The SOLOASSIST II allows the user (surgeon) to control movements of a rigid endoscope directly by using the joystick. The SOLOASSIST II is intended for use in minimally invasive abdominal surgery, thoracic surgery, urology surgery or gynecology surgery where a rigid endoscope is intended for use. The SOLOASSIST II gives the advantage of an image without jerking.
Device Story
Robotic, computer-driven system for holding/positioning rigid laparoscopes/endoscopes in ORs. Surgeon operates via joystick or manual release button on control panel. System uses three motorized axes to simulate arm movement with multiple degrees of freedom. Input: user joystick commands; Trocar point defined by tracer pin at body opening. Output: motorized endoscope positioning. Software calculates required axis movements based on Trocar point and desired view to provide stable, jerk-free visualization. Enables surgeon to convert to open surgery via quick-fastener. Benefits: stable endoscopic visualization, reduced surgeon fatigue, precise control.
Clinical Evidence
No clinical data. Evidence consists of bench testing (mechanical, acoustic, lifetime, force, shock/impact/drop resistance), electrical safety/EMC (IEC 60601 series), biocompatibility (ISO 10993), sterilization validation, and human factors testing (15 operations).
Technological Characteristics
Motorized arm system with 3 axes; 11.5kg weight; 1kg safe working load. Materials: sterilizable components (universal joint, joystick, clamp, tension sleeve). Energy: electric. Connectivity: joystick wired to control unit. Software: moderate level of concern. Sterilization: steam pressure. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 62304 (software), ISO 10993 (biocompatibility).
Indications for Use
Indicated for minimally invasive interventions requiring rigid laparoscope/endoscope visualization, including laparoscopic cholecystectomy, hernia repair, appendectomy, pelvic lymph node dissection, assisted hysterectomy, laparoscopic/thoracoscopic procedures, decompression fixation, wedge resection, lung/pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for CABG, and cardiac chamber examination during valve replacement. Users: general surgeons, gynecologists, cardiac/thoracic surgeons, and urologists.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance.
September 21, 2018
AKTORmed GmbH % Mr. Martin Dumberger Managing Director Micro-Epsilon America 8120 Browneigh Road Raleigh, North Carolina 93092
### Re: K171947
Trade/Device Name: SOLOASSIST II Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: August 20, 2018 Received: August 22, 2018
### Dear Mr. Dumberger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| 510(k) Number (if known) | K171947 |
|--------------------------------|---------------|
| Device Name | SOLOASSIST II |
| Indications for Use (Describe) | |
The intended use of the SOLOASSIST II is a robotic computer driven system whose function is to hold and position a rigid laparoscope / endoscope.
The SOLOASSIST II is indicated for use in minimally invasive interventions where a rigid laparoscope / endoscope is indicated for use. Surgeries, SOLOASSIST II is used, are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectomy, laparoscopic & thorascopic, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the SOLOASSIST II are general surgeons, gynecologists, cardiac surgeon, thoracic surgeon and urologists.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)<br>☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
្រុក
C
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for AKTORmed Robotic Surgery. The logo features the letters 'am' in a blue circle on the left. To the right of the circle is the word 'AKTORmed' in a larger, bold font, with the letters 'TM' in a smaller font in the upper right corner. Below 'AKTORmed' are the words 'ROBOTIC SURGERY' in a smaller font.
# 510(k) Summary for SOLOASSIST II
The 510(k) summary information is being submitted in accordance with the requirements of the Code of Federal Regulations Title 21 section 807.92.
