 mm diameter. Software-controlled via robotic system interface.
Indications for Use
Indicated for resection and transection of vasculature and tissue and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery using a compatible da Vinci Surgical system.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 10, 2021
Intuitive Surgical, Inc. Amrit Jaggi Senior Regulatory Specialist 1266 Kifer Road Sunnyvale, California 94086
Re: K211997
Trade/Device Name: 8 mm SureForm 30 Curved-Tip Stapler, 8 mm SureForm 30 Stapler, SureForm 30 Reloads Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY, GDW Dated: November 5, 2021 Received: November 10, 2021
Dear Amrit Jaggi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mark Trumbore, Ph.D. Assistant Director THT4A1: Robotically-assissted Surgical Devices Team DHT4A: Division of Sugical Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Form FDA 3881 – Indications for Use statement 6
There is 1 page in this section following this cover sheet.
Image /page/2/Picture/4 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "U". The font is sans-serif and the letters are evenly spaced.
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) New device
Device Name
8 mm SureForm 30 Curved-Tip Stapler 8 mm SureForm 30 Stapler 8 mm SureForm 30 Reloads
#### Indications for Use (Describe)
The Intuitive Surgical 8 mm SureForm 30 staplers and reloads and accessories are intended to be used with a compatible da Vinci Surgical system for resection of vasculature and tissue and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CER 801 Subpart D) </b></span> | Over-The-Counter Use (21 CER 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
#### [As Required by 21 CFR 807.92(c)]
| Submitter: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|-------------------|-------------------------------------------------------------------------------------------------------|
| Official Contact: | Amrit Jaggi<br>Senior Regulatory Specialist<br>Phone Number: 949-690-8799<br>Fax Number: 408-523-8907 |
| Date Prepared: | June 25, 2021 |
| Trade Name: | 8 mm SureForm™ 30 Curved-Tip Stapler, 8 mm SureForm 30<br>Stapler, 8 mm SureForm 30 Reloads |
| Common Name: | System, surgical, computer controlled instrument |
| Classification: | Class II<br>21 CFR 876.1500, Endoscope and Accessories<br>21 CFR 878.4750, Implantable Staple |
| Product Codes: | NAY (Endoscope and accessories)<br>GDW (Implantable Staple) |
Predicate Device: SureForm 45 Staplers and Reloads (K190999)
## Device Description
The Intuitive Surgical 8 mm SureForm 30 Curved-Tip Stapler and the 8 mm SureForm 30 Stapler are disposable, fully wristed, articulating, surgical staplers and are designed for use exclusively with the Intuitive Surgical da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200 Systems). The staplers are controlled by the surgeon using the Surgeon Console of the IS4000/IS4200 Systems. They are intended for resection, transection and/or creation of anastomoses in surgery. The staplers achieve their intended use by placing multiple staggered rows of implantable staples in the target tissue (stapling) followed by cutting of the target tissue along the middle of the staple line (transection). The 8 mm SureForm 30 Reloads consist of a single-use cartridge that contains four staggered rows of implantable titanium alloy (Ti3A12.5V) staples. The reloads are offered in three configurations (gray, white, and blue). Each color represents a different staple leg height and tissue gap for use with various tissue thicknesses. Table 1 outlines the specifications of the reloads.
Image /page/4/Picture/7 description: The image shows the word "INTUITIVE" in blue font. There is a small blue circle above the letter "U". The font is sans-serif and the letters are evenly spaced.
{5}------------------------------------------------
| | 8 mm SureForm 30 Reloads | | |
|--------------------------------------------------------|----------------------------------------------------|----------------------------------------------------|----------------------------------------------------|
| Attribute | Gray | White | Blue |
| No. of staple rows<br>and staple line<br>configuration | 4 rows total;<br>2 on each side of the<br>cut line | 4 rows total;<br>2 on each side of the<br>cut line | 4 rows total;<br>2 on each side of the<br>cut line |
| No. of Staples | 34 staples | 34 staples | 34 staples |
| Unformed staple<br>leg length | 2.0 mm | 2.5 mm | 3.5 mm |
| Image | Image: gray stapler | Image: white stapler | Image: blue stapler |
Table 1 8 mm SureForm 30 Reloads Specifications
The reloads are single use devices and are shipped sterile to the surgeon with a retainer and bottom cover that protects the staples during shipping and transportation. The 8 mm SureForm 30 Reloads are not compatible with any other Intuitive Surgical stapler instruments and likewise, the existing Intuitive Surgical stapler reloads are not compatible with the 8 mm SureForm 30 Staplers.
