← Product Code [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY) · K180351

# EndoWrist Mercury Bipolar Grasper (K180351)

_Intuitive Surgical, Inc. · NAY · Aug 7, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K180351

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [NAY](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/NAY.md)
- **Decision Date:** Aug 7, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The EndoWrist Mercury Bipolar Grasper is intended to be used with compatible systems for endoscopic manipulation of tissue, including dissection, grasping, retraction, and bipolar coagulation of tissue.

## Device Story

EndoWrist Mercury Bipolar Grasper is a multiple-use, electrosurgical endoscopic instrument for use with da Vinci Xi (IS4000) and da Vinci X (IS4200) surgical systems and an external electrosurgical unit (ESU). Device comprises housing, shaft, wrist, and grasping tip. Operates via robotic control to interact with patient tissue; allows user to select between two grip strength modes (DEFAULT and STRONG). Used by surgeons in operating rooms to perform tissue grasping, dissection, retraction, and bipolar coagulation. Provides enhanced control and grip force options compared to previous iterations, facilitating precise tissue manipulation during minimally invasive procedures.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (physical, mechanical, electrical, user interface, and software verification) and design validation using animal and cadaver models to confirm functional performance and compatibility with da Vinci surgical systems.

## Technological Characteristics

Multiple-use electrosurgical endoscopic instrument; consists of housing, shaft, wrist, and grasping tip. Features two selectable grip strength modes (DEFAULT and STRONG). Compatible with da Vinci Xi (IS4000) and da Vinci X (IS4200) systems and external ESU. Robotic-controlled articulation.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- EndoWrist Fenestrated Bipolar Forceps ([K131861](/device/K131861.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 7, 2018

Intuitive Surgical, Inc. Mr. Vishal Kanani Sr. Regulatory Affairs Specialist 1266 Kifer Road Sunnyvale, California 94086

Re: K180351

Trade/Device Name: EndoWrist Mercury Bipolar Grasper Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: NAY Dated: June 29, 2018 Received: July 2, 2018

Dear Mr. Kanani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/ofpmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Jennifer R. Stevenson For Binita S. Ashar, M.D., M.B.A., F.A.C.S

Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES         | Form Approved: OMB No. 0910-0120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration                    | Expiration Date: 06/30/2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for Use                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                 | See PRA Statement below.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 510(k) Number (if known)                        | K180351                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device Name                                     | EndoWrist Mercury Bipolar Grasper                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Indications for Use (Describe)                  | The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System, Model IS4000) is intended to             assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp             endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and             accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation,             ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and             accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic             surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures.             The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac             revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an             operating room environment in accordance with the representative, specific procedures set forth in the Professional             Instructions for Use. |
| Type of Use (Select one or both, as applicable) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                 | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                 | CONTINUE ON A SEPARATE PAGE IF NEEDED.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                 | This section applies only to requirements of the Paperwork Reduction Act of 1995.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                 | <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                 | The burden time for this collection of information is estimated to average 79 hours per response, including the             time to review instructions, search existing data sources, gather and maintain the data needed and complete             and review the collection of information. Send comments regarding this burden estimate or any other aspect             of this information collection, including suggestions for reducing this burden, to:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                 | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                 | <i>"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of             information unless it displays a currently valid OMB number."</i>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

Image /page/2/Picture/3 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is in a smaller font and also in a light gray color, with the registered trademark symbol next to it.

{3}------------------------------------------------

|                                                 | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020 |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
|                                                 | Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | See PRA Statement below.                                        |
| 510(k) Number (if known)                        | K180351                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                 |
| Device Name                                     | EndoWrist Mercury Bipolar Grasper                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                 |
| Indications for Use (Describe)                  | The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model IS4200) is intended to             assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp             endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and             accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation,             ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and             accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopie             surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The             system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac             revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an             operating room environment in accordance with the representative, specific procedures set forth in the Professional             Instructions for Use. |                                                                 |
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                 |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

Image /page/3/Picture/3 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in a larger, sans-serif font, with a yellow dot above the "I". Below that, in a smaller font, is the word "SURGICAL" with the registered trademark symbol.

{4}------------------------------------------------

#### 510(k) Summary 6

[As Required by 21 CFR 807.92(c)]

| March 1, 2018 |
|---------------|
|               |

- Submitter: Intuitive Surgical, Inc. 1266 Kifer Road Sunnyvale, CA 94086
- Official Contact: Vishal Kanani Sr. Regulatory Affairs Specialist Ph: 408-523-2035 Fax: 408-523-8907
- Trade Name: EndoWrist® Mercury Bipolar Grasper
- Common Name: Endoscopic instrument control system, endoscopic instruments and accessories
- Endoscope and accessories, 21 CFR 876.1500, NAY Classification:

Predicate Device: EndoWrist® Fenestrated Bipolar Forceps (K131861)

# Device Description:

The EndoWrist Mercury Bipolar Grasper is a multiple-use, electrosurgical endoscopic instrument with a grasping tip to be used in conjunction with the da Vinci Xi (IS4000) and da Vinci X (IS4200) Systems and an external electrosurgical unit (ESU). Similar to the predicate device, the EndoWrist Mercury Bipolar Grasper consists of the housing, shaft, wrist, and tip. The shaft and wrist allow for different axes of rotation, and the instrument tip is used to interact with the patient tissue. In addition, the subject Mercury Bipolar Grasper allows a user to select between two grip strength modes, DEFAULT and STRONG.

# Intended Use:

The EndoWrist Mercury Bipolar Grasper is intended to be used with compatible systems for endoscopic manipulation of tissue, including dissection, grasping, retraction, and bipolar coagulation of tissue.

Image /page/4/Picture/15 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is also in all caps and in a smaller font size, with a registered trademark symbol next to it.

{5}------------------------------------------------

#### Indications for Use:

### da Vinci Xi (IS4000) System:

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

#### da Vinci X (IS4200) System:

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

# Technological Characteristics:

The subject Mercury Bipolar Grasper is equivalent to its predicate, Fenestrated Bipolar Grasper in terms of its indications for use, intended use, design, technology, and performance specifications. Modifications from the predicate device include providing a user the ability to

Image /page/5/Picture/9 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is also in all caps and in a smaller font size, with a registered trademark symbol to the right.

{6}------------------------------------------------

select from two grip modes, DEFAULT and STRONG, and minor changes in the jaw design. These modifications do not affect the intended use or the fundamental technology of the device.

# Performance Data:

Design Verification and Validation were performed on the subject Mercury Bipolar Grasper to demonstrate that the design outputs meet the design inputs and the device performs as intended.

Design Verification: The subject device was subjected to a series of bench tests to evaluate performance and to demonstrate that the design outputs meet the design input requirements. Testing was performed with a compatible da Vinci surgical system. The design verification testing included confirmation that the device meets the physical, mechanical, electrical, user interface and software specifications.

# Design Validation:

The design validation testing summarized in this submission validates general, functional, and interaction (compatibility) requirements for the subject device. Tests with an animal model and a cadaver were performed to confirm that the subject device functions in accordance with its intended use.

# Summary:

Based on the intended use, technical characteristics, and performance data, the subject Mercury Bipolar Grasper is equivalent to the predicate device in terms of safety, effectiveness, and performance.

Image /page/6/Picture/10 description: The image shows the logo for Intuitive Surgical. The logo consists of the word "INTUITIVE" in all caps, with a small yellow dot above the "I". Below the word "INTUITIVE" is the word "SURGICAL" in smaller, all-caps letters with a trademark symbol.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K180351](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/NAY/K180351)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