## 1. Prepared by
| Name and date<br>510(k) Summary was prepared | Author name: Andreas Mohr<br>Date created: 2018-JUL-19 |
|----------------------------------------------|--------------------------------------------------------|
|----------------------------------------------|--------------------------------------------------------|
## 2. 510 (k) Submitter
| Manufacturer / 510(k) Submitter<br>(Company Name, Address, Web) | AKTORmed GmbH<br>Borsigstraße 13<br>D-93092 Barbing<br>Germany<br>http://aktormed.info/en/homepage |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person for 510 (k)<br>(Name, Title, Address, Phone & Fax Number, Email) | Andreas Mohr<br>M.Sc.<br>Borsigstraße 13<br>D-93092 Barbing<br>Germany<br>Tel: +49 9401 9320 150<br>Fax: +49 9401 9320 115<br>andreas.mohr@aktormed.com |
## 3. New Device
| Device Trade Name & Model Number(s)<br>(The trade name is the name under which the device will be<br>marketed) | SOLOASSIST II |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Device Common or Usual Name<br>(is the name of the device as it is commonly known e.g., syringe, hip<br>implant.) | Active endoscope holder |
| Device Medical specialty panels and Regulation<br>Part Number<br>(such as Anesthesiology 21CFR868, Cardiovascular Part<br>21CFR870, Chemistry 21CFR862, Dental 21CFR872 ) | Gastroenterology-Urology 21 CFR 876 |
| Device Regulation Number<br>(such as 21CFR 868.1075 ) | 21 CFR 876.1500 |
| Device Regulation Name<br>(such as Argon gas analyzer ) | Endoscope and accessories |
| Device Classification Product Code<br>(3 character unique product identifier) | NAY |
| Device Class<br>(i.e., whether it is unclassified or a class I, II, or III device) | Class II |
{4}------------------------------------------------
## 4. Predicate Device
| Manufacturer / 510(k)Submitter<br>(Company Name) | COMPUTER MOTION, INC. |
|----------------------------------------------------------------------------------------------------------------|-----------------------|
| Device Trade Name & Model Number(s)<br>(The trade name is the name under which the device will be<br>marketed) | AESOP 3000 System |
| Device Regulation Number<br>(such as 21 CFR 868.1075..) | 21 CFR 876.1500 |
| Device Classification Product Code<br>(3 character unique product identifier) | GCJ |
| Device Class<br>(i.e., whether it is unclassified or a class I, II, or III device) | Class II |
| Predicate Device 510(k) Number | K972699 |
## 5. Description of the Device
### Device Identification:
- Key device: .
- Soloassist II ।
- Device components
- Mains adapter with connector and connecting cable -
- Endoscope clamp l
- -Tension sleeve Ø4 - 12,7mm
- -Sterilisation Tray
- Joystick -
- . Patient contacting components
- Universal joint (tracer pin) -
- Device accessories ●
- Transport trolley -
- -Sterile Cover
### Device Characteristics
- Weight: 11,5kg ●
- Dimensions: 1153 x 401 x 270mm ●
- Safe working load: ● 1kg
- 3 motorized axes of the arm system ●
- Software based movement control of the arm system .
- Controlled by joystick ●
- Quick-fastener
- Sterilization method: steam pressure sterilization (Universal joint, joystick, endoscope ● clamp, tension sleeve)
### Environment of Use
- healthcare facility/hospital: .
- Operating room / operation table ।
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for AKTORmed Robotic Surgery. The logo features the letters "am" in a blue circle on the left. To the right of the circle is the word "AKTORmed" in gray, with the letters in a bold sans-serif font. Below the word "AKTORmed" is the phrase "ROBOTIC SURGERY" in a smaller, sans-serif font.
#### Brief Written Description of the Device:
The SOLOASSIST II allows the user to hold and control the movements of a rigid endoscope by using a joystick or manually by pushing the release button on the Control Panel that is located on the SOLOASSIST II.
The joystick is linked to related connection of the SOLOASSIST II by a wire. The joystick is mounted on the surgical instrument of the surgeon. The movement will be controlled by three motorized axes inside the SOLOASSIST II. The SOLOASSIST II simulates an arm working in several degrees of freedom. The Control Panel of the SOLOASSIST II has 2 buttons and 5 indications. One button is used to set the TROCAR POINT and the other button is used to move the SOLOASSIST manually by pushing the button while moving the arm.
The Tension Sleeve and the Endoscope clamp are loosely screwed together. The combination of Tension Sleeve and Endoscope clamp are slide over the endoscope and the tension sleeve is tightened so the endoscope can't move out but still can turn. It is important to use the right tension sleeve for the right diameter of the endoscope. On the rigid endoscope normally there is a camera linked that sends a video to a monitor. Endoscope and camera as well as the monitor are not part of the SOLOASSIST II.