## Intended Use
The 8 mm SureForm 30 Staplers and Reloads are intended to resect, transect and/or create anastomoses in surgery.
## Indications for Use
The Intuitive Surgical 8 mm SureForm 30 Staplers and Reloads and accessories are intended to be used with a compatible da Vinci Surgical system for resection and transection of vasculature and tissue and/or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.
## Comparison of Technological Characteristics
The subject device, 8 mm SureForm 30 Staplers and the predicate device, SureForm 45 Staplers and Reloads (K190999) are regulated under the same regulation number, product code, and classification. They have the same intended use, sterility characteristics (EO sterilization), and principles of operation. The subject device, 8 mm SureForm 30 Staplers and Reloads, and the predicate SureForm 45 Staplers and Reloads
Image /page/5/Picture/10 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the word. The font is sans-serif and appears to be a clean, modern design. The overall impression is simple and straightforward.
{6}------------------------------------------------
differ in some design attributes (the subject device has a narrower diameter and shorter staple line length), indications for use (the subject device is indicated for use on vasculature and tissue), and patient-contacting materials (some same, some different).
#### Performance Data
Performance test data (bench and animal tests) for the 8 mm SureForm 30 Staplers and Reloads demonstrate that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements, user needs, and intended use. The testing is summarized below.
#### Design Verification (bench):
The 8 mm SureForm 30 Staplers and Reloads were subject to full design verification testing included:
- Physical Specifications ●
- Mechanical Requirements ●
- Electrical Requirements ●
- User Interface Requirements ●
- Equipment Interface Requirements ●
- Reliability ●
- Packaging and labeling
The 8 mm SureForm 30 Staplers and Reloads met all of the bench testing acceptance criteria, demonstrating that that the design output meets the design input requirements.
#### Design Validation (animal):
A series of acute and chronic clinical validation studies were performed using simulated clinical models (animal) to evaluate the performance of the subject device. Acute clinical validation studies included Staple Line Performance, Hemostasis Testing, Leak Onset Testing, Maximum Torque Evaluation, and Design Validation Testing. Chronic survival studies included Lung Lobectomy Study, Lung Wedge Resection Study, Gastrectomy Study, Small Bowel Anastomosis, and Nephrectomy Study. A side-by-side comparison between the subject and predicate device was performed in the Staple Line Performance, Hemostasis Testing, Leak Onset Testing, Maximum Torque Evaluation, and all Chronic Survival Studies to demonstrate substantial equivalence between the subject and predicate devices. Design Validation Testing demonstrated that the design outputs of the subject device fulfill the design input requirements and that user needs and intended uses are met. A summary of the animal validation studies is provided in Table 2 below.
Image /page/6/Picture/16 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "I" in the word. The font is sans-serif and the letters are evenly spaced. The image is simple and clean, with a focus on the word itself.
{7}------------------------------------------------
| Study Name | Study Purpose | Animal<br>Model | Study Outcome |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Acute Testing: | | | |
| Staple Line<br>Performance | Assess staple line<br>performance and staple<br>formation of the subject<br>device compared to the<br>predicate | 3 canine<br>models<br>1 porcine<br>model | The subject device met all<br>acceptance criteria and<br>exhibited acceptable pass<br>rates in the areas of<br>transection, tissue layer<br>approximation, hemostasis,<br>and staple formation |
| Hemostasis | Assess hemostasis<br>performance for the subject<br>device as compared to the<br>predicate device. | 2 porcine<br>models | The subject device<br>demonstrated equivalent<br>hemostasis performance to<br>the predicate device. |
| Leak Onset<br>Pressure | Assess leak onset pressure<br>performance on ex vivo<br>tissue of the subject device<br>as compared to the predicate<br>device. | 1 porcine<br>model | All staple lines met all<br>acceptance criteria and<br>exhibited acceptable pass<br>rates in leak onset pressure<br>on thick (stomach) tissue<br>and thin (vein) tissue. |
| Maximum<br>Torque | Evaluate staple line<br>performance at maximum<br>SmartFire torque limits of<br>the subject device as<br>compared to the predicate<br>device. | 2 canine<br>models<br>1 porcine<br>model | The 8 mm SureForm 30<br>Stapler and Reloads<br>demonstrated clinically<br>acceptable performance and<br>met all design<br>specifications. The subject<br>device demonstrated<br>equivalent clinical<br>performance when<br>compared to the predicate<br>device in all adjacent fires.<br>No new of issues of safety<br>and efficacy were raised. |