The starting point of the movement for an operation will be saved as the TROCAR POINT. The Universal Joint has to be mounted on the SOLOASSIST II. A mounted Universal Joint can be removed by using the release slider. To define the Trocar point the Universal Joint has a small tracer pin. The tracer pin is placed near the body opening for the endoscope. The TROCAR POINT has to be saved by pushing the button on the control panel of SOLOASSIST
The Endoscope Clamp (together with Tension Sleeve and Endoscope) is mounted on the Universal Joint, so the SOLOASSIST II is ready for operation.
Based on the trocar point and the desired view of the software of the device calculates the required individual movements of the axes in order to achieve the desired total movement. For example by pushing the left button on the joystick the image on the monitor moves left, while the SOLOASSIST II is moving to produce the desired view. By pushing the buttons on the joystick it is possible to move the image on the monitor to the left, right, up, down, zoom in and zoom out.
Additionally it is possible to move the arm of the SOLOASSIST II without the use of the iovstick. Therefore the user has to hold the control panel on the SOLOASSIST II and push the unlocking button (release button) while moving the arm system.
To allow the surgeon to convert to an open surgery in case of an emergency, the SOLOASSIST II can be completely mounted or removed by a quick fastener.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters 'am' inside, followed by the word 'AKTORmed' in a larger, bolder font. Below 'AKTORmed' are the words 'ROBOTIC SURGERY' in a smaller font.
#### Key Performance Specifications/Characteristics of the Device
Stable fixation of an endoscope: The system holds an endoscope in a fixed position, which was adjusted by the user. This must also be assured when power is lost.
Enabling the repositioning of an endoscope: The system enables the user (surgeon) to reposition the endoscope by use of a joystick or alternative input device. In case of power loss or other failures it is acceptable that this function is not available.
Manual repositioning of the system arm: The system enables the user (surgeon) to reposition the endoscope by pushing a button. In case of power loss or other failures it is acceptable that this function is not available.
## 6. Intended Use and Indication for Use
## 6.1. Intended Use
The SOLOASSIST II allows the user (surgeon) to control movements of a rigid endoscope directly by using the joystick. The SOLOASSIST II is intended for use in minimally invasive abdominal surgery, thoracic surgery, urology surgery or gynecology surgery where a rigid endoscope is intended for use. The SOLOASSIST II gives the advantage of an image without jerking.
## 6.2. Indication for Use
The intended use of the SOLOASSIST II is a robotic computer driven system whose function is to hold and position a rigid laparoscope / endoscope.
The SOLOASSIST II is indicated for use in minimally invasive interventions where a rigid laparoscope / endoscope is indicated for use. Surgeries, SOLOASSIST II is used, are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic lymph node dissection, laparoscopically assisted hysterectomy, laparoscopic & thorascopic, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement.
The users of the SOLOASSIST II are general surgeons, gynecologists, cardiac surgeon, thoracic surgeon and urologists.
## 6.3. Comparison to Predicate Device
The AKTORmed SOLOASSIST II is substantially equivalent in intended use and technology to the predicate device, COMPUTER MOTION, INC. AESOP 3000 System (K972699).
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" in white on the left side. To the right of the circle is the word "AKTORmed" in gray, with the letters in a bold, sans-serif font. Below "AKTORmed" is the phrase "ROBOTIC SURGERY" in a smaller, lighter gray font.
## 7. Comparison of Technological Characteristics with the Predicate Device
Holding and controlling the movement of a rigid endoscope by motorized axes is the technological principle for both, the SOLOASSIST II and the predicate devices AESOP 3000. Both devices are used for minimally invasive interventions with a software based assistance system.
At a high level, SOLOASSIST II and the predicate devices AESOP 3000 are based on the following same technological elements:
- Used for minimally invasive surgery ●
- Software based control of moving and holding of a rigid endoscope .
- Joystick to control the device ●
- Electricity as energy source
- . Compatible with most of the common endoscopes
- Similar accessory
- . Invariant trocar point to position the arm system
The following technological differences exist between SOLOASSIST II and the predicate devices:
- Different assembly of the motorized axes ●
- Joystick of SOLOASSIST II can be mounted on the MIT instrument of the surgeon, so ● the surgeon can move the system while operating.