| Design<br>Validation | Design validation testing of<br>the subject device was<br>performed in a clinical<br>laboratory setting closely<br>approximating an | 2 canine<br>models<br>2 porcine<br>models | The subject device met all<br>acceptance criteria. |
| Study Name | Study Purpose | Animal<br>Model | Study Outcome |
| | intraoperative use situation. | | |
| Chronic Testing: | | | |
| Lung<br>Lobectomy | Assess subject device<br>performance as compared to<br>the predicate device in a<br>lung lobectomy procedure. | 8 canine<br>models | All staple lines passed<br>assessments for leaks intra-<br>operatively. All animals<br>survived through the 28 day<br>+ survival period. During<br>the terminal procedures,<br>there were no signs of<br>bleeding or leakage at the<br>staple lines. Staple lines<br>were well-healed at the end<br>of the survival period for<br>both the subject and<br>predicate devices. |
| Lung Wedge<br>Resection | Assess subject device<br>performance as compared to<br>the predicate device in a<br>lung wedge resection<br>procedure. | 8 canine<br>models | All staple lines passed<br>assessments for leaks intra-<br>operatively. All animals<br>survived through the 7 day +<br>survival period. During the<br>terminal procedures, there<br>were no signs of bleeding or<br>leakage at the staple lines.<br>Staple lines were well-<br>healed at the end of the<br>survival period for both the<br>subject and predicate<br>devices. |
| Small Bowel<br>Anastomosis | Assess subject device<br>performance as compared to<br>the predicate device in a<br>small bowel anastomosis<br>procedure. | 8 porcine<br>models | All staple lines passed<br>assessments for leaks intra-<br>operatively. All animals<br>survived through the 14 day<br>+ survival period. During<br>the terminal procedures,<br>there were no signs of<br>bleeding or leakage at the<br>staple lines. Staple lines<br>were well-healed at the end<br>of the survival period for<br>both the subject and predicate devices. |
| Study Name | Study Purpose | Animal<br>Model | Study Outcome |
| Nephrectomy | Assess subject device<br>performance as compared to<br>the predicate device in a<br>nephrectomy procedure. | 8 porcine<br>models | All staple lines passed<br>assessments for leaks intra-<br>operatively. All animals<br>survived through the 28 day<br>+ survival period. During<br>the terminal procedures,<br>there were no signs of<br>bleeding or leakage at the<br>staple lines. Staple lines<br>were well-healed at the end<br>of the survival period for<br>both the subject and<br>predicate devices. |
| Gastrectomy | Assess subject device<br>performance as compared to<br>the predicate device in a<br>gastrectomy procedure. | 8 canine<br>models | All staple lines passed<br>assessments for leaks intra-<br>operatively. All animals<br>survived through the 14 day<br>+ survival period. During the terminal procedures,<br>there were no signs of<br>bleeding or leakage at the<br>staple lines. Staple lines were well-healed at the end of the survival period for both the subject and predicate devices. |
## Table 2 Summary of Animal Validation Studies
Image /page/7/Picture/3 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "I". The font is sans-serif and appears to be a bold typeface.
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "I" in the word. The font is sans-serif and the letters are evenly spaced. The image is simple and clean, with a white background.
{9}------------------------------------------------
## Human Factors Evaluation:
As part of the Usability Engineering Process for the 8 mm SureForm 30 Staplers and 8 mm SureForm 30 Reloads, the Usability Risk Analysis was updated to identify any new usability characteristics related to safety, as well as foreseeable hazards and hazardous situations. Human factors evaluation was conducted on the 8 mm SureForm 30 Staplers and 8 mm SureForm 30 Reloads. Based on the results of those studies, the 8 mm SureForm 30 Staplers and Reloads has been found to be safe and effective for the intended users, uses, and use environments.
Image /page/9/Picture/4 description: The image shows the word "INTUITIVE" in blue font. There is a small circle above the letter "I". The font is sans-serif and the text is horizontally aligned.
{10}------------------------------------------------
Based on the indications for use, technological characteristics, and Summary: performance data, the subject device, 8 mm SureForm 30 Staplers and Reloads are substantially equivalent to the predicate devices, the SureForm 45 Staplers and SureForm Reloads.
Image /page/10/Picture/2 description: The image shows the word "INTUITIVE" in blue font. There is a small blue circle above the letter "I". The font is sans-serif and the letters are evenly spaced. The background is white.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.