- Additional input devices for the AESOP 3000: Voice control and foot pedals ● → Not implementing these input devices in the SOLOASSIST do not affect the safety and efficiency
- Number of motorized axes .
- Different dimensions: SOLOASSIST II is smaller than AESOP 3000 ●
- Safe working load: SOLOASSIST II can carry a load of 1kg, AESOP 3000 only 500g ●
## 8. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
## 8.1. Summary Non-Clinical Performance Data
### 8.1.1.Biocompatibility testing
The biocompatibility evaluation for the SOLOASSIST II device was conducted according to Directive 93/42/EEC, 90/385/EEC, DIN EN ISO/IEC 17025: 2005 (DAkkS D-PL-13392-01-00 accredited, ZLG-AP-311.10.26 recognized), and Good Laboratory Practices (GLP).
Following tests were conducted:
Cytotoxicity, L 929-Proliferation EN ISO 10993-5, -12, LM P 4-06, LM SOP 4-06-01 Chemical analysis (characterization of organic leachables/extractables) EN ISO 10993-1, -12, -18, LM P8-01, LM SOP 8-01-01
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" inside, followed by the word "AKTORmed" in gray. Below the word "AKTORmed" is the phrase "ROBOTIC SURGERY" in a smaller, gray font. The logo is clean and modern, and it conveys a sense of precision and technology.
### 8.1.2.Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted for the SOLOASSIST II device. The system complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC. They system also complies with IEC 60601-1-4 standard, IEC 60601-1-6 and IEC 60601-2-18.
#### 8.1.3.Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the patient or operator.
The SOLOASSIST II complies with the IEC 62366 standard and the IEC 62304.
#### 8.1.4.Mechanical, acoustic and bench testing
Mechanical, acoustic and bench testing were conducted for the SOLOASSIST II.
Following tests were conducted:
#### Quick release connector and angle joint: -
In this test the tilting of the SOLOASSIST II mounted on rails was tested (0°, 45°, 90°); also the engage of the angle joint of the SOLOASSIST II was tested according to functionality
#### -Temperature test:
Testing the use of the device for its temperature range for use and testing the status of the device for its storage temperature range
#### -Force test:
Testing the forces that are needed to move the arm system while in release status
#### -Lifetime test:
35.000 movements in x-direction, 55.000 movements in y-direction, 35.000 movements in z-direction (movement length 30cm)
Motor lifetime for at least 1000h (500h with cycle movement with radius 15cm and 28cm)
#### -Testing moving after fixation:
- Shifting after manual positioning with 1kg load; 3 turns; averaged । measurement 2,93mm
- Falling speed after 2 hours: max. 0,0024mm / min -
#### Tensile stress: -
(Mechanic according to EN 60601-1:2016; Chapter 9.8.2): Examination weight 8,1kg (8x normal load)
#### -Sound test:
(Mechanic according to EN 60601-1:2006, Chapter 9.6.2.1): 64.0 dBA (pushing release button); 60,2 dBA (during active operation and 1kg load)
#### -Shock resistance:
(Mechanic according to EN 60601-1:2006, Chapter 15.3.2): continuous force of 240N for 5s (contact with surfaces by a 30mm diameter circular area)
#### -Impact resistance:
(Mechanic according to EN 60601-1:2006; Chapter 15.3.3): Steel ball, weight 500q: Height of steel ball: 1,3m
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image is a logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" inside, followed by the word "AKTORmed" in a bold, sans-serif font. Below the word "AKTORmed" is the phrase "ROBOTIC SURGERY" in a smaller, sans-serif font. The TM symbol is located to the right of the word "AKTORmed".
#### -Drop resistance:
(Mechanic according to EN 60601-1:2006, Chapter 15.3.4.2): SOLOASSIST II is hold over a wooden plate (d = 50mm +/- 5mm) in a height of 3cm. Dropped down 3 times out of each start situation that the SOLOASSIST II can be during the intended use.
#### -Tightening test:
(Mechanic according to EN 60601-1:2006, Chapter 15.4.6.1): Turning star know clock and anti-clockwise till stop and hold the position for 2s with a force of 4Nm. Repeated 10 times.
#### 8.1.5.Animal Study
No animal study was conducted for this device
#### 8.1.6.Cybersecurity risk assessment
Cybersecurity risk assessment was conducted based on FDA quideline "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"
#### 8.1.7.Sterilization
The Sterilization tests for the SOLOASSIST II device was conducted according to Directive 93/42/EEC. 90/385/EEC. DIN EN ISO/IEC 17025: 2005 (DAkkS D-PL-13392-01-00 accredited, ZLG-AP-311.10.26 recognized), and Good Laboratory Practices (GLP).
Following tests were conducted:
- Microbiological efficiency control of steam sterilization, fractionated vacuum cycle l / dynamic-air-removal cycle, product group sterilizable arm components EN ISO 17664, EN ISO 17665-1 (overkill method, partial cycle), ANSI/AAMI ST79, ANSI/AAMI ST81, ANSI/AAMI TIR12, ASTM E 1766 mod., EN 556-1, ANSI/AAMI ST67, KRINKO/RKI/BfArM guideline, LM P 2-11-04, LM SOP 2-11-04
- -Microbiological efficiency control of the automated cleaning and disinfection. product group sterilizable arm components EN ISO 17664, ANSI/AAMI ST81, ANSI/AAMI TIR12 mod., AAMI TIR 30 mod., ASTM E 2314 mod., ASTM E 1837 mod., KRINKO/RKI/BfArM quideline, LM SOP 2-11-01
- Efficiency control for cleaning (optimized protein test) during automated cleaning/Disinfection, product group sterilizable arm components EN ISO 17664, ANSI/AAMI ST81, ANSI/AAMI TIR 12 mod., AAMI TIR 30, ASTM E 2314 mod. , KRINKI/RKI/BfArM guideline, FDA guidance, LM SOP 2-11-01
- -Microbiological efficiency control of the manual cleaning and disinfection, product group sterilizable arm components EN ISO 17664, ANSI/AAMI ST81, ANSI/AAMI TIR12 mod., AAMI TIR 30 mod., ASTM E 2314 mod., ASTM E 1837 mod., KRINKO/RKI/BfArM guideline, LM SOP 2-11-02
- -Efficiency control for manual cleaning (optimized protein test), product group sterilizable arm components
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image contains the logo for AKTORmed Robotic Surgery. The logo features a blue circle with the letters "am" inside, followed by the word "AKTORmed" in a larger, bold font. Below the word "AKTORmed" are the words "ROBOTIC SURGERY" in a smaller font.
EN ISO 17664, ANSI/AAMI ST81, ANSI/AAMI TIR 12 mod., AAMI TIR 30, ASTM E 2314 mod. , KRINKI/RKI/BfArM guideline, FDA guidance, LM SOP 2-11-01
- Efficiency control of the cleaning (TOC method) during automated cleaning / disinfection, product group sterilizable arm components EN ISO 17664, ANSI/AAMI/ISO 17664, ANSI/AAMI TIR12 mod., AAMI TIR 30, ASTM E 2314 mod, KRINKO/RKI/BfArM guideline, RDS 007, FDA guidance, LM SOP 2-11-01
#### 8.1.8.Human Factor Testing
Human Factor Testing for the SOLOASSIST II was conducted according to FDA quideline "Applying Human Factors and Usability Engineering to Medical Devices" (issued on February 3, 2016)
Human factor test has been conducted for the SOLOASSIST II. There has been 15 operations conducted. The test was performed successfully. No unexpected device related events or patient related adverse events occurred during the Human Factor Testing.
#### 8.1.9.Packaging
Package drop test was conducted for the SOLOASSIST II and its trollev according to ISTA 1A. SOLOASSIST II and trolley have been packages in wooden box with filling materials. Drop height for mass 45 to 68kg was set to 200mm. All surfaces and relevant edges as well as a corner was tested. No damage of the SOLOASSIST II occurred.
### 8.2. Summary Clinical Performance Data
No clinical evaluation was conducted for this device.
### 8.3. Conclusion Performance Data
Based on the nonclinical performance data, we conclude that the SOLOASSIST II is substantially equivalent to the predicated device under the Federal Food, Drug and Cosmetic Act.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.